Hanna Instruments HI98191 Portable pH/ORP/ISE Meter

Description

Hanna HI98191 waterproof portable pH, ORP and ISE meter: what it is

The Hanna Instruments HI98191 is a heavy-duty, handheld meter that measures pH, ORP (redox, shown as mV) and ion-selective electrode (ISE) concentration, together with temperature. It is designed to bring bench-level resolution to plants, utilities and field sites: pH can be displayed to three decimals, calibration can use up to five points, and the case is rated IP67. The Hanna HI98191 is the ISE-capable member of a two-model pair; its sibling, the HI98190, covers pH, mV and temperature only. Seven Star LLC is an independent supplier in Muscat and quotes the Hanna HI98191 for customers in Oman and the GCC who need a rugged pH meter with diagnostics.

In practice the HI98191 suits operators who have outgrown a pocket tester but do not want to tie a laboratory benchtop meter to a process area. It accepts a standard BNC pH or ORP electrode, a separate reference input and a temperature probe input, so the sensor can be swapped for the application instead of replacing the whole meter.

Reading the HI98191 model name and what is in the carry case

Hanna uses the HI98190 and HI98191 numbers for one platform with two feature levels. The HI98191 is supplied with the HI72911B titanium-body pH electrode with a built-in temperature sensor, so automatic temperature compensation works from the first measurement. The titanium body resists aggressive chemicals and is described by the manufacturer as virtually unbreakable, which is a useful property in a plant environment where glass-body electrodes are often broken.

  • HI72911B titanium-body pH electrode with BNC connector and 1 m cable
  • HI7004M pH 4.01 and HI7007M pH 7.01 buffer solutions, 230 mL each
  • HI700601 electrode cleaning solution sachets, 2 x 20 mL, and two 100 mL plastic beakers
  • 1.5 V AA batteries (4), HI920015 micro USB cable and HI92000 PC software
  • HI720191 hard carrying case, instruction manual, quick start guide and quality certificate

For ORP work the manufacturer lists the HI3131B and HI3230B electrodes; they are ordered separately.

HI98191 technical specifications

Parameter Specification
Manufacturer / part number Hanna Instruments / HI98191
pH range and resolution –2.000 to 20.000 pH, selectable 0.1, 0.01 or 0.001 pH
pH accuracy at 25 °C ±0.1 / ±0.01 / ±0.002 pH, matching the three resolutions
pH calibration Up to five points; seven standard buffers (1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45) plus five custom buffers
mV range, resolution, accuracy ±2000.0 mV, 0.1 mV, ±0.2 mV
Temperature –20.0 to 120.0 °C, 0.1 °C resolution, ±0.4 °C excluding probe error; manual or automatic compensation
ISE range 1.00 × 10–7 to 9.99 × 1010 concentration, three significant digits
ISE accuracy ±0.5% of reading for monovalent ions, ±1% for divalent ions
ISE calibration Up to five points from six standards (0.1, 1, 10, 100, 1000, 10000 ppm); 15 ISE types preprogrammed
Data logging Log on demand, up to 300 samples (100 each for pH, mV and ISE)
Interface Opto-isolated micro USB for logged data and GLP records, with HI92000 software
Power 4 x 1.5 V AA, about 200 hours without backlight and about 50 hours with backlight
Environment 6 to 50 °C, up to 100% RH, IP67 enclosure

How the HI98191 measures and diagnoses the electrode

A pH electrode produces a small voltage that changes with hydrogen-ion activity; the HI98191 converts that voltage with a high-impedance input and applies the slope and offset found at calibration. The meter’s distinguishing feature is its Cal Check diagnostics. After a calibration it compares the new offset and slope with earlier data, warns when the electrode needs cleaning or when a buffer may be contaminated, and shows overall probe condition as a percentage in steps of 10%. An optional out-of-calibration-range warning tells the operator when a reading falls outside the calibrated pH bracket, and a user-set calibration time-out reminds the team when a new calibration is due.

