Brand: Hanna Instruments
Hanna HI97714 Cyanide Portable Photometer 0-0.200 mg/L
Description
Hanna HI97714 Cyanide portable photometer
The Hanna HI97714 is a pocket-sized, single-parameter photometer built for free cyanide screening in industrial effluent and plating or mining related waters. Its 0.000 to 0.200 mg/L window with 0.001 mg/L resolution is aimed at compliance-level monitoring, where cyanide must be shown to be present only at trace levels. Seven Star LLC is an independent supplier in Muscat and is not affiliated with the manufacturer; we quote the HI97714 to industrial and laboratory buyers across Oman and the GCC.
The model number follows Hanna’s HI977xx pattern for dedicated photometers: the last digits identify the chemistry, so HI97714 always means cyanide and cannot be reconfigured for another analyte. Reagents carry their own HI93xxx codes, listed below, and must be ordered separately from the meter.
HI97714 technical specifications
| Parameter | Specification |
|---|---|
| Manufacturer / part number | Hanna Instruments HI97714 |
| Parameter | Cyanide |
| Range | 0.000 to 0.200 mg/L as CN– |
| Resolution | 0.001 mg/L |
| Accuracy at 25 °C | ±0.005 mg/L ±3 % of reading |
| Method | Adaptation of the pyridine-pyrazolone method |
| Light source | LED with narrow-band interference filter at 610 nm |
| Filter bandwidth / wavelength accuracy | 8 nm / ±1.0 nm |
| Detector | Silicon photocell |
| Cuvette | Round, 24.6 mm diameter (22 mm inside) |
| Data logging | 50 readings (auto logging) |
| Display | 128 × 64 pixel monochrome LCD with backlight |
| Power | Three 1.5 V AA alkaline batteries; more than 800 measurements without backlight |
| Operating environment | 0 to 50 °C; 0 to 100 % RH |
| Ingress protection | IP67, floating case |
| Dimensions / weight | 142.5 × 102.5 × 50.5 mm / 380 g with batteries |
| Warranty | Two years |
How the HI97714 measurement works
Cyanide is converted to cyanogen chloride and then reacts with pyridine-pyrazolone to form a blue dye. The meter reads the colour at 610 nm, so the result tracks the cyanide in the sample provided that oxidants and reductants have been dealt with.
Inside the case a reference detector tracks the LED output, so battery sag or a warm ambient does not move the blank. A lens gathers the light leaving the cuvette, which is why scratches and small glass imperfections matter less than with a simple colorimeter, and the cap notch locks the cuvette in the same orientation every time. A built-in tutorial mode walks the operator through sampling, reagent quantity and timers on screen, so the instrument can be handed to a technician who is not a chemist.
Reagents, procedure and interferences for HI97714
Fill the cuvette with 10 mL of sample and zero, add one spoon of HI93714A-0, then reagents B and C packets, with an initial 30 second countdown and then a 25 minute development time before reading. Undissolved powder does not affect accuracy.
| Code | Description | Quantity per test |
|---|---|---|
| HI93714A-0 | Cyanide Reagent A | 1 spoon |
| HI93714B-0 | Cyanide Reagent B | 1 packet |
| HI93714C-0 | Cyanide Reagent C | 1 packet |
Reagent sets are supplied as HI93714-01 (100 tests) and HI93714-03 (300 tests). Follow the zeroing order given in the manual for this method, and read within the time the manual allows after the countdown ends.
Known interferences listed for this method:
- Large amounts of turbidity give high readings
- Oxidising agents such as chlorine and reducing agents such as sulfide or sulfur dioxide interfere; distillation removes them
- Samples of high pH should be adjusted to about pH 7 before testing
- Cyanides and the hydrogen cyanide liberated by acids are very poisonous, so follow the manual safety instructions
Where the HI97714 sits among Hanna photometers
The HI97714 sits apart from the ammonia, chlorine and phosphorus models because of its toxic analyte and long 25 minute development. It is typically a compliance-screening tool backed by laboratory distillation, not a replacement for it.
| Model | Parameter | Range | LED filter |
|---|---|---|---|
| HI97700 | Ammonia low range | 0.00 to 3.00 mg/L as NH3-N | 420 nm |
| HI97715 | Ammonia medium range | 0.00 to 10.00 mg/L as NH3-N | 420 nm |
| HI97733 | Ammonia high range | 0.0 to 100.0 mg/L as NH4+ | 420 nm |
| HI97701 | Free chlorine | 0.00 to 5.00 mg/L as Cl2 | 525 nm |
| HI97762 | Free chlorine ultra low range | 0.000 to 0.500 mg/L as Cl2 | 525 nm |
| HI97734 | Free and total chlorine, high range | 0.00 to 10.00 mg/L as Cl2 | 525 nm |
| HI97725 | Chlorine, cyanuric acid and pH | 0.00 to 5.00 mg/L Cl2; 0 to 80 mg/L CYA; 6.5 to 8.5 pH | 525 nm |
| HI97706 | Phosphorus | 0.0 to 15.0 mg/L as P | 525 nm |
| HI97714 | Cyanide | 0.000 to 0.200 mg/L as CN– | 610 nm |
| HI97718 | Iodine | 0.0 to 12.5 mg/L as I2 | 525 nm |
| HI97731 | Zinc | 0.00 to 3.00 mg/L as Zn | 575 nm |
| HI97737 | Silver | 0.000 to 1.000 mg/L as Ag | 575 nm |
Selection and ordering checklist for HI97714
- Confirm you need free cyanide at 0 to 0.200 mg/L
- Order HI93714-01 or HI93714-03; check safe storage and disposal for the reagents
- Plan sample pre-treatment for oxidants, sulfide and high pH
- Account for the 25 minute reaction time in sampling routines
- Add CAL Check HI97714-11 for validation
Validation, handling and maintenance of HI97714
Because cyanide is acutely toxic, use eye protection, work in ventilated areas, and dispose of reacted samples through a licensed waste route. Validate with HI97714-11 and keep cuvettes dedicated to this method where practical.
