Brand: JUMO
JUMO tecLine pH Process pH Combination Electrode
Description
A pH electrode series built around the medium, not just the glass
The JUMO tecLine pH (type 201020) is a series of pH combination electrodes in which the measuring electrode and the silver-free reference system share one shaft. JUMO offers it in glass or PEI/PSU plastic shafts, with ceramic, glass-fibre, PTFE ring, annular-gap or perforated diaphragms, and with an optional integrated Pt100 or Pt1000 temperature probe. The same datasheet covers the redox sibling, JUMO tecLine Rd (type 201025), with platinum or gold tips. Every electrode is individually tested, uses FDA-listed physiologically safe materials and lead-free shaft glass, and the reference electrolyte contains no silver ions, which avoids diaphragm blocking by silver sulphide in many industrial liquids.
Technical specifications
| Parameter | Specification |
|---|---|
| Basic type | 201020 JUMO tecLine pH (pH); sibling 201025 JUMO tecLine Rd (redox, platinum or gold tip, ±2000 mV) |
| Shaft extensions | 51 glass shaft; 72 PEI plastic with basket; 73 PSU plastic without basket; 74 PSU plastic with basket (cartridge-style conduction system) |
| Active glass | UW pH 0 to 12 (briefly 14), -5 to +80 °C; HA pH 0 to 14, -5 to +80 °C; HT pH 0 to 14, 0 to 135 °C; DS pH 0 to 12, steam sterilisable; C pH 0 to 11, -5 to +50 °C, fluoride resistant to 1000 mg/l HF |
| Diaphragm codes | 05 glass filament; 07 1x zirconium dioxide; 09 3x zirconium dioxide; 04 PTFE ring; 10 annular gap; 11 perforated (gel, polymerised solid electrolyte) |
| Minimum medium conductivity | 1x ceramic 100 µS/cm (50 µS/cm with salt reserve); 3x ceramic 50 µS/cm; glass fibre 150 µS/cm; PTFE ring 100 µS/cm; annular gap or perforated 500 µS/cm |
| Connection | Pg13.5 screw head (code 22) or VP Pg13.5 plug head (code 18, for Pt100/Pt1000 versions); thread TR12.9 x 3 P1.5 |
| Fitting length | 120 mm standard, 225 mm optional, others on request |
| Pressure and temperature | Pressure-resistant versions up to 10 bar (50 °C); temperature up to -5 to +80 °C, or -30 to +30 °C for the TT version |
| Extra codes | 837 salt reserve; 838 two-chamber system (DOKA) with KCl/KCl bridge; 840 Pt100; 841 Pt1000 |
| Materials | Shaft glass to DIN 19263 or PSU/PEI; ring PSU; O-ring 10 x 3.5 FPM70; diaphragm zirconium dioxide Ø 1 mm |
| Stock example | 201020/51-18-07-22-120/837, part no. 00300151 (glass shaft, ceramic diaphragm, salt reserve, 120 mm) |
| Stock example | 201020/74-18-05-22-225/000, part no. 00354295 (PSU shaft with basket, glass filament diaphragm, 225 mm) |
| Delivery times | Stock versions 3 working days after order; production versions 10 working days |
Matching the glass and diaphragm to the medium
The type code reads 201020/shaft – glass – diaphragm – connection – length /extra codes, for example 201020/51-18-07-22-120/837. The universal UW glass (code 18) covers water, wastewater and low-ion media. HA glass (17) reduces alkaline error above pH 12, C glass (11) survives hydrofluoric acid up to 1000 mg/l, and the high-temperature HT and steam-sterilisable DS glasses serve hot or sterile processes on other series members. A single ceramic diaphragm suits general water engineering, while three diaphragms or the two-chamber DOKA system (838) are chosen where electrode poisons such as sulphides or cyanides are present. For oily wastewater, paints and suspensions the PTFE ring (04) or open annular-gap and perforated diaphragms (10, 11) resist clogging, though the open types are not suitable for very pure or low-ion liquids. The datasheet stresses that the type code is not a free modular system, so uncommon combinations need technical release by JUMO.
Applications, handling and companions
Typical fields include municipal and industrial water and wastewater, neutralisation plants, electroplating, drinking and well water, boiler feed water, paper and drinks production and cooling systems down to -30 °C. The salt-reserve option adds four salt rings to extend life in low-ion water or at high flow, and the datasheet notes that crystallisation on the rings is intended, not a defect. Pair the electrode with a transmitter such as the published dTRANS pH 02 or AQUIS 500 pH, and with a flow, immersion or retractable fitting from the JUMO accessory range. A glass electrode from the same family is also listed as the JUMO pH glass electrode. Browse the liquid analysis sensors category for further probes and fittings.
A good enquiry names the medium, its conductivity, the temperature and pressure at the electrode, and whether the process contains solids, oils or sulphides that would block a ceramic diaphragm. Those answers decide the active glass type, the diaphragm code and whether a salt reserve is wanted. Please also state the fitting length, 120 mm or 225 mm, the connection head, and whether a Pt100 or Pt1000 temperature sensor is built in. The datasheet gives three working days for stock versions and ten working days for production versions, and the stock examples already cover the glass shaft and the PSU shaft with basket.
