The JUMO dTRANS p20 DELTA (type 403022) measures differential pressure of gases, steam and liquids with a silicon sensor behind stainless steel separating diaphragms. It offers a 4 to 20 mA two-wire output, optionally with HART 7, an integrated two-line LCD with bar graph, and a rotary knob for on-site operation. The datasheet states linearity down to 0.07 % and a turndown of up to 400:1 on the -1 to +1 bar range. A version with ATEX Ex ia protection allows installation up to zone 0, and a SIL version evaluated by exida according to IEC 61508 is suitable for monitoring process level and pressure up to SIL2. Diaphragm seals can be fitted for hygienic or aggressive processes.

Technical specifications

Parameter Specification
Nominal ranges (differential) -10 to +10 mbar (not SIL); -1 to +1 bar; 0 to 1 bar; -1 to +6 bar; -1 to +100 bar
Nominal pressure PN 2 (10 mbar); PN 210 (1 to 6 bar ranges); PN 210, optionally PN 420 (100 bar range, extra code 694)
Smallest span 1 mbar and 5 mbar (small ranges); 0.350 bar; 2.5 bar
Turndown Up to r = 20, 400, 200, 20 and 40 respectively
Non-linearity (reference) 0.1 % (10 mbar); 0.07 % (1 bar ranges); 0.07 % (6 bar, r up to 5); r x 0.007 to 0.05 % at higher turndown
Accuracy at 20 degC 0.2 % (10 mbar) or 0.1 % of set span at low turndown; r x 0.01 % to r x 0.1 % at higher turndown
Long-term stability 0.6 % per year (10 mbar); 0.1 % per year; 0.2 % per year
Static pressure influence From 1 % (10 mbar) to P x 0.0003 % per bar of static pressure, by range
Output 4 to 20 mA two-wire (405); with HART version 7 (410); damping 0 to 100 s; T60 190 ms or less
Burden (UB – 12 V) divided by 0.022 A; HART additionally 250 to 1100 ohm
Supply DC 12 to 36 V; Ex ia version DC 12 to 28 V (Ui 28 V, Ii 115 mA, Pi 750 mW, Ci 6 nF, Li 105 uH)
Display LCD, 2 lines plus bar graph; units include bar, mbar, psi, kPa, MPa, mmH2O, inH2O and more; min/max pressure, sensor temperature and operating hours
Operation / interfaces Rotary knob (turn and press); JUMO interface (not in hazardous area); HART interface
Medium temperature Up to 110 degC standard; up to 100 degC for Ex versions
Ambient temperature -40 to +85 degC standard (-50 degC with extra code 681); Ex versions -40 to +60 degC; limited function below -20 degC
Protection IP66/67 (no explosion protection) or IP66 (ATEX Ex ia)
Materials Membrane 316L, Hastelloy C-276 or tantalum; flange 316; seal PTFE; housing 1.4404 or 1.4408 precision cast
Weight About 3.0 kg (short housing), 3.3 kg (long), 4.0 kg (precision casting); +3.8 kg with extra code 694
Approvals as printed ATEX SEV 09 ATEX 0138 X (II 1G Ex ia IIC T4 Ga; II 1D Ex ia IIIC T200 125 degC Da); UKEX EPS 22 UKEX 1 369 X with extra 085; SIL exida JUMO 2203088 C001 for basic extension 2
Stock versions 403022/0-1-1-36-20-1-1-530-410-511-20-01/000 part no. 00569267; 403022/0-1-1-36-20-1-1-532-410-511-20-01/000 part no. 00573301

Housings, options and ordering restrictions

Three housings are offered: a short stainless steel case with M12 plug connection (type 1), a long stainless steel case with cable fitting (type 2) and a precision-cast case with an M20 cable fitting (type 3). Process connections are two 1/4-18 NPT connections to DIN EN 837 or a screwed diaphragm-seal version, with wetted material in stainless steel, tantalum or NiMo and either silicone oil or halogenated oil for oxygen service as the filling medium. Extra codes add UKEX (085), customer-specific factory settings (100), oil and grease free (624), mounting brackets for 2 inch pipe (633), a TAG number (634), a NACE declaration (635), expanded ambient temperature (681) and PN 420 (694).

The datasheet lists limits: the SIL version needs output 410 and the display, and is not available with the 10 mbar range, a diaphragm-seal connection or extra code 681. The Ex ia version needs output 410 and is not available with the plastic cable fitting or plastic cover. The 10 mbar and 1 bar bipolar ranges are only available with stainless steel process connection material.

Setup, accessories and companions

Besides the knob, the device is set up with the JUMO dTRANS p20 setup program (00537577) over a USB/TTL interface cable (00456352), with a HART modem (00443447), or with the DTM from data sheet 403025 (00738288). Other accessories include M12 line sockets of 2 m (00404585 straight, 00409334 angled), a mounting bracket (00543777), an Ex-i repeater power supply (00577948) and on-site calibration including DAkkS (00758363). Valve blocks and diaphragm seals are described in data sheets 409706 and 409772 to 409786.

Display scaling, a linear or square-root characteristic for flow measurement and a current-generator function support commissioning. For related items see the published dTRANS p30 for gauge pressure and the pressure instrumentation category. State the differential range, static pressure, medium and temperature, output, hazardous-area and SIL requirements, housing and diaphragm-seal details when you enquire.

