Hanna HI8733 Multi-Range Portable EC Meter 0-199.9 mS/cm

Description

Hanna HI8733 multi-range portable conductivity meter

The Hanna HI8733 is an analogue-knob portable conductivity meter with four ranges from 0.0 to 199.9 mS/cm, covering everything from low-conductivity process water to seawater and brine. It is supplied with the HI76302W conductivity probe and a calibration sachet. Seven Star LLC is an independent supplier in Muscat and is not affiliated with the manufacturer.

The HI8733 shares its instruction manual with the three-range HI8734 TDS meter and the related HI8633; the part number identifies the EC version with the widest span.

HI8733 technical specifications

Parameter Specification
Manufacturer / part number Hanna Instruments HI8733
EC ranges 0.0 to 199.9 µS/cm; 0 to 1999 µS/cm; 0.00 to 19.99 mS/cm; 0.0 to 199.9 mS/cm
Resolution 0.1 µS/cm; 1 µS/cm; 0.01 mS/cm; 0.1 mS/cm
Accuracy at 25 °C ±1 % of full scale, excluding probe error
Calibration Manual, one point with the EC knob
Temperature compensation Automatic from 0 to 50 °C with beta adjustable from 0 to 2.5 %/°C
Probe HI76302W conductivity probe with internal temperature sensor, DIN connector and 1 m cable
Power Two 9 V alkaline batteries; about 100 hours of continuous use
Environment 0 to 50 °C; RH up to 100 %
Size / weight 146 × 80 × 36 mm; 231 g
Supplied with HI76302W probe, HI70030 12880 µS/cm calibration sachet, battery, instructions
Warranty Two years

How the HI8733 works

The probe applies an alternating signal across its electrodes and the meter measures the current that flows through the sample, which is proportional to the number of dissolved ions. Because conductivity rises with temperature, the internal temperature sensor of the HI76302W adjusts the reading to 25 °C using a beta coefficient that the user can set between 0 and 2.5 % per degree.

Calibration is manual: with the probe in the HI70030 12880 µS/cm standard, the operator turns the EC knob until the display matches the standard. The four range switch positions allow a single instrument to span ultrapure to seawater-class conductivity.

Where HI8733 fits: sensors, accessories and related models

The HI76302W is the supplied probe, with a DIN connector and a 1 m cable. Calibration standards and storage solutions are separate consumables; the 12880 µS/cm sachet included is a good match for brackish and high-conductivity work, while lower ranges benefit from additional standards.

The family comparison shows how the HI8733 differs from the HI8734 TDS model that uses the same case design.

HI8733 family comparison

Model Reads Ranges Probe
HI8733 EC 0.0 to 199.9 µS/cm up to 0.0 to 199.9 mS/cm (4 ranges) HI76302W, auto temperature compensation
HI8734 TDS 0.0 to 199.9 mg/L up to 0.00 to 19.99 g/L (3 ranges) HI76301D four-ring, manual temperature compensation

Choose HI8733 when the specification asks for conductivity in µS/cm or mS/cm. Choose HI8734 when the specification asks for TDS in mg/L or g/L with a 0.5 conversion factor.

Selection and ordering checklist for HI8733

  • Confirm EC is the required parameter and that the four ranges cover the expected values
  • Quote extra calibration solutions to suit each range
  • Decide on a spare HI76302W probe
  • Plan for two 9 V alkaline batteries
  • Check whether a TDS reading (HI8734) is actually what the procedure needs

Installation, calibration and maintenance of HI8733

Rinse the probe with deionised water and then with the sample before calibration, and immerse it fully so that the holes near the sensing rings are covered. Select the suitable range, calibrate at one point with the EC knob, and tap the probe lightly to dislodge bubbles.

After use, rinse and store the probe dry or in the protective cap. If readings become unstable, clean the electrodes gently with a soft brush or the cleaning solution, then recalibrate.

Because calibration is a knob adjustment rather than a software routine, users should recalibrate at the start of every shift and whenever the range selection changes. Record the standard value and temperature used for each calibration so that audits can see the reference.

For repeatable comparisons between sites, keep the same compensation coefficient and record the reading temperature alongside the conductivity value.

Because the display range is selected by the operator, a common mistake is reading a low value on a high range and losing resolution. Start on the highest range when the conductivity is unknown and step down until the reading is in the upper half of the scale. Also check the supplied standard against the range you will use, since a 12880 µS/cm calibration point is not ideal for readings of a few µS/cm, where an additional low standard improves accuracy.

Typical applications of HI8733 in Oman and the GCC

Typical uses in Oman are desalination feed, permeate and brine conductivity spot checks, cooling-tower and boiler blowdown control, irrigation and aquaculture water screening, and wastewater process monitoring. The 199.9 mS/cm top range handles seawater and concentrated brine without dilution.

The rugged analogue-knob design is popular with maintenance crews who want a simple instrument that can be calibrated in the field without menus.

