- Automation & ControlView all 1,651 products
- AC Drives
- Industrial Communications
- PLC & I/O Cards
- Vibration Monitoring
- Automation Accessories
- Controllers & Data Acquisition
- CCTV & Security
- Safety Switches & Guard Locking
- Encoders
- Industrial Computers
- Servo Drives
- Motor Soft Starters
- HMI Displays
- Isolator & Barrier
- Pneumatic Cylinders
- Automation Software
- Machine Vision
- Robotics
Brand: Sonatest
Sonatest T-Gage V Series Ultrasonic Thickness Gauge
Description
A 230 g thickness gauge with a go/no-go keypad
The T-Gage V series is Sonatest’s family of pocket-size ultrasonic thickness gauges, built for rough industrial conditions. Monochrome displays are standard and upgrade options are available throughout the gauge’s life. Three base models cover the range: the T-Gage VB is a basic variable-velocity gauge for single-sided steel structures, the T-Gage V adds echo-to-echo and alarms, and the T-Gage VDL is the top model with a data logger, B-scan and a probe-wear indicator.
Technical specifications
| Parameter | Specification |
|---|---|
| Models | T-Gage VB (basic), T-Gage V (mid-range), T-Gage VDL (data logger); letter suffixes C colour display, W A/B-scan, DL data logger |
| Thickness range | 0.20 to 508 mm (0.008 to 20 in), depending on material temperature and transducer |
| Resolution | 0.01 mm (0.001 in) |
| Bandwidth | 0.5 to 20 MHz (-3 dB) |
| Measurement rate | 20 readings per second; 4 per second in fast mode |
| Units | Inches, millimetres, microseconds |
| Gain | Low, standard or high |
| Display | 128 x 64 monochrome graphic LCD; colour models 220 x 170 TFT |
| Size and weight | 76.2 x 127 x 31.8 mm; 230 g |
| Casing | Splash-proof IP54 high-impact plastic with rubber; illuminated keypad for go/no-go testing |
| Operating temperature | -20 to 50 deg C |
| Material surface temperature | -20 to 537 deg C, depending on probe used |
| Battery | 2 x AA alkaline; 50 h (20 h with backlight) |
| Backlight | LED, on/off or auto on valid reading or key press |
| Datalogger (VDL) | Up to 50,000 readings, B-scan and Data XLS software |
| Warranty | 2 year limited, parts and labour |
What each model adds
| Feature | Available on |
|---|---|
| Delay-line zero measurement | VB, V, VDL |
| Scan, hold and freeze modes | VB, V, VDL |
| Differential mode | V, VDL |
| Alarms (max depth, vibration, beeps, flashing display, keypad illumination) | V, VDL |
| Echo-to-echo (measures metal only, ignores paint) | V, VDL |
| Automatic probe wear indicator | VDL |
| Non-encoded B-scan | VDL |
| Data XLS logging software | VDL |
Further model codes in the brochure are V C (echo-to-echo, colour display), V W (echo-to-echo, A-scan), V CW, V DLC, V DLW and V DLCW; C means colour display and W means A/B scan.
Using it on site
The three function keys light red, yellow and green so that an operator can see go or no-go without reading the number. Display orientation can be switched for left or right hand use, and the gauge sets delay-line zero automatically at power-up, which helps with delay-line wear and acoustic drift at elevated temperatures. Scan mode shows minimum or maximum thickness at 20 measurements per second as the probe is dragged across a surface.
With a 5 MHz dual-element probe, the gauge suits tanks, pipework and plate where only one side is reachable. Note that the specification quotes IP54 splash protection: wipe off couplant and avoid immersion.
Selecting the right model
Pick the VB when the task is plain thickness on bare steel and the budget is tight. Pick the V when painted or coated surfaces are common, because echo-to-echo ignores the coating and reads the metal only, and when alarms are needed to flag readings below a minimum depth. Pick the VDL when readings must be recorded for a survey: it keeps up to 50,000 values, draws a non-encoded B-scan cross section of the test piece, reminds the operator when the transducer is worn and moves data to a PC with Data XLS. Colour or A/B-scan versions are added by the suffix letters listed above, and software upgrades are available during the life of the gauge.
