Hanna HI83399 Multiparameter COD Photometer and pH Meter

Description

Hanna HI83399 water and wastewater multiparameter photometer with COD and pH

The Hanna HI83399 is a benchtop and portable multiparameter photometer for water and wastewater with 98 methods, including four chemical oxygen demand (COD) ranges, and a digital pH probe input. It is the broadest single instrument in the Hanna photometer range covered on this page. Seven Star LLC is an independent supplier in Muscat and is not affiliated with the manufacturer.

The part number identifies the COD-capable model with five optical channels and support for both round 24.6 mm cuvettes and 16 mm vials. Reagents, COD vials, the reactor for digestion and the digital pH electrode are separate items.

HI83399 technical specifications

Parameter Specification
Manufacturer / part number Hanna Instruments HI83399
Methods 98 photometer methods plus pH probe mode
COD (EPA 410.4, 16 mm vial) 0 to 150 mg/L; 0 to 1500 mg/L; 0 to 15000 mg/L; 0 to 60.0 g/L as O2
Light source Five LEDs with narrow-band interference filters at 420, 466, 525, 575 and 610 nm
Absorbance 0.000 to 4.000 Abs; 0.001 resolution; ±0.003 Abs at 1.000 Abs
Bandwidth / wavelength accuracy 8 nm / ±1 nm
Detector / cuvettes Silicon photocell; round 24.6 mm and 16 mm vials
pH probe mode -2.00 to 16.00 pH (±1000.0 mV); 0.01 pH; ±0.01 pH; digital pH/temperature electrode (not included)
Logging 1000 readings (photometer and electrode), optional user and sample ID
Connectivity USB-A host for flash drive; micro-USB-B for power and computer
Display 128 × 64 pixel LCD with backlight
Power 3.7 V Li-polymer rechargeable; more than 500 photometer measurements or 50 hours of pH; 5 V USB adapter
Environment 0 to 50 °C; 0 to 95 % RH non-condensing
Size / weight 206 × 177 × 97 mm; 1.0 kg
Warranty Two years

How the HI83399 works

In photometer mode the HI83399 passes LED light through a reacted sample and calculates concentration from absorbance for the chosen method; five wavelength channels allow a wide library covering colorimetric, turbidimetric and digestion-based tests. COD follows an EPA 410.4 adaptation in which sample is digested in a vial in a heated reactor and then read by the meter.

In probe mode, a digital pH electrode with intelligent identification gives pH from -2.00 to 16.00 with two-point calibration from five available buffers, so one instrument can cover wet-chemistry and pH checks on the same bench.

Where HI83399 fits: sensors, accessories and related models

A typical COD set-up includes the Hanna reactor (the manual references the HI839800 at 105 °C and its HI740217 safety shield for digestion methods), 16 mm digestion vials of the right COD range, pipettes and a compliant waste route for used digestion vials as described in the vial safety data sheet. The CAL Check feature validates the optics.

The smaller HI83326 is the pool-focused cousin with 17 methods and three channels.

HI83399 family comparison

Model Methods Optical channels COD
HI83326 17 3 No
HI83399 98 5 Yes, four ranges

If your laboratory only needs pool parameters the HI83326 is enough; for wastewater COD, silica, heavy metals and nutrient work choose the HI83399.

Selection and ordering checklist for HI83399

  • List all parameters needed and verify them against the 98-method library
  • Choose COD ranges (0 to 150, 1500, 15000 mg/L or 60.0 g/L) and the corresponding 16 mm vials
  • Quote the digestion reactor and safety accessories
  • Add a digital pH electrode, reagent sets and CAL Check cuvettes
  • Plan a waste route for digested COD vials
  • Confirm USB flash or PC export needs

Installation, calibration and maintenance of HI83399

Position the meter on a stable bench away from the reactor to avoid heat and vibration, charge the battery or use the USB adapter, and follow the tutorial for each method. For COD, preheat the reactor to the temperature stated in the vial method, digest, cool the vial and read it only after it has reached ambient.

Clean cuvettes and vials carefully, wipe them dry, and never read a vial with condensation. Calibrate the pH electrode with fresh buffers, keep it in storage solution and replace it when response degrades. Validate with CAL Check cuvettes on a schedule.

Digestion methods carry real safety obligations: use eye protection, handle hot vials with the safety shield and tongs, and dispose of used vials as hazardous waste. Label vials with sample IDs before heating because contents can change colour after digestion.

Where results will be reported to a regulator, keep the log, calibration data and reagent lot numbers together so that audits can reconstruct how a value was produced.

Typical applications of HI83399 in Oman and the GCC

In Oman and the GCC the HI83399 is used in sewage and industrial effluent laboratories that need COD, ammonia, nitrate, phosphate and metals on one bench, in desalination and power-plant laboratories, in environmental consultancies and in university teaching labs. Its 98 methods save buying separate single-parameter photometers.

Because the same meter handles COD and routine nutrients, a small laboratory can standardise on one calibration and validation regime.