Related JUMO products: JUMO dTRANS p02 DELTA Differential Pressure Transmitter, JUMO dTRANS p20 Process Pressure Transmitter with HART, JUMO SIRAS P21 DP SIL Differential Pressure Transmitter, JUMO MIDAS DP10 Differential Pressure Transmitter, JUMO dTRANS p33 Ex-Area Pressure Transmitter and JUMO dTRANS p20 Ex d Flameproof Pressure Transmitter.

Documents and quotation

Hosted for reference is the official JUMO data sheet 403022, version 7.00 dated 2025-09-08, fifteen pages in English, with accuracy tables, approvals, dimensions, wiring and the complete order code. Seven Star LLC is an independent supplier in Muscat; send the order code and approval needs and we will prepare a quotation.

Datasheets & Documents
  • PDF Datasheet JUMO dTRANS p20 DELTA data sheet 403022 V7.00 (EN, 15 pages) Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

JUMO is the German maker of temperature sensors, process controllers, pressure transmitters and liquid analysis instruments. Moritz Kurt Juchheim founded it in Fulda in 1948 as M. K. Juchheim KG, making glass thermometers with six employees. The name JUMO comes from the founder's name. It is still a family company headquartered in Fulda, and the parent company has been called JUMO GmbH & Co. KG since 2003.

Seven Star LLC is an independent supplier in Muscat. We list JUMO RTD and thermocouple probes, thermowells, transmitters, controllers, recorders and water analysis instruments, and supply them to process plants, OEMs, HVAC contractors and water utilities in Oman and the GCC by type number.

JUMO company profile

Established 1948, Fulda, Germany, by Moritz Kurt Juchheim
First products Glass and contact thermometers; platinum glass sensors from 1966
Ownership Family-owned; renamed JUMO GmbH & Co. KG in 2003
Product families RTD and thermocouple probes, thermowells, temperature and pressure transmitters, controllers and thermostats, safety temperature limiters, paperless recorders, liquid analysis and flow and level instruments

Milestones

  • 1948: Moritz Kurt Juchheim founds M. K. Juchheim KG in Fulda, making glass thermometers with six employees.
  • 1966: Platinum glass temperature sensors are added.
  • 2003: The parent company is renamed JUMO GmbH & Co. KG.

JUMO products we list

Family Models Typical use
RTD and thermocouple probes 902020, 902030, 902050, 902120, 902006 2xPt100, PROCESStemp, mineral-insulated RTD, 901020 thermocouple, 901210, 902810 hygienic, 902550 surface, 902520 HVAC, MarineTemp Temperature measurement in pipes, vessels, machines, HVAC and food plants
Thermowells and inserts 909710 thermowells, 909740 measuring inserts Protecting sensors and replacing them without opening the process
Temperature transmitters dTRANS T01, T03, T04, T05, T06 SIL, T07 HART, T09, T1000 IO-Link, DELOS T switch Converting sensor signals to 4-20 mA, HART or IO-Link
Pressure and level dTRANS p20, p30, p31 hygienic, p33 Ex, p02 DELTA, MIDAS S05, MIDAS DP10, DELOS S02, MAERA S25, NESOS R20 Pressure in process, hydraulic and OEM systems; tank and well level
Controllers, limiters and recorders DICON touch, IMAGO 500, dTRON 316, iTRON, cTRON, diraTRON 104, eTRON T, safetyM STB/STW, safetyM TB/TW, LOGOSCREEN 601, mTRON T Furnace, oven and process control, over-temperature protection and recording
Liquid analysis and flow AQUIS touch S, AQUIS touch P, AQUIS 500 pH, AQUIS 500 CR, dTRANS pH 02, pH electrode, digiLine DO, ecoLine NTU, flowTRANS MAG H20, flowTRANS US W01 pH, conductivity, oxygen and turbidity in water treatment; liquid flow

Choosing a temperature sensor starts with the range and accuracy. A Pt100 RTD is accurate and stable from about -200 to 600 °C and is the usual choice in process plants. A thermocouple (type K or J, for example) covers much higher temperatures and responds faster but is less accurate. The wiring also matters: a 3-wire or 4-wire Pt100 connection cancels the resistance of the cable, while a 2-wire connection adds an error that grows with cable length. That is why a head-mounted transmitter is often fitted close to the sensor.

Where JUMO products are used

In Oman and the GCC, JUMO sensors and controllers are used in oil and gas and chemical plants, food and beverage production, plastics and metal heat treatment, HVAC and district cooling, water and wastewater treatment, swimming pools and marine applications.

Ordering JUMO products: what we need

Send the JUMO type number and the full order code from the label (for example 902020/10-...). For probes, add the sensor type (Pt100, Pt1000, type K or J), the number of wires, the insertion length and diameter, and the process connection. For controllers, add the input, outputs and supply voltage.

Facts checked against JUMO's company history, October 2026.

Frequently Asked Questions

Which output does it provide?
4 to 20 mA two-wire (code 405) or 4 to 20 mA with HART version 7 (code 410).
Is there a SIL version?
Yes. Basic type extension 2 is certified by exida to IEC 61508 for monitoring process level and pressure up to SIL2, with output 410 and the display.
What turndown is stated?
Up to r = 400 on the -1 to +1 bar range and r = 200 on the 0 to 1 bar range, with r = 20 to 40 on the other ranges.
Can it go into hazardous areas?
The Ex ia version is marked II 1G Ex ia IIC T4 Ga and II 1D Ex ia IIIC T200 125 degC Da, allowing installation up to zone 0.
Which diaphragm seals can be fitted?
Seals described in data sheets 409772 to 409786 can be connected for specific process applications.