Brand:

JUMO 902044/20-380 Screw-in Pt100 Probe, Pg9 Plug, DNV

Description

JUMO 902044/20-380-1003-1-8-100-104-26/000 is a screw-in Pt100 resistance thermometer with a DIN EN 175301-803 plug-connector head, built on the 8 mm protection tube of basic type 902044/20, the variant that the manufacturer lists with DNV approval. It is a field-replaceable, pressure-rated probe for engines, compressors and plant or ship machinery, and it can be found in a trade-off position between the quick 6 mm stepped tube and heavier thermowell probes. Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer.

Why 902044/20-380-1003-1-8-100-104-26/000 suits machinery and marine duty

The datasheet describes the 902044 range as vibration-resistant, with a locked plug connection and IP65 protection when connected. Basic types 902044/20 and 902044/21 are covered by DNV Class Guideline CG-0339 under certificate TAA00002RV, with vibration resistance of DNV class B up to 250 mm insertion length and class A up to 500 mm. With an insertion length of 100 mm this probe falls within the class B limit. The same pages state a maximum pressure load of 50 bar at 200 degrees Celsius for these types, which fits lube-oil and cooling-water lines on engine rooms and offshore utility modules.

Order code 902044/20-380-1003-1-8-100-104-26/000 explained

Basic type 20 is the screw-in probe with junction box Pg9 according to DIN EN 175301-803, the form formerly known as DIN 43650. Code 380 means minus 50 to plus 200 degrees Celsius. The insert code 1003 is one Pt100 in 2-wire circuit, the 1 is tolerance class B, 8 is the 8 mm diameter protection tube, 100 is the insertion length in mm, 104 is a G 1/2 screw connection, and 26 selects stainless steel 1.4571 for the tube. The closing 000 is no extra code, so neither the SIL and PL compatible options nor the hydrogen-ready option is included.

Specification table for 902044/20-380-1003-1-8-100-104-26/000

Property Value from the manufacturer datasheet
Basic type 902044/20, Pg9 plug connector, DNV approved type
Temperature range minus 50 to plus 200 degrees Celsius
Measuring insert 1x Pt100, DIN EN 60751:2009 / IEC 60751:2008, class B, 2-wire, exchangeable
Protection tube Stainless steel 1.4571, 8 mm
Insertion length 100 mm
Process thread G 1/2, stainless steel
Max pressure 50 bar at 200 degrees Celsius
Response in water 0.4 m/s t0.5 = 15 s, t0.9 = 45 s
Connector rating IP65 connected; ambient minus 40 to plus 125 degrees Celsius; conductor up to 1.5 mm2
Part number 00368414

How 902044/20-380-1003-1-8-100-104-26/000 differs from other 902044 variants

The 15 and 16 versions use a 6 mm tube stepped down to 3.5 mm and respond in under 2 seconds to 50 percent and under 4 seconds to 90 percent; this part, with the 8 mm tube, takes 15 and 45 seconds but withstands the stated 50 bar. Types 21, 26 and 29 swap the Pg9 plug for Pg11, types 25 to 29 integrate a 4 to 20 mA transmitter, and types 80 and 81 have a different 8 mm design with 5 and 12 seconds response. Siblings in stock such as 902044/20-380-1003-1-8-50 and 902044/20-380-1003-1-8-150 only change the insertion length.

Where it is used

Engine cooling circuits, compressor discharge pipes, gearbox oil lines and general plant engineering are the application areas given by the datasheet, together with shipbuilding. In Omani ports, shipyards, power stations with diesel gensets, and gas compression stations, a threaded, DNV-listed Pt100 with a standard Hirschmann-style plug is a practical replacement for older probes in existing pockets.