Information to send with an RFQ for HI8733

  • Quantity and need for spare probes
  • Calibration standards required for each range
  • Expected conductivity range and application
  • Delivery location and lead time
  • Project or tender reference

Storage, handling and spares for the Hanna HI8733

Store the Hanna HI8733 in its case in a clean, dry place. Keep probes, electrodes and reagents as described in the user manual, as many of them need a storage solution or have a limited shelf life. In the GCC climate, keep reagents away from heat and direct sun and check expiry dates before use.

  • Calibrate the instrument with fresh standards at the interval your procedure requires.
  • Rinse probes and sample cells after each use and inspect them for scratches or deposits.
  • Keep spare consumables, such as reagents, standards, cuvettes and electrode solutions, in stock.
  • Record calibration results so that drift or probe ageing can be seen early.

When you order the Hanna HI8733, list the consumables and standards you need in the same RFQ, so the instrument can be used from the day it arrives.

Documents and quotation

The instruction manual for the HI8733 is attached on this page for specifications, calibration procedures and maintenance steps. The same document covers the HI8734.

Send the part number HI8733, the quantity and your delivery location and we will reply with availability and a quotation.

Other Models in the Hanna Ec Tds Meter Range

Seven Star LLC supplies models from the Hanna Ec Tds Meter range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.

ModelCategory
Hanna HI8733 Multi-Range Portable EC Meter 0-199.9 mS/cmLiquid Analytical SystemsYou are viewing this model
Hanna HI8734 Three-Range Portable TDS Meter 0-19.99 g/LLiquid Analytical SystemsView details
Datasheets & Documents
  • PDF Manual Hanna HI8733 Instruction Manual Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Hanna Instruments is a maker of analytical instruments for water and liquid testing, founded in 1978 in the Padua area of Italy by Oscar and Anna Nardo. Its early breakthrough was the pHep, an affordable pocket-sized pH tester. The company is now led by the founders' children, Martino and Pamela Nardo, with its headquarters in Woonsocket, Rhode Island, and offers more than 3,000 products.

Seven Star LLC is an independent supplier in Muscat. We list Hanna instruments for water, wastewater and laboratory users in Oman and the GCC.

Hanna Instruments company profile

Founded 1978, near Padua, Italy, by Oscar and Anna Nardo
Headquarters Woonsocket, Rhode Island, USA
Leadership Martino Nardo and Pamela Nardo
Early product The pHep pocket pH tester
Product families pHep and HI98 waterproof testers, edge benchtop meters, HI83 multiparameter photometers, Checker HC handheld colorimeters, HI931 and HI932 titrators, process controllers, reagents and calibration buffers

Hanna product series

Portable meters include the HI98190 pH and ORP meter, the HI98192 EC, TDS and salinity meter, the HI98193 dissolved oxygen and BOD meter, the HI98199 pH, EC and DO meter and the HI9829 multiparameter water quality meter.

The HI98703 is an EPA compliant portable turbidity meter, the HI97711 a free and total chlorine photometer and the HI83414 a benchtop turbidity and chlorine meter.

Waterproof portable meters such as the HI98190 and HI98192 suit field use, with automatic temperature compensation and GLP data for audits. Turbidity measurement with the HI98703 follows EPA method 180.1, and the HI97711 photometer measures free and total chlorine for disinfection control.

Where Hanna instruments are used

Drinking water and desalination plants, wastewater treatment, environmental field surveys, aquaculture, cooling water and quality control labs.

Hanna Instruments products we list

Family Examples
Portable Test & Measurement Hanna HI97711 Free and Total Chlorine Po…, Hanna HI98190 Waterproof Portable pH and…, Hanna HI98192 Waterproof Portable EC, Hanna HI98193 Waterproof Portable Dissol…, Hanna HI98199 Portable pH, Hanna HI9829 Multiparameter Water Qualit…, Hanna HI98703 Portable Turbidity Meter

Ordering Hanna instruments: what we need

Send the model, the parameters and ranges, the probe and cable length and any reagents or calibration solutions required.

Facts checked against Hanna Instruments' own company story and City of Woonsocket business profile, October 2026.

FAQ

Questions about the Hanna HI8733 Multi-Range Portable EC Meter 0-199.9

Answers from our engineering team on installation, compatibility and integration.

01What ranges does the HI8733 measure?
Four EC ranges: 0.0 to 199.9 uS/cm, 0 to 1999 uS/cm, 0.00 to 19.99 mS/cm and 0.0 to 199.9 mS/cm, with 1 percent of full scale accuracy.
02How is the HI8733 calibrated?
Manually at one point: the probe is placed in the supplied 12880 uS/cm HI70030 standard and the EC knob is turned until the display matches.
03Does it compensate for temperature?
Yes, automatically from 0 to 50 degrees C, with the beta coefficient adjustable from 0 to 2.5 percent per degree C.
04What probe comes with it?
The HI76302W conductivity probe with internal temperature sensor, DIN connector and a 1 m cable.
05What batteries does it use?
Two 9 V alkaline batteries, providing about 100 hours of continuous use.