Endress+Hauser PMP131 Cerabar T Compact Pressure Transducer

Description

Endress+Hauser PMP131 Cerabar T pressure transducer

The Endress+Hauser PMP131 is a compact pressure transducer of the Cerabar T family. Technical Information TI00415P covers three devices: the PMC131 with a ceramic sensor, and the PMP131 and PMP135 with piezoresistive metallic sensors. The PMP131 measures absolute and gauge pressure in gases, vapours, liquids and dusts from 0 to 1 bar up to 0 to 400 bar. It has threaded process connections, a stainless steel 316L (1.4435) wetted part and electrical connections by plug or fixed cable.

In plants the PMP131 is used where a small, fast, robust transducer is enough: machine hydraulics, pump skids, compressor monitoring and utility headers that do not need a display, fieldbus or HART.

PMP131 key facts

  • Piezoresistive measuring cell in AISI 316L (1.4435), 1 to 400 bar.
  • Outputs: 4 to 20 mA two-wire, 0 to 10 V three-wire, or PNP switch (switch frequency up to 10 Hz).
  • Rise time T90 of 2 to 5 ms, much faster than ceramic transducers.
  • Deployed for pressure monitoring up to SIL 2 per IEC 61508 and IEC 61511-1 on the 4-20 mA versions.
  • Process connections include G 1/2, 1/2 MNPT with 1/4 FNPT, G 1/4 and M20 x 1.5.
  • Electrical connections: DIN 43650/A plug, M12 plug, or cable of 5 m and 25 m.

PMP131 technical data

Parameter Value
Measuring range 0 to 1 bar (15 psi) up to 0 to 400 bar (6000 psi)
Output current 4 to 20 mA, 2-wire; maximum load resistance (supply voltage minus 12 V) divided by 0.02 A
Output voltage 0 to 10 V, 3-wire; load at least 5 kilo-ohm; current consumption 6 mA
Switch output PNP; on-state current up to 500 mA; off-state up to 1 mA; up to 10 Hz
Supply voltage 12 to 30 V DC
Reference accuracy 0.5 percent of URL
Switch point deviation 1 percent of URL; non-reproducibility 0.5 percent of URL
Long-term stability 0.15 percent of URL per year
Rise time T90 2 to 5 ms
Process temperature minus 25 to plus 70 degrees Celsius
Ambient temperature minus 25 to plus 70 degrees Celsius; Ex i minus 25 to plus 65
Storage temperature minus 40 to plus 85 degrees Celsius
Weight about 0.24 kg up to 60 bar; about 0.32 kg up to 400 bar
Ingress protection IP 65 with plug; IP 68 (1 m water column for 24 hours) with cable

How PMP131 differs from PMC131 and PMP135

The PMC131 uses a capacitive ceramic cell, measures from minus 1 bar vacuum to 40 bar and supports process temperatures from minus 20 to plus 100 degrees Celsius, but responds in 20 ms and has an accuracy of 0.5 percent of the nominal value times the turn down. The PMP131 uses a metallic piezoresistive cell, reaches 400 bar, responds in 2 to 5 ms and is limited to minus 25 to plus 70 degrees Celsius. The PMP135 is the hygienic version of the same cell with Tri-Clamp, DIN 32676 clamp and SMS connections, 3-A and EHEDG listed with an o-ring seal and welding neck.

So the choice is simple: pick the PMP131 for high pressure and fast response, the PMC131 for vacuum and abrasive media, and the PMP135 where a hygienic connection is mandatory.

PMP131 use in Oman and GCC industry

Typical jobs are hydraulic press and power pack monitoring, compressor discharge, pumping skids for water injection and desalination reverse osmosis feed, and OEM machines delivered to the Gulf. The switch output allows direct pump protection without an extra relay, and the 0 to 10 V version feeds simple building controls and PLC analog cards.

Installing the PMP131

Screw the transducer into a G 1/2 or NPT port with the correct seal. Never tighten using the housing; use the hex on the process connection. Keep the cable gland or plug downward for water drainage and use the 25 m cable version when the PLC is far from the pressure point. Respect the ambient limit of 70 degrees Celsius, and use a siphon or a cooling loop on hot lines.

