Brand:

WIKA TC83 Sapphire-Design High-Temperature Thermocouple

Description

WIKA TC83 sapphire-design thermocouple for hydrogenous atmospheres

The WIKA TC83 is a high-temperature thermocouple in which a gas-tight sapphire tube surrounds the noble-metal sensor. It was developed for atmospheres that contain hydrogen and sulphur compounds, where ceramic tubes slowly let gas through and the thermocouple is poisoned. Because the sapphire has a monocrystalline structure, it protects the noble metal from toxic media without a purge system, which is the main reason to choose the TC83 over the purged TC82. WIKA states high process safety for processes up to 1,700 °C. Seven Star LLC supplies the TC83 in Oman and the GCC as an independent supplier in Muscat.

Where the sapphire design makes sense

The datasheet lists sulphur recovery units, waste-to-energy plants, processes with high hydrogen sulphide content and hydrogen-based direct reduced iron plants. In each, a tube that stays gas-tight, along with hermetically sealed junctions to stop toxic gas escaping the reactor, reduces both unplanned shutdowns and the plant cost of running purge gas. WIKA notes that the sapphire design can significantly improve life expectancy of the thermocouple.

TC83 technical specifications

Parameter Specification
Manufacturer / part number WIKA / TC83 (data sheet TE 65.83)
Type High-temperature thermocouple, sapphire design, non-purge system
Sensor types K, J, E, R, S, B per IEC 60584-1 or ASTM E230
Inner protection tube Sapphire, 8 mm [0.315 in] diameter
Outer protection tube 24 … 26 mm; ceramic C530 and C799 up to 1,600 °C, silicon carbide (Hexoloy) up to 1,650 °C
Operating temperature in air Up to 1,700 °C for the ceramic protection tube
Insertion length U 300 … 1,000 mm [11.8 … 39.4 in]; further lengths on request
Tip design Welded thermocouple (hot junction) with insulation rod, sapphire inner tube and outer tube
Ambient and storage temperature -40 … +80 °C; -60 °C special version
Certificates 2.2 test report and 3.1 inspection certificate per EN 10204
Functional safety SIL 2 with TUV-certified T38 in selected combinations; two thermocouples recommended for SIL 3

Accuracy classes and sensor choice

For base-metal sensors the datasheet repeats the IEC 60584-1 limits: type K class 2 from -40 to +1,200 °C and class 1 to +1,000 °C, type J to +750 °C, type E to +900 °C (class 2) or +800 °C (class 1). For precious-metal sensors type R and S class 2 run 0 … 1,600 °C, and type B class 3 or class 1 runs 600 … 1,700 °C. In sulphur and hydrogen atmospheres noble metal sensors are the usual choice, which is where the sapphire protection of the platinum alloy pays back. ASTM E230 limits are also listed for customers who specify it.

Assembly and variants

Variant Arrangement
Basic module Thermocouple with protection tubes, no head transmitter
Variant 1 Connection head with terminal block
Variant 2 Connection head with built-in temperature transmitter
Variant 3 Field transmitter fitted to the assembly

All variants have a metal support tube, a neck tube and a process connection. The datasheet explains nominal length A, insertion length U and the support tube length X below the process connection, so a specification can be written from a drawing. A ceramic thermowell head design is shown for the basic module.

TC83 compared with TC82 and TC84

Model Principle Use
TC82 Ceramic or SiC tube, purge connection Low-pressure reactors with purge available
TC83 Sapphire inner tube, non-purge Hydrogenous or sulphur-rich atmospheres without purge
TC84 Sapphire with ceramic outer tube and safety chamber High-pressure gasification and GtL reactors

Handling and installation

  • Sapphire and ceramic are brittle: keep transport protection on until installation
  • Heat up and cool down slowly; avoid thermal shock from cold air or water
  • Mount in the direction and with the flange arrangement shown on the drawing, and support long tubes
  • Use matched compensating cable (R, S, B types have their own) or a head transmitter
  • Check head seals and terminals at every shutdown and calibrate on the schedule your site uses

Applications in Oman and the GCC

For regional gas treatment plants and refineries with sulphur recovery trains, the TC83 matches the thermal reactor and waste-heat boiler zones, where purge gas systems are a maintenance burden. It is also suited to waste-to-energy projects and emerging hydrogen-based steel projects, where process gas may contain hydrogen at high temperature.

TC83 ordering checklist

  • Thermocouple type and class (usually R, S or B for the highest range)
  • Outer tube material and diameter, and insertion length U
  • Flange, support tube length X and neck length
  • Head, transmitter option and cable entry
  • Hazardous-area marking and certificate level (2.2 or 3.1)

Practical points when specifying the TC83

A sapphire-tube thermocouple is a long-life, higher-value spare, so treat the specification as a reliability decision rather than a price comparison. Start with the process atmosphere: the datasheet connects the TC83 to hydrogen-rich gas and to gas with a high hydrogen sulphide content, which is exactly where a porous ceramic would allow the gas to reach the noble-metal wires. Then check temperature peaks, not only normal operation, since upset conditions in a thermal reactor are the usual cause of tube failure. Confirm the outer tube material against the gas chemistry, and decide on a certificate level (the datasheet lists a 2.2 test report and a 3.1 inspection certificate per EN 10204) before the order is placed, because certificates cannot be added afterwards.

For the TC83 the wiring side deserves the same care as the sensor. Noble-metal thermocouples produce a small signal, so an installation with a head transmitter, or with matched compensating cable and short runs, keeps errors down. Mark the polarity as the datasheet instructs: the colour code at the positive pole of the instrument decides the correlation of polarity and terminal.

