Hanna Instruments HI98121 pH/ORP/Temperature Tester

Description

Hanna HI98121 pH, ORP and temperature tester: what it is

The Hanna Instruments HI98121 is a pocket tester that reads pH, oxidation-reduction potential (ORP, in mV) and temperature in one waterproof body. Unlike conductivity combo testers, the Hanna HI98121 is aimed at sites where redox control matters as much as acidity, such as disinfection, cyanide or chromium treatment, pool and spa water and aquariums. Seven Star LLC supplies the Hanna HI98121 in Oman and the GCC as an independent supplier in Muscat, typically for operators who need a quick and inexpensive instrument to confirm what an online analyser or test kit is showing.

The tester uses a two-button interface and a case designed to float if dropped. A stability tag appears while the reading settles, and a hold mode freezes the value. The battery indicator warns before power becomes too low for a reliable measurement, and the tester shuts itself off after eight minutes of non-use.

HI98121 sensors: replaceable pH cartridge and fixed ORP sensor

The pH side uses the HI73127 replaceable electrode, which snaps into a stainless-steel round connector and carries an extendable cloth junction. The ORP side is a fixed sensor built into the tip of the tester and is factory calibrated, so there is no ORP calibration routine to perform. Each tester is supplied with two HI70022 470 mV ORP test solution sachets, which are used to check that the redox reading is still correct rather than to adjust it.

HI98121 technical specifications

Parameter Specification
Manufacturer / part number Hanna Instruments / HI98121
pH range, resolution, accuracy –2.00 to 16.00 pH, 0.01 pH, ±0.05 pH at 25 °C
pH calibration Automatic, one or two points, with pH 4.01 / 7.01 / 10.01 or pH 4.01 / 6.86 / 9.18 buffer sets
ORP range, resolution, accuracy ±1000 mV, 1 mV, ±2 mV
ORP calibration Factory calibrated, fixed sensor
Temperature –5.0 to 60.0 °C, 0.1 °C resolution, ±0.5 °C accuracy; automatic compensation for pH
pH electrode HI73127 replaceable cartridge, included
Power 4 x 1.5 V batteries, about 250 hours of continuous use, auto-off after 8 minutes
Operating environment –5 to 50 °C, up to 100% relative humidity

Understanding the HI98121 ORP reading

ORP is a voltage between a platinum sensing surface and a reference, and it indicates the oxidising or reducing tendency of a solution rather than the concentration of any one chemical. In a chlorinated water system the ORP rises with free oxidant, which is why many treatment plants use a target value in mV as a secondary control. The HI98121 gives a fast spot reading of that value to a resolution of 1 mV, which is enough to confirm whether a dosing system is moving in the right direction. It should not be used to infer exact chlorine concentration, since pH, cyanuric acid, temperature and contamination of the sensor all affect the mV figure.

Because pH changes the meaning of an ORP reading, the HI98121 is convenient: the same dip gives both numbers, and ATC ensures the pH value is corrected for temperature.

Choosing the HI98121 among Hanna pocket testers

Related models help to confirm the right selection. The HI98120 is the ORP and temperature version without pH. The HI98129 and HI98130 add conductivity and TDS instead of ORP. If a replaceable ORP electrode, logging, GLP records or a wider temperature range is required, a waterproof meter such as the HI98190 or HI98191 with a separate electrode is a better match. The HI98121 is best where simplicity, portability and low cost outweigh the need for data memory.

Selection and ordering checklist for the HI98121

  • Confirm that ORP is the second parameter required; for conductivity choose the HI98129 or HI98130.
  • State the pH range of your samples; the tester reads –2.00 to 16.00 pH.
  • Select standard or NIST buffers and keep matching buffer stock on site.
  • Order spare HI73127 pH cartridges, HI70022 ORP test solution and electrode cleaning and storage solution.
  • Plan how readings will be recorded, since the tester has no logging memory.

Calibration, checks and care for the HI98121

Calibrate pH at least once a month, when the cartridge is replaced and after testing aggressive samples. Use one buffer for offset or two buffers for offset and slope. To verify ORP, dip the tester into the HI70022 470 mV solution and compare; a drift that cannot be cleared by cleaning with HI700601 solution suggests the platinum surface needs attention or the tester needs service. Rinse after each sample, avoid touching the sensing surfaces and store the tip in storage solution.

HI98121 applications in Oman and the GCC

Local users include swimming-pool and hotel water operators, cooling-tower and chilled-water treatment teams, laboratories in colleges and utilities, wastewater plants, aquaculture and aquarium operators, and plating and metal-finishing shops that monitor redox in treatment baths. High summer temperatures speed up chemical reactions in sample cups, so measure promptly after sampling and keep the tester out of direct sun.

