Hanna Instruments HI98197 Ultra-Pure Water EC/Resistivity Meter

Description

Hanna HI98197 ultra-pure water EC, resistivity and TDS meter: what it is

The Hanna Instruments HI98197 is a rugged handheld meter built to measure the conductivity and resistivity of very clean water. It displays EC, TDS, resistivity, salinity and temperature, and its resolution reaches 0.001 µS/cm for conductivity and 0.1 MΩ·cm for resistivity. The meter is supplied with a four-ring conductivity probe and a stainless-steel flow cell so that ultra-pure water can be measured in a continuous flow, which is the only reliable way to avoid carbon dioxide pickup. Seven Star LLC supplies the Hanna HI98197 in Oman and the GCC as an independent supplier in Muscat for power stations, desalination plants and laboratories that need to verify the quality of demineralised or polished water.

Why ultra-pure water needs a flow cell and a four-ring probe

Pure water reads low conductivity, but it quickly absorbs carbon dioxide from the air. The gas forms carbonic acid, which dissociates into hydrogen and bicarbonate ions and raises conductivity, so a sample in an open beaker gives a misleading answer within seconds. The Hanna HI98197 therefore comes with the HI605453 AISI 316 stainless-steel flow cell, into which the threaded HI763123 probe is screwed, so water passes the sensor without touching the atmosphere. The four-ring construction also reduces the polarisation errors that two-pole cells show, and the probe contains its own temperature sensor. A waterproof quick-connect DIN plug joins the probe to the meter without a threaded collar.

The manufacturer positions the meter for checking mixed-bed resin systems and similar polishers that deliver water close to 18.2 MΩ·cm at 25 °C.

HI98197 technical specifications

Parameter Specification
Manufacturer / part number Hanna Instruments / HI98197
Conductivity range 0 to 400 mS/cm temperature compensated; actual conductivity displayed up to 1000 mS/cm; autoranging from 0.000 µS/cm
EC resolution 0.001, 0.01 and 0.1 µS/cm; 0.001, 0.01 and 0.1 mS/cm by range
EC accuracy at 25 °C ±1% of reading (±0.01 µS/cm or one digit, whichever is greater)
TDS range and accuracy Autoranging 0.00 ppm to 400.0 g/L; ±1% of reading (±0.05 ppm or one digit); conversion factor 0.40 to 1.00
Resistivity range 1.0 Ω·cm to 100.0 MΩ·cm autoranging; ±1% of reading
Salinity scales % NaCl 0.0 to 400.0; seawater scale 0.00 to 80.00 ppt; practical salinity 0.01 to 42.00 PSU
Temperature –20.0 to 120.0 °C, 0.1 °C resolution, ±0.2 °C excluding probe error
Temperature compensation No compensation, linear, or non-linear (natural water); coefficient 0.00 to 10.00 %/°C; reference 15, 20 or 25 °C
Calibration Automatic up to five points from seven memorised standards (0.00 µS/cm, 84.0 µS/cm, 1.413 mS/cm, 5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm, 111.8 mS/cm)
Memory and interface Log on demand and lot logging; opto-isolated USB
Power and protection 4 x 1.5 V AA, about 100 hours without backlight; IP67; 0 to 50 °C, up to 100% RH

Configuring the HI98197 for pharmaceutical and power-plant work

The meter recognises whether a two-ring or four-ring probe is attached and lets the user adjust the nominal cell constant between 0.0 and 10.000. Ten user profiles can store combinations of compensation mode, range, cell constant and TDS factor, so different operators or different water streams can switch quickly. The HI98197 also supports all three stages of the USP <645> method for conductivity of purified water and generates a report when any stage is met, which is useful for laboratories that must document their checks.

Up to five calibration points can be taken with seven stored standards; choose standards close to the conductivity you expect to measure, and use the 1413 µS/cm point for general duty. Both calibration solutions, 84 µS/cm and 1413 µS/cm in 230 mL bottles, are supplied with the meter. Log-on-demand stores up to 400 readings and lot logging records at selected intervals for later review on a PC.

Related meters and where the HI98197 fits

The HI98197 sits in Hanna’s HI9819x family of waterproof portable meters. For general-purpose EC, TDS and salinity of ordinary water, the HI98192 is the easier choice, while multiparameter meters such as HI98195 cover pH, ORP and EC with one cable. The HI98197 is the right meter when the specification quotes conductivity in the sub-µS/cm region or resistivity in MΩ·cm. For permanent online monitoring on a reverse-osmosis or EDI skid, a fixed conductivity analyser is more suitable, with the HI98197 serving as the handheld verification instrument.

Selection checklist for the HI98197

  • Required water quality: resistivity target in MΩ·cm or conductivity in µS/cm, and the temperature at which results are reported.
  • Whether a flow cell connection is possible at the sample point and what tubing, fittings and flow rate are available.
  • Compensation mode and reference temperature required by your procedure or pharmacopoeia.
  • Calibration solutions needed beyond the two supplied bottles.
  • Logging and PC software needs for audit records.

Operation and maintenance of the HI98197

Connect the flow cell to the sample line with clean tubing, flush for several minutes so that the trapped air is displaced and let the reading stabilise before recording. Calibrate with the probe in the flow cell or in a closed vessel, not an open beaker. Keep the four-ring probe clean and free of films; rinse with deionised water after use and avoid touching the cell area. If values drift, recalibrate and inspect the flow-cell seals for air leaks.

