Brand:

Fluke 726 Precision Multifunction Process Calibrator

Description

726 precision multifunction process calibrator with HART mode

The Fluke 726 Precision Multifunction Process Calibrator is a two-channel field calibrator for the process industry. It measures and sources volts, milliamps, RTDs, thermocouples, frequency and resistance, and with Fluke 700Pxx pressure modules it measures or, with a pump, sources pressure. Accuracies are specified to 0.01 %, it calculates transmitter error, stores results and has a HART mode that inserts a 250 ohm resistor. Seven Star LLC supplies the 726 in Oman and the GCC as an independent supplier in Muscat, not affiliated with the manufacturer.

The ordering information lists the “Fluke-726 Precision Multifunction Process Calibrator”. Included are TL75 test leads, AC72 test clips, a pair of stackable test leads, a product overview and a user’s manual on CD-ROM.

726 technical specifications

Parameter Specification
Manufacturer / part number Fluke / 726
Voltage 0 to 100 mV; 0 to 20 V source; 0 to 30 V measure; 0.01 % of reading + 2 LSD
Milliamps 0 to 24 mA, 0.001 mA resolution, 0.01 % of reading + 2 LSD; maximum load 1000 Ω
Thermocouple millivolts -10.00 mV to +75.00 mV, 0.01 % of range + 1 LSD
Resistance 5 Ω to 4000 Ω, 0.015 %
Frequency 2.0 CPM to 15.0 kHz in four ranges, 1 V to 20 V p-p square wave source
Loop supply 24 V dc, 10 %
Thermocouples J, K, T, E, L, N, U, B, R, S, C, BP, XK
RTDs Cu10, Ni120, Pt100 (385, 3916, 3926), Pt200, Pt500, Pt1000, plus custom RTD curves
Pressure Measure with any of 29 Fluke 700Pxx modules; source with pump
Channels Two: measure, source and view simultaneously
Memory Up to 8 calibration results
Operating / storage temperature -10 °C to 55 °C / -20 °C to 71 °C
Batteries Four AA alkaline, about 25 hours typical
Size / weight 200 x 96 x 47 mm / 650 g
Warranty Three years

What the second channel and error calculation do on the 726

Two channels let the 726 source a signal into a transmitter on one channel and measure the output on the other, with both visible on the screen. The transmitter error percentage is calculated for the technician at each test point, so the pass or fail decision does not need a calculator. Up to eight calibration results can be stored in the calibrator and recalled later for analysis, which supports basic record keeping for quality systems.

The integrated pressure switch test captures set, reset and deadband values. Custom RTD curves allow the constants of a certified RTD probe to be entered for better temperature measurement. A frequency totaliser and pulse-train source support flowmeter checks.

The simultaneous-function table of the datasheet shows how the two channels are shared. Milliamps can be measured on channel A while channel B measures or sources milliamps, and a pressure module can be read while milliamps are sourced for an I/P converter. Not every pairing of functions is possible at the same time so check the table in the datasheet before planning a test that needs two unusual signals together.

HART mode and loop power on the 726

Many smart transmitters communicate over the 4 to 20 mA loop. A loop needs at least 250 ohms of resistance for a HART communicator to work, and the HART mode of the 726 inserts that resistor in the mA measure and source functions. The 24 V loop supply can power the transmitter while the milliamp reading is made. The 726 is not itself a HART communicator, so trimming a smart transmitter still needs a communicator or a documenting calibrator such as the 754.

For a fast linearity check the auto step and auto ramp features apply a series of points without keying each value, and the stored test setups can be reused on identical transmitters. This is where the 0.01 % source accuracy pays off: it keeps the calibrator uncertainty small compared with the tolerance of the transmitter under test.

Selecting the 726

  • Choose the 726 over the 725 when you need 0.01 % accuracy, two channels or stored results.
  • Confirm that the 29 Fluke 700Pxx modules cover your pressure ranges.
  • Check that thermocouple types such as C, BP and XK are in your list if you use them.
  • Use the 754 if you need full documenting capability and HART communication.
  • Order a pump and hoses if you plan to source pressure.

Use and maintenance of the 726

The newer input protection design improves reliability against accidental overvoltage, but the maximum input remains 30 V. Use the correct thermocouple wire or miniature connector for thermocouple tests. Keep the instrument in a protective case, charge or replace the four AA batteries before long jobs and send it for periodic calibration against a higher-level standard. Operating temperature is -10 °C to 55 °C, with humidity limits that fall as temperature rises.

Typical applications of the 726 in Oman and the GCC

The 726 is used in plants where instruments must be calibrated to tight tolerances, such as custody and process control loops in oil and gas, power station instruments and desalination control systems. Technicians use it on temperature, pressure, level and flow transmitters, for valve and I/P tests, and for alarm switch checks. High ambient temperatures in plant areas make the 55 °C operating limit useful.

Information to send with an RFQ for the 726

  • Part number 726
  • Quantity and pressure modules, pump and hoses required
  • Instrument types, signal ranges and tolerances
  • Delivery location and time frame
  • Calibration certificate requirement

Storage, handling and spares for the Fluke 726

Keep the Fluke 726 in its carrying case when it is not in use, and store it in a clean, dry place. Remove or charge the batteries as recommended in the user manual before long storage. Field teams in Oman and the GCC should avoid leaving the instrument in a vehicle in the sun, as high cabin temperatures can exceed the storage range and shorten battery life.

  • Check the instrument, cables, hoses or lenses for damage before each job.
  • Clean the unit with a soft, slightly damp cloth only; do not use solvents unless the manual allows them.
  • Plan a calibration interval that matches your quality system, and keep the calibration certificate with the asset record.
  • Order wear parts and accessories, such as batteries, test leads, hoses or adapters, together with the instrument.

