Michell Condumax II CD2-SCR-002 Hydrocarbon Dew-Point Sensor Cell

Description

Product Overview

The Michell Condumax II CD2-SCR-002 is a sensor cell assembly for the Condumax II hydrocarbon dew-point analyzer, used for automatic on-line measurement of hydrocarbon dew point in natural gas. The sensor cell is the heart of the Condumax II sampling system, where the dew-point measurement takes place.

The Condumax II is a proven analyzer used extensively in the natural gas industry for hydrocarbon dew-point measurement. The sensor cell carries an operating pressure rating of 100 barg (1450 psig), and the sampling system adds filtration, pressure regulation and flow control components in a robust enclosure suitable for field installation.

Key Features

  • Sensor cell assembly for the Condumax II hydrocarbon dew-point analyzer
  • For continuous hydrocarbon dew-point measurement in natural gas
  • Safe working pressure up to 100 barg (1450 psig)
  • Robust stainless steel construction for field service
  • Interfaces with the Condumax II electronics and sampling system
  • Proven dark spot optical detection technology
  • Genuine Michell Instruments component

Technical Specifications

Model Condumax II CD2-SCR-002 sensor cell assembly
Analyzer Michell Condumax II hydrocarbon dew-point analyzer
Safe working pressure Sensor cell operating pressure rating 100 barg (1450 psig)
Measurement hydrocarbon dew point
Technology dark spot optical detection on a cooled optical surface
Construction stainless steel sensor cell
Manufacturer Michell Instruments
Sensor cooling Automatic 3-stage Peltier cooler under adaptive control
Measurement depression More than 50 K below the Main Unit operating temperature
Hydrocarbon dew-point accuracy ±0.5 °C
Analysis pressure (HC dew point) 15 to 100 barg (218 to 1450 psig)
Sample flow (HC dew-point channel) 0.03 m³/h (0.5 Nl/min)
Measurement frequency Factory default cycle time 10 minutes; minimum cycle time 10 minutes
Detection Optical detection of condensed hydrocarbons on the cooled optical surface
Recommended replacement interval 24 months from the start of operation

Applications

The CD2-SCR-002 sensor cell is used in natural gas transmission and distribution, gas processing, LNG and custody transfer applications for hydrocarbon dew-point monitoring, protecting pipelines from condensate formation and ensuring gas quality.

Ordering and Selection

When selecting the Condumax II sensor cell, confirm the analyzer version, sample system configuration and pressure rating, because each Condumax II component is built for a specific gas quality monitoring duty. The CD2-SCR-002 sensor cell is the measurement core of the Condumax II analyzer, and the selected Condumax II sensor cell must match the sampling system installed at your station. Our team can help you confirm that the CD2-SCR-002 sensor cell you select is correct for your Condumax II analyzer.

Maintenance and Support

The Condumax II sensor cell should be inspected during routine analyzer service, including the mirror and optical surfaces. Genuine Condumax II spares, including filters and seals for the sensor cell, are available through Seven Star LLC. Contact our team for assistance with the Condumax II, and we will confirm availability and pricing for your project

Industry Applications and Custom Support

The Condumax II sensor cell is trusted by operations and maintenance teams across gas networks and processing plants in Oman, the GCC and the wider Middle East region. Our customers select the Condumax II for reliability, genuine quality and dependable supply. Seven Star LLC provides custom support including technical selection, quotation and after-sales assistance, and we regularly supply this sensor cell as part of preventive maintenance programs and shutdown packages. Whether you need a single unit or a complete set for a project, our team supports procurement and inspection.

The CD2-SCR-002 sensor cell measures the hydrocarbon dew point with the cooled mirror principle, and the cell assembly is the exchange part for the Condumax II analyser of the natural gas transmission.

The cooled mirror cell provides the primary measurement of the hydrocarbon dew point, and the cell exchange restores the accuracy of the analyser.

The CD2-SCR-002 cell is used in the natural gas processing and the transmission, and the cell measures the hydrocarbon dew point for the quality control of the gas.

The cell is installed in the analyser sample loop behind the process isolation valve.

The Michell CD2-SCR-002 hydrocarbon dew point sensor is the exchange cell of the Condumax II analyzer, and the Michell CD2-SCR-002 hydrocarbon dew point sensor measures the dew point of the natural gas at the transmission station.

The Michell CD2-SCR-002 hydrocarbon dew point sensor is exchanged during the Condumax II service, and the CD2-SCR-002 cell restores the dew point measurement.

Service Interval and Related Service Parts

Michell's routine maintenance schedule for the Condumax II names the hydrocarbon dew-point sensor cell by part number CD2-SCR-002 and recommends replacing it 24 months after the start of operation. This is preventive: the cooling capacity of the thermo-electric heat pump can fall over its working life, and the cell assembly also contains the optical surface, light source, photodetector and temperature sensor that the measurement depends on. Michell states that it offers a refurbishment and calibration service for sensor cells returned after removal from operation.

