Endress+Hauser COS22E Memosens Amperometric Oxygen Sensor

Description

Endress+Hauser COS22E Memosens amperometric oxygen sensor overview

Endress+Hauser COS22E is a membrane-covered amperometric oxygen sensor in stainless steel 1.4435 (AISI 316L), engineered to stay stable through repeated steam sterilization. It targets pharmaceutical and biotech fermenters, enzyme production, culture control, and also water treatment, boiler feedwater, WFI and residual-oxygen monitoring. The datasheet describes two builds: a standard sensor for general use and a trace sensor, which is CO2 compatible and aimed at power station and beverage duties.

The Memosens 2.0 head provides permanent self-monitoring and stores calibration. Temperature is read by an NTC 22 kilohm element. Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer.

COS22E technical specifications

Parameter Value
Standard sensor range (code 22) 0 to 60 mg/l; 0 to 600 % SAT; 0 to 1200 hPa; 0 to 100 Vol%
Optimum range (standard) 0 to 20 mg/l; 0 to 200 % SAT; 0 to 400 hPa; 0 to 40 Vol%
Trace sensor range (code 12) 0 to 10 mg/l; 0 to 120 % SAT; 0 to 250 hPa; 0 to 25 Vol%
Optimum range (trace) 0 to 2 mg/l; 0 to 20 % SAT; 0 to 40 hPa; 0 to 4 Vol%
Signal current in air 40 to 100 nA (standard); 210 to 451 nA (trace)
Zero current under 0.1 % (standard); under 0.03 % (trace) of air signal
Measurement error 1 % of value or 10 ppb (standard); 1 % or 1 ppb (trace)
Limit of detection 5 ppb / 1 ppb
Response t90 under 30 s; t98 under 60 s
Long-term drift under 4 % per month; 1 % or less per month below 4 Vol% O2
Process temperature minus 5 to 135 C
Process pressure ambient to 12 bar absolute
Minimum flow 0.02 m/s (standard); 0.1 m/s (trace)
Ambient / storage minus 25 to 70 C / minus 25 to 50 C
Ingress protection IP68 (2 m, 21 C, 24 h), IP69
Surface roughness Ra under 0.38 micrometre
Temperature sensor NTC 22 kilohm
Weight 0.2 to 0.7 kg

How COS22E differs from other oxygen sensors

COS22E, amperometric, needs a minimum flow and consumes oxygen at the cathode, but offers trace-level detection of 1 ppb in the trace version and a high-temperature envelope of 135 C. The optical COS81D covers 0.004 to 26 mg/l at 25 C with t90 under 10 seconds and needs no minimum flow, whereas COS22E trades speed (t90 under 30 s) for trace sensitivity. COS61D is optical too but made for wastewater with a limit of 60 C and 10 bar.

Typical COS22E duties in Oman and the GCC

Power and desalination plants in the Gulf watch dissolved oxygen in boiler feedwater and condensate, where parts-per-billion levels decide corrosion rates; the trace COS22E is made for that. Pharmaceutical and beverage plants use the standard sensor for fermenters and CO2-bearing drinks, and utilities use it for deaerator control. Hazardous-area approvals (ATEX, IECEx, NEPSI, Korea Ex, Japan Ex, UK Ex, INMETRO, and CSA C/US up to Class I Division 1) allow use in refinery-adjacent water systems.

Selection notes for COS22E

Select standard or trace by the oxygen level you expect, and select the process connection material: stainless steel, titanium or Hastelloy. Check the minimum flow at your line: 0.02 m/s or 0.1 m/s. Use cable CYK10 and a Memosens transmitter. The flow assembly CYA21 is offered for sensors with a 12 mm diameter.

Installing and commissioning COS22E

Fit with Pg 13.5 thread at a maximum 3 Nm. Never let the sensor dry out in storage; the datasheet warns of danger of drying and calls for storage with a protective cap. After the first connection, allow the polarisation period before you calibrate, and calibrate against air or air-saturated water.

