Brand: Hanna Instruments
Hanna HI97734 Free and Total Chlorine HR Photometer
Description
Hanna HI97734 Free and total chlorine high range portable photometer
The Hanna HI97734 is a pocket-sized, single-parameter photometer built for high-dose chlorination where both free and total chlorine must be read up to 10 mg/L. This model measures free and total chlorine in a single meter, with a 0.00 to 10.00 mg/L range, so shock-dosed pools, wash-down water and disinfection tanks do not need dilution. Seven Star LLC is an independent supplier in Muscat and is not affiliated with the manufacturer; we quote the HI97734 to industrial and laboratory buyers across Oman and the GCC.
The model number follows Hanna’s HI977xx pattern for dedicated photometers: the last digits identify the chemistry, so HI97734 always means free and total chlorine high range and cannot be reconfigured for another analyte. Reagents carry their own HI93xxx codes, listed below, and must be ordered separately from the meter.
HI97734 technical specifications
| Parameter | Specification |
|---|---|
| Manufacturer / part number | Hanna Instruments HI97734 |
| Parameter | Free and total chlorine high range |
| Range | 0.00 to 10.00 mg/L as Cl2 |
| Resolution | 0.01 mg/L |
| Accuracy at 25 °C | ±0.03 mg/L ±3 % of reading |
| Method | Adaptation of US EPA 330.5, DPD colorimetric |
| Light source | LED with narrow-band interference filter at 525 nm |
| Filter bandwidth / wavelength accuracy | 8 nm / ±1.0 nm |
| Detector | Silicon photocell |
| Cuvette | Round, 24.6 mm diameter (22 mm inside) |
| Data logging | 50 readings (auto logging) |
| Display | 128 × 64 pixel monochrome LCD with backlight |
| Power | Three 1.5 V AA alkaline batteries; more than 800 measurements without backlight |
| Operating environment | 0 to 50 °C; 0 to 100 % RH |
| Ingress protection | IP67, floating case |
| Dimensions / weight | 142.5 × 102.5 × 50.5 mm / 380 g with batteries |
| Warranty | Two years |
How the HI97734 measurement works
The DPD reaction colours the sample pink in proportion to chlorine; total chlorine adds a third reagent that releases combined chlorine. The meter reads 525 nm absorbance and the high-range reagent B supplies buffering for strongly chlorinated samples.
Inside the case a reference detector tracks the LED output, so battery sag or a warm ambient does not move the blank. A lens gathers the light leaving the cuvette, which is why scratches and small glass imperfections matter less than with a simple colorimeter, and the cap notch locks the cuvette in the same orientation every time. A built-in tutorial mode walks the operator through sampling, reagent quantity and timers on screen, so the instrument can be handed to a technician who is not a chemist.
Reagents, procedure and interferences for HI97734
Free chlorine uses one HI93701-0 packet and 5 mL of HI93734B-0 mixed with 5 mL of sample, with a 1 minute countdown. Total chlorine adds 3 drops of HI93734C-0 on a fresh sample and runs a 2 minute 30 second countdown.
| Code | Description | Quantity per test |
|---|---|---|
| HI93701-0 | Free Chlorine Reagent | 1 packet |
| HI93734B-0 | Free and Total Chlorine HR Reagent | 5 mL |
| HI93734C-0 | Total Chlorine Reagent (total only) | 3 drops |
Reagent sets are supplied as HI93734-01 (100 tests) and HI93734-03 (300 tests). Follow the zeroing order given in the manual for this method, and read within the time the manual allows after the countdown ends.
Known interferences listed for this method:
- Bromine, chlorine dioxide, iodine, oxidised chromium and manganese, and ozone
- Alkalinity above 1,000 mg/L CaCO3 as bicarbonate (pH below 8.3), above 25 mg/L as carbonate (pH above 9.0), or acidity above 150 mg/L CaCO3; neutralise with dilute HCl or NaOH
- Hardness above 500 mg/L CaCO3; shake the sample for about 2 minutes after adding powder
Where the HI97734 sits among Hanna photometers
The HI97734 is the high-range, dual-parameter choice. Use the HI97701 for free chlorine only up to 5 mg/L, the HI97762 for traces, and the HI97725 when cyanuric acid and pH are also part of the routine.
| Model | Parameter | Range | LED filter |
|---|---|---|---|
| HI97700 | Ammonia low range | 0.00 to 3.00 mg/L as NH3-N | 420 nm |
| HI97715 | Ammonia medium range | 0.00 to 10.00 mg/L as NH3-N | 420 nm |
| HI97733 | Ammonia high range | 0.0 to 100.0 mg/L as NH4+ | 420 nm |
| HI97701 | Free chlorine | 0.00 to 5.00 mg/L as Cl2 | 525 nm |
| HI97762 | Free chlorine ultra low range | 0.000 to 0.500 mg/L as Cl2 | 525 nm |
| HI97734 | Free and total chlorine, high range | 0.00 to 10.00 mg/L as Cl2 | 525 nm |
| HI97725 | Chlorine, cyanuric acid and pH | 0.00 to 5.00 mg/L Cl2; 0 to 80 mg/L CYA; 6.5 to 8.5 pH | 525 nm |
| HI97706 | Phosphorus | 0.0 to 15.0 mg/L as P | 525 nm |
| HI97714 | Cyanide | 0.000 to 0.200 mg/L as CN– | 610 nm |
| HI97718 | Iodine | 0.0 to 12.5 mg/L as I2 | 525 nm |
| HI97731 | Zinc | 0.00 to 3.00 mg/L as Zn | 575 nm |
| HI97737 | Silver | 0.000 to 1.000 mg/L as Ag | 575 nm |
Selection and ordering checklist for HI97734
- Confirm you need both free and total chlorine
- Verify residuals run above about 5 mg/L, otherwise the HI97701 may be simpler
- Order HI93734-01 or HI93734-03 sets; free and total tests must be run on separate fresh samples
- Add CAL Check HI97734-11 for validation
- Decide how combined chlorine (total minus free) will be recorded
Validation, handling and maintenance of HI97734
Free and total chlorine are read on separate fresh samples, never on the same reacted cuvette. Keep one cuvette for each if throughput matters, and validate the optical system with CAL Check HI97734-11 on a defined schedule.
