Hanna HI97715 Ammonia Medium Range Photometer 0-10 mg/L

Description

Hanna HI97715 Ammonia medium range portable photometer

The Hanna HI97715 is a pocket-sized, single-parameter photometer built for ammonia testing on wastewater and process streams where values run up to about 10 mg/L. With a 0.00 to 10.00 mg/L span it sits between the low-range HI97700 and the 100 mg/L HI97733, which makes it the practical default for raw and partly treated wastewater. Seven Star LLC is an independent supplier in Muscat and is not affiliated with the manufacturer; we quote the HI97715 to industrial and laboratory buyers across Oman and the GCC.

The model number follows Hanna’s HI977xx pattern for dedicated photometers: the last digits identify the chemistry, so HI97715 always means ammonia medium range and cannot be reconfigured for another analyte. Reagents carry their own HI93xxx codes, listed below, and must be ordered separately from the meter.

HI97715 technical specifications

Parameter Specification
Manufacturer / part number Hanna Instruments HI97715
Parameter Ammonia medium range
Range 0.00 to 10.00 mg/L as NH3-N
Resolution 0.01 mg/L
Accuracy at 25 °C ±0.05 mg/L ±5 % of reading
Method Adaptation of ASTM D1426 Nessler method
Light source LED with narrow-band interference filter at 420 nm
Filter bandwidth / wavelength accuracy 8 nm / ±1.0 nm
Detector Silicon photocell
Cuvette Round, 24.6 mm diameter (22 mm inside)
Data logging 50 readings (auto logging)
Display 128 × 64 pixel monochrome LCD with backlight
Power Three 1.5 V AA alkaline batteries; more than 800 measurements without backlight
Operating environment 0 to 50 °C; 0 to 100 % RH
Ingress protection IP67, floating case
Dimensions / weight 142.5 × 102.5 × 50.5 mm / 380 g with batteries
Warranty Two years

How the HI97715 measurement works

The Nessler reagent turns the sample yellow to brown in proportion to ammonia content. The HI97715 measures how much 420 nm light the reacted sample absorbs after being zeroed on the plain sample, so sample colour or slight turbidity present before reagent addition is compensated in the blank.

Inside the case a reference detector tracks the LED output, so battery sag or a warm ambient does not move the blank. A lens gathers the light leaving the cuvette, which is why scratches and small glass imperfections matter less than with a simple colorimeter, and the cap notch locks the cuvette in the same orientation every time. A built-in tutorial mode walks the operator through sampling, reagent quantity and timers on screen, so the instrument can be handed to a technician who is not a chemist.

Reagents, procedure and interferences for HI97715

Ten millilitres of sample go in the cuvette, the meter is zeroed, then four drops each of reagents A and B are added with five inversions after each, followed by a 3 minute 30 second countdown.

Code Description Quantity per test
HI93715A-0 Ammonia MR Reagent A 4 drops
HI93715B-0 Ammonia MR Reagent B 4 drops

Reagent sets are supplied as HI93715-01 (100 tests) and HI93715-03 (300 tests). Follow the zeroing order given in the manual for this method, and read within the time the manual allows after the countdown ends.

Known interferences listed for this method:

  • Hardness above 1 g/L
  • Iron
  • Sulfide, which may cause turbidity
  • Acetone above 0.1 %, alcohols, aldehydes, aliphatic and aromatic amines, chloramines, glycine or urea above 10 mg/L; distillation removes these

Where the HI97715 sits among Hanna photometers

Where a stream swings between a few mg/L and tens of mg/L, many sites keep two meters: the HI97715 for routine work and the HI97733 for concentrated reject streams. The HI97700 is better for final-effluent and receiving-water checks below 3 mg/L.

Model Parameter Range LED filter
HI97700 Ammonia low range 0.00 to 3.00 mg/L as NH3-N 420 nm
HI97715 Ammonia medium range 0.00 to 10.00 mg/L as NH3-N 420 nm
HI97733 Ammonia high range 0.0 to 100.0 mg/L as NH4+ 420 nm
HI97701 Free chlorine 0.00 to 5.00 mg/L as Cl2 525 nm
HI97762 Free chlorine ultra low range 0.000 to 0.500 mg/L as Cl2 525 nm
HI97734 Free and total chlorine, high range 0.00 to 10.00 mg/L as Cl2 525 nm
HI97725 Chlorine, cyanuric acid and pH 0.00 to 5.00 mg/L Cl2; 0 to 80 mg/L CYA; 6.5 to 8.5 pH 525 nm
HI97706 Phosphorus 0.0 to 15.0 mg/L as P 525 nm
HI97714 Cyanide 0.000 to 0.200 mg/L as CN– 610 nm
HI97718 Iodine 0.0 to 12.5 mg/L as I2 525 nm
HI97731 Zinc 0.00 to 3.00 mg/L as Zn 575 nm
HI97737 Silver 0.000 to 1.000 mg/L as Ag 575 nm

Selection and ordering checklist for HI97715

  • Verify normal readings fall between about 0.5 and 10 mg/L
  • Order HI93715-01 or HI93715-03 reagent sets separately
  • Decide on CAL Check cuvettes HI97715-11 for validation records
  • Plan to dilute samples that exceed 10.00 mg/L rather than reading out of range
  • Review sulfide, hardness and organic content of the sample against the interference list

Validation, handling and maintenance of HI97715

The meter supports CAL Check validation with the HI97715-11 kit and stores GLP data. Periodically wipe the cuvette with a lint-free cloth, check that the cap notch seats in the holder groove, and if an ambient-light warning appears, re-seat the cuvette before assuming a fault.