In ISE mode the same input measures ion concentration. Selecting one of the preprogrammed ion types applies the correct charge and molar weight so the meter can display slope percentage, and results can be shown in ppm, ppt, g/L, ppb, µg/L, mg/mL, M, mol/L, mmol/L or % w/v, or in a user-defined unit. Relative mV mode and an AutoHold function, which freezes the first stable reading, complete the measurement toolkit.

Where the HI98191 fits: electrodes, ISE sensors and related models

The HI98191 sits above the pocket testers and below the multiparameter meters in the Hanna portable range. Customers who only need pH, mV and temperature can compare it with the HI98190, which is a pH-only version with an amplified electrode. Customers who need pH plus conductivity or dissolved oxygen in a single probe should look at the HI98195 or HI98196 multiparameter meters instead. Because the HI98191 uses BNC connectors it also works with other Hanna pH, ORP and ISE electrodes; confirm the connector type and cable length before ordering a replacement sensor.

Selection checklist for the HI98191

  • Confirm that ISE measurement is needed; if not, the HI98190 is the simpler choice.
  • Define the sample: temperature, aggressiveness and whether a titanium-body electrode or a different sensor suits the chemistry.
  • State the accuracy needed: 0.1, 0.01 or 0.002 pH determines how many calibration points and buffers you will use.
  • Decide whether logged data will be moved to a PC, which requires the micro USB cable and HI92000 software supplied with the meter.
  • List spare buffers, cleaning solution and storage solution to order with the meter.

Using, calibrating and maintaining the HI98191

Calibrate with fresh buffer in the beakers supplied, rinsing the electrode between solutions and allowing the reading to stabilise. A two- or three-point calibration with 4.01, 7.01 and 10.01 is typical for general work; use 1.68 or 12.45 when the sample sits at an extreme. After calibration read the probe-condition percentage; a falling figure over several calibrations means the electrode needs cleaning with the HI700601 solution or replacement. Keep the electrode wet in storage solution, never in distilled water, and replace the four AA cells before a long field trip.

The GLP key recalls date, time, offset, slope and buffers from the last calibration, which is helpful when an auditor asks for evidence. The log-on-demand memory and USB link make it simple to transfer results to a spreadsheet, while the backlit graphic display and contextual help key reduce the training needed for new operators.

Typical HI98191 applications in Oman and the GCC

In the Gulf the HI98191 is used for pH and ORP checks on desalination pretreatment and post-treatment, cooling-water and boiler-water monitoring, wastewater treatment, plating and surface-treatment baths, chemical manufacturing and food and beverage plants. Heat and dust make the IP67 housing and the chemical-resistant titanium electrode practical advantages when the meter is carried between areas. ISE mode allows spot checks of species such as fluoride, ammonium or nitrate where the right electrode is fitted, subject to the electrode’s own limits and interferences.

HI98191 compared with the HI98190

Item HI98191 HI98190
Parameters pH, ORP (mV), ISE, temperature pH, mV, temperature
ISE measurement Yes, 15 types preprogrammed No
Electrode connector BNC with separate reference and temperature inputs DIN, amplified electrode supplied
Log on demand 300 samples 200 samples

Information to send with your HI98191 RFQ

  • Quantity and whether a titanium-body pH electrode is sufficient or an ORP or ISE electrode is also required.
  • Sample type, temperature range and pH range expected.
  • Accessories needed: extra buffers, storage and cleaning solutions, spare electrode or longer-cable version.
  • Delivery location in Oman or the GCC and any documentation required with the shipment.

Documents and quotation

The instruction manual for the HI98190 and HI98191 is attached on this page and covers specifications, calibration, ISE set-up, GLP and logging in detail.

Send us the part number HI98191 with your quantity and accessory list, and Seven Star LLC will quote with availability and lead time for Oman and the GCC. We are an independent supplier and not affiliated with the manufacturer.

