Hanna HI97731 Zinc Portable Photometer 0-3.00 mg/L
Description
Hanna HI97731 Zinc portable photometer
The Hanna HI97731 is a pocket-sized, single-parameter photometer built for zinc monitoring in effluent, galvanising rinse water and corrosion-inhibitor programmes. The meter reads 0.00 to 3.00 mg/L zinc, covering the typical discharge-limit and treatment-programme region. Seven Star LLC is an independent supplier in Muscat and is not affiliated with the manufacturer; we quote the HI97731 to industrial and laboratory buyers across Oman and the GCC.
The model number follows Hanna’s HI977xx pattern for dedicated photometers: the last digits identify the chemistry, so HI97731 always means zinc and cannot be reconfigured for another analyte. Reagents carry their own HI93xxx codes, listed below, and must be ordered separately from the meter.
HI97731 technical specifications
| Parameter | Specification |
|---|---|
| Manufacturer / part number | Hanna Instruments HI97731 |
| Parameter | Zinc |
| Range | 0.00 to 3.00 mg/L as Zn |
| Resolution | 0.01 mg/L |
| Accuracy at 25 °C | ±0.03 mg/L ±3 % of reading |
| Method | Adaptation of Standard Methods, zincon method |
| Light source | LED with narrow-band interference filter at 575 nm |
| Filter bandwidth / wavelength accuracy | 8 nm / ±1.0 nm |
| Detector | Silicon photocell |
| Cuvette | Round, 24.6 mm diameter (22 mm inside) |
| Data logging | 50 readings (auto logging) |
| Display | 128 × 64 pixel monochrome LCD with backlight |
| Power | Three 1.5 V AA alkaline batteries; more than 800 measurements without backlight |
| Operating environment | 0 to 50 °C; 0 to 100 % RH |
| Ingress protection | IP67, floating case |
| Dimensions / weight | 142.5 × 102.5 × 50.5 mm / 380 g with batteries |
| Warranty | Two years |
How the HI97731 measurement works
Zinc forms a blue complex with zincon at alkaline pH; after masking other metals, the colour intensity measured at 575 nm is proportional to zinc concentration.
Inside the case a reference detector tracks the LED output, so battery sag or a warm ambient does not move the blank. A lens gathers the light leaving the cuvette, which is why scratches and small glass imperfections matter less than with a simple colorimeter, and the cap notch locks the cuvette in the same orientation every time. A built-in tutorial mode walks the operator through sampling, reagent quantity and timers on screen, so the instrument can be handed to a technician who is not a chemist.
Reagents, procedure and interferences for HI97731
Dissolve one HI93731A-0 packet in a 20 mL sample, fill the cuvette with 10 mL of that reacted sample and zero the meter, then add 0.5 mL of HI93731B-0, swirl gently for 15 seconds and press Read for a 3 minute 30 second countdown.
| Code | Description | Quantity per test |
|---|---|---|
| HI93731A-0 | Zinc Reagent A | 1 packet |
| HI93731B-0 | Zinc Reagent B | 0.5 mL |
Reagent sets are supplied as HI93731-01 (100 tests) and HI93731-03 (300 tests). Follow the zeroing order given in the manual for this method, and read within the time the manual allows after the countdown ends.
Known interferences listed for this method:
- Iron above 7 mg/L
- Aluminum above 6 mg/L
- Copper, manganese and nickel above 5 mg/L
- Cadmium above 0.5 mg/L
Where the HI97731 sits among Hanna photometers
The HI97731 is one of the heavy-metal members of the HI977xx range alongside silver (HI97737). Both use the 575 nm LED, but zinc uses a packet plus a liquid and silver uses a four-reagent sequence with a dispersing agent.
| Model | Parameter | Range | LED filter |
|---|---|---|---|
| HI97700 | Ammonia low range | 0.00 to 3.00 mg/L as NH3-N | 420 nm |
| HI97715 | Ammonia medium range | 0.00 to 10.00 mg/L as NH3-N | 420 nm |
| HI97733 | Ammonia high range | 0.0 to 100.0 mg/L as NH4+ | 420 nm |
| HI97701 | Free chlorine | 0.00 to 5.00 mg/L as Cl2 | 525 nm |
| HI97762 | Free chlorine ultra low range | 0.000 to 0.500 mg/L as Cl2 | 525 nm |
| HI97734 | Free and total chlorine, high range | 0.00 to 10.00 mg/L as Cl2 | 525 nm |
| HI97725 | Chlorine, cyanuric acid and pH | 0.00 to 5.00 mg/L Cl2; 0 to 80 mg/L CYA; 6.5 to 8.5 pH | 525 nm |
| HI97706 | Phosphorus | 0.0 to 15.0 mg/L as P | 525 nm |
| HI97714 | Cyanide | 0.000 to 0.200 mg/L as CN– | 610 nm |
| HI97718 | Iodine | 0.0 to 12.5 mg/L as I2 | 525 nm |
| HI97731 | Zinc | 0.00 to 3.00 mg/L as Zn | 575 nm |
| HI97737 | Silver | 0.000 to 1.000 mg/L as Ag | 575 nm |
Selection and ordering checklist for HI97731
- Confirm zinc expected between 0.05 and 3.00 mg/L
- Order HI93731-01 or HI93731-03 reagent sets
- Check iron, aluminum, copper, manganese, nickel and cadmium levels in the sample
- Add HI97731-11 CAL Check cuvettes
- Plan waste handling for the reacted metal-bearing solution
Validation, handling and maintenance of HI97731
Validate with HI97731-11. Metal-bearing blue residue can stain cuvettes, so clean after each series with the Hanna cleaning solution and rinse with deionised water.
