Endress+Hauser PMD75B Deltabar High-Static DP Transmitter

Description

Endress+Hauser PMD75B Deltabar introduction

The Endress+Hauser PMD75B is the high-performance differential pressure transmitter of the Deltabar B series. Technical Information TI01511P lists differential ranges up to 40 bar (600 psi), gauge and absolute measuring cells up to 250 bar (3750 psi), static pressure up to 420 bar (6300 psi) and accuracy up to 0.035 percent. It communicates over 4-20 mA HART, PROFINET over Ethernet-APL or PROFIBUS PA and carries the backlit display, Bluetooth and Heartbeat features of the generation.

It is the model to look at when a process has both a high line pressure and a small differential, such as a flow element on a high-pressure gas line, or when the measured medium attacks stainless steel and requires tantalum or Monel.

Notable properties of PMD75B

  • Wetted parts available in 316L, Alloy C276, tantalum and Monel (Alloy 400).
  • Cells from 10 mbar to 40 bar differential, plus 160 and 250 bar gauge or absolute cells.
  • Ambient range down to minus 40 degrees Celsius standard, minus 50 or minus 54 optional with restrictions.
  • Dual-compartment L-form housing at 1.7 kg for confined installations.
  • PROFIBUS PA operation under FISCO at 9 to 17.5 V DC or entity at 9 to 24 V DC.
  • Industrial surge test level of 1000 V line to earth on DC and signal ports.

PMD75B data

Parameter Value
Differential range up to 40 bar (600 psi)
Gauge and absolute cells up to 250 bar (3750 psi)
Static pressure up to 420 bar (6300 psi)
Accuracy, 10 mbar cell 0.075 percent standard, 0.05 percent Platinum at turn down 1 to 1
Accuracy, 100 mbar cell 0.05 percent standard; 0.04 percent Platinum to turn down 5 to 1
Accuracy, 500 mbar to 40 bar cells 0.05 percent standard to 15 to 1; Platinum 0.035 percent to 5 to 1
Accuracy, 160 and 250 bar cells 0.10 percent for turn down 1 to 5, not available in Platinum
Long-term, 10 and 30 mbar cells 0.20 percent 1 year; 0.28 percent 5 years; 0.31 percent 10 years; 0.34 percent 15 years
Long-term, 100 mbar cell 0.08 percent 1 year; 0.12 percent 5 years; 0.20 percent 10 years
HART supply 10.5 to 35 V DC; Ex i 10.5 to 30 V DC
PROFIBUS PA supply 9 to 32 V DC; nominal current 14 mA
Ambient temperature minus 40 to plus 85 degrees Celsius standard
Housing weight 1.1 kg single; 1.4 kg aluminium dual; 3.3 kg stainless dual; 1.7 kg L-form dual

How PMD75B differs from PMD55B and PMD50

Three numbers separate the PMD75B from its neighbours. Static pressure: 420 bar against 250 bar. Accuracy: 0.035 percent against 0.055 percent for the PMD55B and 0.05 percent for the PMD50. Material options: tantalum and Monel on top of 316L and Alloy C276, while the PMD50 offers only 316L and Alloy C276 with gold coating. The PMD75B also adds PROFIBUS PA, which the PMD55B and PMD50 do not list.

The long-term drift advantage is clearest on the 10 and 30 mbar cells: 0.20 percent in the first year compared with 0.25 percent for the PMD55B. On the 100 mbar cell the PMD75B holds 0.08 percent in the first year against 0.18 percent for the PMD55B, more than twice as stable.

Typical PMD75B applications in Oman and GCC

The PMD75B is chosen for gas export and injection flow measurement across orifice plates, for separator level with pressure overlay at several hundred bar, and for corrosive chemical lines in desalination and petrochemical plants where tantalum or Monel wetted parts are required. It also suits plants with PROFIBUS PA segments that need a high-performance DP transmitter without changing the network.