Documents and quotation
Hosted for reference is JUMO data sheet 201020, covering both the tecLine pH and tecLine Rd series (21 pages, English). Confirm shaft, glass, diaphragm, head and length against your process data before ordering. Seven Star LLC is an independent supplier in Muscat; send the complete type code and medium details and we will quote.
Datasheets & Documents
- Datasheet JUMO tecLine pH data sheet 201020 (EN) Download
Reviews (0)
How would you rate this product? Cancel reply
About brand
JUMO is the German maker of temperature sensors, process controllers, pressure transmitters and liquid analysis instruments. Moritz Kurt Juchheim founded it in Fulda in 1948 as M. K. Juchheim KG, making glass thermometers with six employees. The name JUMO comes from the founder's name. It is still a family company headquartered in Fulda, and the parent company has been called JUMO GmbH & Co. KG since 2003.
Seven Star LLC is an independent supplier in Muscat. We list JUMO RTD and thermocouple probes, thermowells, transmitters, controllers, recorders and water analysis instruments, and supply them to process plants, OEMs, HVAC contractors and water utilities in Oman and the GCC by type number.
JUMO company profile
| Established | 1948, Fulda, Germany, by Moritz Kurt Juchheim |
|---|---|
| First products | Glass and contact thermometers; platinum glass sensors from 1966 |
| Ownership | Family-owned; renamed JUMO GmbH & Co. KG in 2003 |
| Product families | RTD and thermocouple probes, thermowells, temperature and pressure transmitters, controllers and thermostats, safety temperature limiters, paperless recorders, liquid analysis and flow and level instruments |
Milestones
- 1948: Moritz Kurt Juchheim founds M. K. Juchheim KG in Fulda, making glass thermometers with six employees.
- 1966: Platinum glass temperature sensors are added.
- 2003: The parent company is renamed JUMO GmbH & Co. KG.
JUMO products we list
| Family | Models | Typical use |
|---|---|---|
| RTD and thermocouple probes | 902020, 902030, 902050, 902120, 902006 2xPt100, PROCESStemp, mineral-insulated RTD, 901020 thermocouple, 901210, 902810 hygienic, 902550 surface, 902520 HVAC, MarineTemp | Temperature measurement in pipes, vessels, machines, HVAC and food plants |
| Thermowells and inserts | 909710 thermowells, 909740 measuring inserts | Protecting sensors and replacing them without opening the process |
| Temperature transmitters | dTRANS T01, T03, T04, T05, T06 SIL, T07 HART, T09, T1000 IO-Link, DELOS T switch | Converting sensor signals to 4-20 mA, HART or IO-Link |
| Pressure and level | dTRANS p20, p30, p31 hygienic, p33 Ex, p02 DELTA, MIDAS S05, MIDAS DP10, DELOS S02, MAERA S25, NESOS R20 | Pressure in process, hydraulic and OEM systems; tank and well level |
| Controllers, limiters and recorders | DICON touch, IMAGO 500, dTRON 316, iTRON, cTRON, diraTRON 104, eTRON T, safetyM STB/STW, safetyM TB/TW, LOGOSCREEN 601, mTRON T | Furnace, oven and process control, over-temperature protection and recording |
| Liquid analysis and flow | AQUIS touch S, AQUIS touch P, AQUIS 500 pH, AQUIS 500 CR, dTRANS pH 02, pH electrode, digiLine DO, ecoLine NTU, flowTRANS MAG H20, flowTRANS US W01 | pH, conductivity, oxygen and turbidity in water treatment; liquid flow |
Choosing a temperature sensor starts with the range and accuracy. A Pt100 RTD is accurate and stable from about -200 to 600 °C and is the usual choice in process plants. A thermocouple (type K or J, for example) covers much higher temperatures and responds faster but is less accurate. The wiring also matters: a 3-wire or 4-wire Pt100 connection cancels the resistance of the cable, while a 2-wire connection adds an error that grows with cable length. That is why a head-mounted transmitter is often fitted close to the sensor.
Where JUMO products are used
In Oman and the GCC, JUMO sensors and controllers are used in oil and gas and chemical plants, food and beverage production, plastics and metal heat treatment, HVAC and district cooling, water and wastewater treatment, swimming pools and marine applications.
Ordering JUMO products: what we need
Send the JUMO type number and the full order code from the label (for example 902020/10-...). For probes, add the sensor type (Pt100, Pt1000, type K or J), the number of wires, the insertion length and diameter, and the process connection. For controllers, add the input, outputs and supply voltage.
Facts checked against JUMO's company history, October 2026.
Browse JUMO products by category: Temperature Instrumentation, Pressure Instrumentation, Liquid Analysis Sensors, Controllers & Data Acquisition, Level Transmitter.
Related products
Frequently Asked Questions
Data sheet 201020, which also covers the redox electrodes of type 201025 (tecLine Rd).
Pressure-resistant versions are listed up to 10 bar at 50 °C.
It adds a salt reserve of four salt rings to lengthen service life in low-ion media or at high flow rates.
Yes. Extra codes 840 (Pt100) and 841 (Pt1000) add one, with connection code 18 (VP Pg13.5 head).
HA glass (code 17) covers pH 0 to 14 and reduces the alkaline error above pH 12.