Honeywell 2106B1400 Sensepoint H2S Sensor 0-20 ppm M20
Description
Honeywell 2106B1400 is the Sensepoint hydrogen sulphide sensor built for the 0 to 20 ppm span and fitted with an M20 mounting thread. It belongs to the Sensepoint toxic family, a compact two-wire gas detector head that is installed through a suitable Ex d or Ex e junction box and wired straight back to a controller, a DCS or a PLC. Because the 2106B1400 is the lowest-range member of the H2S group, it is the sensor people ask for when personnel exposure limits, not process leak detection, drive the alarm philosophy.
What the 2106B1400 is
The 2106B1400 is a Surecell electrochemical H2S detector with a loop-powered 4 to 20 mA output. The sensing head is moulded in polyphenylene sulphide (PPS), weighs 185 g, and is rated IP65 as supplied and IP67 once the weather protection is fitted. The Sensepoint design keeps the intrinsically safe sensing side separate from the flameproof termination, so the cell can be swapped while the loop is powered, and the sensor does not need a sinter in front of the cell. Honeywell notes that this is what gives it quick access for sticky gases and a fast response.
Technical specifications for 2106B1400
| Parameter | Value |
|---|---|
| Target gas and span | Hydrogen sulphide, 0 to 20 ppm |
| Mounting thread | M20 |
| Signal output | Loop powered 4 to 20 mA |
| Supply | 18 to 30 VDC, 0.9 W maximum |
| Operating temperature | -25 to +50 degrees C |
| Humidity | Continuous 20 to 90 percent RH, intermittent 10 to 99 percent RH, non-condensing |
| Pressure | 90 to 110 kPa |
| Warm-up time | Less than 3 minutes |
| Response time | T20 less than 6 s, T50 less than 15 s, T90 less than 40 s |
| Linearity | Less than 5 percent of full scale deflection |
| Typical operating life | 2 years |
| Zero stability | Less than +/-2 percent fsd per year; less than +/-5 percent fsd over temperature |
| Span stability | Less than +/-2 percent fsd per year; less than +/-25 percent over temperature |
| Weight | 185 g |
| Housing material | Polyphenylene sulphide (PPS) |
| Ingress protection | IP65, IP67 with weather protection |
How 2106B1400 differs from other Sensepoint H2S variants
The H2S group has nine orderable heads. Span is the first difference: 2106B1400 covers 0 to 20 ppm, while 2106B1401 covers 0 to 50 ppm and 2106B1402 covers 0 to 100 ppm, all with the M20 thread. The same three spans are offered with an M25 thread as 2106B1500, 2106B1501 and 2106B1502, and with a three-quarter inch NPT thread as 2106B1700, 2106B1701 and 2106B1702. When this head is paired with an XCD RTD transmitter, the 20 ppm span carries a default low alarm of 4.0 ppm, a default high alarm of 8.0 ppm and a lowest selectable alarm level of 2.0 ppm, whereas the 50 ppm span defaults to 10 and 20 ppm.
Where 2106B1400 is used in the Gulf
A 20 ppm span suits closed or semi-closed rooms where H2S is a toxic exposure hazard rather than a flammable one: sour-water and amine package skids, produced-water and effluent handling, wastewater pumping stations and wet wells, sulphur-recovery sampling rooms and laboratory fume areas. Utilities in Oman and the wider GCC use it in sewage lift stations and desalination pre-treatment buildings, where low-ppm detection protects maintenance crews entering a space. Choose a higher span only if the process can produce sustained leaks in the tens of ppm.
Selection and compatibility notes
- Order the M20 head only if the junction box or enclosure has an M20 entry; M25 and NPT versions are separate part numbers.
- The head outputs a two-wire loop, so the controller input must be a 4 to 20 mA current-sink compatible input with a suitable supply.
- Honeywell lists the Sensepoint toxic heads for use with the XCD RTD transmitter, which accepts them directly or on a remote cable; confirm the match before ordering.
- Ex certification for the toxic heads is II 2 GD Ex d ia IIC T4 Gb for ambient -40 to +65 degrees C with Ex tb IIIC IP67 T135 degrees C Db, under Baseefa08ATEX0263X and IECEx BAS08.0070X.
