Yokogawa FLXA21 2-wire liquid analyzer overview

The Yokogawa FLXA21 is a two-wire, loop-powered liquid analyzer from the FLEXA series. One compact transmitter measures pH/ORP, contacting conductivity, inductive conductivity or dissolved oxygen, depending on the sensor module installed, and it can host two sensor modules of the same type for dual measurement. The analyzer powers itself from the 4 to 20 mA loop and talks HART, so it fits existing two-wire cabling without a separate supply. Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer.

This Yokogawa FLXA21 page covers the base analyzer family. The exact configuration, including approval type, input module, mounting hardware and cable entries, is selected through the suffix and option codes listed below, so a quotation request should always carry the full model string.

FLXA21 key features

  • Four measurement types in one platform: pH/ORP, SC, ISC and DO.
  • Dual sensor measurement on a two-wire analyzer, with calculated values and a redundant mode that switches to the second sensor if the main sensor fails.
  • Touch screen with a simple menu structure in 12 languages, and automatic recognition of the installed sensor module.
  • Quick setup menu that starts a measurement after only date and time, language, basic sensor settings and output are set.
  • Sensor wellness indication, diagnostics, and a logbook of events for maintenance.
  • Hazardous-location approvals for ATEX, IECEx, FM, CSA, NEPSI and KCs, according to the type code selected.

These capabilities let one FLXA21 replace separate transmitters for several measurement points and reduce spare-part variety on a site.

FLXA21 ordering code structure

Position Choices
FLXA21 Model
-D Power supply, always -D
-P Plastic housing
-D Anti-glare LCD
Type -AB general purpose CE/RCM; -AD CSA; -AG KC; -CB intrinsically safe ATEX/IECEx; -CD intrinsically safe US/Canada; -CH NEPSI; -EG KCs; -DD non-incendive US/Canada
1st input -P1 pH/ORP; -C1 conductivity (SC); -C5 inductive conductivity (ISC); -D1 dissolved oxygen
2nd input -NN none, or the same input type as the first
Output -A 4 to 20 mA plus HART
Language -LA English and 11 languages
Country -N global except Japan; -J Japan
Options /UM universal mounting kit, /U pipe and wall hardware, /PM panel hardware, /H6 /H7 /H8 hoods, /SCT tag plate, /CB4 /CD4 /CF4 conduit adapters, /CFC conformal coating

Because so many combinations exist, the FLXA21 should be quoted from a complete code rather than from the family name alone.

FLXA21 general technical specifications

Parameter Value
Measurement objects pH/ORP, contacting conductivity (SC), inductive conductivity (ISC), dissolved oxygen (DO), depending on the sensor module
Sensor modules One or two; two modules must be of the same type (ISC with ISC not available)
Output 4 to 20 mA DC with bi-directional HART, tolerance plus or minus 0.02 mA; burn-out per NAMUR 43
Power Nominal 24 V DC loop powered; 16 to 40 V DC for one SC module
Display LCD with touch screen, black and white, 213 x 160 pixels, 12 message languages
Housing Polycarbonate case, silver gray; IP66 (except Canada), NEMA Type 4X, CSA Type 3S/4X
Cable entry M20 cable glands, 3 holes for one sensor or 4 holes for two sensors
Size and weight 144 x 144 x 151 mm without cable gland; approx. 1 kg
Ambient temperature -20 to +55 C; storage -30 to +70 C
Humidity 10 to 90 %RH at 40 C, non-condensing
EMC EN 61326-1 Class A, Table 2 (industrial locations)

Measurement specifications depend on the input module. For contacting conductivity, for example, the analyzer accepts cell constants from 0.005 to 50.0 per cm and shows plus or minus 0.5 percent F.S. accuracy over most of its range; the pH module offers a pH range of -2 to 16 with 0.01 pH accuracy on simulated input.

Applications in Oman and GCC industry

Water and process quality control is critical in the region. The FLXA21 is used in power station demineralizers and boiler water loops, desalination and seawater reverse osmosis plants, wastewater treatment, chemical dosing and neutralization, and petrochemical cooling water. The intrinsically safe type options suit hazardous areas in oil and gas facilities, where a loop-powered instrument avoids a separate mains supply.

Selection and compatibility of the FLXA21

Choose the input module first: pH/ORP for acidity and redox, SC for conductivity, ISC for inductive conductivity in aggressive media, DO for dissolved oxygen. Then choose the approval type that matches the area classification, and check the supply voltage: one-sensor versions need 16 to 40 V DC, while two-sensor versions need 22.8 to 40 V DC. Other output types, FOUNDATION Fieldbus and PROFIBUS PA, exist under different suffix codes and are covered in separate specifications. Verify sensor compatibility, including cable type and length, against the Yokogawa sensor list.

Installation and commissioning of the FLXA21

  • Mount the analyzer on a pipe, wall or panel with the matching hardware, and fit the hood where direct sun or rain is expected.
  • Install the supplied cable glands or conduit adapters in the unused holes with correct sealing; they are delivered loose.
  • Connect the loop and the sensor cable, observing the grounding sleeve for the plastic case.
  • Let the analyzer recognise the module, run quick setup, then calibrate the sensor against buffer or standard solution.

Maintenance

Maintenance is mostly about the sensor: calibrate at intervals suited to the application, clean electrodes and membranes, and review the logbook. For intrinsically safe types, repairs should be left to personnel named by the manufacturer, and modifications invalidate the certificate. Keep the cover closed and glands tight to preserve IP66 protection.

