Brand:

AMETEK Land SOLOnet Digital Infrared Thermometer

Description

AMETEK Land SOLOnet web-enabled digital infrared thermometer

SOLOnet is a compact digital infrared thermometer from AMETEK Land for stand-alone, single-point or multi-point temperature monitoring. It shows live data on a PC, speaks addressable RS485 and, through an optional interface unit, Ethernet, while still providing the usual 0 or 4 to 20 mA analogue output. Four models cover different materials and temperature spans: SN11, SN21, SN51 and the ratio model SNR1. Seven Star LLC supplies the SOLOnet range in Oman and the GCC as an independent supplier in Muscat.

The model code is simple. SN1 means 1 µm, SN2 means 1.6 µm, SN5 means 5 µm and R indicates the ratio type working near 1 µm, so the part you order is chosen by the surface and temperature you need to read.

SOLOnet technical specifications

Parameter SN11 SN21 SN51 SNR1
Operating range 550 to 1750 °C 250 to 1300 °C 200 to 1100 °C 700 to 1750 °C
Specified range 600 to 1750 °C 300 to 1300 °C 250 to 1100 °C 750 to 1750 °C
Spectral range 1 µm 1.6 µm 5 µm 1 µm ratio
Field of view (95 % energy) 100:1 100:1 50:1 100:1
Response time (0 to 90 %) 10 ms 10 ms 50 ms 10 ms
Absolute accuracy 0.3 % K 2 K 0.35 % K 0.6 % K
Common parameter Specification
Manufacturer / part number AMETEK Land / SOLOnet (SN11, SN21, SN51, SNR1)
Focus Fixed, user-configurable: 250 mm, 500 mm, 1000 mm and infinity
Output 0 or 4 to 20 mA, user selectable, isolated to 50 V
Emissivity setting 0.10 to 1.00; non-greyness 0.80 to 1.25
Time functions Peak picker, track and hold, averager
Alarm outputs Process high, process low, internal temperature, emissivity/NG signal lost, insufficient signal
Optional sighting Laser defining optical axis, class 2, 1 mW, 650 nm, 120 s on, then automatic switch-off
Ambient limits 5 to 60 °C specified; 0 to 70 °C operating
Sealing IP65 / NEMA 4X
Vibration 3 g, 10 to 300 Hz
Power supply 18 to 30 V DC (24 V DC nominal)
Window Sapphire window standard on all models

How SOLOnet communicates and is configured

Each thermometer carries addressable RS485 communication, so it can talk straight to a customer device or through the optional DIN-rail SOLOnet Ethernet Interface Unit SN-W/E. The interface unit lets the installer give the unit an IP address, join the plant network and read data by Ethernet, or switch the front-panel setting to pass RS485 straight through. Current and contact-closure I/O is built into each thermometer. Setup, scaling and data display are carried out with the free SOLOnet and IQ Configurator software on a PC, and field-changeable optics can be fitted without recalibration or extra parts. Calibration is traceable to national standards.

SOLOnet accessories and the wider Land range

The manufacturer lists protective mountings: an air-cooled jacket with air purge, a water-cooled jacket with air purge for ambients above 60 °C or high radiant heat, a mounting plate with air purge for up to 60 °C, single- and dual-axis adjustable brackets, prewired cable assemblies of 3, 15 and 25 m, a DC power supply and the LMi temperature indicator. Compared with UNO, SOLOnet adds digital communications and alarm outputs; compared with System 4 and a Landmark processor it is a compact one-box solution. The SN51 is named by the manufacturer in glass-oven heating and outlet measurement, with other models for glass, aluminium, steel and minerals work.

Selecting a SOLOnet model

  • SN11: 1 µm for the hottest metal and glass work, 550 to 1750 °C.
  • SN21: 1.6 µm, 250 to 1300 °C, for lower temperatures on metals.
  • SN51: 5 µm, 200 to 1100 °C, suited to glass surfaces and the lowest span.
  • SNR1: ratio model for emissivity changes and partly obscured targets, 700 to 1750 °C.
  • Focus: choose 250 mm, 500 mm, 1000 mm or infinity from the working distance.
  • Network: decide whether RS485 direct or Ethernet via SN-W/E is needed.
  • Mounting: specify jacket, purge and bracket for the ambient heat.

Installation, commissioning and maintenance

Fit the thermometer in a mounting that suits ambient temperature; above 60 °C use the water-cooled jacket. Run 24 V DC power and the 0 or 4 to 20 mA loop, and connect RS485 with screened twisted pair if networked. Configure emissivity, span, alarms and time functions by PC, using the SOLOnet configuration guide and quick-start guide. The manufacturer also publishes a laptop-configuration troubleshooting guide, useful for USB or serial adapter issues. Keep the sapphire window clean and note that the optional laser is switched off automatically after 120 s.

Typical applications in Oman and the GCC

The manufacturer lists aluminium, glass, industrial processing, minerals and steel as target industries. In the Gulf this translates into aluminium smelting and casting, glass container and float production, metal heat treatment and minerals processing. IP65 sealing and the sapphire window help where dust is present, and the cooled jacket allows installation near hot equipment in high summer ambient temperatures.