For calibration and test work, it helps to state in your RFQ whether you need an accredited calibration certificate with the Fluke 726, so the right documents can be confirmed with the quotation.

Documents and quotation

A datasheet for the 726 is attached on this page.

Send us the part number 726 and the quantity and we will confirm availability and a quotation.

Other Models in the Fluke Process Calibrator Range

Seven Star LLC supplies models from the Fluke Process Calibrator range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.

ModelCategory
Fluke 725 Multifunction Process CalibratorProcess, Pressure & Temperature CalibratorsView details
Fluke 726 Precision Multifunction Process CalibratorProcess, Pressure & Temperature CalibratorsYou are viewing this model
Datasheets & Documents
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Fluke is the American maker of handheld electronic test tools: digital multimeters, clamp meters, insulation testers, process calibrators, thermometers, vibration meters and power quality analysers. John Fluke founded the John Fluke Engineering Company in Springdale, Connecticut, in June 1948 and moved it to Seattle in 1952. Fluke is headquartered in Everett, Washington, was acquired by Danaher in 1998 and has been part of Fortive since Fortive's spin-off from Danaher in 2016.

Seven Star LLC lists Fluke process calibrators, electrical test tools and bench calibrators, and sources them for maintenance teams and calibration laboratories in Oman.

Fluke company profile

Founded June 1948, Springdale, Connecticut, USA, as the John Fluke Engineering Company
Founder John M. Fluke Sr.
Headquarters Everett, Washington, USA (moved to the Seattle area in 1952)
Ownership Danaher from 1998; part of Fortive (NYSE: FTV) since July 2016
Product families Digital multimeters and clamp meters, insulation resistance testers, process and loop calibrators, infrared thermometers and thermal imagers, vibration meters, power quality analysers, Fluke Calibration laboratory standards

Milestones

  • 1948: John Fluke starts the company in Springdale, Connecticut.
  • 1952: The company moves to Seattle
  • 1987: A joint venture with Philips leads to the ScopeMeter handheld oscilloscope.
  • 1998: Danaher acquires Fluke.
  • 2016: Fluke becomes part of Fortive, spun off from Danaher in July.

Fluke calibrators and test tools listed

The main Fluke products on this page fall into three groups, besides other test tools such as further multimeters, a thermometer, an anemometer, a vibration tester and a portable appliance tester:

  • Process tools: the 754 documenting process calibrator with HART, 789 ProcessMeter, 721 dual sensor pressure calibrator and 729 automatic pressure calibrator
  • Electrical test: the 376 FC true-RMS clamp meter with iFlex probe and the 1587 FC insulation multimeter
  • Fluke Calibration bench standards: 5522A, 5502A, 5500A and 5080A multi-product calibrators, the 5720A multifunction calibrator, 8508A reference multimeter, 6100A electrical power standard and 9640A RF reference source

Several of the bench models belong to earlier generations. When a model is no longer produced, we check the manufacturer's documentation and propose the successor it names.

Where Fluke instruments are used

Fluke process calibrators are used for loop checks, transmitter calibration and valve tests in oil and gas, power and water plants. Clamp meters and insulation testers serve electrical maintenance on motors, switchgear and cables. The bench calibrators and reference multimeters are the working standards of electrical calibration laboratories in the GCC, where they calibrate multimeters, clamp meters, process tools and power analyzers.

Fluke products we list

Family Examples
Portable Test & Measurement 117 Electrician’s True-rms Multimeter with, 1507 Insulation Resistance Tester, 1550C 5 kV Insulation Resistance Tester, 1587 FC Insulation Multimeter, 376 FC True-RMS AC/DC Clamp Meter with iFlex, 5080A Multi-Product Calibrator, 5500A Multi-Product Calibrator, 5502A Multi-Product Calibrator and 17 more
Lubrication & Bearing Maintenance 805 FC Vibration Meter

Ordering Fluke instruments: what we need

Send the exact model with its suffix, since versions differ. A Fluke 721, for example, is defined by its two pressure ranges, and a 729 by its maximum pressure. For bench calibrators, list the options you need, such as oscilloscope calibration or power quality options, the line voltage, and the type of calibration certificate you require from the manufacturer. For Fluke multimeters and clamp meters, add the kit contents wanted, such as probes, leads or cases.

Frequently requested Fluke models: Fluke 5522A Multi-Product Calibrator.

Facts checked against published histories of Fluke Corporation and Fortive, October 2026.

FAQ

Questions about the Fluke 726 Precision Multifunction Process Calibrator

Answers from our engineering team on installation, compatibility and integration.

01What accuracy does the Fluke 726 offer?
The 726 specifies 0.01 % of reading plus 2 least significant digits for voltage and milliamps. Resistance is 0.015 % and thermocouple millivolts 0.01 % of range plus 1 digit.
02Does the Fluke 726 have HART mode?
Yes. HART mode inserts a 250 ohm resistor in the mA measure and source functions. It is not a full HART communicator, so it cannot trim a smart transmitter itself.
03How many results can the Fluke 726 store?
It stores up to 8 calibration results in memory so that data can be recalled from the field for later analysis. The 725 does not offer this.
04Can the Fluke 726 measure pressure?
Yes, using any of 29 Fluke 700Pxx pressure modules. It measures pressure and can source pressure when a pressure pump is connected.
05What is included with the Fluke 726?
The package includes TL75 test leads, AC72 test clips, a pair of stackable test leads, a printed overview and the user manual on CD-ROM.