HI98197 applications in Oman and the GCC

In the Gulf region the HI98197 is relevant to boiler make-up and condensate polishing in power and desalination plants, reverse-osmosis and EDI product water, laboratory water systems, semiconductor and pharmaceutical support utilities, and quality checks on bottled or demineralised water. High feed salinity and warm feedwater are characteristic of the region, so temperature compensation settings should be chosen carefully and documented.

HI98197 compared with other portable conductivity meters

Model Use Sensor
HI98197 Ultra-pure water, 0.001 µS/cm resolution, resistivity to 100.0 MΩ·cm Four-ring probe in stainless-steel flow cell
HI98192 EC, TDS, resistivity and salinity for general water Four-ring probe, no flow cell supplied
HI98195 pH, ORP, EC and temperature in one probe Multiparameter probe

Information to send with your HI98197 RFQ

  • Quantity and whether spare probes, flow cells or calibration solutions are needed.
  • Target resistivity or conductivity and temperature of the process.
  • Pipe or tubing size at the sample point.
  • Delivery location in Oman or the GCC.

Documents and quotation

The HI98197 instruction manual is attached on this page and includes full specifications, calibration, USP guidance and troubleshooting.

Send us the part number HI98197 and your application details and Seven Star LLC will confirm availability and lead time for Oman and the GCC. We are an independent supplier and are not affiliated with the manufacturer.

Datasheets & Documents
  • PDF Manual Hanna Instruments HI98197 Instruction Manual Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Hanna Instruments is a maker of analytical instruments for water and liquid testing, founded in 1978 in the Padua area of Italy by Oscar and Anna Nardo. Its early breakthrough was the pHep, an affordable pocket-sized pH tester. The company is now led by the founders' children, Martino and Pamela Nardo, with its headquarters in Woonsocket, Rhode Island, and offers more than 3,000 products.

Seven Star LLC is an independent supplier in Muscat. We list Hanna instruments for water, wastewater and laboratory users in Oman and the GCC.

Hanna Instruments company profile

Founded 1978, near Padua, Italy, by Oscar and Anna Nardo
Headquarters Woonsocket, Rhode Island, USA
Leadership Martino Nardo and Pamela Nardo
Early product The pHep pocket pH tester
Product families pHep and HI98 waterproof testers, edge benchtop meters, HI83 multiparameter photometers, Checker HC handheld colorimeters, HI931 and HI932 titrators, process controllers, reagents and calibration buffers

Hanna product series

Portable meters include the HI98190 pH and ORP meter, the HI98192 EC, TDS and salinity meter, the HI98193 dissolved oxygen and BOD meter, the HI98199 pH, EC and DO meter and the HI9829 multiparameter water quality meter.

The HI98703 is an EPA compliant portable turbidity meter, the HI97711 a free and total chlorine photometer and the HI83414 a benchtop turbidity and chlorine meter.

Waterproof portable meters such as the HI98190 and HI98192 suit field use, with automatic temperature compensation and GLP data for audits. Turbidity measurement with the HI98703 follows EPA method 180.1, and the HI97711 photometer measures free and total chlorine for disinfection control.

Where Hanna instruments are used

Drinking water and desalination plants, wastewater treatment, environmental field surveys, aquaculture, cooling water and quality control labs.

Hanna Instruments products we list

Family Examples
Portable Test & Measurement Hanna HI97711 Free and Total Chlorine Po…, Hanna HI98190 Waterproof Portable pH and…, Hanna HI98192 Waterproof Portable EC, Hanna HI98193 Waterproof Portable Dissol…, Hanna HI98199 Portable pH, Hanna HI9829 Multiparameter Water Qualit…, Hanna HI98703 Portable Turbidity Meter

Ordering Hanna instruments: what we need

Send the model, the parameters and ranges, the probe and cable length and any reagents or calibration solutions required.

Facts checked against Hanna Instruments' own company story and City of Woonsocket business profile, October 2026.

FAQ

Questions about the Hanna Instruments HI98197 Ultra-Pure Water

Answers from our engineering team on installation, compatibility and integration.

01What resolution does the HI98197 achieve for conductivity?
The HI98197 resolves conductivity to 0.001 µS/cm in its lowest range and resistivity to 0.1 MΩ·cm in its highest range. Accuracy is ±1% of reading, or ±0.01 µS/cm or one digit if that is larger.
02Why is a flow cell supplied with the HI98197?
Ultra-pure water absorbs carbon dioxide from air, which raises conductivity within seconds. The stainless-steel flow cell lets the sample pass the probe in a closed path, so the measurement reflects the true water quality.
03Does the HI98197 support USP 645 testing?
Yes. The meter can perform all three stages of the USP <645> conductivity method for purified water and produces a report when any of the stages is met.
04Which calibration solutions come with the HI98197?
The meter is supplied with 230 mL bottles of 84 µS/cm (HI7033M) and 1413 µS/cm (HI7031M) solution. It stores seven standards and can use up to five calibration points.
05Can the HI98197 use other conductivity probes?
The meter automatically recognises two-ring and four-ring probes and allows the cell constant to be adjusted between 0.0 and 10.000. For ultra-pure water the supplied four-ring HI763123 probe in the flow cell is the intended sensor.