Siemens Maxum II FID/FPD Flame Arrester Hastelloy-C 1460006-001
Overview: A Critical Safety Component for Detector Venting
The Siemens Maxum II Flame Arrester Hastelloy-C, part number 1460006-001, is a safety-critical fitting used on the Flame Ionization Detector (FID) of the Siemens Maxum Edition II process gas chromatograph. Constructed from corrosion-resistant Hastelloy-C, this arrester threads into the FID bottom body, with its insert tube connecting to a union inside the analyzer oven, to prevent flame propagation out of the detector and back along the analyzer’s tubing – a mandatory protective measure on any combustion-based detector operating continuously in a hazardous-area process environment.
As an independent specialized supplier of Siemens Maxum Edition II analyzer spare parts, Seven Star LLC sources and supplies the flame arrester assembly 1460006-001 to refineries, gas plants, and petrochemical complexes across Oman and the wider GCC. Seven Star LLC helps maintenance teams keep the correct genuine Hastelloy-C arrester in stock to avoid safety-related downtime during FID servicing. Note that the Flame Photometric Detector (FPD II) does not use this part: per the Siemens Maxum Edition II Detectors Service Manual, the FPD takes its exhaust from the combustion chamber through a flameproof joint rather than through a flame arrester, and FPD hardware carries separate Siemens part numbers.
Working Principle: Flame Propagation Prevention
A flame arrester works by forcing the gas stream through a network of narrow passages or a porous matrix with dimensions small enough that heat is dissipated faster than a flame front can propagate through them. On the FID, this principle isolates the hydrogen-fuelled detector flame from the tubing that runs back into the analyzer oven – if a flashback event were to occur inside the detector body, the arrester’s narrow internal geometry quenches the flame before it can travel out of the detector and into the surrounding analyzer enclosure.
The choice of Hastelloy-C construction reflects the corrosive and thermally demanding environment inside an operating FID detector body, where the arrester is exposed to combustion byproducts and sustained elevated temperature. Hastelloy-C’s resistance to oxidation and chemical attack ensures the arrester’s narrow flame-quenching geometry remains dimensionally stable and free of corrosion product buildup over long service intervals, preserving both its safety function and its gas flow characteristics.
Engineering Design of the Flame Arrester Assembly
The 1460006-001 flame arrester is machined to the precise internal tolerances required to provide a certified flame-quenching gas path while introducing minimal restriction to the detector’s normal gas flow. Its compact fitting is sized for direct installation into the FID bottom body, matching the Maxum Edition II’s standard detector plumbing without requiring custom adaptation.
Because this arrester sits directly in the flame path of a continuously operating combustion detector, its mechanical integrity is central to the ongoing safety certification of the analyzer installation. Genuine Siemens-specified Hastelloy-C arrester assemblies are manufactured and tested to maintain their flame-quenching performance across the detector’s full operating temperature range, which is a key reason process safety and instrumentation teams standardize on the correct OEM part number rather than substituting generic flame arrester fittings.
Key Features
The Flame Arrester Hastelloy-C 1460006-001 is engineered to protect Flame Ionization Detector installations on the Maxum Edition II. Key characteristics include:
- Direct-fit replacement flame arrester insert assembly for the FID detector module on the Maxum Edition II.
- Hastelloy-C construction for resistance to corrosion and combustion byproduct attack.
- Precision-machined internal geometry engineered to quench flame propagation with minimal flow restriction.
- Rated for the sustained elevated temperatures present inside an operating flame detector body.
- Genuine Siemens Maxum II part number 1460006-001 for correct safety-certified fit.
- Installs and removes with a 9/16 in open-end wrench, seating against its flat washer in the FID bottom body.
Installation and Replacement on the FID
The Siemens Maxum Edition II Detectors Service Manual identifies the flame arrester as three related items within the FID assembly: the Flame Arrestor Insert Assembly that threads into the narrow end of the FID bottom body, the flat washer that seats beneath it, and the Flame Arrestor Bushing and arrester that complete the assembly. Replacement is a mechanical operation carried out with the detector cooled and the hydrogen supply isolated. Tubing is first disconnected from the FID flame arrestor insert tube inside the oven, after which the insert is removed from the FID bottom body using a 9/16 in open-end wrench.