Keep cuvettes clean with the cleaning solution and a lint-free cloth, never touch the lower half of the glass, and store the meter in its carrying case. Replace the three AA cells when the battery icon appears; the meter keeps its logged readings and settings. For long storage between field trips remove the batteries and close the reagent bottles tightly.
Typical applications of HI97731 in Oman and the GCC
Typical Gulf sites include galvanising and metal-fabrication shops, cooling-water systems using zinc-based inhibitors, desalination plants checking corrosion products, and environmental laboratories confirming effluent zinc before discharge.
Heat is a practical factor in the Gulf. Reaction times in the manual are given at 25 °C, so samples that sit in a vehicle at much higher temperature should be allowed to equilibrate, and the meter itself is rated for 0 to 50 °C. The sealed optics and IP67 floating case also cope with dust and spray on plant walkdowns.
Information to send with an RFQ for HI97731
- Quantity of meters and whether reagent sets (HI93731-01 or HI93731-03; CAL Check HI97731-11) and CAL Check cuvettes are also required
- Sample matrix and expected concentration, so we can confirm the range suits the application
- Delivery location in Oman or the GCC and required lead time
- Any need for spare cuvettes, carrying case or cleaning solution
- Project or tender reference, if applicable
Documents and quotation
The instruction manual for the HI97731 is attached on this page and covers specifications, the reagent procedure, CAL Check validation and error messages.
Send the part number HI97731, the quantity and your delivery location and we will reply with availability and a quotation.
Other Models in the Hanna Hi977xx Photometer Range
Seven Star LLC supplies models from the Hanna Hi977xx Photometer range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.
Datasheets & Documents
- Manual Hanna HI97731 Instruction Manual Download
Reviews (0)
How would you rate this product? Cancel reply
About brand
Hanna Instruments is a maker of analytical instruments for water and liquid testing, founded in 1978 in the Padua area of Italy by Oscar and Anna Nardo. Its early breakthrough was the pHep, an affordable pocket-sized pH tester. The company is now led by the founders' children, Martino and Pamela Nardo, with its headquarters in Woonsocket, Rhode Island, and offers more than 3,000 products.
Seven Star LLC is an independent supplier in Muscat. We list Hanna instruments for water, wastewater and laboratory users in Oman and the GCC.
Hanna Instruments company profile
| Founded | 1978, near Padua, Italy, by Oscar and Anna Nardo |
|---|---|
| Headquarters | Woonsocket, Rhode Island, USA |
| Leadership | Martino Nardo and Pamela Nardo |
| Early product | The pHep pocket pH tester |
| Product families | pHep and HI98 waterproof testers, edge benchtop meters, HI83 multiparameter photometers, Checker HC handheld colorimeters, HI931 and HI932 titrators, process controllers, reagents and calibration buffers |
Hanna product series
Portable meters include the HI98190 pH and ORP meter, the HI98192 EC, TDS and salinity meter, the HI98193 dissolved oxygen and BOD meter, the HI98199 pH, EC and DO meter and the HI9829 multiparameter water quality meter.
The HI98703 is an EPA compliant portable turbidity meter, the HI97711 a free and total chlorine photometer and the HI83414 a benchtop turbidity and chlorine meter.
Waterproof portable meters such as the HI98190 and HI98192 suit field use, with automatic temperature compensation and GLP data for audits. Turbidity measurement with the HI98703 follows EPA method 180.1, and the HI97711 photometer measures free and total chlorine for disinfection control.
Where Hanna instruments are used
Drinking water and desalination plants, wastewater treatment, environmental field surveys, aquaculture, cooling water and quality control labs.
Hanna Instruments products we list
Ordering Hanna instruments: what we need
Send the model, the parameters and ranges, the probe and cable length and any reagents or calibration solutions required.
Facts checked against Hanna Instruments' own company story and City of Woonsocket business profile, October 2026.
Browse Hanna Instruments products by category: Portable Test & Measurement.
FAQ
Questions about the Hanna HI97731 Zinc Portable Photometer 0-3.00 mg/L
Answers from our engineering team on installation, compatibility and integration.
01What range does the HI97731 cover?
Zinc from 0.00 to 3.00 mg/L with 0.01 mg/L resolution, accuracy of 0.03 mg/L plus 3 percent of reading at 25 degrees C.
02What method does it use?
An adaptation of the zincon method from Standard Methods, read at 575 nm with an LED and narrow-band filter.
03Which reagents are needed?
HI93731A-0 (one packet) and HI93731B-0 (0.5 mL) per test; sets HI93731-01 for 100 tests and HI93731-03 for 300 tests.
04Which metals interfere?
Iron above 7 mg/L, aluminum above 6 mg/L, copper, manganese and nickel above 5 mg/L, and cadmium above 0.5 mg/L.
05How long is the reaction?
The meter runs a 3 minute 30 second countdown before measuring.