How the COS22E amperometric cell works and what that means for operation

An amperometric oxygen cell reduces dissolved oxygen at a cathode behind a gas-permeable membrane, and the resulting current, measured in nanoamps, is proportional to oxygen partial pressure. The datasheet quotes a signal current in air of 40 to 100 nA for the standard sensor and 210 to 451 nA for the trace sensor, with a zero current under 0.1 percent and under 0.03 percent of the air signal respectively. Because the cell consumes oxygen, the datasheet asks for a minimum flow of 0.02 m/s for the standard sensor and 0.1 m/s for the trace sensor.

The trace design is the reason to choose code 12: its limit of detection is 1 ppb, its limit of quantification 3 ppb and its repeatability 1 ppb, against 5, 15 and 5 ppb for the standard sensor. Measurement errors apply in the optimum range only, not over the entire span, so size the range to the real oxygen level. The sensor can show raw current in nA, which helps diagnose cell condition during troubleshooting.

Quick answers about COS22E

Does COS22E suit CO2-rich beverages? The trace version is described as CO2 compatible. Can it be autoclaved? Yes, and it is sterilizable. Is it hygienic? It is EHEDG certified, complies with ASME-BPE requirements, offers IP69 protection and a surface roughness under 0.38 micrometre. Which materials can it be ordered in? Stainless steel 1.4435 (AISI 316L) throughout, with process-connection options in stainless steel, titanium or Hastelloy. What is the maximum torque? 3 Nm on the Pg 13.5 thread. What is the cable? A Memosens CYK10, and the sensor stores calibration so it can be pre-calibrated in the lab.

Maintenance of COS22E

The consumables are the membrane cap and electrolyte; replace them according to the load history stored in the head. Steam sterilization and autoclaving are permitted.

Documents and quotation

The Technical Information datasheet for COS22E is attached to this page as a PDF. To request a quotation, send your RFQ with the exact COS22E order code and the quantity you need, and we will reply with availability and lead time.

Delivery to Sites in Oman and the GCC

Delivery of the Endress+Hauser COS22E is arranged to suit the site, whether that is a plant in Sohar or Duqm, an oilfield location in the interior, or a project elsewhere in the GCC. Tell us the delivery point, any site access or gate-pass requirements, and the person who will receive the goods. Goods are packed to protect sensitive liquid analysis sensors equipment during road or air transport, and shipping details are shared once the order is dispatched so that your stores team can plan receipt and inspection.

Other Models in the Memosens Oxygen Range

Seven Star LLC supplies models from the Memosens Oxygen range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.

ModelCategory
Endress+Hauser COS61D Oxymax Optical Oxygen SensorLiquid Analysis SensorsView details
Endress+Hauser COS22E Memosens Amperometric Oxygen SensorLiquid Analysis SensorsYou are viewing this model
Endress+Hauser COS81D Memosens Hygienic Optical Oxygen SensorLiquid Analysis SensorsView details
Datasheets & Documents
  • PDF Datasheet Endress+Hauser COS22E Memosens Amperometric Oxygen Sensor Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Endress+Hauser makes process measurement instruments for flow, level, pressure, temperature and liquid analysis. Georg H. Endress and Ludwig Hauser started the business in Lörrach, Germany, on 1 February 1953, first selling level meters, and the Endress+Hauser name has been used since 1957. The group is headquartered in Reinach, Switzerland, and has been wholly owned by the Endress family since 1975.

Seven Star LLC is an independent supplier listing Endress+Hauser flowmeters, level and pressure transmitters, temperature assemblies and liquid analysis products for buyers in Oman and the GCC. Instruments are matched from the full order code on the nameplate.

Endress+Hauser company profile

Origin Founded in Lörrach, Germany, on 1 February 1953 by the Swiss engineer Georg H. Endress and Ludwig Hauser
Headquarters Endress+Hauser AG, Reinach (Basel-Landschaft), Switzerland
Ownership Family owned; the Endress family has been the sole shareholder since 1975, and a family charter commits the company to stay family owned
Size About 16,500 employees and sales of about EUR 3.7 billion (2023, as reported by the company)
Product families Proline flowmeters, Micropilot and Levelflex radar, Liquiphant switches, Cerabar and Deltabar pressure, iTEMP and iTHERM temperature, Liquiline and Memosens liquid analysis

Milestones

  • 1953: Georg H. Endress and Ludwig Hauser start the business in Lörrach on 1 February, first selling level meters.
  • 1957: The Endress+Hauser name comes into use.
  • 1975: The Endress family becomes sole shareholder.
  • 2023: About 16,500 employees and sales of about EUR 3.7 billion, as reported by the company.

