AMETEK Land Gold Cup Reformer Tube Reference Pyrometer

Description

Gold Cup reference pyrometer for reformer tube surface temperature

The AMETEK Land Gold Cup is a water-cooled, hand-held reference thermometer used to measure the true surface temperature of reformer and other fired-heater tubes. A gold-plated hemisphere is pressed against the tube. Inside it, infrared energy is reflected many times before it reaches the detector, so the reading is close to black-body conditions and does not depend on a guessed emissivity. The brochure states that the energy is reflected at least 12 times, which pushes the effective emissivity towards 1.00.

It is not meant to replace permanent temperature measurement. Its purpose is to give plant engineers a trusted number so that portable thermal imagers, infrared thermometers and furnace pyrometers can be checked and corrected. Seven Star LLC supplies it in Oman and the GCC as an independent supplier in Muscat, for refineries, petrochemical plants and fertiliser sites that run fired heaters.

Gold Cup technical specifications

Parameter Specification
Manufacturer / part number AMETEK Land (LAND Instruments International) / Gold Cup
Principle Gold reflecting hemisphere in contact with the tube; black-body cavity measurement
Spectral response 1.6 µm
Output LCD display, 300 to 1300 °C (572 to 2372 °F)
Response time 0.5 s
Resolution 0.1 °C / 1 °F
Repeatability 1 °C / 2 °F
Absolute accuracy 0.25 % + 2 °C
Ambient temperature limits 0 to 60 °C (32 to 140 °F)
Power requirement 1 × 9 V PP3 battery
Maximum cooling water pressure 3 bar
Cooling connections 3/8 in BSP inlet and outlet
Sealing IP67
Weight (with dry protective cover) 6.9 kg (15.2 lb)
Included accessory Rubberised hand-held display unit case
Optional / on request Cooling control panel with hoses; UKAS calibration (details from manufacturer)

How the gold hemisphere removes emissivity and background errors

Ordinary infrared thermometers have two big uncertainties inside a furnace: the unknown emissivity of the tube surface and the hotter radiation reflected from the furnace walls and flames. The hemisphere deals with both at once. The hemisphere touches the tube and shields the measurement area from the hotter surroundings. Its highly reflective gold coating bounces the tube’s own radiation around the cavity so that the optical fibre sees something very close to a black body.

A silicon detector at 1.6 µm (short wavelength) then measures the radiation, and the battery-powered display unit converts it to temperature. The display also powers the thermometer, so no mains supply is needed at the furnace.

Using the probe on a fired heater

The brochure describes the working method. The probe is first connected to its cooling water supply and to the display unit. It is then inserted through an inspection port into the furnace and the gold cup is touched onto the tube, with the probe level and aligned to the centre of the tube. The reading is recorded straight away, before the tube has time to cool under the probe.

  • Check the cooling water supply is below the 3 bar limit and that hoses are fitted to the 3/8 in BSP connections before opening the port.
  • Align the probe level with the tube centre; poor contact lets background radiation in and reduces the benefit of the cavity.
  • Record the display reading immediately and note the tube position and time.
  • Remove the probe and let it cool before the next tube; keep the dry protective cover on during transport.
  • Compare that reading with the reading of the portable imager or pyrometer being checked and derive the correction.

Water-cooled probe compared with the modular version

AMETEK Land also offers a modular version, which uses the same gold-hemisphere principle in a probe that can be assembled to different lengths and bent angles and which is cooled with air rather than water. The classic version described here is the water-cooled version with a fixed probe and a 6.9 kg weight. The table shows how the two differ in the details published by the manufacturer.

Item Gold Cup Modular Gold Cup
Cooling medium Water, max. 3 bar, 3/8 in BSP Air, max. 8 bar(g), 1/2 in BSP/NPT
Configuration Single probe Modular kit: straight or 90/180 kit
Weight / case 6.9 kg with dry cover Complete system under 32 kg in a 1118 × 426 × 202 mm case
Measuring range 300 to 1300 °C 300 to 1300 °C

What to confirm before ordering

  • Cooling supply. This probe needs clean cooling water at the furnace. If only compressed air is available on the platform, the modular air-cooled version is the better fit.
  • Port and tube access. Check port size and the distance from the port to the tube row you want to measure.
  • Temperature range. 300 to 1300 °C covers most reformer tube duties, but confirm the highest metal temperature in your heater.
  • Calibration. UKAS calibration and a cooling control panel are available through the manufacturer; request them with the order if your audit needs a certificate.
  • Spares. Plan a spare 9 V battery and a case; they are handled and stored like any other portable test instrument.

Care and handling

The gold surface is what makes the reading accurate, so keep it clean and avoid contact with anything other than the tube. The probe is sealed to IP67, but the rubberised case and dry protective cover are there to protect the head between measurements. Ensure the cooling water is switched on before the probe enters the heat and left on until it has come out and cooled. Store the instrument in its case with the battery removed if it will not be used for a long time, and arrange recalibration with the manufacturer on the interval your quality system requires.

