Brand: Hanna Instruments
Hanna HI97733 Ammonia High Range Photometer 0-100 mg/L
Description
Hanna HI97733 Ammonia high range portable photometer
The Hanna HI97733 is a pocket-sized, single-parameter photometer built for concentrated ammonia streams that would overrange the low and medium models. Its range is 0.0 to 100.0 mg/L expressed as ammonium, so streams such as sidestream liquors, sour-water related samples and fertiliser-plant condensates can be read without serial dilution. Seven Star LLC is an independent supplier in Muscat and is not affiliated with the manufacturer; we quote the HI97733 to industrial and laboratory buyers across Oman and the GCC.
The model number follows Hanna’s HI977xx pattern for dedicated photometers: the last digits identify the chemistry, so HI97733 always means ammonia high range and cannot be reconfigured for another analyte. Reagents carry their own HI93xxx codes, listed below, and must be ordered separately from the meter.
HI97733 technical specifications
| Parameter | Specification |
|---|---|
| Manufacturer / part number | Hanna Instruments HI97733 |
| Parameter | Ammonia high range |
| Range | 0.0 to 100.0 mg/L as NH4+ |
| Resolution | 0.1 mg/L |
| Accuracy at 25 °C | ±0.5 mg/L ±5 % of reading |
| Method | Adaptation of ASTM D1426 Nessler method |
| Light source | LED with narrow-band interference filter at 420 nm |
| Filter bandwidth / wavelength accuracy | 8 nm / ±1.0 nm |
| Detector | Silicon photocell |
| Cuvette | Round, 24.6 mm diameter (22 mm inside) |
| Data logging | 50 readings (auto logging) |
| Display | 128 × 64 pixel monochrome LCD with backlight |
| Power | Three 1.5 V AA alkaline batteries; more than 800 measurements without backlight |
| Operating environment | 0 to 50 °C; 0 to 100 % RH |
| Ingress protection | IP67, floating case |
| Dimensions / weight | 142.5 × 102.5 × 50.5 mm / 380 g with batteries |
| Warranty | Two years |
How the HI97733 measurement works
The same Nessler reaction as the other ammonia models is used, but the high-range reagent set uses a 9 mL liquid reagent B volume mixed with the sample before zeroing, which sets up the larger concentration window. The reacted colour is measured at 420 nm against the unreacted blank.
Inside the case a reference detector tracks the LED output, so battery sag or a warm ambient does not move the blank. A lens gathers the light leaving the cuvette, which is why scratches and small glass imperfections matter less than with a simple colorimeter, and the cap notch locks the cuvette in the same orientation every time. A built-in tutorial mode walks the operator through sampling, reagent quantity and timers on screen, so the instrument can be handed to a technician who is not a chemist.
Reagents, procedure and interferences for HI97733
The cuvette is filled with 9 mL of HI93733B-0 and topped up to the 10 mL mark with sample, inverted five times and zeroed; four drops of HI93733A-0 are then added, mixed, and the meter runs a 3 minute 30 second countdown before reading.
| Code | Description | Quantity per test |
|---|---|---|
| HI93733A-0 | Ammonia HR Reagent A | 4 drops |
| HI93733B-0 | Ammonia HR Reagent B | 9 mL |
Reagent sets are supplied as HI93733-01 (100 tests) and HI93733-03 (300 tests). Follow the zeroing order given in the manual for this method, and read within the time the manual allows after the countdown ends.
Known interferences listed for this method:
- Hardness above 1 g/L
- Iron
- Sulfide, which may cause turbidity
- Acetone above 0.1 %, alcohols, aldehydes, aliphatic and aromatic amines, chloramines, glycine or urea above 10 mg/L; distillation removes these
Where the HI97733 sits among Hanna photometers
Treat the HI97733 as the screening tool for concentrated ammonia and the HI97715 or HI97700 as the finishing tools once values drop. Because results are shown in NH4+, convert to nitrogen basis if your reporting format requires it.
| Model | Parameter | Range | LED filter |
|---|---|---|---|
| HI97700 | Ammonia low range | 0.00 to 3.00 mg/L as NH3-N | 420 nm |
| HI97715 | Ammonia medium range | 0.00 to 10.00 mg/L as NH3-N | 420 nm |
| HI97733 | Ammonia high range | 0.0 to 100.0 mg/L as NH4+ | 420 nm |
| HI97701 | Free chlorine | 0.00 to 5.00 mg/L as Cl2 | 525 nm |
| HI97762 | Free chlorine ultra low range | 0.000 to 0.500 mg/L as Cl2 | 525 nm |
| HI97734 | Free and total chlorine, high range | 0.00 to 10.00 mg/L as Cl2 | 525 nm |
| HI97725 | Chlorine, cyanuric acid and pH | 0.00 to 5.00 mg/L Cl2; 0 to 80 mg/L CYA; 6.5 to 8.5 pH | 525 nm |
| HI97706 | Phosphorus | 0.0 to 15.0 mg/L as P | 525 nm |
| HI97714 | Cyanide | 0.000 to 0.200 mg/L as CN– | 610 nm |
| HI97718 | Iodine | 0.0 to 12.5 mg/L as I2 | 525 nm |
| HI97731 | Zinc | 0.00 to 3.00 mg/L as Zn | 575 nm |
| HI97737 | Silver | 0.000 to 1.000 mg/L as Ag | 575 nm |
Selection and ordering checklist for HI97733
- Confirm expected ammonium is between 0.1 and 100.0 mg/L
- Note the larger 9 mL reagent B volume, which means reagent sets are consumed faster than on the low-range models
- Order HI93733-01 or HI93733-03 separately
- Consider CAL Check kit HI97733-11 for routine validation
- Confirm waste handling for the reacted Nessler solution under site rules
Validation, handling and maintenance of HI97733
Validate with the HI97733-11 CAL Check cuvettes before critical batches. Rinse the cuvette thoroughly between samples because concentrated ammonia and Nessler residue can carry over, and clean with the Hanna cleaning solution if staining appears.