Documents and quotation

The Yokogawa general specification GS 12A01A02-01E for the FLXA21 is available on request and on this page. Send the RFQ with the complete model code, quantity and tag details so the FLXA21 offer is exact. Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer.

Datasheets & Documents
  • PDF Datasheet Yokogawa FLXA21 2-Wire Liquid Analyzer Transmitter Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Yokogawa is the Japanese measurement and industrial automation company behind the DPharp EJA and EJX pressure transmitters, CENTUM distributed control systems and ProSafe-RS safety systems. Tamisuke Yokogawa set it up in 1915 as an electric meter research institute in Shibuya, Tokyo. It began selling electric meters in 1917, was incorporated as Yokogawa Electric Works in 1920, and has been Yokogawa Electric Corporation since 1986. In 1975 it released CENTUM, which it describes as the world's first distributed control system.

Seven Star LLC is an independent supplier in Muscat. We list Yokogawa transmitters, flowmeters, analysers, controllers, recorders and CENTUM and ProSafe spare modules, and supply them to oil and gas, power, water and process plants in Oman and the GCC by model code.

Yokogawa at a glance

Established 1915, Shibuya, Tokyo, by Tamisuke Yokogawa with Ichiro Yokogawa and Shin Aoki
Milestones Electric meters from 1917; incorporated in 1920; CENTUM DCS in 1975; renamed Yokogawa Electric Corporation in 1986
Headquarters Tokyo, Japan
Product families Pressure and temperature transmitters, magnetic, vortex and Coriolis flowmeters, liquid and gas analysers, controllers and recorders, CENTUM VP DCS, ProSafe-RS safety systems and STARDOM controllers

Yokogawa products we list

Family Models Typical use
Pressure and DP transmitters EJA110E, EJA430E, EJA530E, EJX110A, EJX530A, EJX910A multivariable Pressure, level and DP flow measurement in process plants
Temperature transmitters YTA610, YTA710 RTD and thermocouple signals converted to 4-20 mA, HART or fieldbus
Flowmeters ADMAG AXG, ADMAG AXW, ROTAMASS TI, VY vortex, VF701 Water and conductive liquids, mass flow of hydrocarbons, steam and gas
Analysers TDLS8000 laser gas analyser, FLXA402, FLXA21 DO, E7042UD zirconia cell, K9471DS cover Combustion O2, in-situ gas analysis, pH, conductivity and dissolved oxygen
Controllers and recorders UT35A/UT32A, UT32A-D, UT55A/UT52A, UT75A, UP35A/UP32A, UP55A, YS1700, GX10/GX20, GP10/GP20, GM10, µR20000 Single-loop control, furnace programs and data recording
Control and safety systems CP451-10 processor, ADV161, ADV551, node interface unit, ProSafe-RS, ProSafe-RS Lite, ProSafe-SLS, FCN-500 Spares and expansions for CENTUM DCS, safety systems and STARDOM
Battery packs S9129FA, S9185FA, S9548FA Memory backup batteries for DCS and controller modules

The EJA and EJX transmitters use Yokogawa's DPharp sensor: two silicon resonators whose vibration frequency changes with pressure, so the output is a digital frequency rather than an analogue strain signal. The model code describes the instrument completely. In EJA110E, "110" is a differential pressure transmitter, and the letters after the dash set the span, wetted materials, process connection, electrical output and approvals. When you replace a transmitter, sending the full code from the nameplate avoids a mismatch in range or materials.

Where Yokogawa products are used

In Oman and the GCC, Yokogawa transmitters, analysers and control systems are found in oil and gas production, gas processing and LNG, refineries and petrochemical plants, power and desalination stations, and water treatment. Many plants still run CENTUM CS 3000 and CENTUM VP systems that need spare I/O modules and batteries.

Ordering Yokogawa products: what we need

Send the full model and suffix code from the nameplate (for example EJA110E-JMS5J-912NN/KF22) and the style code. For DCS modules, add the module type and system version. For analysers, send the sensor type, measuring range and process conditions.

Facts checked against Yokogawa's corporate history, October 2026.

FAQ

Questions about the Yokogawa FLXA21 2-Wire Liquid Analyzer Transmitter

Answers from our engineering team on installation, compatibility and integration.

01What is the Yokogawa FLXA21?
The FLXA21 is a two-wire, loop-powered liquid analyzer that measures pH/ORP, contacting conductivity, inductive conductivity or dissolved oxygen, with one or two sensor modules.
02What output does the FLXA21 provide?
One 4 to 20 mA DC output with bi-directional HART communication. FOUNDATION Fieldbus and PROFIBUS PA outputs exist under other suffix codes.
03How is the FLXA21 powered?
It is loop powered at a nominal 24 V DC: 16 to 40 V DC for one sensor module and 22.8 to 40 V DC for two sensor modules, as stated in the specification.
04Is the FLXA21 available for hazardous areas?
Yes. The type code selects approvals such as ATEX and IECEx intrinsic safety (-CB), US and Canada (-CD), NEPSI and KCs. Always quote the full model code.
05How do I request the FLXA21 in Oman?
Send an RFQ with the full model and option code and the quantity. Seven Star LLC is an independent supplier and replies with availability for Oman and the GCC.