Endress+Hauser 50H Promag 50H Hygienic Electromagnetic Flowmeter

Description

50H and the small-bore hygienic niche

The Endress+Hauser 50H serves processes with the smallest flows and demanding hygiene. TI00048D describes it as a flowmeter for smallest flow rates with a modular electronic concept and flexible system integration, with a PFA liner, a stainless steel sensor housing linked to 3-A and EHEDG, and CIP and SIP cleanable wetted materials. The lining material statement lists USP Class VI, FDA 21 CFR 177.1550 and 3A.

The same document describes the Promag 53 variant with a four-line backlit display with touch control and a wider choice of fieldbuses including EtherNet/IP. Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer. We quote the 50H by its root code.

50H feature highlights

  • PFA liner suited to sanitary and chemical fluids
  • Hygienic connection portfolio with O-ring and aseptic molded seals
  • Stainless sensor housing and optional stainless field housing for the transmitter
  • Sizes from DN 2 to DN 150
  • Application-specific Quick Setups through the modular transmitter

50H specification summary

Parameter Value for 50H
Product root 50H, Proline Promag 50H (TI00048D, edition 16.18; same document covers 53H)
Positioning Flowmeter for smallest flow rates with modular electronics and flexible system integration
Liner PFA (USP Class VI, FDA 21 CFR 177.1550, 3A)
Sensor housing Stainless steel 1.4301 (304), 3-A and EHEDG on cover page
Nominal diameters DN 2 to 150 (1/12 to 6 inch)
Medium temperature -20 to +150 degC with EPDM seals (sensor DN 2 to 150)
Connections All connections 1.4404 (316L); PVDF flanges (EN, ANSI, JIS) also listed
Process connection groups O-ring seal DN 2 to 25; aseptic molded seal DN 2 to 25 and DN 40 to 150
Minimum conductivity 5 microsiemens per cm; 20 for demineralized water
Pulse output accuracy (Promag 50) 0.5 % of reading plus 1 mm/s; optional 0.2 % plus 2 mm/s
Supply 85 to 260 V AC; 20 to 55 V AC; 16 to 62 V DC (45 to 65 Hz on AC)
Transmitter housings Compact die-cast aluminum or stainless field housing 1.4301; wall-mounted aluminum
Window Glass or polycarbonate
Ambient temperature, transmitter -20 to +60 degC; optional -40 to +60 degC
Cleaning CIP and SIP cleanable wetted materials

Uses of 50H around the Gulf

Dairy, juice and beverage plants, biotech suites, cosmetics plants and fine-chemical dosing all need hygienic small-bore meters. The 50H can monitor additive injection, flavor dosing, CIP supply and return, and process-water loops. Because the transmitter can be fitted in a stainless field housing, washdown rooms are manageable. In regional pharmaceutical plants the USP Class VI and FDA lining statements support documentation.

How 50H differs from other Promag 50 variants

Only the 50H combines a PFA liner with a stainless sensor housing and the smallest bores, DN 2 to 25, plus larger hygienic sizes to DN 150. The 50P also has PFA but targets chemical process plants at DN 25 to 200 with a 180 degC limit, whereas the 50H stops at 150 degC and relies on hygienic fittings. It is a more capable version of the Promag 10H, which has a plainer transmitter and no fieldbus options.

How to specify 50H

Select diameter, then seal material: the document lists EPDM and other elastomers with a 150 degC upper limit. Select the connection from the O-ring and aseptic lists, and decide whether PVDF flanges are acceptable for the fluid. Specify the housing and communication. Provide the full order code to Seven Star LLC.

Mounting 50H

Keep the sensor filled and orient the electrodes sideways. Ground the sensor according to the document and use the supply options of 85 to 260 V AC, 20 to 55 V AC or 16 to 62 V DC.

Servicing 50H

Replace seals regularly and re-verify the electronics after chemical cleaning cycles. Keep records of CIP and SIP temperatures against the 150 degC limit.