Selecting a PMD75B cell and material

The PMD75B datasheet separates two families of cells. Differential cells run from 10 mbar to 40 bar, and they are used for flow, filter and level duty. Gauge and absolute cells of 160 and 250 bar are offered for plain pressure, and their accuracy is 0.10 percent at turn down 1 to 5, not available in Platinum. Select the differential cell whose upper range limit is just above your largest differential, and then confirm that the static pressure of the line is below 420 bar for the process flange and gasket chosen, and check the flange and gasket limits in the datasheet because they can be lower than the cell rating.

For material, 316L covers most hydrocarbon and water duties. Alloy C276 is selected for chlorides and sour or acidic media, tantalum for strong acids, and Monel (Alloy 400) for hydrofluoric acid and seawater service. Tell us the medium, concentration and temperature in the RFQ so that the correct membrane and flange material are quoted with the PMD75B.

Installation notes for PMD75B

Mount the PMD75B on a manifold rated above the static pressure, and tighten the process flange bolts to the torque specified for the gasket type. Temperature at the transmitter must follow the derating curve that starts at 90 degrees Celsius without manifold. The L-form housing allows the cover to be removed in cramped spaces. After mounting, equalise and zero at operating static pressure.

Care and verification of PMD75B

Use Heartbeat verification to document that the transmitter still meets its specification without removing it from the line, and inspect the manifold and impulse lines for plugging at each turnaround. Because the quoted drift is stated against the upper range limit, re-ranging a large cell to a small span magnifies the relative drift, so choose the smallest cell that covers your span.

Documents and quotation

The Technical Information TI01511P for PMD75B is attached as the datasheet. Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer. Send PMD75B, quantity and configuration details with your RFQ.

Delivery to Sites in Oman and the GCC

Delivery of the Endress+Hauser PMD75B is arranged to suit the site, whether that is a plant in Sohar or Duqm, an oilfield location in the interior, or a project elsewhere in the GCC. Tell us the delivery point, any site access or gate-pass requirements, and the person who will receive the goods. Goods are packed to protect sensitive pressure transmitters equipment during road or air transport, and shipping details are shared once the order is dispatched so that your stores team can plan receipt and inspection.

Datasheets & Documents
  • PDF Datasheet Endress+Hauser PMD75B Deltabar High-Static DP Transmitter Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Endress+Hauser makes process measurement instruments for flow, level, pressure, temperature and liquid analysis. Georg H. Endress and Ludwig Hauser started the business in Lörrach, Germany, on 1 February 1953, first selling level meters, and the Endress+Hauser name has been used since 1957. The group is headquartered in Reinach, Switzerland, and has been wholly owned by the Endress family since 1975.

Seven Star LLC is an independent supplier listing Endress+Hauser flowmeters, level and pressure transmitters, temperature assemblies and liquid analysis products for buyers in Oman and the GCC. Instruments are matched from the full order code on the nameplate.

Endress+Hauser company profile

Origin Founded in Lörrach, Germany, on 1 February 1953 by the Swiss engineer Georg H. Endress and Ludwig Hauser
Headquarters Endress+Hauser AG, Reinach (Basel-Landschaft), Switzerland
Ownership Family owned; the Endress family has been the sole shareholder since 1975, and a family charter commits the company to stay family owned
Size About 16,500 employees and sales of about EUR 3.7 billion (2023, as reported by the company)
Product families Proline flowmeters, Micropilot and Levelflex radar, Liquiphant switches, Cerabar and Deltabar pressure, iTEMP and iTHERM temperature, Liquiline and Memosens liquid analysis

Milestones

  • 1953: Georg H. Endress and Ludwig Hauser start the business in Lörrach on 1 February, first selling level meters.
  • 1957: The Endress+Hauser name comes into use.
  • 1975: The Endress family becomes sole shareholder.
  • 2023: About 16,500 employees and sales of about EUR 3.7 billion, as reported by the company.