Information to send with an RFQ for SOLOnet

  • Model: SN11, SN21, SN51 or SNR1.
  • Target material, temperature range, and working distance.
  • Laser sighting needed or not.
  • Output and network needs (analogue, RS485, Ethernet via SN-W/E).
  • Mounting jacket, purge, bracket and cable length.
  • Ambient temperature at the mounting point.

Site conditions in Oman for the SOLOnet

Field instruments in Oman are exposed to strong sun, high ambient temperature, wind-blown dust and, on coastal plants, salt in the air. When you specify the SOLOnet, check the ambient and process temperature limits, the enclosure rating and the wetted and housing materials against the actual installation point, and consider sunshades for transmitters mounted in direct sunlight. Cable entries should be sealed with suitable glands so that dust and moisture cannot enter the housing.

Calibration and verification planning matters as much as selection. Record the range, configuration and tag data of each SOLOnet at commissioning, define a check interval that suits the criticality of the loop, and keep the certificates with the plant documentation so that audits and shutdown work can proceed without delay.

Documents and quotation

The manufacturer brochure for SOLOnet is attached on this page.

Send the model and your application details to Seven Star LLC and we will confirm availability for Oman and the GCC.

Datasheets & Documents
  • PDF Brochure AMETEK Land SOLOnet Brochure Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

AMETEK is the American maker of electronic instruments and electromechanical devices, headquartered in Berwyn, Pennsylvania. It was founded in 1930 as American Machine and Metals. In process industries it is known through brands such as Western Research (sulfur recovery and H2S analysers), Thermox (oxygen analysers), Land (pyrometers), Drexelbrook (RF level) and Chandler Engineering.

Seven Star LLC is an independent supplier in Muscat. We list AMETEK process analysers, pyrometers, portable testers and spare parts, and supply them to gas plants, refineries and power stations in Oman and the GCC.

AMETEK company profile

Established 1930, USA, as American Machine and Metals; later renamed AMETEK
Headquarters Berwyn, Pennsylvania
Structure Two groups: Electronic Instruments (EIG) and Electromechanical (EMG)
Process brands Western Research, Thermox, Land, Drexelbrook, Chandler Engineering, Ranarex

Milestones

  • 1930: American Machine and Metals is formed in March and listed on the New York Stock Exchange as AME.
  • Early 1960s: The company is renamed AMETEK.
  • 2021: AMETEK acquires Magnetrol.

AMETEK products we list

Product Type Typical use
888, 900, 930 Western Research sulfur recovery analysers Tail gas, air demand and H2S in sulfur pits on Claus units
931, 933, 9900RM/WM UV gas analysers H2S in natural gas and multi-component process streams
5000, 3050-OLV, 5100HD, 5100P Moisture analysers Water content in natural gas and process gases; the 5100 models use TDLAS
241CE II, Chanscope II / Model 13 Dew point analysers and testers Hydrocarbon and water dew point of natural gas
WDG 1200/1210, WDG-V, TM2000 Thermox oxygen analysers Combustion control on heaters and boilers, and trace oxygen
SPOT+, Cyclops L Land pyrometers Non-contact temperature of furnaces, kilns, tubes and hot metal
120HD, Ranarex, Drexelbrook RCT-1210 Portable analysers and level Hydrogen purity in generators, gas specific gravity, RF level measurement

Most of these analysers work on light absorption. UV analysers such as the 933 and 9900 measure H2S and SO2 by their absorption bands, and tail gas analysers use this to keep the H2S to SO2 ratio near 2:1 for the best Claus conversion. TDLAS moisture analysers use a tuned laser on a single water line, so they respond fast and do not suffer from the glycol and methanol contamination that affects sensor-based analysers.

Where AMETEK instruments are used

In Oman and the GCC, AMETEK analysers run on gas processing and sulfur recovery units, gas pipelines and custody points, refinery heaters and boilers, power station generators and steel and cement furnaces.

Ordering AMETEK products: what we need

Send the model and the complete part or sales order number from the nameplate. For a new analyser, send the gas composition, the component and range to be measured, the sample pressure and temperature, and the hazardous area classification. For spare parts, send the part number and the analyser serial number.

Facts checked against AMETEK company history, October 2026.

FAQ

Questions about the AMETEK Land SOLOnet Digital Infrared Thermometer

Answers from our engineering team on installation, compatibility and integration.

01Which SOLOnet models are available?
SN11 (1 µm, 550 to 1750 °C), SN21 (1.6 µm, 250 to 1300 °C), SN51 (5 µm, 200 to 1100 °C) and the ratio model SNR1 (700 to 1750 °C).
02How is SOLOnet connected to a network?
Each thermometer has addressable RS485. The optional SN-W/E Ethernet Interface Unit adds Ethernet and an IP address, or passes RS485 straight through.
03What is the output signal?
A user-selectable 0 or 4 to 20 mA output, isolated to 50 V, scalable to any span within the thermometer range, plus alarm and contact-closure I/O.
04How is SOLOnet configured?
By PC using the free SOLOnet and IQ Configurator software, which also displays live data. Emissivity, alarm limits, time functions and output span can be set.
05What ambient temperature can it take?
5 to 60 °C is the specified ambient and 0 to 70 °C is the operating limit; above 60 °C the manufacturer lists a water-cooled air-purged jacket.