A few practical points help avoid repeat visits. The 50H uses a modular transmitter, so display, board and housing can be swapped without touching the sensor, which stays in the line. The wall-mounted housing is die-cast aluminum, while the compact transmitter can be ordered with a stainless field housing in 1.4301 if the room is washed down. The window is glass or polycarbonate depending on the housing. When the process runs at the top of the range, remember that the transmitter ambient limit is plus 60 degC and that radiant heat from nearby steam lines counts against it. In remote versions, use the specified connecting cable and keep its length within the manual limit, as minimum conductivity depends on it.

Quick answers on 50H

The minimum conductivity is 5 microsiemens per cm, or 20 for demineralized water. The ambient range of the transmitter is minus 20 to plus 60 degC, with an optional minus 40 degC version. The compact transmitter may be aluminum or stainless, which gives an option for hygienic washdown environments.

Buyers sometimes ask how the 50H relates to the 53H. Both are covered by the same technical information. The 53H adds a four-line backlit display with touch control, bidirectional totalizers, software for filling and dosing, density and electrode cleaning, and a wider choice of fieldbuses including EtherNet/IP. The 50H keeps the two-line display with push buttons and covers HART and the common PROFIBUS and Modbus options. Because the sensors are the same, a 50H can be matched with the electronics level the plant prefers, as long as the full order code is specified.

Documents and quotation

Technical Information TI00048D is the datasheet for the 50H. Send an RFQ to Seven Star LLC with part number 50H, nominal diameter, seal, connection and quantity.

Delivery to Sites in Oman and the GCC

Delivery of the Endress+Hauser 50H is arranged to suit the site, whether that is a plant in Sohar or Duqm, an oilfield location in the interior, or a project elsewhere in the GCC. Tell us the delivery point, any site access or gate-pass requirements, and the person who will receive the goods. Goods are packed to protect sensitive ultrasonic flowmeters equipment during road or air transport, and shipping details are shared once the order is dispatched so that your stores team can plan receipt and inspection.

Other Models in the Promag 50 Range

Seven Star LLC supplies models from the Promag 50 range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.

ModelCategory
Endress+Hauser 50W Promag 50W Water Electromagnetic FlowmeterUltrasonic FlowmetersView details
Endress+Hauser 50L Promag 50L Lap-Joint Electromagnetic FlowmeterUltrasonic FlowmetersView details
Endress+Hauser 50E Promag 50E Economy PTFE FlowmeterUltrasonic FlowmetersView details
Endress+Hauser 50D Promag 50D Compact Wafer FlowmeterUltrasonic FlowmetersView details
Endress+Hauser 50H Promag 50H Hygienic Electromagnetic FlowmeterUltrasonic FlowmetersYou are viewing this model
Datasheets & Documents
  • PDF Datasheet Endress+Hauser 50H Promag 50H Hygienic Electromagnetic Flowmeter Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Endress+Hauser makes process measurement instruments for flow, level, pressure, temperature and liquid analysis. Georg H. Endress and Ludwig Hauser started the business in Lörrach, Germany, on 1 February 1953, first selling level meters, and the Endress+Hauser name has been used since 1957. The group is headquartered in Reinach, Switzerland, and has been wholly owned by the Endress family since 1975.

Seven Star LLC is an independent supplier listing Endress+Hauser flowmeters, level and pressure transmitters, temperature assemblies and liquid analysis products for buyers in Oman and the GCC. Instruments are matched from the full order code on the nameplate.

Endress+Hauser company profile

Origin Founded in Lörrach, Germany, on 1 February 1953 by the Swiss engineer Georg H. Endress and Ludwig Hauser
Headquarters Endress+Hauser AG, Reinach (Basel-Landschaft), Switzerland
Ownership Family owned; the Endress family has been the sole shareholder since 1975, and a family charter commits the company to stay family owned
Size About 16,500 employees and sales of about EUR 3.7 billion (2023, as reported by the company)
Product families Proline flowmeters, Micropilot and Levelflex radar, Liquiphant switches, Cerabar and Deltabar pressure, iTEMP and iTHERM temperature, Liquiline and Memosens liquid analysis

Milestones

  • 1953: Georg H. Endress and Ludwig Hauser start the business in Lörrach on 1 February, first selling level meters.
  • 1957: The Endress+Hauser name comes into use.
  • 1975: The Endress family becomes sole shareholder.
  • 2023: About 16,500 employees and sales of about EUR 3.7 billion, as reported by the company.

