Metrohm 885 Compact Oven Sample Changer for KF Titration
Description
Overview
The Metrohm 885 Compact Oven Sample Changer automates thermal sample preparation for Karl Fischer water determination. Samples are weighed into screw-cap vials, loaded on a rack, and the changer takes them one at a time into an oven, where the moisture is driven out and carried by dry gas into a titration cell. Seven Star LLC is an independent supplier serving Oman and the GCC. The notes below follow the Metrohm brochure for the 885 and the product page for article 2.885.0010, and they describe what the oven does, which titrators it works with and how it relates to Metrohm’s other oven products.
The oven method in brief
Some samples cannot be dissolved in the titration medium, release their water only at high temperature, or react with the KF solution. For them, Metrohm recommends the oven method: the sealed vial is heated, a double hollow needle pierces the septum, and a stream of dry carrier gas sweeps the evaporated water into the cell. Because the sample never touches the KF solution, contamination of the oven and the cell is avoided, and carryover or memory effects that could distort results are excluded. Depending on the water content, the titration may be coulometric or volumetric.
Specification of Metrohm 885 Compact Oven Sample Changer
| Item | Value |
|---|---|
| Function | automatic oven sample changer for Karl Fischer titration |
| Rack positions | 18 |
| Oven temperature range | 50 to 250 degrees C |
| Heating rate | typically 15 degrees C per minute |
| Cooling rate | typically 9 degrees C per minute |
| Carrier gas flow | 10 to 150 mL per minute |
| Control | integral keypad; set oven temperature, gas flow and number of samples |
| Vials | 8 mL screw-cap vials with septum, reusable |
| Operating range | 5 to 45 degrees C ambient; humidity below 80 percent |
| Supply | 100 to 120 V or 220 to 240 V, 50 or 60 Hz |
| Power input | 200 W |
| Article number | 2.885.0010 |
Workflow with the Metrohm 885 Compact Oven Sample Changer
After weighing in, the operator closes the vials, places them on the rack and starts the series. Only the oven temperature, the gas flow and the number of samples need to be set on the keypad. If the titrator is linked with a balance, the sample weights do not have to be typed in. Screw-cap vials are used, so after measurement they can be opened, cleaned and reused, and only the pierced septum must be replaced. The brochure lists screw caps with septum, loose septa and 8 mL sample vessels as consumables, and a USB mini cable to connect the unit.
Which titrators work with it
Metrohm states that the changer can be operated with all stand-alone Karl Fischer titrators that have a sample table. The ordering list names the 851 and 852 Titrando coulometers with diaphragm-free generator electrodes and a 900 Touch Control, the 890 Titrando with 900 Touch Control, the 915 KF Ti-Touch, the 917 Coulometer and the Eco Coulometer, with a remote cable for the last of these. For trace water in plastics or oils, the coulometric route is usual; for higher contents, volumetric.
How 885 Compact Oven Sample Changer differs from the 860 and 874
The 860 KF Thermoprep is the manual oven: one vial at a time is placed by hand. The 885 adds an 18-position rack and automatic sequence handling in a small footprint, which Metrohm describes as affordable automation. The 874 Oven Sample Processor is the larger system for high sample throughput. If a laboratory runs fewer than a dozen oven samples a day and always the same temperature, the 860 is enough. The 885 suits batches that need unattended running overnight, for example plastics granules or powders from production lines.
Reproducibility and quality control
Automation improves repeatability because every vial spends the same time in the oven before the needle pierces it, and every transfer follows the same timing. That removes operator influence from the heating stage, which is often the largest source of scatter in manual oven work. Include a certified water standard in each batch, for example a sealed standard vial, to prove the system is working, and run empty vials as blanks at the same temperature so that background moisture can be subtracted. Keep the results with the method settings so an auditor can see that the oven was operated according to procedure.
Where it is used in Oman and the GCC
Polymer and cable manufacturers, pharmaceutical packaging plants, lubricant blenders and food processors all need water data on matrices that do not dissolve. Moisture in plastics affects extrusion quality, moisture in powders affects flow and shelf life, and moisture in oil affects performance. Because the oven avoids special solvents, laboratories also save on hazardous chemical storage.
Choosing sample mass, temperature and gas flow
The three settings on the keypad interact. A higher oven temperature releases water faster but risks decomposing the matrix, a higher gas flow carries the water into the cell faster but dilutes it and can strip the cell solution, and a larger sample mass lengthens the release time. A sensible method development routine fixes the gas flow first, often near 60 to 80 mL per minute, then runs the same material at several temperatures and watches the extraction curve for a plateau. After that the sample mass is chosen so that the water amount fits the measuring range of the titrator. Record these choices in the method together with vial size and septum type, and repeat a check sample at the start of each batch to confirm that the oven, gas and titrator behave as before.
