Brand:

Yokogawa EJA130E-JHS4G-812EK High Static DP, 316L Bolts

Description

Mid-range DP with stainless bolts: EJA130E-JHS4G-812EK

Yokogawa EJA130E-JHS4G-812EK takes the EJA130E platform to the H capsule, a span of 5 to 500 kPa, and fits 316L stainless bolts and nuts together with 1/2 NPT female process connectors. The installation is horizontal piping with the high-pressure side on the right (-8). A digital indicator with a range-setting push button and a 316 SST L-type bracket are fitted, and the output is HART.

Reading EJA130E-JHS4G-812EK

Code position Meaning
-J 4 to 20 mA DC with HART 5 or HART 7 digital communication (selected at ordering)
H Capsule: span 5 to 500 kPa
S Wetted parts: F316 SST cover flange, ASTM CF-8M connector, Hastelloy C-276 diaphragm, Teflon-coated 316L SST gasket, 316 SST plug
4 Process connection: with 1/2 NPT female process connector
G Bolts and nuts: 316L SST (32 MPa MWP)
-8 Installation: horizontal piping, right side high pressure
1 Amplifier housing: cast aluminum alloy
2 Electrical: 1/2 NPT female, two electrical connections, no blind plugs
E Integral indicator: digital indicator with range-setting push button
K Mounting bracket: 316 SST 2-inch pipe bracket, L type (vertical piping)

H capsule numbers

Reference accuracy is plus or minus 0.055 percent of span for spans of 100 kPa or larger, with a term of 0.005 plus 0.01 times URL over span below that; URL is 500 kPa. The ambient temperature effect is plus or minus (0.07 percent of span plus 0.015 percent of URL) per 28 degrees C, the best figure of the three capsules, and static pressure effect on zero is plus or minus 0.028 percent of URL per 6.9 MPa. Stability is plus or minus 0.1 percent of URL over ten years and response time is 150 ms. With 316L bolts (code G) the maximum working pressure is 32 MPa.

Data table, EJA130E-JHS4G-812EK

Item Value
Capsule H 5 to 500 kPa; range minus 500 to 500 kPa
Bolts 316L SST, MWP 32 MPa
Process connection 1/2 NPT female connector
Orientation Horizontal piping, right side high pressure
Indicator E with push button
Bracket K, 316 SST L type
Output HART 5/7
Burst 132 MPa
Weight 6.8 kg

Uses

Gas flow orifice metering at pipeline pressures, differential pressure on high-pressure filters and coalescers, level on high-pressure vessels and drum levels. The H capsule serves most orifice sizing results between 5 and 500 kPa.

Bolts and pressure rating

The datasheet ties maximum working pressure to the bolt code: C, J and G allow 32 MPa, while K and H allow 41.3 MPa. Overpressure effect is plus or minus 0.03 percent of URL for C, J, G and 0.04 percent for K, H. For this part number the 316L stainless bolts G give 32 MPa working pressure.

Fitting

Use the 1/2 NPT female connectors with suitable sealing. Install with the right-hand side high pressure. Attach the K-type L bracket to a vertical two-inch pipe; the transmitter body itself follows the horizontal-piping arrangement indicated by installation code -8.

Service

Keep the impulse lines clean and leak-free, and always equalize through the manifold before a zero trim, since a trapped differential would show up as a zero error on the 5 to 500 kPa span. After each pressure cycle at the limits of the working pressure, check the cover flange bolts for relaxation and the vent and drain plugs for seepage. Because the transmitter also reports static pressure, compare that value with the line gauge as a quick health check, and review the self-diagnostic status through HART at planned shutdowns.

Keep the nameplate data on file: the capsule, bolt code and orientation digit decide which spare you need.

Electrical data and standards for EJA130E-JHS4G-812EK

HART 5 or HART 7 is selected at ordering. The two-wire loop works from 10.5 to 42 V DC with 0 to 1290 ohm of load in operation, and with a 24 V DC supply up to 550 ohm can be used. Output range is 3.6 to 21.6 mA. The 5-digit display and 6-digit unit field can rotate through measured differential pressure, differential pressure in percent, scaled differential pressure and static pressure. Update period is 45 ms for differential pressure and 360 ms for static pressure. Applicable standards include EN 61326-1 Class A and EN 61010-1, and the transmitter is IP66/IP67 with Type 4X. SIL 2 capability applies for the standard HART configuration.

Ambient limits are minus 40 to 85 degrees C and minus 30 to 80 degrees C with the LCD, and process temperature is limited to minus 40 to 120 degrees C. Lower limits are minus 15 degrees C when the 1/4 or 1/2 NPT process connectors are selected, according to the datasheet.

Quick answers about EJA130E-JHS4G-812EK

  • What span does the H capsule give? The H capsule spans 5 to 500 kPa, with a range of minus 500 to 500 kPa.
  • What is the maximum working pressure with code G bolts? 32 MPa (4500 psi), the same as codes C and J.
  • Which connection do I get? Code 4, with a 1/2 NPT female process connector.
  • Is there a push button? Yes, indicator code E is a digital indicator with a range-setting push button.
  • What does -8 mean? Horizontal piping with the right side of the transmitter as the high-pressure side.

Documents and quotation

Say whether the lines are 1/4 inch or 1/2 inch NPT.

The datasheet for this transmitter, Yokogawa GS 01C31B04-01EN, is supplied as a PDF with the listing. For a quotation, send the full part number EJA130E-JHS4G-812EK, the quantity, the tag numbers, the line pressure and the differential pressure range. Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer. Availability and delivery time are confirmed per RFQ.

