Brand:

AMETEK EXSH1 Ex SPOT Flameproof Housing for SPOT+

Description

AMETEK Land EXSH1 Ex SPOT flameproof housing for SPOT+ pyrometers

The AMETEK Land EXSH1 Ex SPOT is a flameproof (Ex d) stainless steel housing that lets Ex SPOT+ infrared pyrometers be installed in hazardous areas with gas or dust atmospheres. The enclosure is certified to ATEX, IECEx and UKEX, so one specification can be used for projects across several regions. It needs no purge air for safety.

In the part name, EXSH1 is the housing and “Ex SPOT” identifies the pyrometer version it is made for. The same EXSH1 housing family also exists for the LWIR-640 thermal imager (listed as EXSH1 LWIR), so the pyrometer version is the one to choose when the instrument inside is an Ex SPOT+ unit. Seven Star LLC supplies the EXSH1 Ex SPOT in Oman and the GCC as an independent supplier in Muscat.

EXSH1 Ex SPOT technical specifications

Parameter Specification
Manufacturer / part number AMETEK Land (LAND Instruments International) / EXSH1 Ex SPOT
Function Flameproof housing for Ex SPOT+ pyrometers
Compatible pyrometers Ex SPOT+ M100, M160, M210, M390, M390 HD, R100, R160, R210 and GS
Dimensions Ø168 mm, 356 mm length
Weight 20 kg maximum
Material 316 / 316L stainless steel; sapphire window
Power rating 19 to 26.4 V DC, 2.3 A maximum
Ingress protection IP66 / IP68 (2 m, 24 hours)
Certified ambient temperature -40 to +70 °C (-40 to 158 °F)
Storage temperature -20 to +70 °C
Approvals ATEX, IECEx and UKEX
Standards listed EN IEC 60079-0, IEC 60079-1, EN 60079-1, IEC 60079-31, EN 60079-31, IEC 61010-1, EN 61326-1 (EMC)

EXSH1 Ex SPOT certification and temperature limits

Zone Marking Functional ambient temperature
Zones 1 and 21 II 2 (1) G D; Ex db [op is Ga] IIC T6 Gb; Ex tb [op is Da] IIIC T85C Db -40 to +50 °C without cooling; to +70 °C with water or air cooling
Zones 2 and 22 II 3 (1) G D; Ex db [op is Ga] IIC T6…T3 Gc; Ex tc [op is Da] IIIC T85…160 C Dc -40 to +50 °C without cooling; to +125 °C air cooled; to +160 °C water cooled

The manufacturer notes that the maximum functional temperature is limited by the SPOT+ pyrometer inside, while the enclosure stays safe at the higher temperatures shown in its markings. Always check the marking plate on the delivered unit against your area classification, gas group and temperature class before installation.

How the flameproof housing works

A flameproof enclosure contains any internal ignition and prevents it from igniting the surrounding atmosphere, instead of keeping explosive gas out. The EXSH1 body is 316 / 316L stainless steel with a sapphire window in front of the pyrometer optics. Because the window is gridless, the field of view is not obstructed by a protective mesh.

Cooling is optional. With no cooling the housing works to +50 °C ambient; air or water cooling extends the functional ambient range as shown in the table. Dust and water protection is IP66 and IP68.

Compatibility with Ex SPOT+ pyrometers

Pyrometer Type / wavelength
Ex M100 Mono, 1.0 µm
Ex M160 Mono, 1.6 µm
Ex M210 Mono, 2.3 µm
Ex M390 Mono, 3.9 µm
Ex M390 HD Mono, 3.9 µm
Ex R100 Ratio and mono, 1.0 and 1.2 µm
Ex R160 Ratio and mono, 1.0 and 1.5 µm
Ex R210 Ratio and mono, 2.1 and 2.4 µm
Ex GS Application-specific, galvanizing and galvannealing

Measuring ranges differ between these pyrometers and are given in the SPOT+ brochure, so confirm the range of the model you select there. An existing SPOT+ pyrometer cannot be retrofitted into the EXSH1 housing. Plan the pyrometer and housing together at the enquiry stage.

Selection and ordering checklist for EXSH1 Ex SPOT

  • Pyrometer model from the compatibility list above
  • Hazardous-area zone, gas group and temperature class (zones 1, 21, 2, 22)
  • Ambient temperature and whether air or water cooling is needed
  • Mounting accessories for the housing
  • Supply of 19 to 26.4 V DC at up to 2.3 A
  • Approval needed: ATEX, IECEx or UKEX
  • Cable glands and conduit entries suited to the Ex d enclosure

Installation and maintenance of EXSH1 Ex SPOT

The 20 kg housing needs a rigid bracket aligned on the target. Use certified glands and follow the flameproof joint rules: do not paint over or damage flame paths and do not open the enclosure under power in a hazardous area. Check the sapphire window periodically and clean it with a suitable cloth when dust or deposit builds up.

Wiring runs from a certified junction box or safe-area cabinet to the housing, with the 19 to 26.4 V DC supply sized for the 2.3 A maximum rating. Record the marking plate, serial number and certificate references in the plant hazardous-area register at commissioning, and keep a spare window seal set and gland kit on the shelf so a routine inspection does not become a long outage. Aiming is done on the bracket before the process is hot, using the pyrometer sighting method described in the SPOT+ documentation.