Checks and care for PMP131

Because the PMP131 has no display or adjustment, verify it against a reference gauge once a year, comparing zero and a mid-point. Expect about 0.15 percent of URL drift per year in stable conditions. Replace the plug seal if the connector is removed often, and protect cables against abrasion.

Choosing the PMP131 range correctly

The reference accuracy is stated relative to the upper range limit, so a transducer with a 400 bar range gives roughly 2 bar of tolerance, which is too coarse for a 40 bar line. Choose the range so that your normal operating pressure sits at 50 to 80 percent of the URL and keep the overload margin for pressure spikes in mind; the ordering section of the datasheet lists the overload limit for each range.

PMP131 order options at a glance

The datasheet organises PMP131 versions by electrical connection letter and by thread. A1 and A2 are the DIN 43650/A plug in M16 or 1/2 NPT form (IP 65), A3 and A4 are 5 m and 25 m cable versions (IP 68), and A5 is a plug M12 or DIN 43650/C. Threads shown are ISO 228 G 1/2, ANSI 1/2 MNPT with 1/4 FNPT, ISO 228 G 1/4, ANSI 1/4 MNPT and M20 x 1.5, plus a flush-mounted G 1/2 version with a DIN 3852-A seal seat in AISI 304. Remember that Ex versions are marked ATEX II 3 G for zone 2 use, and the SIL 2 assessment applies only to the 4 to 20 mA electronics.

Documents and quotation

Technical Information TI00415P for the PMP131 is attached as the datasheet. Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer. Please send PMP131, the quantity and your range and connection with your RFQ.

Selection Notes for the Endress+Hauser PMP131

When comparing the Endress+Hauser PMP131 with alternatives, start from the duty it must perform: the measured or controlled variable, the range and accuracy you need, the environment at the mounting point and the interface to your control system. Check the electrical and mechanical connections against the installed equipment, and confirm any hazardous-area or ingress-protection requirements for the location. If any point is unclear, send a photo of the nameplate and a short description of the application, and we will compare it with the Endress+Hauser (E & H) pressure transmitters range and confirm the correct order code before you commit.

Other Models in the Eh Cerabar T Range

Seven Star LLC supplies models from the Eh Cerabar T range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.

ModelCategory
Endress+Hauser PMC131 Cerabar T Ceramic Pressure TransducerPressure TransmittersView details
Endress+Hauser PMP131 Cerabar T Compact Pressure TransducerPressure TransmittersYou are viewing this model
Datasheets & Documents
  • PDF Datasheet Endress+Hauser PMP131 Cerabar T Compact Pressure Transducer Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Endress+Hauser makes process measurement instruments for flow, level, pressure, temperature and liquid analysis. Georg H. Endress and Ludwig Hauser started the business in Lörrach, Germany, on 1 February 1953, first selling level meters, and the Endress+Hauser name has been used since 1957. The group is headquartered in Reinach, Switzerland, and has been wholly owned by the Endress family since 1975.

Seven Star LLC is an independent supplier listing Endress+Hauser flowmeters, level and pressure transmitters, temperature assemblies and liquid analysis products for buyers in Oman and the GCC. Instruments are matched from the full order code on the nameplate.

Endress+Hauser company profile

Origin Founded in Lörrach, Germany, on 1 February 1953 by the Swiss engineer Georg H. Endress and Ludwig Hauser
Headquarters Endress+Hauser AG, Reinach (Basel-Landschaft), Switzerland
Ownership Family owned; the Endress family has been the sole shareholder since 1975, and a family charter commits the company to stay family owned
Size About 16,500 employees and sales of about EUR 3.7 billion (2023, as reported by the company)
Product families Proline flowmeters, Micropilot and Levelflex radar, Liquiphant switches, Cerabar and Deltabar pressure, iTEMP and iTHERM temperature, Liquiline and Memosens liquid analysis

Milestones

  • 1953: Georg H. Endress and Ludwig Hauser start the business in Lörrach on 1 February, first selling level meters.
  • 1957: The Endress+Hauser name comes into use.
  • 1975: The Endress family becomes sole shareholder.
  • 2023: About 16,500 employees and sales of about EUR 3.7 billion, as reported by the company.