Selection and compatibility notes

If the process needs rapid response, choose the stepped tube; if pressure and robustness matter, as for 902044/20-380-1003-1-8-100-104-26/000, choose the 8 mm tube. A thermowell from datasheet 909710 can protect the tube in aggressive media. The element is an exchangeable measuring insert in the non-transmitter versions, but shorter fitting lengths with poor heat conduction may exceed the tolerance class, a point the datasheet makes explicitly. Confirm the cable entry of the mating socket, Pg9 here.

Installation and commissioning

Seal the G 1/2 thread with a suitable washer, tighten, then fit the plug and fix its screw to lock the contacts. The datasheet warns that the contact marked for earth must not be used as the earth connection. Wire in 2-wire circuit, then compare the loop reading against a reference at a known temperature.

Maintenance of 902044/20-380-1003-1-8-100-104-26/000

Check the connector gasket and the locking screw during service intervals; replace the measuring insert if the element drifts, since it is exchangeable on this version. Keep part number 00368414 on the spares list.

Transmitter option and accuracy remarks for 902044/20-380-1003-1-8-100-104-26/000

The same datasheet explains that basic types 902044/25 to 902044/29 integrate a 4 to 20 mA transmitter with a measuring range limit of minus 50 to plus 260 degrees Celsius, a span of 25 to 310 K, supply of 7.5 to 30 V DC and a transfer accuracy of 0.1 percent. 902044/20-380-1003-1-8-100-104-26/000 is a plain resistance version without that electronics, so the head contains only the plug contacts, and the Pt100 resistance is passed to the instrument through the cable. That is an advantage where the controller or PLC card already linearizes Pt100, because a cabinet-mounted input is easier to calibrate and replace than a transmitter in a hot engine room.

For temperature tolerance the 8 mm tube of 902044/20-380-1003-1-8-100-104-26/000 with class B follows the standard DIN EN 60751 limits; classes A and AA are separate digits in the 902044/20 order table, together with two Pt100 elements in 2-wire circuit and a Pt1000 option. If the application is safety related, the SIL and PL compatible extra codes 658, 659 and 665 would need to be added to the order string and are linked to specific JUMO limiters and transmitters. The probe should not be located right behind a heater element or at a dead leg, where it reports the pocket temperature instead of the process.

Documents and quotation

The JUMO 902044 datasheet is attached. For a quotation on 902044/20-380-1003-1-8-100-104-26/000, send the complete order code, the quantity and your delivery location in Oman or the GCC with your RFQ.

Datasheets & Documents
  • PDF Datasheet JUMO 902044/20-380 Screw-in Pt100 Probe, Pg9 Plug, DNV Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

JUMO is the German maker of temperature sensors, process controllers, pressure transmitters and liquid analysis instruments. Moritz Kurt Juchheim founded it in Fulda in 1948 as M. K. Juchheim KG, making glass thermometers with six employees. The name JUMO comes from the founder's name. It is still a family company headquartered in Fulda, and the parent company has been called JUMO GmbH & Co. KG since 2003.

Seven Star LLC is an independent supplier in Muscat. We list JUMO RTD and thermocouple probes, thermowells, transmitters, controllers, recorders and water analysis instruments, and supply them to process plants, OEMs, HVAC contractors and water utilities in Oman and the GCC by type number.

JUMO company profile

Established 1948, Fulda, Germany, by Moritz Kurt Juchheim
First products Glass and contact thermometers; platinum glass sensors from 1966
Ownership Family-owned; renamed JUMO GmbH & Co. KG in 2003
Product families RTD and thermocouple probes, thermowells, temperature and pressure transmitters, controllers and thermostats, safety temperature limiters, paperless recorders, liquid analysis and flow and level instruments

Milestones

  • 1948: Moritz Kurt Juchheim founds M. K. Juchheim KG in Fulda, making glass thermometers with six employees.
  • 1966: Platinum glass temperature sensors are added.
  • 2003: The parent company is renamed JUMO GmbH & Co. KG.

