Brand:

Honeywell STG735 SmartLine STG700 Gauge Pressure Transmitter

Description

STG735 at a glance

The Honeywell STG735 is the lowest-range gauge model in the STG700 SmartLine line: 50 psi (3.5 bar) at the top, with the lower limit at minus 14.7 psi (minus 1.0 bar) so that mild vacuum can be read as well as positive gauge pressure. With a minimum span of 0.5 psi (0.035 bar), it is the choice for tank blanketing, low-pressure steam headers and air supply regulation, where a 0 to 10 psi calibration needs better resolution than a 500 psi body could give.

Inside is the same piezoresistive sensor technology as the other SmartLine models, delivering terminal-based accuracy up to 0.065 percent of span. The two-wire HART 7 output suits Experion and third-party systems alike. Seven Star LLC is an independent supplier serving Oman and the wider GCC and is not affiliated with the manufacturer.

What stands out on the STG735

  • Range from minus 14.7 psi to 50 psi (minus 1.0 to 3.5 bar).
  • Reference turndown 16.7:1, with the accuracy knee at 3 psi (0.21 bar).
  • Temperature effect 0.070 plus 0.008 times URL over span per 28 degC.
  • Total performance 0.128 percent of span at 10 psi, 5:1 turndown, plus or minus 28 degC shift.
  • Maximum allowable working pressure equal to its range, 50 psi (3.5 bar).

STG735 specification table

Parameter Value
Model STG735 (STG700 SmartLine gauge pressure)
Upper range limit 50 psi (3.5 bar)
Lower range limit -14.7 psi (-1.0 bar)
Minimum span 0.5 psi (0.035 bar)
Maximum turndown 100:1
Reference accuracy 0.065 percent of span (A 0.005, B 0.060)
Accuracy knee C 3 psi (0.21 bar)
Temperature effect 0.070 + 0.008 x URL/span, percent of span per 28 degC
Total performance example 0.128 percent of span at 10 psi
Stability 0.020 percent of URL per year, quoted over ten years
MAWP 50 psi (3.5 bar)
Output / protocol Two-wire 4 to 20 mA with HART 7
Supply voltage and loop load 10.8 to 42.4 VDC at the terminals (intrinsically safe versions limited to 30 VDC); load 0 to 1,440 ohms
Response time 100 ms; turn-on time 2.5 s; damping 0 to 32 s, default 0.5 s
Rated ambient / meter body temperature -40 to 85 degC / -40 to 110 degC (operative limit 125 degC)
Overpressure 1.5 x MAWP of the model
Process connections Dual head: 1/2-inch NPT female. Inline: 1/2-inch NPT female or male, 9/16-18 high-pressure cone and thread female, G1/2-B male
Wetted diaphragm / head 316L stainless steel or Hastelloy C-276 diaphragm; heads in carbon steel (zinc plated), 316 SS or Hastelloy C-276 (dual head), 316L SS or Hastelloy C-276 (inline)
Fill fluid Silicone or CTFE
Housing Polyester powder-coated low-copper aluminium, Type 4X / IP66 / IP67; stainless housing optional
Net weight Dual head 8.3 lb (3.8 kg); inline 3.6 lb (1.6 kg), with aluminium housing

STG735 compared with the rest of the STG700 line

Among the six STG700 models documented here the STG735 is the most sensitive. Its 0.5 psi minimum span is one tenth of the STG745 (5 psi) and one sixtieth of the STG775 (30 psi). It also has the best temperature figures of the family, with a 0.008 E coefficient against 0.013 for the STG745 and STG775, and 0.070 base effect against 0.100 for the STG73S.

Its sibling STG73S shares the same 50 psi (3.5 bar) range, yet the datasheet gives the STG73S a reference turndown of only 8:1 and a knee of 6 psi (0.42 bar). For the same 10 psi application the STG735 delivers 0.128 percent total performance whereas the STG73S delivers 0.187 percent.

