Endress+Hauser TST310 Pt100 Cable Thermometer
Description
Spare-part view of the Endress+Hauser TST310 Pt100 cable thermometer
Endress+Hauser TST310 is a resistance thermometer with a flexible connecting cable and an anti-kink spring at the sensor exit. Technical Information TI00085T describes it for machinery, laboratory equipment and plants working with gases or liquids such as air, water and oil. It is delivered as a screw-in or insertion probe with user-defined insertion length, a single or duplex Pt100 and tolerance Class A, B or AA to IEC 60751. Versions for hazardous locations exist as intrinsically safe (Ex ia) and non-sparking (Ex nA). Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer.
Two probe constructions, two measuring ranges
The TST310 is built either as a bendable probe or as a rigid one, and the range follows from that choice. The bendable type uses mineral-insulated sheathed cable and measures -200 to +600 degrees C. The non-bendable type has insulated sensor wires inside a stainless steel tube and covers -50 to +250 degrees C. The cable resistance may reach 50 ohm per lead. The connecting cable limits the ambient temperature at the cable: PVC/PVC 80 degrees C, PTFE/silicone 180 degrees C and PTFE/PTFE 200 degrees C. Thin-film elements keep Class A from -30 to +200 degrees C.
| Sensor | Pt100, single or duplex, Class A, B or AA (IEC 60751) |
|---|---|
| Range, bendable MI cable | -200 to +600 degrees C |
| Range, rigid wires in tube | -50 to +250 degrees C |
| Tolerance Class AA | +/-(0.1 + 0.0017 |t|) degrees C |
| Tolerance Class A | +/-(0.15 + 0.002 |t|) degrees C |
| Tolerance Class B | +/-(0.3 + 0.005 |t|) degrees C |
| Response, 6 mm MI cable | t50 3.5 s, t90 8 s |
| Response, 3 mm MI cable | t50 2 s, t90 5 s |
| Response, 6 mm insulated wires | t50 9 s, t90 28 s |
| Response, 3 mm insulated wires | t50 6 s, t90 18 s |
| Insulation resistance | At least 100 megohm at 100 V DC |
| Max. static pressure | 75 bar (1 088 psi) |
| Protection | IP65 |
| Vibration | 3 g, 10 to 500 Hz |
| Core cross-section | Up to 0.382 mm2 (AWG 22) with 5 mm end sleeves |
| Weight | From 100 g; 150 g for NL = 100 mm with brazed G1/2 |
Fitting the TST310: threads, ferrules and pressure
Process attachment is by a brazed thread or an adjustable compression fitting called TA50. With the compression fitting the thread options include G1/8 (clamping ring length 35 mm), G1/4, G1/2, NPT 1/8, NPT 1/4, NPT 1/2, M10x1 and M8x1. Stainless steel 316 clamping rings take 40 bar at 20 degrees C up to 800 degrees C but can be used once. PTFE rings allow re-positioning and are rated 10 bar at 20 degrees C and 200 degrees C. The brazed G1/4 version tolerates 75 bar at 20 degrees C. Spare ferrules are sold in packs of ten: order number 60011600 for 6.1 mm PTFE, 60011599 for 6.1 mm SS316 and 60011598 for 3 mm PTFE.
How TST310 differs from the TSC310 thermocouple version
The thermocouple twin TSC310 uses the same cable-probe concept but measures only types J and K, reaches +1 100 degrees C and holds 40 bar, while the TST310 holds 75 bar and tops out at +600 degrees C. Choose the TST310 when a linear, repeatable Pt100 signal below 600 degrees C matters more than the extra top-end range.
Typical duties around Muscat and the Gulf
Practical places for a TST310 are chilled-water return lines, lube-oil and hydraulic skids, gearbox oil sumps, compressor aftercoolers, test benches and laboratory rigs. The long flexible cable lets the probe be tucked into tight machine housings while a transmitter or PLC input sits several metres away. For Ex ia loops, the datasheet instructs readers to follow the separate Ex documentation, so always state the required protection when you ask for an offer.