Keep cuvettes clean with the cleaning solution and a lint-free cloth, never touch the lower half of the glass, and store the meter in its carrying case. Replace the three AA cells when the battery icon appears; the meter keeps its logged readings and settings. For long storage between field trips remove the batteries and close the reagent bottles tightly.
Typical applications of HI97734 in Oman and the GCC
Suitable for hotel and resort pools along the Omani coast, water-tanker and storage-tank disinfection, food-plant sanitation rinses, cooling water shock dosing and wastewater disinfection contact chambers where residuals are well above potable levels.
Heat is a practical factor in the Gulf. Reaction times in the manual are given at 25 °C, so samples that sit in a vehicle at much higher temperature should be allowed to equilibrate, and the meter itself is rated for 0 to 50 °C. The sealed optics and IP67 floating case also cope with dust and spray on plant walkdowns.
Information to send with an RFQ for HI97734
- Quantity of meters and whether reagent sets (HI93734-01 or HI93734-03; CAL Check HI97734-11) and CAL Check cuvettes are also required
- Sample matrix and expected concentration, so we can confirm the range suits the application
- Delivery location in Oman or the GCC and required lead time
- Any need for spare cuvettes, carrying case or cleaning solution
- Project or tender reference, if applicable
Documents and quotation
The instruction manual for the HI97734 is attached on this page and covers specifications, the reagent procedure, CAL Check validation and error messages.
Send the part number HI97734, the quantity and your delivery location and we will reply with availability and a quotation.
Other Models in the Hanna Hi977xx Photometer Range
Seven Star LLC supplies models from the Hanna Hi977xx Photometer range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.
Datasheets & Documents
- Manual Hanna HI97734 Instruction Manual Download
Reviews (0)
How would you rate this product? Cancel reply
About brand
Hanna Instruments is a maker of analytical instruments for water and liquid testing, founded in 1978 in the Padua area of Italy by Oscar and Anna Nardo. Its early breakthrough was the pHep, an affordable pocket-sized pH tester. The company is now led by the founders' children, Martino and Pamela Nardo, with its headquarters in Woonsocket, Rhode Island, and offers more than 3,000 products.
Seven Star LLC is an independent supplier in Muscat. We list Hanna instruments for water, wastewater and laboratory users in Oman and the GCC.
Hanna Instruments company profile
| Founded | 1978, near Padua, Italy, by Oscar and Anna Nardo |
|---|---|
| Headquarters | Woonsocket, Rhode Island, USA |
| Leadership | Martino Nardo and Pamela Nardo |
| Early product | The pHep pocket pH tester |
| Product families | pHep and HI98 waterproof testers, edge benchtop meters, HI83 multiparameter photometers, Checker HC handheld colorimeters, HI931 and HI932 titrators, process controllers, reagents and calibration buffers |
Hanna product series
Portable meters include the HI98190 pH and ORP meter, the HI98192 EC, TDS and salinity meter, the HI98193 dissolved oxygen and BOD meter, the HI98199 pH, EC and DO meter and the HI9829 multiparameter water quality meter.
The HI98703 is an EPA compliant portable turbidity meter, the HI97711 a free and total chlorine photometer and the HI83414 a benchtop turbidity and chlorine meter.
Waterproof portable meters such as the HI98190 and HI98192 suit field use, with automatic temperature compensation and GLP data for audits. Turbidity measurement with the HI98703 follows EPA method 180.1, and the HI97711 photometer measures free and total chlorine for disinfection control.
Where Hanna instruments are used
Drinking water and desalination plants, wastewater treatment, environmental field surveys, aquaculture, cooling water and quality control labs.
Hanna Instruments products we list
Ordering Hanna instruments: what we need
Send the model, the parameters and ranges, the probe and cable length and any reagents or calibration solutions required.
Facts checked against Hanna Instruments' own company story and City of Woonsocket business profile, October 2026.
Browse Hanna Instruments products by category: Portable Test & Measurement.
FAQ
Questions about the Hanna HI97734 Free and Total Chlorine HR Photometer
Answers from our engineering team on installation, compatibility and integration.
01How wide is the HI97734 range?
It reads 0.00 to 10.00 mg/L as chlorine for both free and total chlorine, with 0.01 mg/L resolution and 0.03 mg/L plus 3 percent of reading accuracy.
02Which reagents are needed?
Free chlorine uses HI93701-0 and HI93734B-0; total chlorine uses HI93734B-0 with HI93734C-0 as well. Sets are HI93734-01 and HI93734-03.
03How long do the tests take?
Free chlorine runs a 1 minute countdown and total chlorine a 2 minute 30 second countdown, both after reagent addition.
04Can free and total be measured in one cuvette?
No. Free and total chlorine must be measured separately with fresh unreacted samples following their own procedures.
05What conditions affect accuracy?
Oxidants such as bromine and ozone, hardness above 500 mg/L, and extreme alkalinity or acidity are the listed interferences; the manual gives corrective steps.