Brand:

Normalab NDI 450 Automated Atmospheric Distillation Analyzer

Description

Fifth-generation automated distillation for fuel quality labs

The Normalab NDI 450 is a fully automated atmospheric distillation analyzer for petroleum products. Normalab describes it as the fifth generation of automated distillation it has developed, and it runs the whole boiling-range test from a 10.4 inch colour touch screen and Windows-based software. Heating, condenser temperature, receiver temperature and the volume reading in the receiver are all controlled by the instrument, so the operator loads the sample and starts the programmed method.

It is intended for refinery quality control, fuel terminals and independent laboratories that distil gasoline, aviation turbine fuel, kerosene, gas oil and distillate fuel oil to the ASTM D86 family of methods, and it is ready to use for D86 as delivered.

Technical specifications

Parameter Specification
Manufacturer / model Normalab NDI 450
Order numbers 60502 (AC 230 V, 50/60 Hz, 9 A); 60503 (same unit, AC 115 V, 50/60 Hz, 9 A)
Temperature measurement range Ambient to 450 °C
Temperature accuracy 0.05 °C
Heating element Low voltage, 850 W, automatic positioning, quick cooling of the heater by fan
Cooling system Built-in Peltier system
Condenser temperature 0 to 60 °C
Receiver compartment temperature 0 to 60 °C
Level follower 2 photoelectric forks; accuracy 0.1 mL, resolution 0.05 mL
Barometric pressure correction Automatic
Display and software 10.4 inch colour touch screen, Windows based, preprogrammed standard methods plus user programmable methods
Connections 3 USB, 1 RS232, 1 LAN, 1 VGA; LIMS via RS232 / LAN in XML, ASCII or CSV; e-mail reporting
Sample overheating protection Integrated, at 600 °C
Dimensions and weight (brochure) 540 x 450 x 630 mm (W x D x H); about 70 kg

Methods, calculations and safety

The brochure lists ASTM D86, D850 and D1078, ISO 3405, IP 123, IP 195, GOST B, JIS K2254 and DIN 51751. The unit is ready for D86 and related methods as delivered; an optional kit (P/N 60517) adds ASTM D850 and D1078. The brochure also names additional tests that the instrument can run: ASTM D189, D524 and D4530, the 10 percent distillation residue method using a 100 mL sample. The software calculates the cetane index to ISO 4264, and the unit is described as ready for biodiesel analysis.

Safety is built in: automatic fire detection, an automated nitrogen fire extinguisher with the N2 bottle delivered with the unit, an extinguisher push button, and sensors that confirm the receiver is in position, the compartment door is closed, the drip deflector and PT 100 probe are in place, the coolant level is sufficient and the nitrogen pressure is adequate before a test is allowed.

Scope of delivery and ordering

The 60502 package includes a 125 mL distillation flask (P/N 24019), a 100 mL receiver (26111), 38 mm and 50 mm heating plates (60507, 60508), a 5 mL residue receiver (12609EC), a PT 100 probe (60510) with centring device, a dry point thermocouple (60509), a bag of pumice (60514), a condenser cleaning cable (60513) and three stoppers (60515, 40065, 40064). The product page also lists the distillation flask, 25 mm and 32 mm heater plates, a Teflon plug for the probe and thermocouple, an RIS clamp for the centre plug and a USB colour printer as options.

Spares for the coolant circuit are available separately; see the Normalab 60606 circulation pump for the NDI 450.

Related Normalab products: Normalab NDI Classic Semi-Automatic Distillation Analyzer, Normalab NDI Basic Manual Atmospheric Distillation Unit, Normalab NSB Tech Automatic Salt in Crude Oil Analyzer, Normalab NAE 440 Automatic Twin Aniline Point Tester, Normalab NBA 450 Automatic Ring and Ball Softening Point Tester and Normalab NMC 460 Automatic Micro Carbon Residue Tester.