Model URL Min span Total perf. example
STG735 50 psi (3.5 bar) 0.5 psi 0.128% at 10 psi
STG73S 50 psi (3.5 bar) 0.5 psi 0.187% at 10 psi
STG745 500 psi (35 bar) 5 psi 0.154% at 100 psi

Process situations suited to STG735

Typical placements in Omani plants include fuel-gas knock-out drum pressure, nitrogen blanketing on storage tanks, instrument air receivers and low-pressure flare headers. In water treatment it monitors backwash air blowers and pump suction. Industrial estates in the GCC use it on steam condensate return lines and boiler feedwater deaerators where pressure is a few psi above atmosphere.

Because the lower range limit is negative, the STG735 can also be calibrated for compound ranges such as minus 5 to plus 20 psi.

Fitting STG735 into an existing instrument list

Choose the STG735 when peak pressure, including relief set point margins, stays under 50 psi. Remember that the overpressure rating is 1.5 times MAWP, so a 50 psi unit is not appropriate on a line with 150 psi relief. The datasheet lists dual-head and inline arrangements, with 1/2-inch NPT female for the dual head and several inline thread options including 9/16-18 cone and thread and G1/2-B male.

Check existing installed brackets: carbon steel, 304 and 316 stainless steel mounting brackets are available for two-inch pipes.

Piping up and starting STG735

Mount above or at the tap with a drain valve in the impulse line for steam service and include a siphon where condensate would hot-soak the diaphragm; meter body temperature is rated up to 110 degC. Make electrical connections with 16 AWG or smaller conductors, close covers to maintain the IP66 and IP67 rating, and set damping from the default 0.5 second to suit pulsation.

Zero the unit against atmosphere with the process vented.

Care, checks and spare modules

Annual checks are usually adequate for blanketing service, with calibration verification at two years per the datasheet. Look at the display for write-protect indication and critical diagnostics. Keep the electronics module and the display as site spares.

STG735: practical questions answered

What gauge range does the Honeywell STG735 cover? The STG735 covers minus 14.7 psi to 50 psi (minus 1.0 to 3.5 bar) with a minimum span of 0.5 psi (0.035 bar).

How accurate is the STG735? Reference accuracy is up to 0.065 percent of span; the total performance example is 0.128 percent of span at 10 psi with 5:1 turndown and a 28 degC shift.

Can the STG735 read vacuum? Yes, the lower range limit is minus 14.7 psi (minus 1.0 bar), which is full gauge vacuum.

What is the maximum working pressure of the STG735? The datasheet gives a maximum allowable working pressure of 50 psi (3.5 bar), with overpressure of 1.5 times that value.

Which communication protocol does the STG735 use? It uses a two-wire 4 to 20 mA output with HART 7 digital communication.

Documents and quotation

The STG700 specification 34-ST-03-122 includes the STG735 ranges, accuracy tables and approval list, and is supplied with quotations.

To quote the Honeywell STG735, send part number, quantity, calibrated range, process connection and approval requirements to Seven Star LLC.

Other Models in the STG700 Range

Seven Star LLC supplies models from the STG700 range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.

ModelCategory
Honeywell STG735 SmartLine STG700 Gauge Pressure TransmitterPressure TransmittersYou are viewing this model
Honeywell STG745 SmartLine STG700 Gauge Pressure TransmitterPressure TransmittersView details
Datasheets & Documents
  • PDF Datasheet Honeywell STG735 SmartLine STG700 Gauge Pressure Transmitter Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Honeywell is the American industrial technology group whose products on this site cover gas and flame detection, pressure transmitters, flow measurement, gas metering, fire alarm systems and helicopter rotor balancing equipment. Its roots go back to a thermostat business in Minneapolis in 1885 and to Mark Honeywell's heating company, which merged with it in 1927. Several brands we list have joined Honeywell over the years: Gent fire systems of Leicester (part of Honeywell since 2005), Chadwick-Helmuth (a Honeywell business unit from 2002) and Elster gas metering (acquired around the end of 2015).

Seven Star LLC is an independent supplier in Muscat. We list Honeywell gas detectors and sensors, transmitters, Elster gas meters and flow computers, Gent fire alarm equipment and Chadwick-Helmuth balancers, and supply them and their spares in Oman and the GCC by part number.

