Endress+Hauser CA80AM Liquiline System Ammonium Analyzer

Description

CA80AM: wet-chemical ammonium in almost continuous mode

Endress+Hauser CA80AM is the ammonium analyzer of the Liquiline System CA80 platform. It draws a filtered sample into a reaction cuvette, meters two reagents with a dosing dispenser, lets the indophenol dye form and reads absorption with a multispectral LED photometer that also takes a reference measurement. The result is NH4-N, NH4 plus or equivalent NH3 in mg/l, produced roughly every 9 minutes. Municipal and industrial wastewater plants use it to follow aeration basins, outlet quality and industrial treatment steps.

Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer. The datasheet used is TI01111C, edition 10.26-00.

CA80AM ranges and performance

Parameter Value
Method Indophenol blue per ISO 7150-1 (also GB 7481-87, DIN 38406-5)
Ranges A1/S4 0 to 20; A2/S5 0 to 50; A3/S6 0 to 100 mg/l NH4-N
Dilution option On range A2 (S5): factors 1, 2, 4, 10, 20 giving up to 0 to 1000 mg/l
Repeatability A1 to A3 2 percent of display plus 0.05 mg/l NH4-N
Repeatability S4 / S5 / S6 0.04 mg/l or 2 percent / 0.05 mg/l or 2 percent then 3 percent / 0.3 mg/l or 3 percent
Measuring interval About 9 min continuous; adjustable from 10 min
Reagent use About 70 ul per reagent per measurement, about 250 ml per reagent per month at 15 min
Supply 100 to 120 / 200 to 240 V AC or 24 V DC, 50 or 60 Hz; 130 VA plus 660 VA per hose heater, up to 1450 VA with cooling
Ambient 5 to 40 degrees C; outdoor version minus 20 to 40 degrees C
Sample 4 to 40 degrees C, under 50 NTU, unpressurised
Protection IP55, Type 3R
Weight Cabinet 42 kg with cooling, 39.5 kg without; open installation 34 or 31.5 kg

Interference limits listed for ammonium

The datasheet checked individual ions and saw no cross-sensitivity at 1000 mg/l sulfate, 500 mg/l sodium and potassium, 250 mg/l nitrate-nitrogen and phosphate, 50 mg/l chromium(III) and zinc, 30 mg/l nitrite-nitrogen and 2 mg/l silver. A combined effect was not studied, so unusual industrial streams should be tested.

How CA80AM differs from the other CA80 analyzers

Ammonium is the only CA80 variant in this group with a cooled version aimed at longer reagent life together with several range orders, A1 to A3 and S4 to S6. CA80AL measures aluminium in micrograms per litre, CA80FE iron, CA80HA total hardness and CA80SI silica in ultrapure water. CA80AM is also unusual in offering a continuous 9 minute cycle, and in having the reagent consumption stated per month for a 15 minute interval, which is the figure to use when budgeting reagent supply.

Sample preparation for CA80AM

  • CAT810 microfiltration with a 50 um sieve and backflushing for plant outlets.
  • CAT820 ceramic membrane, 0.1 or 0.4 um cartridge or 0.04 um plate filter, with compressed air backflushing for aeration basins and surface water.
  • CAT860 ceramic candle with automatic backflush using cleaning solution and compressed air, for plant inlets.
  • CAT820 and CAT860 communicate by Memosens and are controlled from the analyzer; installation is simple using a Flexdip CYH112 holder.

Controller, communication and layout of CA80AM

The analyzer includes a controller with up to two measuring channels, which means a second ammonium channel (for example a second sampling point) can be handled by the same cabinet. Fieldbus options listed are PROFINET, PROFIBUS DP, Modbus TCP, Modbus RS485 and EtherNet/IP with a web server. The cabinet version, the open-installation version and an analyzer stand are offered, and the weights run from 31.5 kg for the open version without cooling to 42 kg for the cabinet with cooling. A cooled version slows reagent decay so replacement intervals can be longer, which matters in hot climates. Hose heaters at 660 VA each protect sample hoses in cold locations and must be added to the power budget; in the Gulf they are rarely needed, while an air-conditioned shelter often is.