Endress+Hauser product lines

Line Measuring principle Typical use
Promass F 300, Promass 80F, 83F/84F Coriolis mass flow Direct mass flow and density of liquids and gases, hydrocarbon transfer and chemical dosing
Promag P 300, P 500, H 300, W 400, W 10, 50P Electromagnetic flow Water, wastewater, slurries and other conductive liquids
Prowirl F 200, 72F Vortex flow Steam, gas and low-viscosity liquids
Prosonic Flow 92F, t-mass 65F Ultrasonic and thermal mass flow Water lines; compressed air and industrial gases
Micropilot FMR62, FMR67, FMR60, FMR10B, FMR20, FMR50, FMR52 Free-space radar level, 80 GHz and 26 GHz Tanks, process vessels, silos and open water channels
Levelflex FMP51, FMP50, FMP54, FMP55 Guided wave radar level Level and interface in vessels, separators and bypass chambers
Liquiphant FTL51, Prosonic S FMU90 Vibronic point level; ultrasonic level High and low level alarms, pump protection, open channel flow
Cerabar PMC71B, PMP51B, PMP71, Deltabar PMD78B, PMD75 Pressure and differential pressure Process pressure, hydrostatic level and DP flow
iTEMP TMT82, TMT162, iTHERM TM101, TR66, TR10 Temperature transmitters and thermometers RTD and thermocouple points with thermowells
Liquiline CM444, CM442, Memosens CPS11E, Stamolys reagents Liquid analysis pH, ORP, conductivity, dissolved oxygen and nutrient analysis in water treatment and process

Where Endress+Hauser instruments are used

In Oman and the GCC, Endress+Hauser instruments are common in oil and gas production and processing, refineries and petrochemicals, desalination and water networks, wastewater treatment, power plants and food and beverage plants. Coriolis meters measure hydrocarbon and chemical flows, electromagnetic meters cover water and wastewater, radar transmitters handle tank and silo level, and Liquiline analysers monitor water quality. Proline 300 and 500 flowmeters include Heartbeat Technology for in-situ verification, which reduces the need to remove a meter for checks.

Ordering Endress+Hauser instruments: what we need

The order code on the nameplate (for example PMP71-1QA1S11RDADU) defines every option: measuring range, process connection, materials, output, approvals and display. Send it with the serial number; Endress+Hauser can retrieve the original configuration from the serial number. For a new instrument add the medium, operating and design pressure and temperature, range, line size or tank geometry, process connection, wetted materials, output or protocol (4-20 mA HART, PROFIBUS, FOUNDATION Fieldbus, Modbus) and the hazardous-area approval required. Older models such as Promass 80/83 or Micropilot M are replaced by current Proline or 80 GHz Micropilot devices.

Facts checked against Endress+Hauser company history and product information, October 2026.

FAQ

Questions about the Endress+Hauser 50H Promag 50H Hygienic

Answers from our engineering team on installation, compatibility and integration.

01What liner does the 50H use?
PFA, with the lining statement referring to USP Class VI, FDA 21 CFR 177.1550 and 3A in TI00048D.
02What sizes are available for the 50H?
DN 2 to 150 (1/12 to 6 inch), with separate connection tables for DN 2 to 25 and DN 40 to 150.
03What medium temperature applies to the 50H?
Minus 20 to plus 150 degC for the sensor and the EPDM seal listed in the datasheet.
04What conductivity does the 50H require?
At least 5 microsiemens per cm, and at least 20 microsiemens per cm for demineralized water.
05How do I quote the 50H?
Send an RFQ to Seven Star LLC with 50H, nominal diameter, seal material, connection and quantity.