Brand:

Yokogawa YTA310 Single-Input Temperature Transmitter

Description

When a temperature point needs one sensor, HART or BRAIN communication and a SIL 2 rated transmitter, the Yokogawa YTA310 is the model to specify. It is the single-input version described in General Specifications GS 01C50B02-00EN (28th edition, June 2017), shared with the dual-input YTA320. The YTA310 converts thermocouple, RTD, ohm or millivolt signals into a two-wire 4 to 20 mA signal and supports BRAIN or HART communication.

The YTA310 at a glance

Think of the YTA310 as a YTA320 with one sensor channel. It uses the same dual-compartment housing, the same sensor library and the same diagnostics, and in its standard BRAIN or HART configuration it is certified by TUV NORD CERT for SIL 2 capability with a single transmitter and SIL 3 with two. FOUNDATION Fieldbus is not available on the YTA310, because the fieldbus code -F applies only to the YTA320.

Technical data for YTA310

Model YTA310 temperature transmitter with single sensor input
Output signal codes -D BRAIN; -E HART
Output Two-wire 4 to 20 mA DC, 3.68 to 20.8 mA range
Update time About 0.5 second for a single sensor
Turn-on time About 5 seconds
Damping 0 to 99 seconds selectable
Supply voltage 10.5 to 42 V DC; 10.5 to 30 V DC for intrinsically safe; 10.5 to 32 V DC with option /A
Load 0 to 1335 ohm; 250 to 600 ohm for communication
Isolation 500 V DC between input, output and ground
Ambient temperature -40 to 85 degrees C; -30 to 80 degrees C with indicator
Humidity 5 to 100 % RH at 40 degrees C
Housing Low-copper aluminium or SCS14A stainless steel; IP66/IP67, NEMA4X
Weight 1.2 kg; indicator 0.2 kg; brackets 0.3 kg horizontal, 1.0 kg vertical
Pt100 range -200 to 850 degrees C; JPt100 -200 to 500; Cu -70 to 150
mV / ohm -10 to 100 mV, min span 3 mV; 0 to 2000 ohm, min span 20 ohm

Features that justify choosing YTA310

  • Sensor matching: the Callender-Van Dusen coefficients of a specific RTD can be programmed to improve system accuracy
  • Self-diagnostics for loss of input, ambient temperature, EEPROM and CPU faults
  • Manual output function for loop checks
  • Failure output up-scale at 110 percent by default; down-scale at minus 5 percent with /C1 or /C2; the C2 and C3 NAMUR NE43 options limit the signal to 3.8 to 20.5 mA
  • Lightning protector /A rated for 6000 A (1 x 40 microseconds) and 1000 A repeated 100 times
  • Optional LCD with digital reading and bargraph; vibration effect limited to a 0.21 mm displacement from 10 to 60 Hz and 3 G from 60 to 2000 Hz

Using the YTA310 in oil, gas and utilities

Pipeline operators and gas plants in Oman use the YTA310 where a point needs HART diagnostics and a long-term drift specification, since RTD stability is within 0.1 percent of reading or 0.1 degrees C over two years. Utilities fit it to boiler feedwater and turbine casing points. Because the case is rated IP66/IP67, it can be installed outdoors at wellheads and tank farms in Gulf conditions; the maximum ambient is 85 degrees C, or 80 degrees C if the indicator is fitted.

How YTA310 differs from YTA110 and YTA320

Relative to the YTA110, the YTA310 sits in the GS 01C50B02 datasheet, which also covers the YTA320 and gives an update time of 0.5 second for one sensor and 0.8 second for two. The YTA320 adds a second sensor input, average and differential outputs, sensor backup and the FOUNDATION Fieldbus option. When only one sensor is wired, the YTA310 is the simpler model and avoids the extra parameters of the dual unit.

Wiring and commissioning of YTA310

Choose connection code 0, 2, 3 or 4 for G1/2, 1/2 NPT, Pg 13.5 or M20 female entries. Keep RTD lead resistance under 10 ohm per wire. Set the sensor type, range, burnout direction and tag by BT200 or HART communicator, and confirm the loop with a 12 mA manual output. The HART minimum voltage of 16.4 V DC at the terminals must be respected when the loop resistor and barrier are sized.

YTA310 option codes and enclosure choices

The suffix and option list of the YTA310 is long, so it helps to know what each group changes. Painting options include epoxy resin coating (X1), black (P1), jade green (P2) or metallic silver (P7) amplifier covers and a red finish for both covers (PR). Option HC turns the nameplate, tag plate and screws on the amplifier housing into SUS316 stainless steel, while E1 changes the housing itself to SCS14A cast stainless. D2 sets the display and calibration unit to degrees F or degrees R. N4 adds a wired stainless tag plate. Hazardous-area approvals are separate option codes for IECEx, CSA, FM and others, each with its own ambient limit, so state the approval at order time. Mounting brackets come in SUS304 (codes B and D) or SUS316 (J and K) for horizontal or vertical 2-inch pipes.

Thermocouple and RTD coverage of YTA310

Thermocouple ranges are B 100 to 1820, E -200 to 1000, J -200 to 1200, K -200 to 1372, N -200 to 1300, R and S -50 to 1768, T -200 to 400, W3 and W5 0 to 2300, L -200 to 900 and U -200 to 600 degrees C. RTD choices are Pt100, Pt200 and Pt500 to 850 degrees C, JPt100 to 500 degrees C and Cu to 150 degrees C. Recommended minimum spans are 25 degrees C for thermocouples and 10 degrees C for RTDs, which sets how tightly a range can be re-ranged in service.

Check wiring terminals for moisture after humid summer periods, keep the covers tight and confirm calibration annually against a dry-block. A YTA310 replacement loaded with the stored configuration can be swapped quickly.