Information to send with an RFQ

Provide the process gas composition, temperature, pressure, flange details, insertion length and approvals so we can confirm the tube and sensor combination with you.

Documents and quotation

The WIKA data sheet TE 65.83 for the TC83 is attached on this page. Send us the specification and quantity for a quotation from Muscat as an independent supplier.

Other Models in the Wika Tc8x Range

Seven Star LLC supplies models from the Wika Tc8x range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.

ModelCategory
WIKA TC80 High-Temperature Thermocouple, DIN 50446Temperature InstrumentationView details
WIKA TC82 High-Temperature Thermocouple with PurgeTemperature InstrumentationView details
WIKA TC83 Sapphire-Design High-Temperature ThermocoupleTemperature InstrumentationYou are viewing this model
Datasheets & Documents
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

WIKA is the German manufacturer of pressure, temperature, level and flow instruments, best known for its pressure gauges. Alexander Wiegand and Philipp Kachel founded it on 2 January 1946 in Klingenberg am Main, Bavaria. The name comes from the first two letters of their surnames. WIKA is still run by the Wiegand family and is headquartered in Klingenberg.

Seven Star LLC is an independent supplier in Muscat. We list WIKA pressure gauges, diaphragm seals, transmitters, switches, thermometers and thermowells, and supply them to plants, contractors and panel builders in Oman and the GCC by model and order code.

WIKA company profile

Established 2 January 1946, Klingenberg am Main, Germany, by Alexander Wiegand and Philipp Kachel
First products Mechanical pressure and temperature gauges
Ownership Family-owned; managed by Alexander Wiegand, grandson of the founder, since 1996
Product families Pressure gauges, diaphragm seals, pressure transmitters and switches, thermometers, RTDs, thermocouples, thermowells, calibration equipment, level and flow instruments

Milestones

  • 1946: Alexander Wiegand and Philipp Kachel found the company in Klingenberg am Main on 2 January; pressure gauge production starts at the end of the year. The name WIKA comes from the first letters of their surnames.
  • 1951: Konrad Wiegand, the founder's son, takes over management.
  • 1967: Ursula Wiegand leads the company after her husband's death.
  • 1986: The company is reunited with the Kachel thermometer factory, which Philipp Kachel had set up after leaving in 1956.
  • 1994: Certification to DIN ISO 9001.
  • 1996: Alexander Wiegand, the founder's grandson, becomes head of the company.

WIKA products we list

Family Models Typical use
Bourdon tube pressure gauges 111.10, 111.12, 212.20, 213.53, 232.50, 233.50, 232.30/233.30 safety pattern HVAC and utilities (111 series) to process plants (23x series in stainless steel)
Diaphragm and differential pressure gauges 432.50, 732.51, 700.01, DPGT43 Low pressures, viscous or dirty media, and filter differential pressure
Diaphragm seals 990.10, 990.27 Separating the gauge from corrosive, hot or clogging media
Transmitters, switches and test gauges A-10, UPT-20, PSM01, CPG1500 4-20 mA pressure signals, switching, and precision calibration checks
Thermometers TG54, Model 55, A52, 73 gas-actuated Local temperature indication on pipes, vessels and HVAC
Temperature sensors and thermowells TR10, TR12-B, TR33, TC10-B, TW10 flanged, TW15 threaded RTD and thermocouple measurement in process lines

The WIKA model number tells you the build. In 232.50, the first digit 2 means stainless steel wetted parts, the second digit 3 means a liquid-fillable case, and ".50" is the case style. The 111 series uses copper alloy wetted parts and suits water and air. Where the medium is corrosive, very hot or would block the gauge, a diaphragm seal such as the 990.27 keeps it away from the Bourdon tube. For temperature, a thermowell like the TW10 lets you remove the sensor without opening the process.

Where WIKA instruments are used

In Oman and the GCC, WIKA gauges, transmitters and thermometers are used in oil and gas, refining and petrochemical plants, power and desalination, water networks, district cooling and HVAC, and on skids and machines from OEMs.

Ordering WIKA instruments: what we need

Send the model, dial size (for example 63, 100 or 160 mm), range and unit, connection size and position (lower or back mount), accuracy class and whether liquid filling is needed. For diaphragm seals and thermowells, add the process connection, materials and insertion length.

Facts checked against WIKA company history, October 2026.

FAQ

Questions about the WIKA TC83 Sapphire-Design High-Temperature

Answers from our engineering team on installation, compatibility and integration.

01What makes the TC83 different from the TC82?
The TC83 uses a gas-tight sapphire inner tube that protects the noble-metal thermocouple without a purge system. The TC82 relies on ceramic or silicon carbide with an optional purge.
02What atmospheres is it designed for?
Hydrogenous atmospheres and processes with high hydrogen sulphide content, such as sulphur recovery units, waste-to-energy plants and hydrogen-based DRI plants.
03What is the sapphire tube size?
The sapphire inner protection tube has a diameter of 8 mm. The outer protection tube is 24 to 26 mm.
04What is the maximum temperature?
High process safety is stated for processes up to 1,700 degrees C. Ceramic outer tubes are rated 1,600 degrees C and silicon carbide 1,650 degrees C.
05Is SIL use possible?
Selected versions with a certified T38 transmitter are suitable for SIL 2. Two thermocouples with one certified T38 are recommended for SIL 3.