The measurement itself depends on the probe, the couplant and the surface. Calibrate on a block of the same material and similar temperature, check the velocity setting, and confirm the zero before a survey. These are general good practice points and not requirements printed in the brochure.
Kit contents
The standard T-Gage V and VDL kits include the gauge, echo-to-echo probe DG537 (5 MHz, 9.52 mm), operator manual, cable, couplant and carry case; the VDL also includes a pouch, Data XLS and a USB cable. The VB kit uses a DKS-537 probe. A hard shipping case and a moulded pouch with wrist strap are optional.
Choosing a T-Gage V model for work in Oman and the GCC
Pocket thickness gauges are used daily in the region on storage tanks, pipelines, structural steel, ship hulls and process piping, where corrosion and erosion must be tracked over time. For quick spot checks on bare steel, the basic VB model is often enough. Where paint or coating must be ignored, the echo-to-echo mode of the higher models allows the steel thickness to be measured without removing the coating. For corrosion surveys with many points that are repeated every year, the VDL data logger and B-scan save time and reduce reporting errors.
Hot surfaces are common on operating plant. Choose the probe and couplant for the actual surface temperature, and remember that readings on hot steel may need a velocity correction.
Information to send with an enquiry
- Model: VB, V or VDL, and whether a colour display or A/B-scan is needed
- Probe type and frequency, including high-temperature probes
- Test block, couplant and spare cables
- Quantity and certificate requirements
Documents and quotation
The Sonatest T-Gage V series brochure with specifications is attached. See more corrosion thickness gauges. Seven Star LLC supplies Sonatest gauges on request; tell us the model suffix, probe and quantity.
Datasheets & Documents
- Datasheet Sonatest T-Gage V series brochure with specifications, REV03_0223 (PDF) Download
Reviews (0)
How would you rate this product? Cancel reply
About brand
Sonatest is a British manufacturer of ultrasonic non-destructive testing equipment, established in 1958 in Wolverton and based in Milton Keynes. Its UK head office handles design, manufacturing and global sales, while a team in Canada develops phased array instruments. Sonatest makes flaw detectors, phased array systems, thickness gauges and transducers.
Sonatest company profile
| Founded | 1958, Wolverton, England |
|---|---|
| Headquarters | Old Wolverton, Milton Keynes |
| Incorporated | Sonatest Ltd, 30 November 1965 |
| R&D | Phased array development in Canada |
| Product families | WAVE touchscreen flaw detectors, Veo3 phased array and TFM systems, XEO phased array tablet, Sitescan digital flaw detectors, Railscan rail flaw detectors, RSflite composite inspection, thickness gauges, transducers |
Milestones
- 1958: Sonatest is established in Wolverton.
- 1965: Sonatest Ltd is incorporated.
Product ranges
The Sonatest catalogue covers conventional flaw detectors such as the Sitescan, Masterscan and Railscan families, the Wave touchscreen detector, phased array and TFM instruments including Veo3 and XEO, corrosion and precision thickness gauges, WheelProbe array probes, transducers, calibration blocks and couplants.
Industries served
Sonatest lists aerospace, rail, oil and gas, chemical and petrochemical, power generation, maritime, mining and infrastructure among the sectors that use its equipment. Seven Star LLC is an independent supplier in Muscat that sources genuine Sonatest products for customers in Oman and the GCC on request.
Sonatest products we list
Facts checked against Companies House records and BINDT member profile, October 2026.
FAQ
Questions about the Sonatest T-Gage V Series Ultrasonic Thickness Gauge
Answers from our engineering team on installation, compatibility and integration.
01Which T-Gage V models can ignore paint and coatings?
The T-Gage V and VDL have echo-to-echo measurement, which reads metal thickness only; the VB does not.
02What probe is supplied?
The standard kit includes a DG537 twin-crystal 5 MHz 0.375 in (9.52 mm) potted probe, or DKS-537 with the VB.
03How hot can the surface be?
The specification states -20 to 537 deg C material surface temperature, depending on the probe used.
04How many readings can the VDL store?
The brochure says it is a logger taking up to 50,000 readings.
05What is the protection class?
The casing is splash-proof, IP54, not a submersible rating.


















Reviews