JUMO products we list

Family Models Typical use
RTD and thermocouple probes 902020, 902030, 902050, 902120, 902006 2xPt100, PROCESStemp, mineral-insulated RTD, 901020 thermocouple, 901210, 902810 hygienic, 902550 surface, 902520 HVAC, MarineTemp Temperature measurement in pipes, vessels, machines, HVAC and food plants
Thermowells and inserts 909710 thermowells, 909740 measuring inserts Protecting sensors and replacing them without opening the process
Temperature transmitters dTRANS T01, T03, T04, T05, T06 SIL, T07 HART, T09, T1000 IO-Link, DELOS T switch Converting sensor signals to 4-20 mA, HART or IO-Link
Pressure and level dTRANS p20, p30, p31 hygienic, p33 Ex, p02 DELTA, MIDAS S05, MIDAS DP10, DELOS S02, MAERA S25, NESOS R20 Pressure in process, hydraulic and OEM systems; tank and well level
Controllers, limiters and recorders DICON touch, IMAGO 500, dTRON 316, iTRON, cTRON, diraTRON 104, eTRON T, safetyM STB/STW, safetyM TB/TW, LOGOSCREEN 601, mTRON T Furnace, oven and process control, over-temperature protection and recording
Liquid analysis and flow AQUIS touch S, AQUIS touch P, AQUIS 500 pH, AQUIS 500 CR, dTRANS pH 02, pH electrode, digiLine DO, ecoLine NTU, flowTRANS MAG H20, flowTRANS US W01 pH, conductivity, oxygen and turbidity in water treatment; liquid flow

Choosing a temperature sensor starts with the range and accuracy. A Pt100 RTD is accurate and stable from about -200 to 600 °C and is the usual choice in process plants. A thermocouple (type K or J, for example) covers much higher temperatures and responds faster but is less accurate. The wiring also matters: a 3-wire or 4-wire Pt100 connection cancels the resistance of the cable, while a 2-wire connection adds an error that grows with cable length. That is why a head-mounted transmitter is often fitted close to the sensor.

Where JUMO products are used

In Oman and the GCC, JUMO sensors and controllers are used in oil and gas and chemical plants, food and beverage production, plastics and metal heat treatment, HVAC and district cooling, water and wastewater treatment, swimming pools and marine applications.

Ordering JUMO products: what we need

Send the JUMO type number and the full order code from the label (for example 902020/10-...). For probes, add the sensor type (Pt100, Pt1000, type K or J), the number of wires, the insertion length and diameter, and the process connection. For controllers, add the input, outputs and supply voltage.

Facts checked against JUMO's company history, October 2026.

FAQ

Questions about the JUMO 902044/20-380 Screw-in Pt100 Probe, Pg9 Plug,

Answers from our engineering team on installation, compatibility and integration.

01Is 902044/20-380-1003-1-8-100-104-26/000 the DNV approved type?
Yes. The datasheet lists basic types 902044/20 and 902044/21 under DNV Class Guideline CG-0339, certificate TAA00002RV, for insertion lengths up to 500 mm.
02How fast is 902044/20-380-1003-1-8-100-104-26/000?
With the 8 mm protection tube the datasheet gives t0.5 of 15 s and t0.9 of 45 s in water at 0.4 m/s, much slower than the 6 mm stepped versions.
03What pressure can 902044/20-380-1003-1-8-100-104-26/000 take?
The datasheet states a maximum pressure load of 50 bar at 200 degrees Celsius for basic types 902044/20 and 902044/21.
04Which plug does 902044/20-380-1003-1-8-100-104-26/000 use?
Basic type 20 has a DIN EN 175301-803 plug connector with Pg9 cable entry, locked for contact, IP65 when connected, with conductors up to 1.5 mm2.
05Can the sensor in 902044/20-380-1003-1-8-100-104-26/000 be exchanged?
The measuring insert is exchangeable in the versions without a transmitter, and this code has no transmitter; shorter fitting lengths can exceed the tolerance class.