Brand:

Yokogawa PH450G-A-A pH/ORP Converter, AC

Description

Introduction to the Yokogawa PH450G-A-A

The pH loop is the place where a converter earns its keep: glass impedance, long cables and wet terminal boxes all conspire against the signal. Yokogawa PH450G-A-A is the AC-powered general purpose pH and ORP converter of the EXAxt 450 line, designed with a dual high-impedance amplifier and a liquid earth so that the measuring electrode and the reference are treated symmetrically.

We list it because maintenance teams in Oman tend to look for the PH450G code by its exact string when a field unit fails; the suffix pair A-A identifies the mains version with cable glands.

Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer.

What PH450G-A-A can read

Parameter Value
pH input range -2 to 16 pH
ORP input -1500 to 1500 mV
rH 0 to 100 rH
Input impedance Dual high impedance, at least 10 to the 12 ohm
Temperature elements Pt1000, Pt100, 350 ohm (DKK), 5k1, 6k8, PTC 10k, 3k Balco: -30 to 140 C; NTC 8k55: -10 to 120 C
Accuracy pH within 0.01; ORP within 1 mV; temperature within 0.3 C
Response Less than 4 s for 90 percent (pH 7 to pH 4)

Outputs, relays and wash

  • Two isolated 4-20 mA outputs, common negative, maximum load 600 ohm, with bi-directional HART 5 on mA1.
  • Output function linear or a 21-step table for pH, temperature, ORP or rH, with PID control and burn-up 21.0 mA or burn-down 3.6 mA per NAMUR NE43.
  • Four SPDT relays rated 100 VA at 250 V AC 5 A; they can signal high or low alarms, run on-off, duty-cycle or pulse-frequency control, trigger wash cycles, show hold, and S4 serves as the fail-safe contact.

Why the diagnostics matter

The converter performs sensor diagnostics by monitoring impedance, which means a cracked glass bulb or a blocked reference junction is flagged by the instrument, not discovered during a shift audit. Temperature compensation follows the Nernst equation with processing options: a configurable temperature coefficient, NEN6411 for water or strong acids and bases, or a programmable matrix. A hold function freezes the mA outputs during calibration, and a logbook records events for retrieval at the screen or through HART.

How PH450G-A-A differs from PH450G-D-U

PH450G-A-A takes 100 to 240 V AC and ships with IP66 cable glands, whereas PH450G-D-U is powered from 12 to 24 V DC (10 W) and has FM nonincendive approval, Class I Division 2, Groups A to D, T6, with NEMA 4X plugs in unused entries. The measurement specification, the 144 by 144 mm housing and the roughly 1.5 kg weight are identical. Option codes /U, /PM, /H5, /AFTG, /ANSI and /AM20 apply only to the general purpose type, which is therefore the more flexible build for mounting.

Applications and compatibility in Oman

Common duties are effluent neutralisation, scrubber and flue-gas desulphurisation liquor control, boiler feedwater conditioning, swimming-pool and potable water dosing, and chemical plant reactors. The PH450G family pairs with pH sensors that use the G measuring-system code, for example PH8ERP and PH8EFP in the G variant, and with the WTB10 terminal boxes for long runs. Tell us your sensor code so we can check the pairing.

Compatible sensors and cables for PH450G-A-A

The input is a glass-electrode pH input, so the sensor must be a pH or ORP sensor with a standard high-impedance signal and a temperature element from the supported list. Cable length and quality matter more here than on a conductivity loop; high impedance means moisture in a junction box becomes a measurement error. Where the converter and sensor are far apart, use a terminal box and keep cable runs short. Provide the sensor code with your RFQ.

The converter also supports rH, the combined function of pH and ORP, which is helpful when a process such as cyanide destruction or chrome reduction is tracked by redox potential and pH together. With ORP input accuracy within 1 mV and a range of plus or minus 1500 mV, the instrument covers common redox ranges.

Technical environment limits

Ambient temperature -20 to +55 C, storage -30 to +70 C, humidity 10 to 90 percent RH at 40 C non-condensing, enclosure IP66, NEMA 4X, CSA Type 3S, installation altitude up to 2000 m, installation category II and pollution degree 2. Safety standards listed include EN 61010-1 and UL 61010-1; EMC is EN 61326-1 Class A, which means the unit is intended for industrial environments. Configuration and logbook are stored in EEPROM, a lithium cell backs the clock, and a watchdog timer checks the microprocessor, returning to measurement after a power-down reset. If nobody touches the screen for 10 minutes the unit returns automatically to the measuring mode.

Installation of PH450G-A-A

Mount the unit on pipe, wall or panel, route the sensor cable on its own gland, and connect the liquid earth as described in the manual. During commissioning choose pH, ORP or rH as the primary variable, set the temperature element, and calibrate with two buffers. Keep the display area out of long direct sun using the /H5 hood if the unit is outdoors. For PH450G-A-A the six cable entries carry the supplied IP66 glands, so sensor, power, mA outputs and relay wiring can each have a separate entry. Run a short dry-fit check of the terminal wiring before power-up, confirm that the supply sits between 90 and 264 V AC, and then verify the impedance diagnostics show a healthy glass and reference before you accept the first calibration. Record the buffer temperatures and the slope result in the logbook so that later drift can be compared against the commissioning values.

Documents and quotation

The datasheet for PH450G-A-A is attached. For an offer, send an RFQ to Seven Star LLC giving the part number PH450G-A-A, quantity, any slash options and your delivery location.

