LAR QuickCODultra Online COD (TOD) Analyzer

Description

Overview

The LAR QuickCODultra is an online COD (TOD) water analyzer developed by LAR Process Analysers in Berlin, now part of Process Insights. It measures chemical oxygen demand through total oxygen demand, by oxidizing the complete water sample at 1,200 °C and detecting the oxygen consumed. The method follows ASTM D6238 and needs no catalysts and no additional hazardous chemicals, unlike the dichromate reflux method used in laboratories. A COD (TOD) result is available within 1 to 2 minutes, and the analyzer can handle sticky, fatty, salty and particle-laden samples, including salt concentrations up to 600 g/l.

Seven Star LLC supplies the LAR QuickCODultra to industrial wastewater treatment plants, refineries, petrochemical complexes and municipal operators in Oman and the GCC. From Muscat we supply consumables on request and provide engineering support with application questionnaires, sample preparation and integration into plant control systems.

Measurement Principle

In the LAR QuickCODultra a robotic injection system doses a defined sample volume into a high-temperature reactor held at 1,200 °C, where all oxidizable organic and inorganic compounds are converted completely. A carrier gas with a known oxygen content flows through the reactor, and the analyzer measures the oxygen consumed by the oxidation, which is the total oxygen demand of the sample. Because oxidation is thermal rather than chemical, results are free of chloride interference thanks to LAR process management, and there is no reagent handling or chromium and mercury waste.

In general, TOD and COD correlate closely for most wastewater matrices, which is why a TOD-based analyzer is widely used for continuous COD monitoring and discharge control. The LAR QuickCODultra works on a batch principle, and optional sum parameters such as TOC and TNb can be measured in combination with COD (TOD), giving a fuller picture of organic and nitrogen loads from one instrument.

Engineering Design of the LAR QuickCODultra

Sample preparation is handled by the FlowSampler, a maintenance-free particle separator, with an optional homogenizer for continuous homogenization; particles below 800 µm can be measured. The analyzer can handle up to six sample streams. The standard housing is powder-coated steel rated IP54, measuring W 848 x H 1,100 x D 635 mm and weighing from 115 kg. For hazardous areas, Process Insights lists ATEX/UKEX Zone 2 T4 and ATEX/UKEX Zone 1/IECEx T4 versions in IP54 stainless steel housings, the Zone 1 analyzer measuring W 1,080 x H 1,293 x D 700 mm with a separate signal box.

Inflow and outflow use 4.8 mm, 8 mm and 12 mm ID tubing. Power is 230 or 115 V AC at 50 or 60 Hz with 16 A external protection. Outputs include 0/4-20 mA analog signals and an RS 232 serial interface, and remote control through TCP/IP is offered as an option. Permissible sample temperature is 5 to 60 °C, and the recommended ambient temperature is 10 to 30 °C at up to 80% relative humidity, which in the GCC usually means installation in an air-conditioned analyzer shelter.

As general background, the classic laboratory COD test digests a sample with potassium dichromate and sulfuric acid for about two hours, using silver and mercury salts, and produces hazardous waste. A continuous thermal method such as the LAR QuickCODultra replaces that with an automatic measurement every few minutes and no reagent waste, which changes COD from a delayed compliance number into a live process variable. Many plants still run periodic laboratory tests to confirm the correlation between the online TOD reading and their reported COD, and the LAR QuickCODultra result can be scaled accordingly in the control system.

Key Features

The LAR QuickCODultra stands out for:

  • High-temperature thermal oxidation at 1,200 °C, with complete conversion of the sample.
  • A catalyst-free, reagent-free technique according to ASTM D6238.
  • COD (TOD) cycle and response time of 1 to 2 minutes for fast process control.
  • Measurement ranges of 150 to 30,000 or 20,000 to 250,000 ppm COD (TOD), with dilution up to 1,000,000 possible.
  • Accuracy and repeatability of maximum 2% of full scale range.
  • Tolerance of salt concentrations up to 600 g/l and particles below 800 µm.
  • Up to six sample streams and optional TOC and TNb parameters.
  • Optional ATEX, UKEX and IECEx housings for hazardous areas.

Industrial Applications

Process Insights lists discharge control, industrial wastewater, de-icing water, water influent, water effluent and process water as applications for the LAR QuickCODultra. In Oman and the GCC this covers refinery and petrochemical effluent treatment in Sohar and Duqm, produced water and oily water treatment at oil and gas facilities, influent and effluent monitoring at municipal wastewater plants, airport de-icing runoff, and food and beverage wastewater. Fast results allow operators to divert high-load streams, protect biological treatment and document discharge compliance.