Endress+Hauser product lines

Line Measuring principle Typical use
Promass F 300, Promass 80F, 83F/84F Coriolis mass flow Direct mass flow and density of liquids and gases, hydrocarbon transfer and chemical dosing
Promag P 300, P 500, H 300, W 400, W 10, 50P Electromagnetic flow Water, wastewater, slurries and other conductive liquids
Prowirl F 200, 72F Vortex flow Steam, gas and low-viscosity liquids
Prosonic Flow 92F, t-mass 65F Ultrasonic and thermal mass flow Water lines; compressed air and industrial gases
Micropilot FMR62, FMR67, FMR60, FMR10B, FMR20, FMR50, FMR52 Free-space radar level, 80 GHz and 26 GHz Tanks, process vessels, silos and open water channels
Levelflex FMP51, FMP50, FMP54, FMP55 Guided wave radar level Level and interface in vessels, separators and bypass chambers
Liquiphant FTL51, Prosonic S FMU90 Vibronic point level; ultrasonic level High and low level alarms, pump protection, open channel flow
Cerabar PMC71B, PMP51B, PMP71, Deltabar PMD78B, PMD75 Pressure and differential pressure Process pressure, hydrostatic level and DP flow
iTEMP TMT82, TMT162, iTHERM TM101, TR66, TR10 Temperature transmitters and thermometers RTD and thermocouple points with thermowells
Liquiline CM444, CM442, Memosens CPS11E, Stamolys reagents Liquid analysis pH, ORP, conductivity, dissolved oxygen and nutrient analysis in water treatment and process

Where Endress+Hauser instruments are used

In Oman and the GCC, Endress+Hauser instruments are common in oil and gas production and processing, refineries and petrochemicals, desalination and water networks, wastewater treatment, power plants and food and beverage plants. Coriolis meters measure hydrocarbon and chemical flows, electromagnetic meters cover water and wastewater, radar transmitters handle tank and silo level, and Liquiline analysers monitor water quality. Proline 300 and 500 flowmeters include Heartbeat Technology for in-situ verification, which reduces the need to remove a meter for checks.

Ordering Endress+Hauser instruments: what we need

The order code on the nameplate (for example PMP71-1QA1S11RDADU) defines every option: measuring range, process connection, materials, output, approvals and display. Send it with the serial number; Endress+Hauser can retrieve the original configuration from the serial number. For a new instrument add the medium, operating and design pressure and temperature, range, line size or tank geometry, process connection, wetted materials, output or protocol (4-20 mA HART, PROFIBUS, FOUNDATION Fieldbus, Modbus) and the hazardous-area approval required. Older models such as Promass 80/83 or Micropilot M are replaced by current Proline or 80 GHz Micropilot devices.

Facts checked against Endress+Hauser company history and product information, October 2026.

FAQ

Questions about the Endress+Hauser COS22E Memosens Amperometric Oxygen

Answers from our engineering team on installation, compatibility and integration.

01What are the two versions of Endress+Hauser COS22E?
The code 22 is the standard sensor, measuring 0 to 60 mg/l, and code 12 is the trace sensor, measuring 0 to 10 mg/l and detecting down to 1 ppb, including a CO2-compatible design.
02How fast does COS22E respond?
From air to nitrogen at reference conditions t90 is under 30 seconds and t98 is under 60 seconds.
03What is the detection limit of COS22E?
5 ppb for the standard sensor and 1 ppb for the trace sensor, with limits of quantification of 15 ppb and 3 ppb.
04What process limits apply to COS22E?
Minus 5 to 135 C and ambient pressure up to 12 bar absolute; Ex versions may differ, so follow the safety instructions.
05How much drift does COS22E have?
Less than 4 percent per month under reference conditions and 1 percent or less per month in operation below 4 Vol percent oxygen.