Installation, wiring and checks for TST310
Wire the flying leads straight to a transmitter or input card: colour code red/red and white/white pairs for a four-wire Pt100, with two Pt100 in two-wire for duplex. Do not exceed the bending radius of the cable and do not move the compression fitting once a stainless ring has been tightened. Immerse the probe deeper than its minimum length so that conduction along the sheath does not distort the reading. Self-heating is small, but keep the measuring current at the transmitter’s specified value.
Accuracy planning with TST310
Because tolerances grow with temperature, the class is worth deciding early. At 100 degrees C a Class AA element allows +/-0.27 degrees C, Class A +/-0.35 degrees C and Class B +/-0.8 degrees C, all from the IEC 60751 formulas printed in the datasheet. At 200 degrees C the same classes give +/-0.44, +/-0.55 and +/-1.3 degrees C. On a loop that feeds energy metering or a heat-balance calculation, the better class pays back quickly. For plain monitoring of oil temperature on a gearbox, Class B is normally sufficient. Remember that the cable can add error: with a two-wire connection the lead resistance is included in the reading, so use three- or four-wire connections on long cable runs.
Maintenance
This type is not field-repairable. Check insulation resistance against the 100 megohm value, look for kinks at the spring, and verify with a dry-block. If a stainless ring has been loosened, replace it with a spare ferrule from the order numbers above.
Documents and quotation
The TI00085T datasheet is attached for the TST310. Please send an RFQ to Seven Star LLC listing the part number TST310, the order code or the needed insertion length, cable material, thread, quantity and delivery destination.
Typical Questions From Maintenance Teams
Maintenance engineers replacing a Endress+Hauser TST310 usually ask whether the new unit is a direct fit, whether settings can be copied from the old one, and whether any accessories must be ordered separately. The answer depends on the exact order code and the generation of the installed rtd & thermocouple sensors, so the nameplate details are the best starting point. Send them with your request and we will check the match, point out any differences in connections or options, and list the parts needed so that the job can be finished in one visit.
Other Models in the Cable Probes Range
Seven Star LLC supplies models from the Cable Probes range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.
| Model | Category | |
|---|---|---|
| Endress+Hauser TSC310 Thermocouple Cable Probe | RTD & Thermocouple Sensors | View details |
| Endress+Hauser TST310 Pt100 Cable Thermometer | RTD & Thermocouple Sensors | You are viewing this model |
Datasheets & Documents
- Datasheet Endress+Hauser TST310 Pt100 Cable Thermometer Datasheet Download
Reviews (0)
How would you rate this product? Cancel reply
About brand
Endress+Hauser makes process measurement instruments for flow, level, pressure, temperature and liquid analysis. Georg H. Endress and Ludwig Hauser started the business in Lörrach, Germany, on 1 February 1953, first selling level meters, and the Endress+Hauser name has been used since 1957. The group is headquartered in Reinach, Switzerland, and has been wholly owned by the Endress family since 1975.
Seven Star LLC is an independent supplier listing Endress+Hauser flowmeters, level and pressure transmitters, temperature assemblies and liquid analysis products for buyers in Oman and the GCC. Instruments are matched from the full order code on the nameplate.
Endress+Hauser company profile
| Origin | Founded in Lörrach, Germany, on 1 February 1953 by the Swiss engineer Georg H. Endress and Ludwig Hauser |
|---|---|
| Headquarters | Endress+Hauser AG, Reinach (Basel-Landschaft), Switzerland |
| Ownership | Family owned; the Endress family has been the sole shareholder since 1975, and a family charter commits the company to stay family owned |
| Size | About 16,500 employees and sales of about EUR 3.7 billion (2023, as reported by the company) |
| Product families | Proline flowmeters, Micropilot and Levelflex radar, Liquiphant switches, Cerabar and Deltabar pressure, iTEMP and iTHERM temperature, Liquiline and Memosens liquid analysis |
Milestones
- 1953: Georg H. Endress and Ludwig Hauser start the business in Lörrach on 1 February, first selling level meters.
- 1957: The Endress+Hauser name comes into use.
- 1975: The Endress family becomes sole shareholder.
- 2023: About 16,500 employees and sales of about EUR 3.7 billion, as reported by the company.