- Operating temperature for the H2S cell is -25 to +50 degrees C, narrower than the housing certification, so cold-store and hot-roof installations need a check.
Installing the 2106B1400 head
Follow the Honeywell manual for head orientation. Terminate in the junction box: the red wire is the positive supply, the blue wire is the negative, and the cable screen is landed at one end only. Honeywell supplies a standard junction box with an earth continuity plate for added RFI protection (part 00780-A-0100, a Bartec box with one 25 mm and three 20 mm entries), and it lists a nylon 20 mm cable gland and cap set under 225-262-138. Fit the weatherproof cap outdoors to reach IP67.
Commissioning and routine care
Allow the three-minute warm-up before judging the reading, then apply span gas with a toxic calibration kit. The calibration kit in the Sensepoint range, 2106B2114, includes a calibration cover, a removal tool and a potentiometer adjustment tool. Because the typical cell life is two years and span stability is quoted as less than plus or minus 2 percent of full scale per year, plan a calibration check at intervals set by your site procedure and a cell exchange when response degrades. The H2S replacement is sold as exchange kit 2106B1549.
Related items for 2106B1400
Commonly ordered alongside the head are the junction box 00780-A-0100, the weather protection cap, the flow adaptor for calibration, the duct mounting kit and the collecting cone, all named in the Sensepoint accessory list. An M36/M40 thread adaptor is also available under 2106D2055 where an enclosure has a larger entry. For a head that measures a different toxic gas, see the Sensepoint CO, NH3 and Cl2 listings on this site.
Frequently asked points about 2106B1400
Buyers most often ask about span, thread and wiring. The span is 0 to 20 ppm H2S, the thread is M20, and the wiring is a two-wire 4 to 20 mA loop supplied at 18 to 30 VDC. Several also ask about life: Honeywell gives a typical operating life of two years for the H2S cell and six months of storage life, so ordering a spare far in advance of installation is not advisable.
Documents and quotation
Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer. Request the Sensepoint datasheet with your quotation, and send an RFQ naming part number 2106B1400, the quantity and the delivery location. Tell us the transmitter or controller it will connect to and the thread of your junction box so that we can confirm that the M20 version is the correct choice.
Datasheets & Documents
- Datasheet Sensepoint Gas Detector Datasheet Download
Reviews (0)
How would you rate this product? Cancel reply
About brand
Honeywell is the American industrial technology group whose products on this site cover gas and flame detection, pressure transmitters, flow measurement, gas metering, fire alarm systems and helicopter rotor balancing equipment. Its roots go back to a thermostat business in Minneapolis in 1885 and to Mark Honeywell's heating company, which merged with it in 1927. Several brands we list have joined Honeywell over the years: Gent fire systems of Leicester (part of Honeywell since 2005), Chadwick-Helmuth (a Honeywell business unit from 2002) and Elster gas metering (acquired around the end of 2015).
Seven Star LLC is an independent supplier in Muscat. We list Honeywell gas detectors and sensors, transmitters, Elster gas meters and flow computers, Gent fire alarm equipment and Chadwick-Helmuth balancers, and supply them and their spares in Oman and the GCC by part number.
Honeywell company profile
| Origins | 1885, Minneapolis (Butz thermostat business); merged with Mark Honeywell's company in 1927 |
|---|---|
| Brands in our range | Honeywell Analytics gas detection, SmartLine transmitters, Elster gas metering, Gent fire detection, Chadwick-Helmuth rotor track and balance |
| Gent | Founded in Leicester, UK, around 1872 by John Thomas Gent; part of Honeywell since 2005 |
| Elster | Gas, water and electricity metering; bought by Honeywell from Melrose, deal completed around the end of 2015 |
Milestones
- 1885: The Butz thermostat business starts in Minneapolis.
- 1927: It merges with Mark Honeywell's company.
- 2005: Gent fire detection (Leicester, UK) becomes part of Honeywell.
- 2015: Honeywell completes the purchase of Elster from Melrose around the end of the year.
