Endress+Hauser CPF81E Memosens Compact pH Sensor

Description

Endress+Hauser CPF81E: robust compact pH sensor for mining and wastewater

The Endress+Hauser CPF81E is a plastic-bodied Memosens 2.0 compact pH sensor. Instead of a long glass shaft, it has a protective PPS housing with 3/4 inch NPT threads at both ends, so it can be screwed straight into a pipe tee, a flow cell or a submerged mounting pipe. The datasheet places it in flotation, leaching, neutralization, and inlet and outlet monitoring, in mining, industrial water and wastewater treatment.

For Gulf customers this makes the CPF81E a candidate for mineral-processing sites, quarry wash water, and effluent outfalls where glass shafts get broken.

Design highlights of CPF81E

  • KNO3 electrolyte bridge for better protection against sulfide and cyanide poisons
  • Optional flat membrane for improved wear resistance
  • 3/4 inch NPT connection at top and bottom
  • Double-chamber reference system
  • Protection guard against mechanical damage
  • Application L covers pH 0 to 14 and 0 to 110 °C
  • NTC 30K temperature compensation

CPF81E key data

Parameter Value
Application L pH 0 to 14, temperature 0 to 110 °C
Application N pH 0 to 11, temperature 0 to 80 °C
Process pressure 0.8 to 11 bar absolute
Minimum conductivity 50 microsiemens per cm at atmospheric pressure, no flow
Reference Double chamber system
Housing and shaft material PPS
Process connection NPT 3/4 inch
Weight 0.12 to 0.15 kg, depending on version, without cable
Temperature sensor NTC 30K
Head pressure resistance 16 bar relative
Degree of protection IP 68
Storage 0 to 50 °C

How CPF81E differs from glass-shaft Memosens sensors

A CPS11E or CPS71E is a 12 mm glass tube with Pg 13.5 thread and needs an assembly. The CPF81E is a threaded plastic body that takes damage better, weighs 0.12 to 0.15 kg instead of 40 to 100 g, and is limited to 11 bar absolute instead of 14 to 17. Its range of 0 to 110 °C is lower than the 135 to 140 °C of the chemical glass sensors. It is also not a pressurized-reference or sterilizable design, so it sits in the heavy-duty water segment of the Memosens range.

Typical installations

Mount the CPF81E in a pipe tee, in a flow cell via the NPT thread, or at the end of a pipe in a tank. In neutralization tanks and flotation circuits the flat-membrane version is preferred where abrasive slurry would otherwise wear the dome of the glass.

Compatibility and ordering

A Memosens cable such as CYK10 or CYK20 connects the CPF81E to a Liquiline transmitter. When requesting a quote, state the application (L or N), whether you need the flat membrane, and the cable length of the transmitter system.

Installation hints

Use the NPT thread with suitable sealing, and do not exceed the process pressure and temperature limits. Observe the 50 microsiemens per cm minimum conductivity so the double-chamber reference can operate, and keep flow low and stable during calibration.

Maintenance and replacement

The CPF81E stores calibration and load data like all Memosens sensors, so a pre-calibrated spare can be fitted in minutes. Inspect the protection guard and the membrane for wear and replace the sensor when the slope falls outside your acceptance limits.

pH measurement principle in this sensor

The glass membrane of the CPF81E generates an electrochemical potential in proportion to the pH of the fluid, produced by H+ ions that accumulate on the outer membrane layer. An integrated Ag/AgCl reference serves as the counter electrode, and the voltage is converted to pH by the Nernst equation. The digital Memosens electronics send the result to the transmitter inductively, avoiding corrosion and moisture problems at the plug.

Poison protection explained

Sulfide and cyanide ions can ruin a standard reference in a few weeks. The KNO3 electrolyte bridge and double chamber of the CPF81E slow that attack, which explains why the datasheet highlights it for flotation and leaching duty. It is still good practice to schedule regular calibration checks in such liquids.

Ordering the right variant of the CPF81E

The two applications are not interchangeable. Application L covers the full pH 0 to 14 window and reaches 110 degrees Celsius, while application N covers pH 0 to 11 and stops at 80 degrees Celsius. Choose L for hot flotation circuits or strongly alkaline leach liquors and N for cooler inlet monitoring where the pH is moderate. The membrane shape is the next choice: the standard dome or the flat membrane for abrasive slurry.

Because the body is plastic, check chemical compatibility of PPS with your fluid before ordering. The datasheet lists PPS as the housing material, and PPS is widely used in water treatment, but strong oxidizers at high temperature need a check against the manufacturer’s resistance data.

Installation geometry

With NPT 3/4 inch threads at both ends, the CPF81E can be installed in a pipe from either side, which makes it easy to retrofit. A top thread can hold a conduit or protection tube for the Memosens cable while the bottom thread holds the sensor in the line. Use PTFE tape or an equivalent sealant that suits your fluid, avoid over-tightening plastic threads, and orient the body so the membrane is always wetted.

Keep the cable gland and the Memosens plug-in head dry during installation. Although the sensor has IP 68 protection, a clean and dry connector helps the first connection.

For spares planning, record the application type and membrane shape of each installed CPF81E so that the replacement matches exactly, and keep one pre-calibrated unit on the shelf for fast exchange.

Documents and quotation

The datasheet for the CPF81E is attached. Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer. Send an RFQ giving the part number CPF81E, application and quantity.

Other Models in the Memosens Ph Range

Seven Star LLC supplies models from the Memosens Ph range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.