Ordering YTA310

Give the model YTA310, the output code, connection code, indicator and bracket codes and any approval or option codes with quantity. Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer.

Documents and quotation

The manufacturer datasheet GS 01C50B02-00EN covering the YTA310 is attached. Send an RFQ stating part number YTA310, the full code and the quantity.

Other Models in the Yokogawa Yta Field Range

Seven Star LLC supplies models from the Yokogawa Yta Field range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.

ModelCategory
Yokogawa YTA110 Temperature Transmitter SIL 2Temperature TransmittersView details
Yokogawa YTA310 Single-Input Temperature TransmitterTemperature TransmittersYou are viewing this model
Datasheets & Documents
  • PDF Datasheet Yokogawa YTA310 Single-Input Temperature Transmitter Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Yokogawa is the Japanese measurement and industrial automation company behind the DPharp EJA and EJX pressure transmitters, CENTUM distributed control systems and ProSafe-RS safety systems. Tamisuke Yokogawa set it up in 1915 as an electric meter research institute in Shibuya, Tokyo. It began selling electric meters in 1917, was incorporated as Yokogawa Electric Works in 1920, and has been Yokogawa Electric Corporation since 1986. In 1975 it released CENTUM, which it describes as the world's first distributed control system.

Seven Star LLC is an independent supplier in Muscat. We list Yokogawa transmitters, flowmeters, analysers, controllers, recorders and CENTUM and ProSafe spare modules, and supply them to oil and gas, power, water and process plants in Oman and the GCC by model code.

Yokogawa at a glance

Established 1915, Shibuya, Tokyo, by Tamisuke Yokogawa with Ichiro Yokogawa and Shin Aoki
Milestones Electric meters from 1917; incorporated in 1920; CENTUM DCS in 1975; renamed Yokogawa Electric Corporation in 1986
Headquarters Tokyo, Japan
Product families Pressure and temperature transmitters, magnetic, vortex and Coriolis flowmeters, liquid and gas analysers, controllers and recorders, CENTUM VP DCS, ProSafe-RS safety systems and STARDOM controllers

Yokogawa products we list

Family Models Typical use
Pressure and DP transmitters EJA110E, EJA430E, EJA530E, EJX110A, EJX530A, EJX910A multivariable Pressure, level and DP flow measurement in process plants
Temperature transmitters YTA610, YTA710 RTD and thermocouple signals converted to 4-20 mA, HART or fieldbus
Flowmeters ADMAG AXG, ADMAG AXW, ROTAMASS TI, VY vortex, VF701 Water and conductive liquids, mass flow of hydrocarbons, steam and gas
Analysers TDLS8000 laser gas analyser, FLXA402, FLXA21 DO, E7042UD zirconia cell, K9471DS cover Combustion O2, in-situ gas analysis, pH, conductivity and dissolved oxygen
Controllers and recorders UT35A/UT32A, UT32A-D, UT55A/UT52A, UT75A, UP35A/UP32A, UP55A, YS1700, GX10/GX20, GP10/GP20, GM10, µR20000 Single-loop control, furnace programs and data recording
Control and safety systems CP451-10 processor, ADV161, ADV551, node interface unit, ProSafe-RS, ProSafe-RS Lite, ProSafe-SLS, FCN-500 Spares and expansions for CENTUM DCS, safety systems and STARDOM
Battery packs S9129FA, S9185FA, S9548FA Memory backup batteries for DCS and controller modules

The EJA and EJX transmitters use Yokogawa's DPharp sensor: two silicon resonators whose vibration frequency changes with pressure, so the output is a digital frequency rather than an analogue strain signal. The model code describes the instrument completely. In EJA110E, "110" is a differential pressure transmitter, and the letters after the dash set the span, wetted materials, process connection, electrical output and approvals. When you replace a transmitter, sending the full code from the nameplate avoids a mismatch in range or materials.

Where Yokogawa products are used

In Oman and the GCC, Yokogawa transmitters, analysers and control systems are found in oil and gas production, gas processing and LNG, refineries and petrochemical plants, power and desalination stations, and water treatment. Many plants still run CENTUM CS 3000 and CENTUM VP systems that need spare I/O modules and batteries.

Ordering Yokogawa products: what we need

Send the full model and suffix code from the nameplate (for example EJA110E-JMS5J-912NN/KF22) and the style code. For DCS modules, add the module type and system version. For analysers, send the sensor type, measuring range and process conditions.

Facts checked against Yokogawa's corporate history, October 2026.

FAQ

Questions about the Yokogawa YTA310 Single-Input Temperature Transmitter

Answers from our engineering team on installation, compatibility and integration.

01What is the difference between the Yokogawa YTA310 and the YTA320?
The YTA310 has one sensor input and BRAIN or HART output. The YTA320 has two sensor inputs, sensor backup, average and differential measurement, and an optional FOUNDATION Fieldbus code.
02Is the YTA310 SIL certified?
Yes, in the standard BRAIN or HART configuration it is certified by TUV NORD CERT for SIL 2 capability with a single transmitter and SIL 3 with dual transmitters under IEC 61508.
03How fast does the YTA310 update its output?
The update time is approximately 0.5 second for a single sensor and the turn-on time is about 5 seconds; damping can be set from 0 to 99 seconds.
04What are the stability figures of the YTA310?
For RTDs stability is plus/minus 0.1 percent of reading or 0.1 degrees C per two years, and for thermocouples the same values per year; five-year RTD stability is 0.2 percent or 0.2 degrees C.
05Can the YTA310 use a lightning protector?
Yes, option /A adds a lightning protector with a maximum 6000 A surge rating and limits the supply to 10.5 to 32 V DC.