Brand:

Normalab NGT Classic Existent Gum Tester for Fuels

Description

Existent gum testing by air or steam jet evaporation

The NGT Classic is a five-station Normalab gum tester. A measured portion of fuel is evaporated from a tared beaker under a controlled jet of heated air or superheated steam; the residue left after evaporation is the existent gum, which is weighed and reported. High gum content signals deposits that can foul fuel systems, intake valves and injectors, so the test matters for aviation fuel, motor gasoline and other volatile distillates.

Normalab lists ASTM D381, IP 131, IP 540, ISO 6246 and DIN 51784 and related methods. The scope covers the existent gum content of aviation fuels and the gum content of motor gasolines or other volatile distillates in their finished form at the time of test. A switch selects air or steam, and the device carries digital controllers for air temperature and for both air and steam flows.

Technical specifications

Parameter Specification
Manufacturer / part number Normalab (France) / 941320
Standards listed ASTM D381, IP 131, IP 540, ISO 6246, DIN 51784
Stations 5, with easy installation of 5 injectors
Method Air jet or steam jet evaporation, selectable
Temperature range Ambient to 260 °C
Air temperature 155 °C ± 0.5 °C
Steam temperature 232 °C ± 2 °C
Air and steam flow 1000 ml/s ± 15 %
Control Digital temperature control and digital flowmeter
Steam method Integrated superheater
Safety Power cut in case of overheating
Required pressure 2.15 bar max
Power supply 230 V AC, 50/60 Hz, 16 A
Dimensions W x D x H 580 x 350 x 440 mm (630 mm height with grip)
Weight 40 kg

Delivered items, accessories and calibration

The NGT Classic is delivered with an ASTM 3C thermometer (P/N 11491, certificate on request), five beakers (P/N 19035), five conical outlets (P/N 941306), forceps for handling (P/N 10686) and a pair of gloves (P/N 9417444). Accessories and options must be ordered separately according to the application.

Listed accessories are a two-stage air compressor with 300 litre capacity (ref. 9413080, 400 V three-phase), a steam generator (ref. 941309, 400 V three-phase) with automatic water feeding, water level control, a pressure switch, a safety valve and a manometer, and a 120 g analytical balance with 0.1 mg readability (ref. 24248). The web page also lists 150 ml filter funnels with 50 mm frits, a class A 50 ml graduated beaker, a dryer and a porcelain plate for the desiccator.

For quality control, Normalab offers a calibration kit for air and a calibration kit for steam, which help verify the flow and temperature conditions the method requires.

Planning the installation

The steam method needs a steam generator and the air method needs a compressor of adequate capacity, and both of those listed accessories need a 400 V three-phase supply, unlike the 230 V tester itself. Confirm the utilities before ordering, together with the beakers and funnels, a balance, a drying oven or desiccator, and wash solvents.

Choose the NGT Classic when the laboratory tests jet fuel or gasoline gum regularly and wants digital control of the air and steam conditions. Related equipment is grouped under liquid analytical systems.

Documents and quotation

The Normalab NGT Classic leaflet is attached. Send the fuels tested, method (air or steam) and accessories for a quotation; availability and lead time are confirmed before ordering.

Datasheets & Documents
  • PDF Brochure Normalab NGT Classic leaflet (EN) Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Normalab is a French manufacturer of laboratory instruments for petroleum product testing, established in 1963 and based in Valliquerville, Normandy. The company started out as glassblowers and still runs a glassblowing workshop. For more than 55 years it has designed and built test instruments that follow ASTM, ISO and IP methods, including its NPM 450, the fifth generation of its automated flash point testers.

Seven Star LLC, an independent supplier in Oman, lists Normalab spare parts: the 60606 circulation pump for the coolant circuit of the NDI 450 automatic distillation analyzer, and the 60448 thermal detection thermocouple. We source Normalab parts by their five-digit reference number.

Normalab company profile

Founded 1963, France
Headquarters Valliquerville, Normandy, France
Origin Laboratory glassblowing
Certification ISO 9001
Product families NPM 450 automated flash point testers, automatic distillation analysers, cloud and pour point testers, kinematic viscosity baths, copper corrosion and oxidation stability testers, laboratory glassware

Normalab instrument families

The manufacturer arranges its instruments by the property tested: flammability, volatility, viscosity, cold flow properties, cleanliness and purity, lubricants, and bitumen, wax and grease. Its glassware line supports the same tests. The methods referenced are ASTM, ISO and IP. Refinery and fuel quality laboratories in the GCC run these instruments daily, which is why the usual request is for a replacement part rather than a complete instrument.

Normalab products we list

Family Examples
Liquid Analytical Systems 13860 Manual Houillon Viscosity Bath, CPP Classic Cloud, FIA Classic Hydrocarbon Types Analyzer (…, NAB 440 Automatic Abel Closed-Cup Flash…, NAB Tech Semi-Automatic Abel Closed-Cup…, NAE 440 Automatic Twin Aniline Point Tes…, NBA 450 Automatic Ring and Ball Softenin…, NCL 120 Manual Cleveland Open-Cup Flash… and 17 more
Temperature Instrumentation 60448
Pumps 60606 Circulation Pump for NDI 450 Coola…

Ordering Normalab parts: what we need

Send the Normalab reference number of the part, the instrument model and its serial number, and a photo of the instrument nameplate. Where the part is a sensor or pump, a photo of the old item with its connector helps the manufacturer confirm the version fitted to your serial number. If you need a complete instrument, name the test method and the products you test.

Facts checked against Normalab's product catalogue and company profiles, October 2026.

Frequently Asked Questions

Which standards does the NGT Classic cover?
ASTM D381, IP 131, IP 540, ISO 6246 and DIN 51784 and related methods.
How many samples can it evaporate at once?
It has five stations with five injectors.
Can it use both air and steam?
Yes, a switch selects air or steam and an integrated superheater supports the steam method.
What flow and temperatures does it control?
Air at 155 °C ± 0.5 °C, steam at 232 °C ± 2 °C and a flow of 1000 ml/s ± 15 %.
Is an air compressor included?
No, the compressor and steam generator are listed as separate accessories.