CA80AM in Oman treatment plants

Nitrification control in aeration basins is the typical task: the analyzer shows how much ammonium remains so that blowers can be tuned. Plant outlets in the Gulf use it for discharge compliance and as an early warning of upset. Up to four additional Memosens sensors (for example oxygen or pH) can be plugged into the same controller, turning it into a measuring station.

Ordering notes for CA80AM

Reagents, cleaner and standard solution are ordered separately, and sample conditioning is optional. Please tell us the range code (A1 to A3 or S4 to S6), dilution, cooling, outdoor housing, voltage, fieldbus and sample preparation when you request a quote.

CA80AM questions answered briefly

Which method does Endress+Hauser CA80AM use for ammonium? The indophenol blue method with sodium dichloroisocyanurate and sodium salicylate in alkaline solution, defined in ISO 7150-1, measured with a multispectral photometer.

What measuring ranges can CA80AM be ordered with? 0 to 20, 0 to 50 or 0 to 100 mg/l NH4-N in the A1, A2 and A3 versions, with S4, S5 and S6 variants, and an optional dilution function on the 0 to 50 range reaching 1000 mg/l.

How often does CA80AM measure? Continuously at about 9 minutes per cycle, or adjustable with a minimum of 10 minutes; dilution adds about 3 minutes.

What sample must CA80AM receive? Aqueous, homogenised, low-solids sample under 50 NTU turbidity, at 4 to 40 degrees C, unpressurised.

Is there an outdoor CA80AM? Yes, a cabinet outdoor version operates from minus 20 to 40 degrees C, while other housings are rated 5 to 40 degrees C.

Documents and quotation

The datasheet for CA80AM is attached to this page. Send an RFQ with CA80AM and the configuration, plus the quantity of analyzers and reagent sets required.

Other Models in the Liquiline Ca80 Range

Seven Star LLC supplies models from the Liquiline Ca80 range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.

ModelCategory
Endress+Hauser CA80FE Liquiline Dissolved Iron AnalyzerLiquid Analytical SystemsView details
Endress+Hauser CA80SI Liquiline Silica Analyzer for Boiler WaterLiquid Analytical SystemsView details
Endress+Hauser CA80HA Liquiline Water Hardness AnalyzerLiquid Analytical SystemsView details
Endress+Hauser CA80AM Liquiline System Ammonium AnalyzerLiquid Analytical SystemsYou are viewing this model
Endress+Hauser CA80AL Liquiline Aluminum Analyzer for Water TreatmentLiquid Analytical SystemsView details
Datasheets & Documents
  • PDF Datasheet Endress+Hauser CA80AM Liquiline System Ammonium Analyzer Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Endress+Hauser makes process measurement instruments for flow, level, pressure, temperature and liquid analysis. Georg H. Endress and Ludwig Hauser started the business in Lörrach, Germany, on 1 February 1953, first selling level meters, and the Endress+Hauser name has been used since 1957. The group is headquartered in Reinach, Switzerland, and has been wholly owned by the Endress family since 1975.

Seven Star LLC is an independent supplier listing Endress+Hauser flowmeters, level and pressure transmitters, temperature assemblies and liquid analysis products for buyers in Oman and the GCC. Instruments are matched from the full order code on the nameplate.

Endress+Hauser company profile

Origin Founded in Lörrach, Germany, on 1 February 1953 by the Swiss engineer Georg H. Endress and Ludwig Hauser
Headquarters Endress+Hauser AG, Reinach (Basel-Landschaft), Switzerland
Ownership Family owned; the Endress family has been the sole shareholder since 1975, and a family charter commits the company to stay family owned
Size About 16,500 employees and sales of about EUR 3.7 billion (2023, as reported by the company)
Product families Proline flowmeters, Micropilot and Levelflex radar, Liquiphant switches, Cerabar and Deltabar pressure, iTEMP and iTHERM temperature, Liquiline and Memosens liquid analysis

Milestones

  • 1953: Georg H. Endress and Ludwig Hauser start the business in Lörrach on 1 February, first selling level meters.
  • 1957: The Endress+Hauser name comes into use.
  • 1975: The Endress family becomes sole shareholder.
  • 2023: About 16,500 employees and sales of about EUR 3.7 billion, as reported by the company.