Honeywell products we list
| Family | Models | Typical use |
|---|---|---|
| Fixed gas detectors | Searchline Excel, Searchline Excel Plus, 02104-N-XSSA, Sensepoint XCD RTD, Sensepoint XCD IR, XCD H2, Sensepoint XCL, Sensepoint XRL, OmniPoint, RAEGuard 2 PID | Flammable, toxic and VOC gas detection in process areas |
| Sensors, flame detection and control | SR-W-MP75C LEL sensor, SR-X2V O2 sensor, FS24X Plus flame detector, R4343E flame switch, System 57 card, 5704FS panel | Replacement sensors, fire detection and gas detection control panels |
| Transmitters and flow | SmartLine STG700, STG800, STR82D remote seal, VersaFlow Coriolis | Process pressure and mass flow measurement |
| Elster gas metering | Q.Sonic plus, TRZ2, Quantometer, RABO, IRM-3 DUO, EK280, EC 350, enCore FC1, enCore ZM1, gas-net F1, gas-net D2, GasLab Q2, USM display | Gas meters, volume correction and fiscal flow computing |
| Gent fire alarm systems | Vigilon Plus, Vigilon Compact Plus, Nano, Xenex, S-Quad detector, S-Cubed, ALD302 base, S4-34420 interfaces, loop card, repeat panel, mimic panel, XENS-850 call point, 74430 bell | Addressable and conventional fire alarm systems in buildings and plants |
| Chadwick-Helmuth rotor track and balance | 8500C, 8500C Plus, Vibrex 2000, Model 11, photocell system, printer, bracket, protractor | Helicopter rotor track and balance and propeller balancing |
Gas detection uses two basic approaches. Point detectors such as the Sensepoint XCD measure the gas at one location, with a catalytic or infrared sensor for flammable gas and an electrochemical cell for toxic gas such as H2S. Open path detectors such as the Searchline Excel send an infrared beam across an area up to a few hundred metres long and detect a gas cloud anywhere along it. Infrared sensors cannot be poisoned the way catalytic beads can, which matters on sites with silicones or H2S.
Where Honeywell products are used
In Oman and the GCC, Honeywell gas detectors protect wellheads, gas plants, refineries and tank farms. Elster meters and flow computers measure gas at city gate stations and in industry. Gent systems are installed in commercial buildings, hospitals and hotels, and Chadwick-Helmuth balancers are used by helicopter operators.
Ordering Honeywell products: what we need
Send the part number from the label (for example 02104-N-XSSA or SPXCDASMTXF). For gas detectors, add the gas, range, output and certification. For gas meters, add size (DN), flow range and pressure rating. For Gent equipment, add the panel type and loop protocol.
Facts checked against Honeywell press releases and SEC filings, and published histories of Gent and Chadwick-Helmuth, October 2026.
Browse Honeywell products by category: Fire & Gas Detector, Vibration, Flowmeters, Pressure Instrumentation, Temperature Instrumentation.
FAQ
Questions about the Honeywell 2106B1400 Sensepoint H2S Sensor 0-20 ppm
Answers from our engineering team on installation, compatibility and integration.
01What gas and range does Honeywell 2106B1400 detect?
The 2106B1400 is a Sensepoint electrochemical sensor for hydrogen sulphide with a 0 to 20 ppm measuring range and an M20 mounting thread.
02What is the difference between 2106B1400 and 2106B1401?
The 2106B1400 measures 0 to 20 ppm, while the 2106B1401 measures 0 to 50 ppm. Both use the M20 thread and the same loop-powered 4 to 20 mA output.
03How is the 2106B1400 powered and wired?
It is a two-wire loop-powered detector fed from 18 to 30 VDC at 0.9 W maximum. The red wire is positive, the blue wire is negative, and the screen is earthed at one end.
04How long does the 2106B1400 cell last?
Honeywell quotes a typical operating life of two years for the H2S cell in the Sensepoint range, with a storage life of six months.
05What is the response time of the 2106B1400?
The datasheet lists T20 below 6 seconds, T50 below 15 seconds and T90 below 40 seconds for the Sensepoint H2S sensor, after a warm-up under 3 minutes.