Keep cuvettes clean with the cleaning solution and a lint-free cloth, never touch the lower half of the glass, and store the meter in its carrying case. Replace the three AA cells when the battery icon appears; the meter keeps its logged readings and settings. For long storage between field trips remove the batteries and close the reagent bottles tightly.
Typical applications of HI97733 in Oman and the GCC
Fertiliser and ammonia-related process units, refinery sour-water and wastewater treatment, aquaculture hatchery sumps and food or beverage plants in the GCC benefit from a meter that reads up to 100 mg/L directly. In desalination chemical-cleaning waste and in industrial evaporation ponds it provides a quick on-site number before laboratory confirmation.
Heat is a practical factor in the Gulf. Reaction times in the manual are given at 25 °C, so samples that sit in a vehicle at much higher temperature should be allowed to equilibrate, and the meter itself is rated for 0 to 50 °C. The sealed optics and IP67 floating case also cope with dust and spray on plant walkdowns.
Information to send with an RFQ for HI97733
- Quantity of meters and whether reagent sets (HI93733-01 or HI93733-03; CAL Check HI97733-11) and CAL Check cuvettes are also required
- Sample matrix and expected concentration, so we can confirm the range suits the application
- Delivery location in Oman or the GCC and required lead time
- Any need for spare cuvettes, carrying case or cleaning solution
- Project or tender reference, if applicable
Documents and quotation
The instruction manual for the HI97733 is attached on this page and covers specifications, the reagent procedure, CAL Check validation and error messages.
Send the part number HI97733, the quantity and your delivery location and we will reply with availability and a quotation.
Other Models in the Hanna Hi977xx Photometer Range
Seven Star LLC supplies models from the Hanna Hi977xx Photometer range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.
| Model | Category | |
|---|---|---|
| Hanna HI97700 Ammonia Low Range Photometer 0-3 mg/L | Liquid Analytical Systems | View details |
| Hanna HI97715 Ammonia Medium Range Photometer 0-10 mg/L | Liquid Analytical Systems | View details |
| Hanna HI97733 Ammonia High Range Photometer 0-100 mg/L | Liquid Analytical Systems | You are viewing this model |
Datasheets & Documents
- Manual Hanna HI97733 Instruction Manual Download
Reviews (0)
How would you rate this product? Cancel reply
About brand
Hanna Instruments is a maker of analytical instruments for water and liquid testing, founded in 1978 in the Padua area of Italy by Oscar and Anna Nardo. Its early breakthrough was the pHep, an affordable pocket-sized pH tester. The company is now led by the founders' children, Martino and Pamela Nardo, with its headquarters in Woonsocket, Rhode Island, and offers more than 3,000 products.
Seven Star LLC is an independent supplier in Muscat. We list Hanna instruments for water, wastewater and laboratory users in Oman and the GCC.
Hanna Instruments company profile
| Founded | 1978, near Padua, Italy, by Oscar and Anna Nardo |
|---|---|
| Headquarters | Woonsocket, Rhode Island, USA |
| Leadership | Martino Nardo and Pamela Nardo |
| Early product | The pHep pocket pH tester |
| Product families | pHep and HI98 waterproof testers, edge benchtop meters, HI83 multiparameter photometers, Checker HC handheld colorimeters, HI931 and HI932 titrators, process controllers, reagents and calibration buffers |
Hanna product series
Portable meters include the HI98190 pH and ORP meter, the HI98192 EC, TDS and salinity meter, the HI98193 dissolved oxygen and BOD meter, the HI98199 pH, EC and DO meter and the HI9829 multiparameter water quality meter.
The HI98703 is an EPA compliant portable turbidity meter, the HI97711 a free and total chlorine photometer and the HI83414 a benchtop turbidity and chlorine meter.
Waterproof portable meters such as the HI98190 and HI98192 suit field use, with automatic temperature compensation and GLP data for audits. Turbidity measurement with the HI98703 follows EPA method 180.1, and the HI97711 photometer measures free and total chlorine for disinfection control.
Where Hanna instruments are used
Drinking water and desalination plants, wastewater treatment, environmental field surveys, aquaculture, cooling water and quality control labs.
Hanna Instruments products we list
Ordering Hanna instruments: what we need
Send the model, the parameters and ranges, the probe and cable length and any reagents or calibration solutions required.
Facts checked against Hanna Instruments' own company story and City of Woonsocket business profile, October 2026.
Browse Hanna Instruments products by category: Portable Test & Measurement.
FAQ
Questions about the Hanna HI97733 Ammonia High Range Photometer 0-100
Answers from our engineering team on installation, compatibility and integration.
01What does the HI97733 read?
Ammonia from 0.0 to 100.0 mg/L, displayed as ammonium (NH4+), with 0.1 mg/L resolution and accuracy of 0.5 mg/L plus 5 percent of reading at 25 degrees C.
02How is it different from HI97700 and HI97715?
The chemistry family is the same Nessler method, but HI97733 spans a much higher range and uses a 9 mL reagent B volume, whereas the others use drop reagents.
03Which reagent codes are used?
HI93733A-0 at four drops and HI93733B-0 at 9 mL, supplied as HI93733-01 for 100 tests or HI93733-03 for 300 tests.
04Does the meter log readings?
Yes, it stores 50 readings with auto logging and keeps GLP information, including CAL Check validation records, in the instrument.
05What power does it need?
Three 1.5 V AA alkaline batteries, giving more than 800 measurements without the backlight.