Set-up and care
Connect the carrier gas through a drying cartridge, set the flow within 10 to 150 mL per minute, link the changer with the titrator, and run blank vials to find the background. Choose the oven temperature from a test series: high enough to release the water, below the decomposition point of the sample. Replace septa regularly, keep the needle clean and inspect tubing for leaks. Allow for the heating rate of about 15 degrees C per minute when planning the first sample of the day.
Related items you may also need: Metrohm 2035 Process Analyzer | Metrohm 916 Ti-Touch Compact Potentiometric Titrator | Metrohm 905 Titrando Potentiometric Titrator | All Metrohm products.
Before you order the Metrohm 885 Compact Oven Sample Changer
- To avoid a wrong delivery, compare the code on the nameplate of the installed unit with the Metrohm 885 Compact Oven Sample Changer code on this page.
- The quotation for the Metrohm 885 Compact Oven Sample Changer states the confirmed code and the lead time to site.
- The attached manufacturer datasheet covers the Metrohm 885 Compact Oven Sample Changer ratings, dimensions and installation details.
- Delivery of the Metrohm 885 Compact Oven Sample Changer is arranged to project sites in Oman and across the GCC, alongside the rest of our liquid analytical systems range.
- A spare Metrohm 885 Compact Oven Sample Changer on the shelf is a simple way to protect critical service from long lead times.
Documents and quotation
The Metrohm brochure for the Metrohm 885 Compact Oven Sample Changer is attached and includes technical specifications and ordering information. Send your RFQ with the model, the titrator you will combine it with, spare vials and septa needed, quantity, and delivery location in Oman or the GCC.
Other Models in the Oven Range
Seven Star LLC supplies models from the Oven range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.
| Model | Category | |
|---|---|---|
| Metrohm 860 KF Thermoprep Karl Fischer Sample Oven | Liquid Analytical Systems | View details |
| Metrohm 885 Compact Oven Sample Changer for KF Titration | Liquid Analytical Systems | You are viewing this model |
Datasheets & Documents
- Datasheet Metrohm 885 Compact Oven Sample Changer for KF Titration Datasheet Download
Reviews (0)
How would you rate this product? Cancel reply
About brand
Metrohm is a Swiss maker of instruments for chemical analysis, founded in 1943 by Bertold Suhner in Herisau, where it is still headquartered. It is best known for titration and also makes ion chromatography, electrochemistry, spectroscopy and process analysers. Metrohm is owned by a foundation.
Seven Star LLC lists the Metrohm 905 Titrando, the 916 Ti-Touch and the 2035 Process Analyzer, and supplies Metrohm instruments and accessories to laboratories and plants in Oman and the GCC.
Metrohm company profile
| Founded | 1943, Herisau, Switzerland |
|---|---|
| Founder | Bertold Suhner |
| Headquarters | Herisau, Switzerland |
| Ownership | Metrohm Foundation |
| Product families | Titrando and Ti-Touch titrators, ion chromatography systems, 2035 and 2060 process analysers, electrodes |
Metrohm instruments listed here
The 905 Titrando is a high-end potentiometric titrator for use with Dosino dosing units. It runs dynamic (DET), monotonic (MET) and set endpoint (SET) titrations, takes up to four 800 Dosino units, recognizes intelligent iTrode electrodes, and is operated from tiamo or OMNIS software or a Touch Control. The 2035 Process Analyzer is an online wet-chemistry analyzer for automatic analysis of samples drawn from the process.
Metrohm products we list
| Family | Examples |
|---|---|
| Liquid Analytical Systems | 2035 Process Analyzer, 905 Titrando Potentiometric Titrator, 916 Ti-Touch Compact Potentiometric Titr… |
Ordering Metrohm equipment: what we need
Metrohm article numbers follow a fixed pattern (the 905 Titrando is 2.905.0010), and that number is the best reference for instruments, electrodes and accessories. For a new application, tell us the sample, the parameter and its concentration range, the method or standard you follow, and for a process analyzer the number of sample streams and the signals your control system needs.
Facts checked against Metrohm's published company history, October 2026.
Browse Metrohm products by category: Liquid Analytical Systems.
FAQ
Questions about the Metrohm 885 Compact Oven Sample Changer for KF
Answers from our engineering team on installation, compatibility and integration.
01How many samples fit on the Metrohm 885 Compact Oven Sample Changer?
The rack holds 18 positions, and the samples are processed one after another automatically once the series is started.
02What temperature does the 885 oven reach?
The temperature range is 50 to 250 degrees C with a typical heating rate of 15 degrees C per minute and cooling rate of 9 degrees C per minute.
03What carrier gas flow can be set on the 885?
The gas flow range is 10 to 150 mL per minute.
04Which titrators can be combined with the 885?
All stand-alone Karl Fischer titrators with a sample table, including the 915 KF Ti-Touch, 917 Coulometer, 890 Titrando and 851 or 852 Titrando coulometers.
05Are the vials of the 885 reusable?
Yes. Screw-cap vials can be opened and cleaned after measuring, and only the pierced septum needs replacing.