Datasheets & Documents
  • PDF Manual Hanna Instruments HI98191 Instruction Manual Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Hanna Instruments is a maker of analytical instruments for water and liquid testing, founded in 1978 in the Padua area of Italy by Oscar and Anna Nardo. Its early breakthrough was the pHep, an affordable pocket-sized pH tester. The company is now led by the founders' children, Martino and Pamela Nardo, with its headquarters in Woonsocket, Rhode Island, and offers more than 3,000 products.

Seven Star LLC is an independent supplier in Muscat. We list Hanna instruments for water, wastewater and laboratory users in Oman and the GCC.

Hanna Instruments company profile

Founded 1978, near Padua, Italy, by Oscar and Anna Nardo
Headquarters Woonsocket, Rhode Island, USA
Leadership Martino Nardo and Pamela Nardo
Early product The pHep pocket pH tester
Product families pHep and HI98 waterproof testers, edge benchtop meters, HI83 multiparameter photometers, Checker HC handheld colorimeters, HI931 and HI932 titrators, process controllers, reagents and calibration buffers

Hanna product series

Portable meters include the HI98190 pH and ORP meter, the HI98192 EC, TDS and salinity meter, the HI98193 dissolved oxygen and BOD meter, the HI98199 pH, EC and DO meter and the HI9829 multiparameter water quality meter.

The HI98703 is an EPA compliant portable turbidity meter, the HI97711 a free and total chlorine photometer and the HI83414 a benchtop turbidity and chlorine meter.

Waterproof portable meters such as the HI98190 and HI98192 suit field use, with automatic temperature compensation and GLP data for audits. Turbidity measurement with the HI98703 follows EPA method 180.1, and the HI97711 photometer measures free and total chlorine for disinfection control.

Where Hanna instruments are used

Drinking water and desalination plants, wastewater treatment, environmental field surveys, aquaculture, cooling water and quality control labs.

Hanna Instruments products we list

Family Examples
Portable Test & Measurement Hanna HI97711 Free and Total Chlorine Po…, Hanna HI98190 Waterproof Portable pH and…, Hanna HI98192 Waterproof Portable EC, Hanna HI98193 Waterproof Portable Dissol…, Hanna HI98199 Portable pH, Hanna HI9829 Multiparameter Water Qualit…, Hanna HI98703 Portable Turbidity Meter

Ordering Hanna instruments: what we need

Send the model, the parameters and ranges, the probe and cable length and any reagents or calibration solutions required.

Facts checked against Hanna Instruments' own company story and City of Woonsocket business profile, October 2026.

FAQ

Questions about the Hanna Instruments HI98191 Portable pH/ORP/ISE Meter

Answers from our engineering team on installation, compatibility and integration.

01What does the HI98191 measure that the HI98190 does not?
The HI98191 adds ion-selective electrode (ISE) measurement with 15 preprogrammed ion types and concentration units, and uses a BNC input with separate reference and temperature connections. The HI98190 is limited to pH, mV and temperature.
02How many points can the HI98191 use for pH calibration?
It supports up to five calibration points drawn from seven standard buffers (1.68, 4.01, 6.86, 7.01, 9.18, 10.01 and 12.45) and five custom buffers. After calibration it reports electrode condition as a percentage.
03What accuracy and resolution does the HI98191 give for pH?
At 25 °C the accuracy is ±0.1, ±0.01 or ±0.002 pH depending on whether 0.1, 0.01 or 0.001 pH resolution is selected, over a range of –2.000 to 20.000 pH.
04Is the HI98191 waterproof?
The enclosure is rated IP67. That protects against dust and temporary immersion, so the meter can be used in wet areas, but the electrode connector and cables should still be kept clean and dry when stored.
05How is logged data and calibration history retrieved from the HI98191?
Up to 300 samples can be stored on demand and transferred over the sealed micro USB port with the HI92000 software. The GLP key shows the date, offset, slope and buffers of the last calibration on the display.