Reassembly is where most field errors occur. Siemens is explicit that the threads of the FID flame arrestor must be clean and free of lubricant or any other substance before the insert and its washer are placed into the FID bottom body and tightened, because a contaminated thread can compromise the flameproof path that the arrester exists to maintain. The insert tubing is then reconnected to the correct connection union in the oven. Because the Siemens Maxum II Flame Arrester Hastelloy-C is part of the analyzer’s flameproof containment rather than a consumable measuring component, instrumentation teams across Oman and the GCC generally schedule its inspection and replacement into planned detector overhauls and turnaround safety inspections rather than waiting for a symptom to appear.
Industrial Applications
The flame arrester assembly is a required maintenance item wherever a Maxum Edition II analyzer is fitted with an FID, including refinery hydrocarbon monitoring, natural gas quality analysis, and total hydrocarbon measurement on petrochemical process streams. Because it is a safety-critical component, it is typically replaced as part of scheduled detector overhaul and safety inspection intervals rather than only on failure, making reliable stock availability important for planned turnaround maintenance.
Seven Star LLC supports refinery instrumentation teams, gas processing operators, and petrochemical maintenance departments across Oman, Muscat, Sohar, and the broader GCC region with sourcing of this and other Maxum Edition II detector safety hardware, helping analyzer houses maintain both measurement continuity and process safety compliance.
Technical Specifications
The table below summarizes the core specifications of the Siemens Maxum II Flame Arrester Hastelloy-C as supplied by Seven Star LLC.
| Parameter | Specification |
|---|---|
| Model / Part Number | Flame Arrester Assembly, P/N 1460006-001 |
| Material | Hastelloy-C |
| Function | Flame propagation prevention (flashback protection) |
| Compatible Detector | Flame Ionization Detector (FID). Not used on the FPD II, which vents through a flameproof joint |
| Target Platform | Siemens Maxum Edition II Process Gas Chromatograph |
| Typical Location | Detector vent / inlet gas tubing |
| Hardware Condition | New, genuine Siemens part number |
| Assembly Items (per Siemens FID assembly) | Flame Arrestor Insert Assembly, flat washer, Flame Arrestor Bushing, Flame Arrestor |
| Mounting | Threads into the narrow end of the FID bottom body; insert tube connects to an oven union |
| Service Tool | 9/16 in open-end wrench |
| Installation Requirement | Threads must be clean and free of lubricant or other substances |
| Reference Document | Siemens Maxum Edition II Detectors Service Manual, A5E42019847001 |
Request a Quote for Siemens Maxum II FID/FPD Flame Arrester Hastelloy-C 1460006-001
Seven Star LLC is a dedicated supplier of Siemens Maxum Edition II process gas chromatograph spare parts and detector safety hardware throughout Oman and the GCC. To request a formal B2B quotation, verify current availability and lead time, or confirm compatibility of the Flame Arrester Hastelloy-C 1460006-001 with your FID detector configuration, contact our instrumentation sourcing team today.
Related Products
- Siemens Maxum II SYSCON Replacement Kit Compact Flash 2022018-001
- Dräger C-7540 Dissolved Oxygen Comparator (0–40 ppb)
- LAR 501-40-000215 Peristaltic Pump 5-Channel 24VAC 10RPM
- Siemens Maxum II Heater Cartridge A5E01673207
- Mitsubishi TN6RPC Gas Tube Dryer Set for NSX-2100 Sulfur Analyzer
- Yokogawa E7042UD Universal Zirconia Cell Assembly Kit for ZO21D/ZR22/ZR202
- Instruction Manual Siemens Maxum II FID/FPD Flame Arrester Hastelloy-C 1460006-001 Manual Download

Reviews
There are no reviews yet.