Keep cuvettes clean with the cleaning solution and a lint-free cloth, never touch the lower half of the glass, and store the meter in its carrying case. Replace the three AA cells when the battery icon appears; the meter keeps its logged readings and settings. For long storage between field trips remove the batteries and close the reagent bottles tightly.
Typical applications of HI97714 in Oman and the GCC
Relevant to metal-finishing and plating shops, mining and mineral processing support, certain petrochemical and gas-processing streams, and environmental laboratories in Oman and the GCC that screen effluent before discharge. Heat shortens sample stability, so test promptly.
Heat is a practical factor in the Gulf. Reaction times in the manual are given at 25 °C, so samples that sit in a vehicle at much higher temperature should be allowed to equilibrate, and the meter itself is rated for 0 to 50 °C. The sealed optics and IP67 floating case also cope with dust and spray on plant walkdowns.
Information to send with an RFQ for HI97714
- Quantity of meters and whether reagent sets (HI93714-01 or HI93714-03; CAL Check HI97714-11) and CAL Check cuvettes are also required
- Sample matrix and expected concentration, so we can confirm the range suits the application
- Delivery location in Oman or the GCC and required lead time
- Any need for spare cuvettes, carrying case or cleaning solution
- Project or tender reference, if applicable
Documents and quotation
The instruction manual for the HI97714 is attached on this page and covers specifications, the reagent procedure, CAL Check validation and error messages.
Send the part number HI97714, the quantity and your delivery location and we will reply with availability and a quotation.
Other Models in the Hanna Hi977xx Photometer Range
Seven Star LLC supplies models from the Hanna Hi977xx Photometer range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.
Datasheets & Documents
- Manual Hanna HI97714 Instruction Manual Download
Reviews (0)
How would you rate this product? Cancel reply
About brand
Hanna Instruments is a maker of analytical instruments for water and liquid testing, founded in 1978 in the Padua area of Italy by Oscar and Anna Nardo. Its early breakthrough was the pHep, an affordable pocket-sized pH tester. The company is now led by the founders' children, Martino and Pamela Nardo, with its headquarters in Woonsocket, Rhode Island, and offers more than 3,000 products.
Seven Star LLC is an independent supplier in Muscat. We list Hanna instruments for water, wastewater and laboratory users in Oman and the GCC.
Hanna Instruments company profile
| Founded | 1978, near Padua, Italy, by Oscar and Anna Nardo |
|---|---|
| Headquarters | Woonsocket, Rhode Island, USA |
| Leadership | Martino Nardo and Pamela Nardo |
| Early product | The pHep pocket pH tester |
| Product families | pHep and HI98 waterproof testers, edge benchtop meters, HI83 multiparameter photometers, Checker HC handheld colorimeters, HI931 and HI932 titrators, process controllers, reagents and calibration buffers |
Hanna product series
Portable meters include the HI98190 pH and ORP meter, the HI98192 EC, TDS and salinity meter, the HI98193 dissolved oxygen and BOD meter, the HI98199 pH, EC and DO meter and the HI9829 multiparameter water quality meter.
The HI98703 is an EPA compliant portable turbidity meter, the HI97711 a free and total chlorine photometer and the HI83414 a benchtop turbidity and chlorine meter.
Waterproof portable meters such as the HI98190 and HI98192 suit field use, with automatic temperature compensation and GLP data for audits. Turbidity measurement with the HI98703 follows EPA method 180.1, and the HI97711 photometer measures free and total chlorine for disinfection control.
Where Hanna instruments are used
Drinking water and desalination plants, wastewater treatment, environmental field surveys, aquaculture, cooling water and quality control labs.
Hanna Instruments products we list
Ordering Hanna instruments: what we need
Send the model, the parameters and ranges, the probe and cable length and any reagents or calibration solutions required.
Facts checked against Hanna Instruments' own company story and City of Woonsocket business profile, October 2026.
Browse Hanna Instruments products by category: Portable Test & Measurement.
FAQ
Questions about the Hanna HI97714 Cyanide Portable Photometer 0-0.200
Answers from our engineering team on installation, compatibility and integration.
01What range does the HI97714 cover?
Cyanide from 0.000 to 0.200 mg/L as CN-, with 0.001 mg/L resolution and 0.005 mg/L plus 3 percent of reading accuracy at 25 degrees C.
02What method is used and at what wavelength?
An adaptation of the pyridine-pyrazolone method, measured with an LED and narrow-band interference filter at 610 nm.
03How long does a test take?
A 30 second countdown during preparation is followed by a 25 minute development countdown before the meter reads.
04Does undissolved reagent powder matter?
No. The manual states accuracy is not affected by undissolved reagent powder, so there is no need to wait for complete dissolution.
05What safety points apply?