Brand:

Rosemount 02120-7000-0003 2120 NAMUR Cassette

Description

Overview of 02120-7000-0003: the NAMUR cassette for the 2120

Where a control system already owns an intrinsically safe NAMUR isolating amplifier, the 2120 level switch is ordered with electronic type K, and the module that supplies that behaviour is Rosemount 02120-7000-0003. The manufacturer gives it a light blue label. It is a plug-in replacement cassette for 2120 units shipped since June 2013, so a failed or damaged module does not mean retiring a fork that is still sound in a vessel in the hazardous area.

NAMUR output means the switch does not drive a load; it presents a defined current signature that a separate isolating amplifier reads, which is why this cassette suits zone 0, 1 and 2 style duty where energy must be limited.

Signal and wiring facts from the datasheet

Parameter Value
Part number 02120-7000-0003
Description Replacement cassette: NAMUR
Label colour Light blue
Electronic type code K
Output type NAMUR, two terminals marked minus and plus
Interface device Certified intrinsically safe isolating amplifier to IEC 60947-5-6
Certification choices All except the Korean IP code (the K row reads ‘All except IP’)
Surge protection Yes, to IEC 61326 (all electronics)
Safety options QS and QT are available (not blocked for K)
Switching delay 0.3, 1, 3, 10 or 30 s selectable
Mode Dry = on or Wet = on
Process temperature range of the 2120 minus 40 to 302 deg F
Housing choices Nylon A or D, aluminium X or Y, stainless S or T

How 02120-7000-0003 differs from the other 2120 cassettes

Compared with the red two-wire module 02120-7000-0001, 02120-7000-0003 cannot switch a load by itself, but it is the only module besides the 8/16 mA type that can carry the whole family of intrinsically safe approvals: I1 ATEX, I2 Brazil, I3 China (K only), I4 Japan, I5 USA, I6 Canada, I7 IECEx, IM EAC and IW India. The 8/16 mA cassette 02120-7000-0005 adds a second current level for line fault detection; the NAMUR module offers the simpler classic signature. In relation to 02120-7000-0002 (PNP, 20 to 60 Vdc, yellow) the NAMUR module is the choice when no field supply is available at the tank.

How the NAMUR level switch behaves in service

The heartbeat LED, visible through the lens of non-metal housings, flashes when the output is off and glows steadily when on; different flash rates are used to flag a product malfunction. The fork responds to the change in oscillation frequency between air and liquid, with a switching point 0.5 in (13 mm) from the tip in vertical mounting and 0.1 in (2.5 mm) hysteresis in water, so the signature changes at a well-defined height. The fork material depends on the build: 316/316L stainless steel (dual certified 1.4401/1.4404) for the D code, Alloy C-276 for the C code, or ECTFE-coated steel on flanged versions. Liquids down to a density of 600 kg/m3 and up to 10000 cP viscosity are within the specified range, and the fork shape is a fast-drip design so that viscous product clears quickly from the tines and the output returns to dry without a long lag. None of these fork data change when the NAMUR cassette is exchanged, which is why this part can be swapped without recalibration: the 2120 needs no calibration.

Hazardous-area guidance

  • Replace an IS cassette only with an IS cassette of the same type; the datasheet is explicit about this rule.
  • Match the cassette type to the amplifier channel, because a NAMUR input and an 8/16 mA input are set up for different signatures.
  • Keep the amplifier outside the hazardous area or ensure its own certificate allows zone installation.
  • Keep the control drawing for the installed certification code (for example I5 or I1) with the loop file.

When the plant carries a Gulf refinery loop-check dossier, record the new cassette part number there so auditors can see the module was changed on a certified basis.

Applications across Oman and the GCC

NAMUR 2120 switches are common as independent high-level alarms on crude, condensate and chemical tanks, where a separate overfill layer must not share the primary gauge. Because the cassette is certified at module level, tank-farm engineers keep the same fork in service for years and rotate the electronics. In desalination chemical dosing rooms the same part gives a low-level signal that stops a dosing pump. Further typical homes include lube-oil reservoirs on compressor skids, slop tanks at marine terminals, chilled-water expansion vessels, and sump pits in power stations, anywhere a hazardous-area point level signal is wanted from a robust fork that is not upset by foam, bubbles, vibration or coating. The datasheet notes that the switch is virtually unaffected by flow, bubbles, turbulence, foam, vibration, solids content and liquid properties, which explains its reputation as a fit-and-forget instrument for dirty duties.

Fitting 02120-7000-0003 and checking the loop

De-energise the amplifier channel, open the cover (allow 1.2 in or 30 mm clearance), lift out the old cassette and fit 02120-7000-0003. Observe the polarity marked on the module, because this electronics type is not on the datasheet list of polarity-insensitive types. Restore the loop, check the LED heartbeat, then hold a magnet at the side test point; the amplifier channel should toggle. Record the dry and wet signature currents seen by the amplifier for the maintenance file.

Maintenance and shelf stock

Fork surfaces are smooth, with no moving parts, so cleaning with plant-approved solvents is enough when a coating is a concern; avoid bridging between the two tines with solids larger than 0.2 in (5 mm). Because the cassette is a sealed module, keep spares in their pack and away from heat; after a lightning event, replace rather than repair.

Documents and quotation

The datasheet containing the 02120-7000-0003 entry is attached for reference. Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer. Please send your RFQ quoting part number 02120-7000-0003, the required quantity and the 2120 nameplate model code, together with the delivery location in Oman or the GCC, so that we can confirm availability.

Other Models in the Rosemount 2120 Spares Range

Seven Star LLC supplies models from the Rosemount 2120 Spares range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.