Information to send with an RFQ for HI83399

  • Quantity and the list of parameters to be measured
  • COD range(s) and number of tests
  • Reactor, reagents, vials and CAL Check cuvettes
  • Digital pH electrode and buffers
  • Delivery location and lead time
  • Project or tender reference

Storage, handling and spares for the Hanna HI83399

Store the Hanna HI83399 in its case in a clean, dry place. Keep probes, electrodes and reagents as described in the user manual, as many of them need a storage solution or have a limited shelf life. In the GCC climate, keep reagents away from heat and direct sun and check expiry dates before use.

  • Calibrate the instrument with fresh standards at the interval your procedure requires.
  • Rinse probes and sample cells after each use and inspect them for scratches or deposits.
  • Keep spare consumables, such as reagents, standards, cuvettes and electrode solutions, in stock.
  • Record calibration results so that drift or probe ageing can be seen early.

When you order the Hanna HI83399, list the consumables and standards you need in the same RFQ, so the instrument can be used from the day it arrives.

Documents and quotation

The manual for the HI83399 is a large document. Send us the part number and we will confirm which documents are available for your project.

Send the part number HI83399, the quantity and your delivery location and we will reply with availability and a quotation.

Other Models in the Hanna Benchtop Photometer Range

Seven Star LLC supplies models from the Hanna Benchtop Photometer range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.

ModelCategory
Hanna HI83326 Pool and Spa Photometer with pH InputLiquid Analytical SystemsView details
Hanna HI83399 Multiparameter COD Photometer and pH MeterLiquid Analytical SystemsYou are viewing this model
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Hanna Instruments is a maker of analytical instruments for water and liquid testing, founded in 1978 in the Padua area of Italy by Oscar and Anna Nardo. Its early breakthrough was the pHep, an affordable pocket-sized pH tester. The company is now led by the founders' children, Martino and Pamela Nardo, with its headquarters in Woonsocket, Rhode Island, and offers more than 3,000 products.

Seven Star LLC is an independent supplier in Muscat. We list Hanna instruments for water, wastewater and laboratory users in Oman and the GCC.

Hanna Instruments company profile

Founded 1978, near Padua, Italy, by Oscar and Anna Nardo
Headquarters Woonsocket, Rhode Island, USA
Leadership Martino Nardo and Pamela Nardo
Early product The pHep pocket pH tester
Product families pHep and HI98 waterproof testers, edge benchtop meters, HI83 multiparameter photometers, Checker HC handheld colorimeters, HI931 and HI932 titrators, process controllers, reagents and calibration buffers

Hanna product series

Portable meters include the HI98190 pH and ORP meter, the HI98192 EC, TDS and salinity meter, the HI98193 dissolved oxygen and BOD meter, the HI98199 pH, EC and DO meter and the HI9829 multiparameter water quality meter.

The HI98703 is an EPA compliant portable turbidity meter, the HI97711 a free and total chlorine photometer and the HI83414 a benchtop turbidity and chlorine meter.

Waterproof portable meters such as the HI98190 and HI98192 suit field use, with automatic temperature compensation and GLP data for audits. Turbidity measurement with the HI98703 follows EPA method 180.1, and the HI97711 photometer measures free and total chlorine for disinfection control.

Where Hanna instruments are used

Drinking water and desalination plants, wastewater treatment, environmental field surveys, aquaculture, cooling water and quality control labs.

Hanna Instruments products we list

Family Examples
Portable Test & Measurement Hanna HI97711 Free and Total Chlorine Po…, Hanna HI98190 Waterproof Portable pH and…, Hanna HI98192 Waterproof Portable EC, Hanna HI98193 Waterproof Portable Dissol…, Hanna HI98199 Portable pH, Hanna HI9829 Multiparameter Water Qualit…, Hanna HI98703 Portable Turbidity Meter

Ordering Hanna instruments: what we need

Send the model, the parameters and ranges, the probe and cable length and any reagents or calibration solutions required.

Facts checked against Hanna Instruments' own company story and City of Woonsocket business profile, October 2026.

FAQ

Questions about the Hanna HI83399 Multiparameter COD Photometer and pH

Answers from our engineering team on installation, compatibility and integration.

01How many methods does the HI83399 have?
It has 98 photometer methods, including four COD ranges, plus a digital pH probe mode with two-point calibration.
02What COD ranges are available?
0 to 150 mg/L, 0 to 1500 mg/L, 0 to 15000 mg/L and 0 to 60.0 g/L as O2, using 16 mm vials and an adaptation of EPA 410.4.
03How many optical channels does it have?
Five LED channels with narrow-band filters at 420, 466, 525, 575 and 610 nm, 8 nm bandwidth, plus or minus 1 nm wavelength accuracy.
04Does it need a separate reactor for COD?
Yes. COD is digested in vials in a heated reactor, which is a separate item, and the vial is read after cooling.
05How does it power and log?
A 3.7 V Li-polymer battery or 5 V USB adapter; it logs 1000 readings and exports through USB flash drive or computer.