Brand:

JUMO flowTRANS MAG H10 Hygienic Magnetic Flowmeter, 406061

Description

A stainless steel magmeter for food, beverage and pharmaceutical lines

The JUMO flowTRANS MAG H10 (type 406061) is an electromagnetic flowmeter designed for hygienic applications in the food and pharmaceutical industries. The sensor is made completely of stainless steel with a vacuum-proof PFA liner, which makes it easy to clean, and the datasheet states that it meets the requirements of the 3A sanitary standards and uses FDA-compliant materials. A variable connection concept combines one standardised sensor with many process connections, so the same meter can be fitted with a Tri-Clamp, a dairy threaded union, weld stubs, a flange or a wafer body. It measures conductive liquids from 20 µS/cm, including demineralised water, and is available as an integral-mount meter or with a remote transmitter on a signal cable of up to 50 m.

Technical specifications

Parameter Specification
Type 406061/1 integral mount, 406061/2 remote mount; remote transmitter flowTRANS MAG 10 (406067/2)
Nominal diameter DN 3 to DN 100 (1/10 to 4 in.)
Accuracy 0.5 % of rate; pulse output ±0.5 % of value ±0.02 % of QmaxDN; reproducibility up to 0.11 % of value
Minimum conductivity Above 20 µS/cm, 20 µS/cm also for demineralised water
Medium temperature -25 to +130 °C; flange and variable connections up to 95 °C at ambient 60 °C, up to 130 °C at ambient 45 °C
Nominal pressure PN 10 to PN 40, ASME CL 150 and CL 300, JIS 10K
Process connections Wafer, DIN/ASME/JIS flange, DIN 11851 threaded, weld stubs, Tri-Clamp DIN 32676 (PN 10 to 16) and ASME BPE, external thread ISO 228 DN 3 to 25
Wetted materials PFA lining, vacuum-proof; electrodes 1.4539 (904L) standard, or 1.4571, Hastelloy B or C, platinum-iridium, tantalum, titanium
Gaskets EPDM or silicone with FDA approval; PTFE with FDA approval for DN 3 to 8
Housing and meter tube Deep-drawn stainless steel 1.4301 or 1.4308 housing, stainless steel tube; painted aluminium terminal box RAL 5013
Flow range Full scale 0.02 to 2 x QmaxDN, for example 4 l/min for DN 3, 200 l/min for DN 25 and 240 m3/h for DN 100
Cleaning Steam up to 150 °C and cleaning fluid up to 140 °C for at most 60 minutes
Supply 100 to 240 V AC (-15 %/+10 %) or 24 to 48 V DC (-10 %/+10 %)
Outputs Active 4 to 20 mA; two passive digital outputs for pulse, frequency or switching
Display Optional rotatable LCD with capacitive buttons and plain-text diagnostics
Ambient temperature -20 to +60 °C; storage -30 to +70 °C
Protection IP65/IP67, IP68 for remote sensors, NEMA 4X
Approvals Pressure Equipment Directive 2014/68/EU category III fluid group 1; FDA-compliant materials; 3A requirements stated

Process connections and pressure limits

The wafer body allows 40 bar up to DN 50 and 16 bar for DN 65 to 100 at -25 to +130 °C. Weld stubs to DIN 2463, ISO 1127 or DIN 11850 allow 40 bar up to DN 40, 16 bar for DN 50 and DN 80 and 10 bar for DN 65 and DN 100, while SMS 1145 weld stubs are limited to 6 bar. Threaded connections to DIN 11851 are rated 40 bar up to DN 40. Because limits depend on the liner and the connection used, check the name plate values and the table in the datasheet before specifying the system.

Electrical details match the larger MAG S10 family: a 5 m signal cable ships with remote meters, a longer one is part number 00645914, and a pairing number links each sensor to its transmitter.

Hygienic installation and use

For 3A-compliant installations the datasheet limits the choice to weld stubs and Tri-Clamp connections, and it sets further rules: do not mount the meter with the terminal box or transmitter housing pointing downward, keep the leakage hole of the process connection at the lowest point, and note that the angel bracket option is not 3A compliant. A vertical position is preferred, and a horizontal meter must be installed so that it drains itself. The housing cover must be sealed without gaps.

Fit the sensor with the flow direction marked, 10 x DN of straight run upstream and 5 x DN downstream, and make sure that gaskets do not protrude into the flow, since turbulence affects accuracy. A conductive deposit can form behind graphite gaskets, so avoid them. CIP and SIP cleaning above the normal medium limit is allowed for a short time only, and the temperature is reduced when the ambient temperature rises above 25 °C. Typical locations are dairy and beverage filling, juice and syrup dosing, pharmaceutical purified-water loops and clean utilities, where the housing resists washdown at IP65, IP67 or NEMA 4X.

Using the flowTRANS MAG H10 in Oman and the GCC

The food and beverage industry in the region, including dairy, juice, bottled water and soft drink plants, uses hygienic magnetic flowmeters for dosing, batching, filling and CIP return lines. Pharmaceutical and personal care plants have similar needs. The all-stainless sensor with a PFA liner can be cleaned in place, and the variable connection concept means the meter can be fitted to the clamp or threaded connections already used in the plant.

For accurate batching, install the meter where the pipe is always full, such as a vertical line with upward flow, and keep the recommended straight runs. Check that cleaning temperatures and chemicals stay within the datasheet limits. Where plants run in hot climates, consider the location of the transmitter, keeping it out of direct sun and away from steam during cleaning.

Seven Star can supply the meter together with matching hygienic process connections and seals, so the whole assembly arrives ready for installation. For new lines, sending the piping drawing or a photo of the existing connection helps confirm the right fitting before the order is placed.