Typical applications in Oman and the GCC

The manufacturer lists continuous annealing lines, semiconductor growth and hydrocarbon processing and chemical applications. In Oman and the GCC, the relevant use is high-temperature process measurement in refineries, petrochemical units and gas plants where a standard pyrometer is not allowed in the classified area. The -40 to +70 °C certified range suits hot, sunlit outdoor installations.

Information to send with your EXSH1 Ex SPOT RFQ

  • The Ex SPOT+ model you plan to fit
  • Zone, gas and dust group, and temperature class
  • Process temperature and target distance
  • Ambient range and cooling choice
  • Quantity and delivery location

Documents and quotation

The AMETEK Land EXSH1 Ex SPOT housing brochure is attached on this page.

Send the EXSH1 Ex SPOT part number with your area classification and pyrometer model and we will prepare a quotation. Seven Star LLC is an independent supplier in Muscat.

Other Models in the Ametek Land Exsh1 Range

Seven Star LLC supplies models from the Ametek Land Exsh1 range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.

ModelCategory
AMETEK EXSH1 Ex SPOT Flameproof Housing for SPOT+Temperature InstrumentationYou are viewing this model
AMETEK EXSH1 LWIR Flameproof Housing for LWIR-640Temperature InstrumentationView details
Datasheets & Documents
  • PDF Brochure AMETEK Land EXSH1 Ex SPOT Brochure Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

AMETEK is the American maker of electronic instruments and electromechanical devices, headquartered in Berwyn, Pennsylvania. It was founded in 1930 as American Machine and Metals. In process industries it is known through brands such as Western Research (sulfur recovery and H2S analysers), Thermox (oxygen analysers), Land (pyrometers), Drexelbrook (RF level) and Chandler Engineering.

Seven Star LLC is an independent supplier in Muscat. We list AMETEK process analysers, pyrometers, portable testers and spare parts, and supply them to gas plants, refineries and power stations in Oman and the GCC.

AMETEK company profile

Established 1930, USA, as American Machine and Metals; later renamed AMETEK
Headquarters Berwyn, Pennsylvania
Structure Two groups: Electronic Instruments (EIG) and Electromechanical (EMG)
Process brands Western Research, Thermox, Land, Drexelbrook, Chandler Engineering, Ranarex

Milestones

  • 1930: American Machine and Metals is formed in March and listed on the New York Stock Exchange as AME.
  • Early 1960s: The company is renamed AMETEK.
  • 2021: AMETEK acquires Magnetrol.

AMETEK products we list

Product Type Typical use
888, 900, 930 Western Research sulfur recovery analysers Tail gas, air demand and H2S in sulfur pits on Claus units
931, 933, 9900RM/WM UV gas analysers H2S in natural gas and multi-component process streams
5000, 3050-OLV, 5100HD, 5100P Moisture analysers Water content in natural gas and process gases; the 5100 models use TDLAS
241CE II, Chanscope II / Model 13 Dew point analysers and testers Hydrocarbon and water dew point of natural gas
WDG 1200/1210, WDG-V, TM2000 Thermox oxygen analysers Combustion control on heaters and boilers, and trace oxygen
SPOT+, Cyclops L Land pyrometers Non-contact temperature of furnaces, kilns, tubes and hot metal
120HD, Ranarex, Drexelbrook RCT-1210 Portable analysers and level Hydrogen purity in generators, gas specific gravity, RF level measurement

Most of these analysers work on light absorption. UV analysers such as the 933 and 9900 measure H2S and SO2 by their absorption bands, and tail gas analysers use this to keep the H2S to SO2 ratio near 2:1 for the best Claus conversion. TDLAS moisture analysers use a tuned laser on a single water line, so they respond fast and do not suffer from the glycol and methanol contamination that affects sensor-based analysers.

Where AMETEK instruments are used

In Oman and the GCC, AMETEK analysers run on gas processing and sulfur recovery units, gas pipelines and custody points, refinery heaters and boilers, power station generators and steel and cement furnaces.

Ordering AMETEK products: what we need

Send the model and the complete part or sales order number from the nameplate. For a new analyser, send the gas composition, the component and range to be measured, the sample pressure and temperature, and the hazardous area classification. For spare parts, send the part number and the analyser serial number.

Facts checked against AMETEK company history, October 2026.

FAQ

Questions about the AMETEK EXSH1 Ex SPOT Flameproof Housing for SPOT+

Answers from our engineering team on installation, compatibility and integration.

01What is the EXSH1 Ex SPOT?
It is a flameproof 316 / 316L stainless steel housing that allows Ex SPOT+ pyrometers to be used in hazardous gas and dust areas.
02Which approvals does the EXSH1 Ex SPOT carry?
It is certified to ATEX, IECEx and UKEX for zones 1, 21, 2 and 22. Check the marking plate against your classification.
03What ambient temperature does it cover?
The certified range is -40 to +70 °C. Without cooling the functional limit is +50 °C; cooling extends it.
04Can I fit an existing SPOT+ pyrometer into it?
No. The manufacturer states existing SPOT+ pyrometers cannot be retrofitted to the EXSH1 housing; the Ex SPOT+ versions are used.
05Does the housing need purge air?
No. The enclosure does not require purge air for safety, and its window does not need a protective grid.