Honeywell company profile

Origins 1885, Minneapolis (Butz thermostat business); merged with Mark Honeywell's company in 1927
Brands in our range Honeywell Analytics gas detection, SmartLine transmitters, Elster gas metering, Gent fire detection, Chadwick-Helmuth rotor track and balance
Gent Founded in Leicester, UK, around 1872 by John Thomas Gent; part of Honeywell since 2005
Elster Gas, water and electricity metering; bought by Honeywell from Melrose, deal completed around the end of 2015

Milestones

  • 1885: The Butz thermostat business starts in Minneapolis.
  • 1927: It merges with Mark Honeywell's company.
  • 2005: Gent fire detection (Leicester, UK) becomes part of Honeywell.
  • 2015: Honeywell completes the purchase of Elster from Melrose around the end of the year.

Honeywell products we list

Family Models Typical use
Fixed gas detectors Searchline Excel, Searchline Excel Plus, 02104-N-XSSA, Sensepoint XCD RTD, Sensepoint XCD IR, XCD H2, Sensepoint XCL, Sensepoint XRL, OmniPoint, RAEGuard 2 PID Flammable, toxic and VOC gas detection in process areas
Sensors, flame detection and control SR-W-MP75C LEL sensor, SR-X2V O2 sensor, FS24X Plus flame detector, R4343E flame switch, System 57 card, 5704FS panel Replacement sensors, fire detection and gas detection control panels
Transmitters and flow SmartLine STG700, STG800, STR82D remote seal, VersaFlow Coriolis Process pressure and mass flow measurement
Elster gas metering Q.Sonic plus, TRZ2, Quantometer, RABO, IRM-3 DUO, EK280, EC 350, enCore FC1, enCore ZM1, gas-net F1, gas-net D2, GasLab Q2, USM display Gas meters, volume correction and fiscal flow computing
Gent fire alarm systems Vigilon Plus, Vigilon Compact Plus, Nano, Xenex, S-Quad detector, S-Cubed, ALD302 base, S4-34420 interfaces, loop card, repeat panel, mimic panel, XENS-850 call point, 74430 bell Addressable and conventional fire alarm systems in buildings and plants
Chadwick-Helmuth rotor track and balance 8500C, 8500C Plus, Vibrex 2000, Model 11, photocell system, printer, bracket, protractor Helicopter rotor track and balance and propeller balancing

Gas detection uses two basic approaches. Point detectors such as the Sensepoint XCD measure the gas at one location, with a catalytic or infrared sensor for flammable gas and an electrochemical cell for toxic gas such as H2S. Open path detectors such as the Searchline Excel send an infrared beam across an area up to a few hundred metres long and detect a gas cloud anywhere along it. Infrared sensors cannot be poisoned the way catalytic beads can, which matters on sites with silicones or H2S.

Where Honeywell products are used

In Oman and the GCC, Honeywell gas detectors protect wellheads, gas plants, refineries and tank farms. Elster meters and flow computers measure gas at city gate stations and in industry. Gent systems are installed in commercial buildings, hospitals and hotels, and Chadwick-Helmuth balancers are used by helicopter operators.

Ordering Honeywell products: what we need

Send the part number from the label (for example 02104-N-XSSA or SPXCDASMTXF). For gas detectors, add the gas, range, output and certification. For gas meters, add size (DN), flow range and pressure rating. For Gent equipment, add the panel type and loop protocol.

Facts checked against Honeywell press releases and SEC filings, and published histories of Gent and Chadwick-Helmuth, October 2026.

FAQ

Questions about the Honeywell STG735 SmartLine STG700 Gauge Pressure

Answers from our engineering team on installation, compatibility and integration.

01What gauge range does the Honeywell STG735 cover?
The STG735 covers minus 14.7 psi to 50 psi (minus 1.0 to 3.5 bar) with a minimum span of 0.5 psi (0.035 bar).
02How accurate is the STG735?
Reference accuracy is up to 0.065 percent of span; the total performance example is 0.128 percent of span at 10 psi with 5:1 turndown and a 28 degC shift.
03Can the STG735 read vacuum?
Yes, the lower range limit is minus 14.7 psi (minus 1.0 bar), which is full gauge vacuum.
04What is the maximum working pressure of the STG735?
The datasheet gives a maximum allowable working pressure of 50 psi (3.5 bar), with overpressure of 1.5 times that value.
05Which communication protocol does the STG735 use?
It uses a two-wire 4 to 20 mA output with HART 7 digital communication.