Keep cuvettes clean with the cleaning solution and a lint-free cloth, never touch the lower half of the glass, and store the meter in its carrying case. Replace the three AA cells when the battery icon appears; the meter keeps its logged readings and settings. For long storage between field trips remove the batteries and close the reagent bottles tightly.

Typical applications of HI97715 in Oman and the GCC

Typical Gulf users include sewage treatment works checking aeration performance, industrial effluent teams at petrochemical and refinery sites, and aquaculture or hatchery operators who watch ammonia in recirculating systems. Desalination and utility labs use the same meter for brine and cooling-water investigations.

Heat is a practical factor in the Gulf. Reaction times in the manual are given at 25 °C, so samples that sit in a vehicle at much higher temperature should be allowed to equilibrate, and the meter itself is rated for 0 to 50 °C. The sealed optics and IP67 floating case also cope with dust and spray on plant walkdowns.

Information to send with an RFQ for HI97715

  • Quantity of meters and whether reagent sets (HI93715-01 or HI93715-03; CAL Check HI97715-11) and CAL Check cuvettes are also required
  • Sample matrix and expected concentration, so we can confirm the range suits the application
  • Delivery location in Oman or the GCC and required lead time
  • Any need for spare cuvettes, carrying case or cleaning solution
  • Project or tender reference, if applicable

Documents and quotation

The instruction manual for the HI97715 is attached on this page and covers specifications, the reagent procedure, CAL Check validation and error messages.

Send the part number HI97715, the quantity and your delivery location and we will reply with availability and a quotation.

Other Models in the Hanna Hi977xx Photometer Range

Seven Star LLC supplies models from the Hanna Hi977xx Photometer range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.

ModelCategory
Hanna HI97700 Ammonia Low Range Photometer 0-3 mg/LLiquid Analytical SystemsView details
Hanna HI97715 Ammonia Medium Range Photometer 0-10 mg/LLiquid Analytical SystemsYou are viewing this model
Datasheets & Documents
  • PDF Manual Hanna HI97715 Instruction Manual Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Hanna Instruments is a maker of analytical instruments for water and liquid testing, founded in 1978 in the Padua area of Italy by Oscar and Anna Nardo. Its early breakthrough was the pHep, an affordable pocket-sized pH tester. The company is now led by the founders' children, Martino and Pamela Nardo, with its headquarters in Woonsocket, Rhode Island, and offers more than 3,000 products.

Seven Star LLC is an independent supplier in Muscat. We list Hanna instruments for water, wastewater and laboratory users in Oman and the GCC.

Hanna Instruments company profile

Founded 1978, near Padua, Italy, by Oscar and Anna Nardo
Headquarters Woonsocket, Rhode Island, USA
Leadership Martino Nardo and Pamela Nardo
Early product The pHep pocket pH tester
Product families pHep and HI98 waterproof testers, edge benchtop meters, HI83 multiparameter photometers, Checker HC handheld colorimeters, HI931 and HI932 titrators, process controllers, reagents and calibration buffers

Hanna product series

Portable meters include the HI98190 pH and ORP meter, the HI98192 EC, TDS and salinity meter, the HI98193 dissolved oxygen and BOD meter, the HI98199 pH, EC and DO meter and the HI9829 multiparameter water quality meter.

The HI98703 is an EPA compliant portable turbidity meter, the HI97711 a free and total chlorine photometer and the HI83414 a benchtop turbidity and chlorine meter.

Waterproof portable meters such as the HI98190 and HI98192 suit field use, with automatic temperature compensation and GLP data for audits. Turbidity measurement with the HI98703 follows EPA method 180.1, and the HI97711 photometer measures free and total chlorine for disinfection control.

Where Hanna instruments are used

Drinking water and desalination plants, wastewater treatment, environmental field surveys, aquaculture, cooling water and quality control labs.

Hanna Instruments products we list

Family Examples
Portable Test & Measurement Hanna HI97711 Free and Total Chlorine Po…, Hanna HI98190 Waterproof Portable pH and…, Hanna HI98192 Waterproof Portable EC, Hanna HI98193 Waterproof Portable Dissol…, Hanna HI98199 Portable pH, Hanna HI9829 Multiparameter Water Qualit…, Hanna HI98703 Portable Turbidity Meter

Ordering Hanna instruments: what we need

Send the model, the parameters and ranges, the probe and cable length and any reagents or calibration solutions required.

Facts checked against Hanna Instruments' own company story and City of Woonsocket business profile, October 2026.

FAQ

Questions about the Hanna HI97715 Ammonia Medium Range Photometer 0-10

Answers from our engineering team on installation, compatibility and integration.

01What range does the HI97715 cover?
It reads 0.00 to 10.00 mg/L ammonia nitrogen with 0.01 mg/L resolution and an accuracy of 0.05 mg/L plus 5 percent of reading at 25 degrees C.
02Which reagent codes does it use?
It uses HI93715A-0 and HI93715B-0, four drops of each. Sets are sold as HI93715-01 for 100 tests and HI93715-03 for 300 tests.
03What is the wavelength and light source?
An LED with a 420 nm narrow-band interference filter, 8 nm bandwidth, read by a silicon photocell inside a sealed optical block.
04What if my sample is above 10 mg/L?
Dilute the sample with ammonia-free water and multiply the result, or move to the HI97733 high-range photometer, which reads up to 100.0 mg/L as ammonium.
05Is the meter splash and dust resistant?
Yes. The case is rated IP67 and floats, and the optical system is sealed from dust, dirt and water.