Documents and quotation

The Normalab NDI 450 brochure is linked on this page. Seven Star LLC, an independent supplier in Muscat, quotes this analyzer within our liquid analytical systems range. Before ordering, confirm the mains version (60502 or 60503), whether the D850 and D1078 kit is needed, the nitrogen supply for the extinguisher and the installation dimensions, because the brochure and the web product page give different footprint figures.

Datasheets & Documents
  • PDF Brochure Normalab NDI 450 brochure (EN, PDF) Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Normalab is a French manufacturer of laboratory instruments for petroleum product testing, established in 1963 and based in Valliquerville, Normandy. The company started out as glassblowers and still runs a glassblowing workshop. For more than 55 years it has designed and built test instruments that follow ASTM, ISO and IP methods, including its NPM 450, the fifth generation of its automated flash point testers.

Seven Star LLC, an independent supplier in Oman, lists Normalab spare parts: the 60606 circulation pump for the coolant circuit of the NDI 450 automatic distillation analyzer, and the 60448 thermal detection thermocouple. We source Normalab parts by their five-digit reference number.

Normalab company profile

Founded 1963, France
Headquarters Valliquerville, Normandy, France
Origin Laboratory glassblowing
Certification ISO 9001
Product families NPM 450 automated flash point testers, automatic distillation analysers, cloud and pour point testers, kinematic viscosity baths, copper corrosion and oxidation stability testers, laboratory glassware

Normalab instrument families

The manufacturer arranges its instruments by the property tested: flammability, volatility, viscosity, cold flow properties, cleanliness and purity, lubricants, and bitumen, wax and grease. Its glassware line supports the same tests. The methods referenced are ASTM, ISO and IP. Refinery and fuel quality laboratories in the GCC run these instruments daily, which is why the usual request is for a replacement part rather than a complete instrument.

Normalab products we list

Family Examples
Liquid Analytical Systems 13860 Manual Houillon Viscosity Bath, CPP Classic Cloud, FIA Classic Hydrocarbon Types Analyzer (…, NAB 440 Automatic Abel Closed-Cup Flash…, NAB Tech Semi-Automatic Abel Closed-Cup…, NAE 440 Automatic Twin Aniline Point Tes…, NBA 450 Automatic Ring and Ball Softenin…, NCL 120 Manual Cleveland Open-Cup Flash… and 17 more
Temperature Instrumentation 60448
Pumps 60606 Circulation Pump for NDI 450 Coola…

Ordering Normalab parts: what we need

Send the Normalab reference number of the part, the instrument model and its serial number, and a photo of the instrument nameplate. Where the part is a sensor or pump, a photo of the old item with its connector helps the manufacturer confirm the version fitted to your serial number. If you need a complete instrument, name the test method and the products you test.

Facts checked against Normalab's product catalogue and company profiles, October 2026.

FAQ

Questions about the Normalab NDI 450 Automated Atmospheric Distillation

Answers from our engineering team on installation, compatibility and integration.

01Which standards does the NDI 450 cover?
The brochure lists ASTM D86, D850, D1078, ISO 3405, IP 123, IP 195, GOST B, JIS K2254 and DIN 51751. It is ready to use for D86 and related methods, with an optional kit for D850 and D1078.
02What is the difference between 60502 and 60503?
Both are the same NDI 450 analyzer. 60502 is for AC 230 V and 60503 is for AC 115 V, both 50/60 Hz and 9 A.
03How does the instrument cool the condenser?
It uses a built-in Peltier cooling system with a condenser temperature of 0 to 60 °C and a receiver compartment temperature of 0 to 60 °C.
04Does it calculate cetane index?
Yes. The brochure states automatic cetane index calculation per ISO 4264 and automatic barometric pressure correction.
05What fire protection is fitted?
Automatic fire detection, an automated nitrogen extinguisher with the N2 bottle delivered with the unit, an extinguisher push button and sample overheating protection at 600 °C.