Other Models in the PH450G Range

Seven Star LLC supplies models from the PH450G range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.

ModelCategory
Yokogawa PH450G-A-A pH/ORP Converter, ACLiquid Analytical SystemsYou are viewing this model
Yokogawa PH450G-D-U FM pH/ORP Converter, 24 V DCLiquid Analytical SystemsView details
Datasheets & Documents
  • PDF Datasheet Yokogawa PH450G-A-A pH/ORP Converter, AC Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Yokogawa is the Japanese measurement and industrial automation company behind the DPharp EJA and EJX pressure transmitters, CENTUM distributed control systems and ProSafe-RS safety systems. Tamisuke Yokogawa set it up in 1915 as an electric meter research institute in Shibuya, Tokyo. It began selling electric meters in 1917, was incorporated as Yokogawa Electric Works in 1920, and has been Yokogawa Electric Corporation since 1986. In 1975 it released CENTUM, which it describes as the world's first distributed control system.

Seven Star LLC is an independent supplier in Muscat. We list Yokogawa transmitters, flowmeters, analysers, controllers, recorders and CENTUM and ProSafe spare modules, and supply them to oil and gas, power, water and process plants in Oman and the GCC by model code.

Yokogawa at a glance

Established 1915, Shibuya, Tokyo, by Tamisuke Yokogawa with Ichiro Yokogawa and Shin Aoki
Milestones Electric meters from 1917; incorporated in 1920; CENTUM DCS in 1975; renamed Yokogawa Electric Corporation in 1986
Headquarters Tokyo, Japan
Product families Pressure and temperature transmitters, magnetic, vortex and Coriolis flowmeters, liquid and gas analysers, controllers and recorders, CENTUM VP DCS, ProSafe-RS safety systems and STARDOM controllers

Yokogawa products we list

Family Models Typical use
Pressure and DP transmitters EJA110E, EJA430E, EJA530E, EJX110A, EJX530A, EJX910A multivariable Pressure, level and DP flow measurement in process plants
Temperature transmitters YTA610, YTA710 RTD and thermocouple signals converted to 4-20 mA, HART or fieldbus
Flowmeters ADMAG AXG, ADMAG AXW, ROTAMASS TI, VY vortex, VF701 Water and conductive liquids, mass flow of hydrocarbons, steam and gas
Analysers TDLS8000 laser gas analyser, FLXA402, FLXA21 DO, E7042UD zirconia cell, K9471DS cover Combustion O2, in-situ gas analysis, pH, conductivity and dissolved oxygen
Controllers and recorders UT35A/UT32A, UT32A-D, UT55A/UT52A, UT75A, UP35A/UP32A, UP55A, YS1700, GX10/GX20, GP10/GP20, GM10, µR20000 Single-loop control, furnace programs and data recording
Control and safety systems CP451-10 processor, ADV161, ADV551, node interface unit, ProSafe-RS, ProSafe-RS Lite, ProSafe-SLS, FCN-500 Spares and expansions for CENTUM DCS, safety systems and STARDOM
Battery packs S9129FA, S9185FA, S9548FA Memory backup batteries for DCS and controller modules

The EJA and EJX transmitters use Yokogawa's DPharp sensor: two silicon resonators whose vibration frequency changes with pressure, so the output is a digital frequency rather than an analogue strain signal. The model code describes the instrument completely. In EJA110E, "110" is a differential pressure transmitter, and the letters after the dash set the span, wetted materials, process connection, electrical output and approvals. When you replace a transmitter, sending the full code from the nameplate avoids a mismatch in range or materials.

Where Yokogawa products are used

In Oman and the GCC, Yokogawa transmitters, analysers and control systems are found in oil and gas production, gas processing and LNG, refineries and petrochemical plants, power and desalination stations, and water treatment. Many plants still run CENTUM CS 3000 and CENTUM VP systems that need spare I/O modules and batteries.

Ordering Yokogawa products: what we need

Send the full model and suffix code from the nameplate (for example EJA110E-JMS5J-912NN/KF22) and the style code. For DCS modules, add the module type and system version. For analysers, send the sensor type, measuring range and process conditions.

Facts checked against Yokogawa's corporate history, October 2026.

FAQ

Questions about the Yokogawa PH450G-A-A pH/ORP Converter, AC

Answers from our engineering team on installation, compatibility and integration.

01What does PH450G-A-A measure?
It measures pH from -2 to 16, ORP from -1500 to 1500 mV and rH from 0 to 100, together with temperature from -30 to 140 C on most element types, using a dual high impedance input of at least 10 to the 12 ohm.
02How accurate is PH450G-A-A?
pH input accuracy is within 0.01 pH, ORP input within 1 mV and temperature within 0.3 C (0.4 C for Pt100), with a step response under 4 seconds for 90 percent from pH 7 to pH 4. The mA output adds 0.02 mA tolerance.
03Can PH450G-A-A control a sensor washer?
Yes. A relay contact can start manual or interval wash cycles, a contact input can start a remote wash, and the main screen shows WASH and HOLD states.
04Which buffers can PH450G-A-A calibrate with?
Semi-automatic one- or two-point calibration with preconfigured NIST, US and DIN buffer tables for pH 4, 7 and 9, or user-defined buffer tables, with an automatic stability check.
05Is PH450G-A-A AC or DC powered?
AC. The first suffix A is 100 to 240 V AC (90 to 264 V AC), 15 VA. The D version is 12 to 24 V DC at 10 W.