Brand:

JUMO TDA-300 and TDA-3000 Handheld Thermometer with Data Logger

Description

A handheld thermometer with probes and a data logger for field checks

The JUMO TDA-300 (type 702540) and its sister model TDA-3000 (type 702541) are handheld digital temperature indicators for use with precision thermocouples or RTDs. They measure temperatures on surfaces, in liquids, in melts and in soft materials, and both have a data logging function. The measurement input accepts Pt100, NiCr-Ni K, Fe-Con J and Cu-Con T probes, so one meter serves many tasks, and probes are interchangeable for quick adaptation. The TDA-3000 adds 9999 stored readings, up to 99 measurement points and users, and a USB interface that exports ASCII files in CSV format, which any spreadsheet can read without special software. The plastic housing is shock- and break-proof and resistant to corrosive substances, and the meters, probes and accessories are available from stock.

Technical specifications

Parameter Specification
Type 702540 TDA-300; sister type 702541 TDA-3000
Inputs Pt100 to EN 60751; thermocouples NiCr-Ni K, Fe-Con J and Cu-Con T to EN 60584
Range Pt100 -200 to +850 °C (resolution 1 °C); -199.9 to +850.0 °C (0.1 °C)
Range K -200 to +1372 °C; J -200 to +1200 °C; T -50 to +400 °C
Display accuracy ±(0.1 % + 1 digit) or ±0.3 °C, whichever is larger, at 23 °C
Cold junction ±0.5 °C at 5 to 40 °C; ±1 °C at -20 to +5 °C and +40 to +50 °C
Sampling rate 0.5 s; digital filter 0 to 100 s; offset ±99.9 °C; unit °C or °F
Probe monitoring Symbol for short circuit, break or wrong connection
Data logger TDA-300 99 values in SRAM; TDA-3000 9999 values in EEPROM
Recording Manual or automatic, 1 to 3600 s interval; data include temperature, TAG, user, limits, date and time
Measurement points TDA-300 5 TAG numbers, 1 user; TDA-3000 99 TAG numbers and 99 users, 11 characters each
Limit monitoring Upper and lower limit per measurement point; peak and bottom value storage
Interface TDA-3000: USB 2.0 Mini-B, cable included, supplied by the PC
Supply One AA alkaline cell (LR6); about 400 h continuous use; 10 mW average
Enclosure TDA-300 IP67; TDA-3000 IP54
Operating temperature -20 to +50 °C; error 0.01 % of span between 5 and 40 °C
Housing 57 x 152 x 46 mm; about 150 g with battery
Display FSTN LCD, 4-digit temperature, 11-character message window, 3-stage battery indicator

Probes and accessories

The datasheet shows a range of probes. Pt100 probes include an immersion probe 702546/01-100 for liquids up to 250 °C and an insertion probe 702546/02-100 for food and soft materials. Type K thermocouple probes include a flexible mineral-insulated immersion probe for up to 1150 °C, an insertion probe for soft materials, and surface probes such as 702545/03-004 for small uneven objects up to 400 °C and 702545/03-015 for flat surfaces up to 500 °C, whose transversely linked spring bands make the reading largely independent of contact pressure and angle. Probe cables are about 1500 mm long and the insertion probes are protected to IP67. Adapters 702546/04-000 for Pt100 and 702545/04-000 for type K, each 1700 mm long and usable up to 100 °C ambient, connect existing probes with Mini-Flach connectors. The Pt100 probes are class A and the thermocouple probes class 2. A DKD calibration certificate is available with the instrument, which matters for audit trails in food and process plants.

Order the meter as 702540/88-000 or 702541/88-000 and the probes by code, for example 702545/01-200 for a 200 mm immersion thermocouple. Standard delivery includes the manual, a battery, a carrying hook and, for the TDA-3000, a USB cable. A 20 ml tube of silicone heat transfer compound (94091460, to 200 °C) and a carrying case for the meter and two probes (00453912) are accessories.

Choosing the model and using it

Choose the TDA-300 when a rugged IP67 meter with 99 stored readings is enough for spot checks, and the TDA-3000 when records must be downloaded to a PC. Both log either on demand or automatically, and the logging symbol lights while recording. If the battery is flat, the TDA-300 loses its SRAM data, but the TDA-3000 EEPROM keeps data for about 10 years and can be rewritten about 100,000 times. Typical jobs are surface checks on pipes and motors, core temperature checks on food, melt checks, and commissioning of ovens and baths. Keep the meter inside the stated -20 to +50 °C operating range, because the temperature error rises from 0.01 % to 0.02 % of span outside 5 to 40 °C.

See portable test and measurement for related instruments.

Using the TDA-300 and TDA-3000 in Oman and the GCC

Handheld thermometers are used in the region by maintenance teams, quality inspectors and commissioning engineers. Typical tasks include checking surface temperatures on motors, bearings and pipes, verifying process temperatures against installed sensors, checking food temperatures in cold chain and catering, and measuring temperatures in molten materials or ovens with suitable probes.

Choose the probe type and range to suit each task: surface probes for pipes and motors, immersion probes for liquids, and penetration probes for soft materials. Thermocouple probes cover high temperatures, while Pt100 probes give better accuracy in the lower range. The data logging function of the TDA-3000 helps when readings at many measurement points must be recorded and reported.

For quality and food safety audits, a calibration certificate for the instrument and probes is often required, so state this in the enquiry. Spare probes are also worth ordering at the same time, because probes are the parts most likely to be damaged in daily use.