Endress+Hauser product lines

Line Measuring principle Typical use
Promass F 300, Promass 80F, 83F/84F Coriolis mass flow Direct mass flow and density of liquids and gases, hydrocarbon transfer and chemical dosing
Promag P 300, P 500, H 300, W 400, W 10, 50P Electromagnetic flow Water, wastewater, slurries and other conductive liquids
Prowirl F 200, 72F Vortex flow Steam, gas and low-viscosity liquids
Prosonic Flow 92F, t-mass 65F Ultrasonic and thermal mass flow Water lines; compressed air and industrial gases
Micropilot FMR62, FMR67, FMR60, FMR10B, FMR20, FMR50, FMR52 Free-space radar level, 80 GHz and 26 GHz Tanks, process vessels, silos and open water channels
Levelflex FMP51, FMP50, FMP54, FMP55 Guided wave radar level Level and interface in vessels, separators and bypass chambers
Liquiphant FTL51, Prosonic S FMU90 Vibronic point level; ultrasonic level High and low level alarms, pump protection, open channel flow
Cerabar PMC71B, PMP51B, PMP71, Deltabar PMD78B, PMD75 Pressure and differential pressure Process pressure, hydrostatic level and DP flow
iTEMP TMT82, TMT162, iTHERM TM101, TR66, TR10 Temperature transmitters and thermometers RTD and thermocouple points with thermowells
Liquiline CM444, CM442, Memosens CPS11E, Stamolys reagents Liquid analysis pH, ORP, conductivity, dissolved oxygen and nutrient analysis in water treatment and process

Where Endress+Hauser instruments are used

In Oman and the GCC, Endress+Hauser instruments are common in oil and gas production and processing, refineries and petrochemicals, desalination and water networks, wastewater treatment, power plants and food and beverage plants. Coriolis meters measure hydrocarbon and chemical flows, electromagnetic meters cover water and wastewater, radar transmitters handle tank and silo level, and Liquiline analysers monitor water quality. Proline 300 and 500 flowmeters include Heartbeat Technology for in-situ verification, which reduces the need to remove a meter for checks.

Ordering Endress+Hauser instruments: what we need

The order code on the nameplate (for example PMP71-1QA1S11RDADU) defines every option: measuring range, process connection, materials, output, approvals and display. Send it with the serial number; Endress+Hauser can retrieve the original configuration from the serial number. For a new instrument add the medium, operating and design pressure and temperature, range, line size or tank geometry, process connection, wetted materials, output or protocol (4-20 mA HART, PROFIBUS, FOUNDATION Fieldbus, Modbus) and the hazardous-area approval required. Older models such as Promass 80/83 or Micropilot M are replaced by current Proline or 80 GHz Micropilot devices.

Facts checked against Endress+Hauser company history and product information, October 2026.

FAQ

Questions about the Endress+Hauser PMD75B Deltabar High-Static DP

Answers from our engineering team on installation, compatibility and integration.

01How high is the static pressure rating of the PMD75B?
The datasheet front page gives a static pressure of up to 420 bar (6300 psi), compared with 250 bar on the PMD55B and PMD50.
02What accuracy does the PMD75B reach?
Up to 0.035 percent is stated; the 500 mbar to 40 bar standard cells hold 0.05 percent and Platinum versions hold 0.035 to 0.04 percent in the listed turn down ranges.
03Which wetted materials can the PMD75B have?
Membranes and flange raised faces can be 316L (1.4435), Alloy C276, tantalum or Monel (Alloy 400), which suits aggressive media.
04What housing options exist for the PMD75B?
Single compartment at 1.1 kg, dual compartment in aluminium at 1.4 kg or stainless steel at 3.3 kg, and an L-form dual compartment at 1.7 kg.
05What do you need for a PMD75B quotation?
Send PMD75B with the differential cell, static pressure, material, process flange, protocol (HART, APL or PROFIBUS PA), approvals and quantity.