For gases with more than 2,000 ppmV hydrogen sulphide (up to 0.5 mol %), the manufacturer says the 24-month interval should be strictly observed and that the internal solenoid valve (CD2-21027) and flow switch (CD2-21362H, or CD2-21362W for the optional water dew-point channel) should be replaced at the same service. The quartz window O-ring (CD2-28469) must be renewed every time the cell optical housing is re-assembled. Analyzers with the water dew-point option also have a separate sensor, CD2-WDC-002, with a 12-month replacement interval, so both parts are often ordered together.

What to Plan for When Exchanging the Cell

The cell is fixed to the enclosure base by four M6 screws and connected by 1/8 inch sample tubes and its electrical connectors. Before fitting the new cell, a light, even layer of thermally conductive paste is applied to the full underside of the base plate, and both mating surfaces must be free of particles. After any work on the cell, Michell requires a gas pressure test and leak check at 1.5 times the maximum operating pressure before the analyzer is re-energised, so plan for a suitable test gas supply.

After a new cell is installed the analyzer must run a full sensitivity calibration, and the Cell Constant has to be set to the value on the calibration certificate supplied with the cell, so keep that certificate with the analyzer records. If the optical surface only needs cleaning, the manufacturer specifies laboratory-grade acetone or dichloromethane applied with very light pressure, and warns that the cell must not be dismantled further because this invalidates its calibration. A badly contaminated or damaged optical surface means the cell should be replaced.

Why Seven Star LLC?

Seven Star LLC supplies genuine Michell Instruments analyzer components to customers in Oman and the GCC. We verify part numbers, ensuring the correct sensor cell for your Condumax II system.

Our engineers help you verify the exact model and revision before ordering, avoiding costly mismatches.

Request a Quote for Michell Condumax II CD2-SCR-002 Hydrocarbon Dew-Point Sensor Cell

For the CD2-SCR-002 or any related control and instrumentation requirement, contact the Seven Star LLC team. Submit a formal inquiry through our RFQ page with your configuration and application details, and our specialists will respond with a detailed commercial proposal and the official CD2-SCR-002 documentation.

Related Products

Datasheets & Documents
  • PDF Instruction Manual Michell Condumax II CD2-SCR-002 Hydrocarbon Dew-Point Sensor Cell Manual Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Moisture in gas is a small number with large consequences, and measuring it has been the work of Michell Instruments since 1974. Based in Ely, United Kingdom, the company makes dew-point transmitters, hygrometers and process analyzers for moisture, hydrocarbon dew point and oxygen.

Seven Star LLC lists Michell dew-point transmitters, the Condumax II analyzer and its replacement sensor parts, and supplies them to gas plants and compressed air users in Oman by exact order code.

Michell product families on this page

The Condumax II is an automatic hydrocarbon dew-point analyzer for natural gas. Several of its parts are listed separately: the CD2-SCR-002 hydrocarbon dew-point sensor cell, the CD2-WDC-002 water dew-point sensor assembly and the CD2-SX-WDC-002, which is the exchange version of that water dew-point sensor. The Easidew line is represented by the Easidew EA2 dew point transmitter and by the Easidew PRO I.S., an intrinsically safe moisture transmitter for hazardous areas. A ceramic impedance moisture sensor module is also listed, the sensing element behind these transmitters.

Where Michell analyzers are used

Hydrocarbon and water dew point are contract quality parameters for natural gas, so the Condumax II is found at gas processing plants, pipeline entry points and custody transfer stations. Easidew transmitters monitor dryers for instrument air and industrial gases, glove boxes, and dry gas feeds to processes that cannot tolerate water.

Ordering Michell parts: what we need

For Condumax II parts send the analyzer serial number and the part code, such as CD2-SCR-002, and state if a new assembly or the exchange item is wanted. For Easidew transmitters we need the measuring span in dew point or ppm, the process connection thread, the output signal, the operating pressure, the gas and the area certification. A photo of the label on the old transmitter usually answers most of these in one go.

Frequently Asked Questions

What is the CD2-SCR-002 sensor cell?
It is the sensor cell assembly for the Michell Condumax II hydrocarbon dew-point analyzer.
What is the maximum working pressure?
The sensor cell carries an operating pressure rating of 100 barg (1450 psig).
How does the Condumax II measure dew point?
It uses dark spot optical detection of hydrocarbon condensate on a cooled optical surface.
Where is the Condumax II used?
It is used for hydrocarbon dew-point monitoring in natural gas networks and processing plants.
What details should I include in my inquiry for the Michell CD2-SCR-002 hydrocarbon dew point sensor?
Send the part number, your system details and the required quantity through the RFQ page. Seven Star LLC confirms the exact configuration, the matching accessories and the commercial terms for your project.