Information to send with an enquiry

  • Model: TDA-300 or TDA-3000
  • Probe types and temperature ranges
  • Calibration certificate requirements
  • Quantity

Related JUMO products: JUMO P700 Series Precision Temperature Indicating Device and JUMO 202710 Handheld Meters for pH, Redox and Conductivity.

Documents and quotation

The hosted document is JUMO data sheet 702540 (6 pages, English). It covers both the TDA-300 and the TDA-3000 and lists the probe types. Confirm model, probe type and calibration certificate before ordering. Seven Star LLC is an independent supplier in Muscat; send the temperature range and probe needs for a quotation. To help us prepare an accurate quotation, please include the temperature range, probe types, logging needs and whether a calibration certificate is required, together with the quantity needed and the delivery location. We will then compare these details with the order information in the datasheet.

Datasheets & Documents
  • PDF Datasheet JUMO 702540 data sheet 702540 (EN) Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

JUMO is the German maker of temperature sensors, process controllers, pressure transmitters and liquid analysis instruments. Moritz Kurt Juchheim founded it in Fulda in 1948 as M. K. Juchheim KG, making glass thermometers with six employees. The name JUMO comes from the founder's name. It is still a family company headquartered in Fulda, and the parent company has been called JUMO GmbH & Co. KG since 2003.

Seven Star LLC is an independent supplier in Muscat. We list JUMO RTD and thermocouple probes, thermowells, transmitters, controllers, recorders and water analysis instruments, and supply them to process plants, OEMs, HVAC contractors and water utilities in Oman and the GCC by type number.

JUMO company profile

Established 1948, Fulda, Germany, by Moritz Kurt Juchheim
First products Glass and contact thermometers; platinum glass sensors from 1966
Ownership Family-owned; renamed JUMO GmbH & Co. KG in 2003
Product families RTD and thermocouple probes, thermowells, temperature and pressure transmitters, controllers and thermostats, safety temperature limiters, paperless recorders, liquid analysis and flow and level instruments

Milestones

  • 1948: Moritz Kurt Juchheim founds M. K. Juchheim KG in Fulda, making glass thermometers with six employees.
  • 1966: Platinum glass temperature sensors are added.
  • 2003: The parent company is renamed JUMO GmbH & Co. KG.

JUMO products we list

Family Models Typical use
RTD and thermocouple probes 902020, 902030, 902050, 902120, 902006 2xPt100, PROCESStemp, mineral-insulated RTD, 901020 thermocouple, 901210, 902810 hygienic, 902550 surface, 902520 HVAC, MarineTemp Temperature measurement in pipes, vessels, machines, HVAC and food plants
Thermowells and inserts 909710 thermowells, 909740 measuring inserts Protecting sensors and replacing them without opening the process
Temperature transmitters dTRANS T01, T03, T04, T05, T06 SIL, T07 HART, T09, T1000 IO-Link, DELOS T switch Converting sensor signals to 4-20 mA, HART or IO-Link
Pressure and level dTRANS p20, p30, p31 hygienic, p33 Ex, p02 DELTA, MIDAS S05, MIDAS DP10, DELOS S02, MAERA S25, NESOS R20 Pressure in process, hydraulic and OEM systems; tank and well level
Controllers, limiters and recorders DICON touch, IMAGO 500, dTRON 316, iTRON, cTRON, diraTRON 104, eTRON T, safetyM STB/STW, safetyM TB/TW, LOGOSCREEN 601, mTRON T Furnace, oven and process control, over-temperature protection and recording
Liquid analysis and flow AQUIS touch S, AQUIS touch P, AQUIS 500 pH, AQUIS 500 CR, dTRANS pH 02, pH electrode, digiLine DO, ecoLine NTU, flowTRANS MAG H20, flowTRANS US W01 pH, conductivity, oxygen and turbidity in water treatment; liquid flow

Choosing a temperature sensor starts with the range and accuracy. A Pt100 RTD is accurate and stable from about -200 to 600 °C and is the usual choice in process plants. A thermocouple (type K or J, for example) covers much higher temperatures and responds faster but is less accurate. The wiring also matters: a 3-wire or 4-wire Pt100 connection cancels the resistance of the cable, while a 2-wire connection adds an error that grows with cable length. That is why a head-mounted transmitter is often fitted close to the sensor.

Where JUMO products are used

In Oman and the GCC, JUMO sensors and controllers are used in oil and gas and chemical plants, food and beverage production, plastics and metal heat treatment, HVAC and district cooling, water and wastewater treatment, swimming pools and marine applications.

Ordering JUMO products: what we need

Send the JUMO type number and the full order code from the label (for example 902020/10-...). For probes, add the sensor type (Pt100, Pt1000, type K or J), the number of wires, the insertion length and diameter, and the process connection. For controllers, add the input, outputs and supply voltage.

Facts checked against JUMO's company history, October 2026.

FAQ

Questions about the JUMO TDA-300 and TDA-3000 Handheld Thermometer with

Answers from our engineering team on installation, compatibility and integration.

01What is the difference between TDA-300 and TDA-3000?
The TDA-300 stores 99 readings in IP67 housing, and the TDA-3000 stores 9999 readings with USB export in an IP54 housing.
02Which probes can be connected?
Pt100 RTDs and K, J and T thermocouples.
03What is the display accuracy?
±(0.1 % + 1 digit) or ±0.3 °C, whichever is larger, at 23 °C ambient.
04What battery does it use?
One AA alkaline cell, giving about 400 hours of continuous operation.
05How are logged data exported?
The TDA-3000 sends CSV files over USB, which open in spreadsheet programs.