HI98121 against other Hanna pocket testers

Model Measures Note
HI98121 pH, ORP, temperature Replaceable pH cartridge, fixed ORP sensor
HI98120 ORP, temperature No pH measurement
HI98129 pH, EC, TDS, temperature Low-range conductivity
HI98130 pH, EC, TDS, temperature High-range conductivity

Information to send with your HI98121 RFQ

  • Quantity and spare cartridges or solutions required.
  • Application (disinfection, pool, process bath, laboratory) and expected pH and mV range.
  • Delivery location in Oman or the GCC.

Documents and quotation

The HI98121 instruction manual is attached on this page for reference and describes specifications, calibration and set-up.

Send us the part number HI98121 with your quantity and Seven Star LLC will confirm availability and lead time for Oman and the GCC, as an independent supplier that is not affiliated with the manufacturer.

Other Models in the Hanna Pocket Tester Range

Seven Star LLC supplies models from the Hanna Pocket Tester range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.

ModelCategory
Hanna Instruments HI98129 pH/EC/TDS/Temperature TesterAutomation AccessoriesView details
Hanna Instruments HI98121 pH/ORP/Temperature TesterAutomation AccessoriesYou are viewing this model
Datasheets & Documents
  • PDF Manual Hanna Instruments HI98121 Instruction Manual Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Hanna Instruments is a maker of analytical instruments for water and liquid testing, founded in 1978 in the Padua area of Italy by Oscar and Anna Nardo. Its early breakthrough was the pHep, an affordable pocket-sized pH tester. The company is now led by the founders' children, Martino and Pamela Nardo, with its headquarters in Woonsocket, Rhode Island, and offers more than 3,000 products.

Seven Star LLC is an independent supplier in Muscat. We list Hanna instruments for water, wastewater and laboratory users in Oman and the GCC.

Hanna Instruments company profile

Founded 1978, near Padua, Italy, by Oscar and Anna Nardo
Headquarters Woonsocket, Rhode Island, USA
Leadership Martino Nardo and Pamela Nardo
Early product The pHep pocket pH tester
Product families pHep and HI98 waterproof testers, edge benchtop meters, HI83 multiparameter photometers, Checker HC handheld colorimeters, HI931 and HI932 titrators, process controllers, reagents and calibration buffers

Hanna product series

Portable meters include the HI98190 pH and ORP meter, the HI98192 EC, TDS and salinity meter, the HI98193 dissolved oxygen and BOD meter, the HI98199 pH, EC and DO meter and the HI9829 multiparameter water quality meter.

The HI98703 is an EPA compliant portable turbidity meter, the HI97711 a free and total chlorine photometer and the HI83414 a benchtop turbidity and chlorine meter.

Waterproof portable meters such as the HI98190 and HI98192 suit field use, with automatic temperature compensation and GLP data for audits. Turbidity measurement with the HI98703 follows EPA method 180.1, and the HI97711 photometer measures free and total chlorine for disinfection control.

Where Hanna instruments are used

Drinking water and desalination plants, wastewater treatment, environmental field surveys, aquaculture, cooling water and quality control labs.

Hanna Instruments products we list

Family Examples
Portable Test & Measurement Hanna HI97711 Free and Total Chlorine Po…, Hanna HI98190 Waterproof Portable pH and…, Hanna HI98192 Waterproof Portable EC, Hanna HI98193 Waterproof Portable Dissol…, Hanna HI98199 Portable pH, Hanna HI9829 Multiparameter Water Qualit…, Hanna HI98703 Portable Turbidity Meter

Ordering Hanna instruments: what we need

Send the model, the parameters and ranges, the probe and cable length and any reagents or calibration solutions required.

Facts checked against Hanna Instruments' own company story and City of Woonsocket business profile, October 2026.

FAQ

Questions about the Hanna Instruments HI98121 pH/ORP/Temperature Tester

Answers from our engineering team on installation, compatibility and integration.

01Does the HI98121 need ORP calibration?
No. The ORP sensor is fixed and factory calibrated, so the tester has no ORP calibration routine. A 470 mV test solution sachet is supplied so the user can check that the reading remains correct.
02What is the measuring range of the HI98121?
The tester measures pH from –2.00 to 16.00 with 0.01 resolution, ORP over ±1000 mV with 1 mV resolution and temperature from –5.0 to 60.0 °C with 0.1 °C resolution.
03Can the pH electrode of the HI98121 be changed?
Yes. The HI73127 pH cartridge is replaceable using the HI73128 removal tool. The ORP sensor is a fixed part of the tester body, so only the pH side is serviceable by the user.
04How long do the batteries of the HI98121 last?
The manufacturer states about 250 hours of continuous use on four 1.5 V batteries. The tester switches itself off after eight minutes of non-use and warns when battery capacity becomes low.
05Is the HI98121 suitable for measuring conductivity?
No. It measures pH, ORP and temperature only. For conductivity and TDS together with pH the HI98129 or HI98130 combo testers in the same housing should be selected.