Endress+Hauser product lines

Line Measuring principle Typical use
Promass F 300, Promass 80F, 83F/84F Coriolis mass flow Direct mass flow and density of liquids and gases, hydrocarbon transfer and chemical dosing
Promag P 300, P 500, H 300, W 400, W 10, 50P Electromagnetic flow Water, wastewater, slurries and other conductive liquids
Prowirl F 200, 72F Vortex flow Steam, gas and low-viscosity liquids
Prosonic Flow 92F, t-mass 65F Ultrasonic and thermal mass flow Water lines; compressed air and industrial gases
Micropilot FMR62, FMR67, FMR60, FMR10B, FMR20, FMR50, FMR52 Free-space radar level, 80 GHz and 26 GHz Tanks, process vessels, silos and open water channels
Levelflex FMP51, FMP50, FMP54, FMP55 Guided wave radar level Level and interface in vessels, separators and bypass chambers
Liquiphant FTL51, Prosonic S FMU90 Vibronic point level; ultrasonic level High and low level alarms, pump protection, open channel flow
Cerabar PMC71B, PMP51B, PMP71, Deltabar PMD78B, PMD75 Pressure and differential pressure Process pressure, hydrostatic level and DP flow
iTEMP TMT82, TMT162, iTHERM TM101, TR66, TR10 Temperature transmitters and thermometers RTD and thermocouple points with thermowells
Liquiline CM444, CM442, Memosens CPS11E, Stamolys reagents Liquid analysis pH, ORP, conductivity, dissolved oxygen and nutrient analysis in water treatment and process

Where Endress+Hauser instruments are used

In Oman and the GCC, Endress+Hauser instruments are common in oil and gas production and processing, refineries and petrochemicals, desalination and water networks, wastewater treatment, power plants and food and beverage plants. Coriolis meters measure hydrocarbon and chemical flows, electromagnetic meters cover water and wastewater, radar transmitters handle tank and silo level, and Liquiline analysers monitor water quality. Proline 300 and 500 flowmeters include Heartbeat Technology for in-situ verification, which reduces the need to remove a meter for checks.

Ordering Endress+Hauser instruments: what we need

The order code on the nameplate (for example PMP71-1QA1S11RDADU) defines every option: measuring range, process connection, materials, output, approvals and display. Send it with the serial number; Endress+Hauser can retrieve the original configuration from the serial number. For a new instrument add the medium, operating and design pressure and temperature, range, line size or tank geometry, process connection, wetted materials, output or protocol (4-20 mA HART, PROFIBUS, FOUNDATION Fieldbus, Modbus) and the hazardous-area approval required. Older models such as Promass 80/83 or Micropilot M are replaced by current Proline or 80 GHz Micropilot devices.

Facts checked against Endress+Hauser company history and product information, October 2026.

FAQ

Questions about the Endress+Hauser PMP131 Cerabar T Compact Pressure

Answers from our engineering team on installation, compatibility and integration.

01What ranges does the PMP131 cover?
The piezoresistive PMP131 measures from 0 to 1 bar (15 psi) up to 0 to 400 bar (6000 psi), in absolute or gauge forms, with 316L wetted parts.
02What outputs can the PMP131 give?
A 2-wire 4 to 20 mA current output, a 3-wire 0 to 10 V voltage output with a 5 kilo-ohm maximum load, or a PNP switch output with 500 mA maximum on-state current.
03How accurate is the PMP131?
The reference accuracy for the PMP131 and PMP135 is 0.5 percent of the upper range limit, with long-term stability of 0.15 percent of URL per year, as listed in the datasheet.
04How fast is the PMP131?
The T90 rise time is 2 to 5 ms for the PMP131 and PMP135, compared with 20 ms for the ceramic PMC131.
05What should my RFQ for a PMP131 include?
State PMP131, the pressure range, absolute or gauge, thread, electrical connection (plug or cable), output type, approvals and quantity.