ModelCategory
Rosemount 02120-7000-0007 2120 Relay 9-30 Vdc CassetteLevel SwitchesView details
Rosemount 02120-7000-0001 2120 Direct Load CassetteLevel SwitchesView details
Rosemount 02120-7000-0005 2120 8/16 mA CassetteLevel SwitchesView details
Rosemount 02100-1060-0001 2120 Quick Release KitLevel SwitchesView details
Rosemount 02120-7000-0003 2120 NAMUR CassetteLevel SwitchesYou are viewing this model
Rosemount 02100-1010-0001 Adaptor Boss 1-in. BSPP Tri ClampLevel SwitchesView details
Datasheets & Documents
  • PDF Datasheet Rosemount 02120-7000-0003 2120 NAMUR Cassette Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Rosemount is Emerson's brand for pressure, temperature, level and flow measurement, best known for the 3051 pressure transmitter. Frank Werner, Robert Keppel and Vernon Heath founded it in Minnesota in 1956 as Rosemount Engineering Company, first making temperature sensors for aerospace. It moved into the process industries, took the name Rosemount Inc. in 1966, and has been part of Emerson since 1976.
Seven Star LLC is an independent supplier in Muscat. We list Rosemount transmitters, level instruments, flowmeters and analytical sensors, and supply them and their spares to oil and gas, power, water and process plants in Oman and the GCC by model code.
Rosemount company profile

Established1956, Minnesota, USA, as Rosemount Engineering Company
FoundersFrank D. Werner, Robert E. Keppel and Vernon H. Heath
OwnershipAcquired by Emerson in 1976; now an Emerson brand
Product familiesPressure, DP flow and level transmitters, temperature sensors and transmitters, radar and guided wave radar level, vibrating fork switches, tank gauging, magnetic, vortex and ultrasonic flowmeters, liquid analysis sensors and wireless instruments

Milestones

1956: Robert Keppel, Vernon Heath and Frank Werner incorporate Rosemount Engineering Company in Minnesota; early work is aerospace sensors.
1969: The Rosemount 1151 analog pressure transmitter is introduced.
1976: Emerson acquires Rosemount.
1989: The Rosemount 3051 pressure transmitter, built on digital technology, is launched.
2001: The 3051S generation follows.

Rosemount products we list

FamilyModelsTypical use

Pressure transmitters3051, 3051C coplanar, 3051T in-line, 3051S, 3051S series, 3051SMV multivariable, 2051, 2088, 1199 diaphragm sealsPressure, differential pressure and remote seal measurement
DP flow and DP level3051CF flowmeter, 3051L, 3051SALFlow through primary elements, and level in pressurised vessels
Temperature644, 248, 3144P, 848T, 648 wireless, 214C RTD sensor, 2240S multi-inputProcess temperature, high-density multipoint and tank temperature
Radar and level5408, 5400, 5300 GWR, 3300 GWR, 3308A wireless, 5900S, 5900C tank gaugesLevel in tanks, separators and custody transfer tank gauging
Level switches2120, 2140, 2160 wireless, 702 discrete transmitterHigh and low level alarms, overfill protection and pump protection
Flowmeters8700 magnetic, 8800D vortex, SeniorSonic 3414, 3415/3416, 3417, JuniorSonic 3411/3412, 3814 liquidWater, steam and gas flow, and fiscal gas and liquid metering
Liquid analysis396P TUpH, 396PVP, 499ACL chlorine sensorpH, ORP and free chlorine in water and process streams

Many Rosemount transmitters share one platform, so the model code tells you what you have. On a 3051, the letter after the number sets the type: C for coplanar (differential or gauge), T for in-line, L for liquid level and CF for a complete flowmeter. The next characters set the measurement type, the range code, the output (4-20 mA HART, FOUNDATION Fieldbus or WirelessHART) and the wetted materials. A transmitter with the wrong range code may work, but with poor accuracy at the point that matters.
Where Rosemount instruments are used
In Oman and the GCC, Rosemount instruments are used across upstream oil and gas, gas plants and LNG, refineries, tank farms and terminals, power and desalination stations, and water networks. Ultrasonic gas meters and tank gauges are used for fiscal and custody transfer measurement.
Ordering Rosemount instruments: what we need
Send the full model code from the nameplate (for example 3051CD2A22A1AM5B4) and the tag if it is a replacement. For a new instrument, send the measuring range, process connection, wetted materials, output protocol and hazardous area approval.
Facts checked against Emerson's Rosemount history and published company histories, October 2026.
Browse Rosemount products by category: Pressure Instrumentation, Temperature Instrumentation, Level Transmitter, Liquid Analysis Sensors, Flowmeters.

FAQ

Questions about the Rosemount 02120-7000-0003 2120 NAMUR Cassette

Answers from our engineering team on installation, compatibility and integration.

01What is Rosemount 02120-7000-0003?
It is the replacement NAMUR electronics cassette (light blue label, electronic type K) for the Rosemount 2120 vibrating fork level switch.
02What device must be connected to 02120-7000-0003?
A certified intrinsically safe isolating amplifier to IEC 60947-5-6, because the NAMUR output is a current signature rather than a power switch.
03Can 02120-7000-0003 be swapped with an 8/16 mA cassette?
Only if the existing amplifier and certification support it; IS cassettes may be replaced only by the same type, so keep NAMUR for NAMUR.
04Is a NAMUR cassette available with intrinsic safety approvals?
Yes. Electronic type K is listed for all certification codes except the Korean IP code, including ATEX I1, USA I5 and IECEx I7.
05Does the cassette have a switching delay?
Yes, the 2120 offers selectable 0.3, 1, 3, 10 or 30 second delays and a Dry-on or Wet-on mode.