Typical applications in Oman and the GCC

Steam methane reformers in ammonia, methanol and hydrogen plants, and ethylene or refinery fired heaters, operate close to the metallurgical limit of their tubes. Overestimating tube temperature wastes capacity, while underestimating it shortens tube life. Operators in the Gulf use surface checks during turnarounds and as part of regular campaigns to calibrate the imagers and infrared thermometers that they use in between. High ambient temperature outdoors and strong furnace radiation also make an emissivity-independent reference useful, because it lets different technicians arrive at the same number.

Information to send with an RFQ

  • Furnace type, tube diameter and expected tube metal temperature.
  • Port size and probe reach needed.
  • Cooling water availability and pressure (or air only).
  • Whether a cooling control panel and UKAS calibration are required.
  • Quantity and delivery location in Oman or the GCC.

Documents and quotation

The manufacturer brochure with the specification sheet is attached on this page.

Send us the part number, your application and any options, and Seven Star LLC will confirm availability and lead time for delivery in Oman and the GCC.

Datasheets & Documents
  • PDF Datasheet AMETEK Land Gold Cup Brochure and Specification Sheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

AMETEK is the American maker of electronic instruments and electromechanical devices, headquartered in Berwyn, Pennsylvania. It was founded in 1930 as American Machine and Metals. In process industries it is known through brands such as Western Research (sulfur recovery and H2S analysers), Thermox (oxygen analysers), Land (pyrometers), Drexelbrook (RF level) and Chandler Engineering.

Seven Star LLC is an independent supplier in Muscat. We list AMETEK process analysers, pyrometers, portable testers and spare parts, and supply them to gas plants, refineries and power stations in Oman and the GCC.

AMETEK company profile

Established 1930, USA, as American Machine and Metals; later renamed AMETEK
Headquarters Berwyn, Pennsylvania
Structure Two groups: Electronic Instruments (EIG) and Electromechanical (EMG)
Process brands Western Research, Thermox, Land, Drexelbrook, Chandler Engineering, Ranarex

Milestones

  • 1930: American Machine and Metals is formed in March and listed on the New York Stock Exchange as AME.
  • Early 1960s: The company is renamed AMETEK.
  • 2021: AMETEK acquires Magnetrol.

AMETEK products we list

Product Type Typical use
888, 900, 930 Western Research sulfur recovery analysers Tail gas, air demand and H2S in sulfur pits on Claus units
931, 933, 9900RM/WM UV gas analysers H2S in natural gas and multi-component process streams
5000, 3050-OLV, 5100HD, 5100P Moisture analysers Water content in natural gas and process gases; the 5100 models use TDLAS
241CE II, Chanscope II / Model 13 Dew point analysers and testers Hydrocarbon and water dew point of natural gas
WDG 1200/1210, WDG-V, TM2000 Thermox oxygen analysers Combustion control on heaters and boilers, and trace oxygen
SPOT+, Cyclops L Land pyrometers Non-contact temperature of furnaces, kilns, tubes and hot metal
120HD, Ranarex, Drexelbrook RCT-1210 Portable analysers and level Hydrogen purity in generators, gas specific gravity, RF level measurement

Most of these analysers work on light absorption. UV analysers such as the 933 and 9900 measure H2S and SO2 by their absorption bands, and tail gas analysers use this to keep the H2S to SO2 ratio near 2:1 for the best Claus conversion. TDLAS moisture analysers use a tuned laser on a single water line, so they respond fast and do not suffer from the glycol and methanol contamination that affects sensor-based analysers.

Where AMETEK instruments are used

In Oman and the GCC, AMETEK analysers run on gas processing and sulfur recovery units, gas pipelines and custody points, refinery heaters and boilers, power station generators and steel and cement furnaces.

Ordering AMETEK products: what we need

Send the model and the complete part or sales order number from the nameplate. For a new analyser, send the gas composition, the component and range to be measured, the sample pressure and temperature, and the hazardous area classification. For spare parts, send the part number and the analyser serial number.

Facts checked against AMETEK company history, October 2026.

FAQ

Questions about the AMETEK Land Gold Cup Reformer Tube Reference

Answers from our engineering team on installation, compatibility and integration.

01What does the Gold Cup measure?
It measures the true surface temperature of reformer and fired-heater tubes from 300 to 1300 °C using a gold reflecting hemisphere that touches the tube, so emissivity and background radiation are largely cancelled.
02What is the accuracy of the Gold Cup?
The absolute accuracy is 0.25 % + 2 °C, with 0.1 °C / 1 °F resolution and 1 °C / 2 °F repeatability. Response time is 0.5 s.
03How is the Gold Cup cooled and powered?
It is water cooled, at a maximum of 3 bar through 3/8 in BSP connections, and runs from one 9 V PP3 battery in the hand-held display unit.
04Is the Gold Cup used to replace other pyrometers?
No. The manufacturer positions it as a reference to check and correct portable imagers and infrared thermometers, not as a permanent measurement.
05How does the Gold Cup differ from the Modular Gold Cup?
The Modular Gold Cup is assembled into straight or 90/180 configurations and is cooled with air, while this Gold Cup is the water-cooled probe weighing 6.9 kg.