ModelCategory
Endress+Hauser CPF81E Memosens Compact pH SensorLiquid Analysis SensorsYou are viewing this model
Endress+Hauser CPS96E Memosens pH/ORP Sensor, Open JunctionLiquid Analysis SensorsView details
Datasheets & Documents
  • PDF Datasheet Endress+Hauser CPF81E Memosens Compact pH Sensor Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Endress+Hauser makes process measurement instruments for flow, level, pressure, temperature and liquid analysis. Georg H. Endress and Ludwig Hauser started the business in Lörrach, Germany, on 1 February 1953, first selling level meters, and the Endress+Hauser name has been used since 1957. The group is headquartered in Reinach, Switzerland, and has been wholly owned by the Endress family since 1975.

Seven Star LLC is an independent supplier listing Endress+Hauser flowmeters, level and pressure transmitters, temperature assemblies and liquid analysis products for buyers in Oman and the GCC. Instruments are matched from the full order code on the nameplate.

Endress+Hauser company profile

Origin Founded in Lörrach, Germany, on 1 February 1953 by the Swiss engineer Georg H. Endress and Ludwig Hauser
Headquarters Endress+Hauser AG, Reinach (Basel-Landschaft), Switzerland
Ownership Family owned; the Endress family has been the sole shareholder since 1975, and a family charter commits the company to stay family owned
Size About 16,500 employees and sales of about EUR 3.7 billion (2023, as reported by the company)
Product families Proline flowmeters, Micropilot and Levelflex radar, Liquiphant switches, Cerabar and Deltabar pressure, iTEMP and iTHERM temperature, Liquiline and Memosens liquid analysis

Milestones

  • 1953: Georg H. Endress and Ludwig Hauser start the business in Lörrach on 1 February, first selling level meters.
  • 1957: The Endress+Hauser name comes into use.
  • 1975: The Endress family becomes sole shareholder.
  • 2023: About 16,500 employees and sales of about EUR 3.7 billion, as reported by the company.

Endress+Hauser product lines

Line Measuring principle Typical use
Promass F 300, Promass 80F, 83F/84F Coriolis mass flow Direct mass flow and density of liquids and gases, hydrocarbon transfer and chemical dosing
Promag P 300, P 500, H 300, W 400, W 10, 50P Electromagnetic flow Water, wastewater, slurries and other conductive liquids
Prowirl F 200, 72F Vortex flow Steam, gas and low-viscosity liquids
Prosonic Flow 92F, t-mass 65F Ultrasonic and thermal mass flow Water lines; compressed air and industrial gases
Micropilot FMR62, FMR67, FMR60, FMR10B, FMR20, FMR50, FMR52 Free-space radar level, 80 GHz and 26 GHz Tanks, process vessels, silos and open water channels
Levelflex FMP51, FMP50, FMP54, FMP55 Guided wave radar level Level and interface in vessels, separators and bypass chambers
Liquiphant FTL51, Prosonic S FMU90 Vibronic point level; ultrasonic level High and low level alarms, pump protection, open channel flow
Cerabar PMC71B, PMP51B, PMP71, Deltabar PMD78B, PMD75 Pressure and differential pressure Process pressure, hydrostatic level and DP flow
iTEMP TMT82, TMT162, iTHERM TM101, TR66, TR10 Temperature transmitters and thermometers RTD and thermocouple points with thermowells
Liquiline CM444, CM442, Memosens CPS11E, Stamolys reagents Liquid analysis pH, ORP, conductivity, dissolved oxygen and nutrient analysis in water treatment and process

Where Endress+Hauser instruments are used

In Oman and the GCC, Endress+Hauser instruments are common in oil and gas production and processing, refineries and petrochemicals, desalination and water networks, wastewater treatment, power plants and food and beverage plants. Coriolis meters measure hydrocarbon and chemical flows, electromagnetic meters cover water and wastewater, radar transmitters handle tank and silo level, and Liquiline analysers monitor water quality. Proline 300 and 500 flowmeters include Heartbeat Technology for in-situ verification, which reduces the need to remove a meter for checks.

Ordering Endress+Hauser instruments: what we need

The order code on the nameplate (for example PMP71-1QA1S11RDADU) defines every option: measuring range, process connection, materials, output, approvals and display. Send it with the serial number; Endress+Hauser can retrieve the original configuration from the serial number. For a new instrument add the medium, operating and design pressure and temperature, range, line size or tank geometry, process connection, wetted materials, output or protocol (4-20 mA HART, PROFIBUS, FOUNDATION Fieldbus, Modbus) and the hazardous-area approval required. Older models such as Promass 80/83 or Micropilot M are replaced by current Proline or 80 GHz Micropilot devices.

Facts checked against Endress+Hauser company history and product information, October 2026.

FAQ

Questions about the Endress+Hauser CPF81E Memosens Compact pH Sensor

Answers from our engineering team on installation, compatibility and integration.

01Does the CPF81E have a glass shaft?
No, the housing and shaft are PPS plastic with a protection guard, which suits rough mining and wastewater service.
02What connection does the CPF81E use?
It has NPT 3/4 inch threads at top and bottom for installation in pipes or flow cells.
03What is the process range of the CPF81E?
Application L is pH 0 to 14 and 0 to 110 °C; application N is pH 0 to 11 and 0 to 80 °C; pressure is 0.8 to 11 bar absolute.
04Is there a flat membrane option for CPF81E?
Yes, a flat membrane is optional and improves wear resistance in abrasive media.
05How heavy is the CPF81E?
The weight is 0.12 to 0.15 kg depending on the version, excluding the cable.