Delivery to Sites in Oman and the GCC

Delivery of the Yokogawa EJA130E-JHS4G-812EK is arranged to suit the site, whether that is a plant in Sohar or Duqm, an oilfield location in the interior, or a project elsewhere in the GCC. Tell us the delivery point, any site access or gate-pass requirements, and the person who will receive the goods. Goods are packed to protect sensitive pressure transmitters equipment during road or air transport, and shipping details are shared once the order is dispatched so that your stores team can plan receipt and inspection.

Other Models in the EJA130E Range

Seven Star LLC supplies models from the EJA130E range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.

ModelCategory
Yokogawa EJA130E-JMS5J-912NB High Static Pressure DPPressure TransmittersView details
Yokogawa EJA130E-JHS4G-812EK High Static DP, 316L BoltsPressure TransmittersYou are viewing this model
Datasheets & Documents
  • PDF Datasheet Yokogawa EJA130E-JHS4G-812EK High Static DP, 316L Bolts Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Yokogawa is the Japanese measurement and industrial automation company behind the DPharp EJA and EJX pressure transmitters, CENTUM distributed control systems and ProSafe-RS safety systems. Tamisuke Yokogawa set it up in 1915 as an electric meter research institute in Shibuya, Tokyo. It began selling electric meters in 1917, was incorporated as Yokogawa Electric Works in 1920, and has been Yokogawa Electric Corporation since 1986. In 1975 it released CENTUM, which it describes as the world's first distributed control system.

Seven Star LLC is an independent supplier in Muscat. We list Yokogawa transmitters, flowmeters, analysers, controllers, recorders and CENTUM and ProSafe spare modules, and supply them to oil and gas, power, water and process plants in Oman and the GCC by model code.

Yokogawa at a glance

Established 1915, Shibuya, Tokyo, by Tamisuke Yokogawa with Ichiro Yokogawa and Shin Aoki
Milestones Electric meters from 1917; incorporated in 1920; CENTUM DCS in 1975; renamed Yokogawa Electric Corporation in 1986
Headquarters Tokyo, Japan
Product families Pressure and temperature transmitters, magnetic, vortex and Coriolis flowmeters, liquid and gas analysers, controllers and recorders, CENTUM VP DCS, ProSafe-RS safety systems and STARDOM controllers

Yokogawa products we list

Family Models Typical use
Pressure and DP transmitters EJA110E, EJA430E, EJA530E, EJX110A, EJX530A, EJX910A multivariable Pressure, level and DP flow measurement in process plants
Temperature transmitters YTA610, YTA710 RTD and thermocouple signals converted to 4-20 mA, HART or fieldbus
Flowmeters ADMAG AXG, ADMAG AXW, ROTAMASS TI, VY vortex, VF701 Water and conductive liquids, mass flow of hydrocarbons, steam and gas
Analysers TDLS8000 laser gas analyser, FLXA402, FLXA21 DO, E7042UD zirconia cell, K9471DS cover Combustion O2, in-situ gas analysis, pH, conductivity and dissolved oxygen
Controllers and recorders UT35A/UT32A, UT32A-D, UT55A/UT52A, UT75A, UP35A/UP32A, UP55A, YS1700, GX10/GX20, GP10/GP20, GM10, µR20000 Single-loop control, furnace programs and data recording
Control and safety systems CP451-10 processor, ADV161, ADV551, node interface unit, ProSafe-RS, ProSafe-RS Lite, ProSafe-SLS, FCN-500 Spares and expansions for CENTUM DCS, safety systems and STARDOM
Battery packs S9129FA, S9185FA, S9548FA Memory backup batteries for DCS and controller modules

The EJA and EJX transmitters use Yokogawa's DPharp sensor: two silicon resonators whose vibration frequency changes with pressure, so the output is a digital frequency rather than an analogue strain signal. The model code describes the instrument completely. In EJA110E, "110" is a differential pressure transmitter, and the letters after the dash set the span, wetted materials, process connection, electrical output and approvals. When you replace a transmitter, sending the full code from the nameplate avoids a mismatch in range or materials.

Where Yokogawa products are used

In Oman and the GCC, Yokogawa transmitters, analysers and control systems are found in oil and gas production, gas processing and LNG, refineries and petrochemical plants, power and desalination stations, and water treatment. Many plants still run CENTUM CS 3000 and CENTUM VP systems that need spare I/O modules and batteries.

Ordering Yokogawa products: what we need

Send the full model and suffix code from the nameplate (for example EJA110E-JMS5J-912NN/KF22) and the style code. For DCS modules, add the module type and system version. For analysers, send the sensor type, measuring range and process conditions.

Facts checked against Yokogawa's corporate history, October 2026.

FAQ

Questions about the Yokogawa EJA130E-JHS4G-812EK High Static DP, 316L

Answers from our engineering team on installation, compatibility and integration.

01What span does the H capsule give?
The H capsule spans 5 to 500 kPa, with a range of minus 500 to 500 kPa.
02What is the maximum working pressure with code G bolts?
32 MPa (4500 psi), the same as codes C and J.
03Which connection do I get?
Code 4, with a 1/2 NPT female process connector.
04Is there a push button?
Yes, indicator code E is a digital indicator with a range-setting push button.
05What does -8 mean?
Horizontal piping with the right side of the transmitter as the high-pressure side.