Endress+Hauser product lines
| Line | Measuring principle | Typical use |
|---|---|---|
| Promass F 300, Promass 80F, 83F/84F | Coriolis mass flow | Direct mass flow and density of liquids and gases, hydrocarbon transfer and chemical dosing |
| Promag P 300, P 500, H 300, W 400, W 10, 50P | Electromagnetic flow | Water, wastewater, slurries and other conductive liquids |
| Prowirl F 200, 72F | Vortex flow | Steam, gas and low-viscosity liquids |
| Prosonic Flow 92F, t-mass 65F | Ultrasonic and thermal mass flow | Water lines; compressed air and industrial gases |
| Micropilot FMR62, FMR67, FMR60, FMR10B, FMR20, FMR50, FMR52 | Free-space radar level, 80 GHz and 26 GHz | Tanks, process vessels, silos and open water channels |
| Levelflex FMP51, FMP50, FMP54, FMP55 | Guided wave radar level | Level and interface in vessels, separators and bypass chambers |
| Liquiphant FTL51, Prosonic S FMU90 | Vibronic point level; ultrasonic level | High and low level alarms, pump protection, open channel flow |
| Cerabar PMC71B, PMP51B, PMP71, Deltabar PMD78B, PMD75 | Pressure and differential pressure | Process pressure, hydrostatic level and DP flow |
| iTEMP TMT82, TMT162, iTHERM TM101, TR66, TR10 | Temperature transmitters and thermometers | RTD and thermocouple points with thermowells |
| Liquiline CM444, CM442, Memosens CPS11E, Stamolys reagents | Liquid analysis | pH, ORP, conductivity, dissolved oxygen and nutrient analysis in water treatment and process |
Where Endress+Hauser instruments are used
In Oman and the GCC, Endress+Hauser instruments are common in oil and gas production and processing, refineries and petrochemicals, desalination and water networks, wastewater treatment, power plants and food and beverage plants. Coriolis meters measure hydrocarbon and chemical flows, electromagnetic meters cover water and wastewater, radar transmitters handle tank and silo level, and Liquiline analysers monitor water quality. Proline 300 and 500 flowmeters include Heartbeat Technology for in-situ verification, which reduces the need to remove a meter for checks.
Ordering Endress+Hauser instruments: what we need
The order code on the nameplate (for example PMP71-1QA1S11RDADU) defines every option: measuring range, process connection, materials, output, approvals and display. Send it with the serial number; Endress+Hauser can retrieve the original configuration from the serial number. For a new instrument add the medium, operating and design pressure and temperature, range, line size or tank geometry, process connection, wetted materials, output or protocol (4-20 mA HART, PROFIBUS, FOUNDATION Fieldbus, Modbus) and the hazardous-area approval required. Older models such as Promass 80/83 or Micropilot M are replaced by current Proline or 80 GHz Micropilot devices.
Facts checked against Endress+Hauser company history and product information, October 2026.
Browse Endress+Hauser products by category: Flowmeters, Level Transmitter, Pressure Instrumentation, Temperature, Liquid Analysis Sensors, Liquid Analytical Systems.
FAQ
Questions about the Endress+Hauser TST310 Pt100 Cable Thermometer
Answers from our engineering team on installation, compatibility and integration.
01What temperature range does the Endress+Hauser TST310 cover?
The bendable mineral-insulated version covers -200 to +600 degrees C, while the rigid version with insulated wires in a steel tube covers -50 to +250 degrees C.
02What pressure can a TST310 withstand?
The datasheet gives a maximum static pressure of 75 bar (1 088 psi), but compression fittings are lower: 40 bar with an SS316 ring and 10 bar with a PTFE ring at 20 degrees C.
03Which tolerance classes are available on the TST310?
Pt100 sensors can be supplied in Class AA, A or B according to IEC 60751, in single or duplex versions.
04Is the TST310 available for hazardous areas?
The datasheet names intrinsically safe (Ex ia) and non-sparking (Ex nA) versions, and says the Ex documentation must be observed during installation.
05Which spare ferrules fit the TST310?
Ten-piece packs are listed: 60011600 PTFE for 6.1 mm, 60011599 SS316 for 6.1 mm and 60011598 PTFE for 3 mm probes.