Information to send with an enquiry

  • Nominal size and process connection standard
  • Product, flow range and temperature, including cleaning temperatures
  • Output signals and power supply
  • Hygienic certificates required
  • Quantity and the line or project name

Related JUMO products: JUMO flowTRANS MAG H20 Hygienic Electromagnetic Flowmeter, JUMO flowTRANS MAG S01 Electromagnetic Flowmeter, JUMO flowTRANS MAG I02 Insertion Electromagnetic Flowmeter, JUMO flowTRANS US W01 Ultrasonic Flowmeter for Liquids and JUMO flowTRANS MAG S10 Electromagnetic Flowmeter, Type 406060.

Documents and quotation

The hosted document is JUMO data sheet 406061 (34 pages, English). It includes dimension tables for every connection. Confirm nominal width, process connection, electrode material and mount design before ordering. Seven Star LLC is an independent supplier in Muscat; send the product, line size and cleaning regime for a quotation.

Datasheets & Documents
  • PDF Datasheet JUMO 406061 data sheet 406061 (EN) Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

JUMO is the German maker of temperature sensors, process controllers, pressure transmitters and liquid analysis instruments. Moritz Kurt Juchheim founded it in Fulda in 1948 as M. K. Juchheim KG, making glass thermometers with six employees. The name JUMO comes from the founder's name. It is still a family company headquartered in Fulda, and the parent company has been called JUMO GmbH & Co. KG since 2003.

Seven Star LLC is an independent supplier in Muscat. We list JUMO RTD and thermocouple probes, thermowells, transmitters, controllers, recorders and water analysis instruments, and supply them to process plants, OEMs, HVAC contractors and water utilities in Oman and the GCC by type number.

JUMO company profile

Established 1948, Fulda, Germany, by Moritz Kurt Juchheim
First products Glass and contact thermometers; platinum glass sensors from 1966
Ownership Family-owned; renamed JUMO GmbH & Co. KG in 2003
Product families RTD and thermocouple probes, thermowells, temperature and pressure transmitters, controllers and thermostats, safety temperature limiters, paperless recorders, liquid analysis and flow and level instruments

Milestones

  • 1948: Moritz Kurt Juchheim founds M. K. Juchheim KG in Fulda, making glass thermometers with six employees.
  • 1966: Platinum glass temperature sensors are added.
  • 2003: The parent company is renamed JUMO GmbH & Co. KG.

JUMO products we list

Family Models Typical use
RTD and thermocouple probes 902020, 902030, 902050, 902120, 902006 2xPt100, PROCESStemp, mineral-insulated RTD, 901020 thermocouple, 901210, 902810 hygienic, 902550 surface, 902520 HVAC, MarineTemp Temperature measurement in pipes, vessels, machines, HVAC and food plants
Thermowells and inserts 909710 thermowells, 909740 measuring inserts Protecting sensors and replacing them without opening the process
Temperature transmitters dTRANS T01, T03, T04, T05, T06 SIL, T07 HART, T09, T1000 IO-Link, DELOS T switch Converting sensor signals to 4-20 mA, HART or IO-Link
Pressure and level dTRANS p20, p30, p31 hygienic, p33 Ex, p02 DELTA, MIDAS S05, MIDAS DP10, DELOS S02, MAERA S25, NESOS R20 Pressure in process, hydraulic and OEM systems; tank and well level
Controllers, limiters and recorders DICON touch, IMAGO 500, dTRON 316, iTRON, cTRON, diraTRON 104, eTRON T, safetyM STB/STW, safetyM TB/TW, LOGOSCREEN 601, mTRON T Furnace, oven and process control, over-temperature protection and recording
Liquid analysis and flow AQUIS touch S, AQUIS touch P, AQUIS 500 pH, AQUIS 500 CR, dTRANS pH 02, pH electrode, digiLine DO, ecoLine NTU, flowTRANS MAG H20, flowTRANS US W01 pH, conductivity, oxygen and turbidity in water treatment; liquid flow

Choosing a temperature sensor starts with the range and accuracy. A Pt100 RTD is accurate and stable from about -200 to 600 °C and is the usual choice in process plants. A thermocouple (type K or J, for example) covers much higher temperatures and responds faster but is less accurate. The wiring also matters: a 3-wire or 4-wire Pt100 connection cancels the resistance of the cable, while a 2-wire connection adds an error that grows with cable length. That is why a head-mounted transmitter is often fitted close to the sensor.

Where JUMO products are used

In Oman and the GCC, JUMO sensors and controllers are used in oil and gas and chemical plants, food and beverage production, plastics and metal heat treatment, HVAC and district cooling, water and wastewater treatment, swimming pools and marine applications.

Ordering JUMO products: what we need

Send the JUMO type number and the full order code from the label (for example 902020/10-...). For probes, add the sensor type (Pt100, Pt1000, type K or J), the number of wires, the insertion length and diameter, and the process connection. For controllers, add the input, outputs and supply voltage.

Facts checked against JUMO's company history, October 2026.

FAQ

Questions about the JUMO flowTRANS MAG H10 Hygienic Magnetic Flowmeter,

Answers from our engineering team on installation, compatibility and integration.

01What makes the MAG H10 hygienic?
It has an all-stainless sensor with a vacuum-proof PFA liner, FDA-compliant materials and hygienic process connections such as Tri-Clamp.
02Which sizes are offered?
Nominal diameters from DN 3 to DN 100.
03Which process connections are available?
Wafer, flange, DIN 11851 thread, weld stubs, Tri-Clamp to DIN 32676 or ASME BPE, and external thread up to DN 25.
04Can it be steam cleaned?
The datasheet permits steam up to 150 °C and cleaning fluid up to 140 °C for a maximum of 60 minutes.
05What is the accuracy?
0.5 % of rate for liquids with a minimum conductivity of 20 µS/cm.