Endress+Hauser product lines

Line Measuring principle Typical use
Promass F 300, Promass 80F, 83F/84F Coriolis mass flow Direct mass flow and density of liquids and gases, hydrocarbon transfer and chemical dosing
Promag P 300, P 500, H 300, W 400, W 10, 50P Electromagnetic flow Water, wastewater, slurries and other conductive liquids
Prowirl F 200, 72F Vortex flow Steam, gas and low-viscosity liquids
Prosonic Flow 92F, t-mass 65F Ultrasonic and thermal mass flow Water lines; compressed air and industrial gases
Micropilot FMR62, FMR67, FMR60, FMR10B, FMR20, FMR50, FMR52 Free-space radar level, 80 GHz and 26 GHz Tanks, process vessels, silos and open water channels
Levelflex FMP51, FMP50, FMP54, FMP55 Guided wave radar level Level and interface in vessels, separators and bypass chambers
Liquiphant FTL51, Prosonic S FMU90 Vibronic point level; ultrasonic level High and low level alarms, pump protection, open channel flow
Cerabar PMC71B, PMP51B, PMP71, Deltabar PMD78B, PMD75 Pressure and differential pressure Process pressure, hydrostatic level and DP flow
iTEMP TMT82, TMT162, iTHERM TM101, TR66, TR10 Temperature transmitters and thermometers RTD and thermocouple points with thermowells
Liquiline CM444, CM442, Memosens CPS11E, Stamolys reagents Liquid analysis pH, ORP, conductivity, dissolved oxygen and nutrient analysis in water treatment and process

Where Endress+Hauser instruments are used

In Oman and the GCC, Endress+Hauser instruments are common in oil and gas production and processing, refineries and petrochemicals, desalination and water networks, wastewater treatment, power plants and food and beverage plants. Coriolis meters measure hydrocarbon and chemical flows, electromagnetic meters cover water and wastewater, radar transmitters handle tank and silo level, and Liquiline analysers monitor water quality. Proline 300 and 500 flowmeters include Heartbeat Technology for in-situ verification, which reduces the need to remove a meter for checks.

Ordering Endress+Hauser instruments: what we need

The order code on the nameplate (for example PMP71-1QA1S11RDADU) defines every option: measuring range, process connection, materials, output, approvals and display. Send it with the serial number; Endress+Hauser can retrieve the original configuration from the serial number. For a new instrument add the medium, operating and design pressure and temperature, range, line size or tank geometry, process connection, wetted materials, output or protocol (4-20 mA HART, PROFIBUS, FOUNDATION Fieldbus, Modbus) and the hazardous-area approval required. Older models such as Promass 80/83 or Micropilot M are replaced by current Proline or 80 GHz Micropilot devices.

Facts checked against Endress+Hauser company history and product information, October 2026.

FAQ

Questions about the Endress+Hauser CA80AM Liquiline System Ammonium

Answers from our engineering team on installation, compatibility and integration.

01Which method does Endress+Hauser CA80AM use for ammonium?
The indophenol blue method with sodium dichloroisocyanurate and sodium salicylate in alkaline solution, defined in ISO 7150-1, measured with a multispectral photometer.
02What measuring ranges can CA80AM be ordered with?
0 to 20, 0 to 50 or 0 to 100 mg/l NH4-N in the A1, A2 and A3 versions, with S4, S5 and S6 variants, and an optional dilution function on the 0 to 50 range reaching 1000 mg/l.
03How often does CA80AM measure?
Continuously at about 9 minutes per cycle, or adjustable with a minimum of 10 minutes; dilution adds about 3 minutes.
04What sample must CA80AM receive?
Aqueous, homogenised, low-solids sample under 50 NTU turbidity, at 4 to 40 degrees C, unpressurised.
05Is there an outdoor CA80AM?
Yes, a cabinet outdoor version operates from minus 20 to 40 degrees C, while other housings are rated 5 to 40 degrees C.