Seven Star LLC supplies the LAR QuickCODultra with application support and spare parts. For related LAR consumables, see the LAR 501-08-000045 gas drying tube; alternative online organic carbon analyzers on this site include the Hach BioTector B3500ul online TOC analyzer and the Forbes Marshall FM TOC101 online TOC analyser.

Technical Specifications

The LAR QuickCODultra data below is taken from the Process Insights datasheet (v25):

Parameter Specification
Model QuickCODultra online COD (TOD) analyzer
Measurement method Thermal oxidation at 1,200 °C, non-catalytic
Measurement compliance ASTM D6238 (TOD analysis)
Measurement ranges 150 to 30,000 or 20,000 to 250,000 ppm COD (TOD); dilution up to 1,000,000 possible
Parameters COD (TOD); combination with TOC and TNb possible
Accuracy / repeatability max. 2% of FSR / max. 2% of FSR
Cycle time / response time 1 to 2 min / 1 to 2 min
Sample preparation FlowSampler particle separator; optional homogenizer
Particle size below 800 µm
Sample streams up to 6
Salt tolerance up to 600 g/l
Sample temperature 5 to 60 °C
Ambient temperature (recommended) 10 to 30 °C, max. 80% RH non-condensing
Standard housing Powder-coated steel, IP54, W 848 x H 1,100 x D 635 mm, from 115 kg
Hazardous area options ATEX/UKEX Zone 2 T4; ATEX/UKEX Zone 1/IECEx T4 (stainless steel, IP54)
Tubing 4.8, 8 and 12 mm ID
Power supply 230/115 V AC, 50/60 Hz, 16 A external protection
Outputs 0/4-20 mA analog, RS 232; TCP/IP remote control optional

How to Order

To quote the LAR QuickCODultra, send Seven Star LLC your sample description, expected COD range, number of streams, area classification and required outputs. The LAR QuickCODultra datasheet can be downloaded from this page. Our Muscat team will review the application with you, confirm availability and support installation and service across Oman and the GCC.

Datasheets & Documents
  • PDF Datasheet LAR QuickCODultra Online COD (TOD) Analyzer Datasheet (Process Insights v25) Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand
Online TOC measurement in difficult water is the specialty of LAR Process Analysers, a maker of water and wastewater analysers founded in 1986 and based in Berlin, Germany. Its QuickTOC analyzers are found on effluent, process water and condensate lines. Seven Star LLC sources and supplies LAR spare parts and consumables for operators in Oman who already run these analyzers. Current LAR analyzers are offered under the Process Insights name.

LAR analyzer spare parts listed here

The four listed items are parts for the UV persulfate oxidation analyzer type, the QuickTOCuv: a glass reactor for the UV method (501-40-000106, formerly 501-04-000021), a low-pressure 15 W ozone-free UV emitter (501-40-000159), a five-channel peristaltic pump (501-40-000215) and a gas drying tube for the carrier gas path (501-08-000045). All of them are wear or consumable parts replaced during routine upkeep.

Ordering LAR parts: what we need

Send the LAR part number in the 501-xx-xxxxxx format, and the old number as well if your manual shows one. Add the analyzer type and serial number from the type plate, because the QuickTOCuv, QuickTOCultra and other models use different reactors, pumps and tubing. A photo of the old part and the manual's spare parts page speed up the check. We supply these parts to water, power and petrochemical sites across the GCC.

Frequently Asked Questions

How does the LAR QuickCODultra measure COD?
It oxidizes the whole sample at 1,200 C and measures the oxygen consumed, which is the total oxygen demand (TOD). The method follows ASTM D6238 and uses no catalysts or hazardous reagents.
How fast is the QuickCODultra?
A COD (TOD) measurement takes 1 to 2 minutes, so short load peaks in wastewater can be detected and used for process control or diversion.
Can the QuickCODultra handle salty or dirty samples?
Yes. The datasheet lists salt concentrations up to 600 g/l and particles below 800 micrometres, with the FlowSampler separator and an optional homogenizer for sample preparation.
Is a hazardous area version available?
Process Insights lists ATEX/UKEX Zone 2 T4 and ATEX/UKEX Zone 1/IECEx T4 versions in IP54 stainless steel housings, in addition to the standard powder-coated steel IP54 housing.
Does Seven Star LLC supply LAR analyzers in Oman?
Seven Star LLC supplies the LAR QuickCODultra and related consumables to customers in Oman and the GCC, with application and engineering support from Muscat.