Beamex 8003187 PGHH vent valve: what the assembly is

Beamex 8003187 is the replacement vent valve assembly for the Beamex PGHH hydraulic calibration pump. It arrives as a preassembled vent valve mechanism with its handwheel already fitted, and the set also carries a bonded seal and a 6 mm stainless steel release valve ball. The maker describes it as an assembly that can simply be swapped into a PGHH pump when the original has broken.

On the pump the vent valve is item 4 in the overview drawing of the PGHH manual, sitting beside the fine adjust and the reservoir filling plug. It is the control that dumps system pressure back into the reservoir, so a worn or cracked unit shows up quickly as an inability to hold or to release pressure cleanly. Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer.

Contents of the set and how it works

  • Preassembled vent valve mechanism with handwheel.
  • One bonded seal for the valve seat.
  • One 6 mm stainless steel release valve ball.

The PGHH manual lists the seal at this position as a 1/8 inch bonded seal (usit) for the vent valve, and it names the release valve tappet inside the top of the pump as the part the valve acts on. Turning the handwheel clockwise closes the valve for pressure build-up; turning it counterclockwise opens the release path. Because 8003187 is sold as one assembly, you do not have to rebuild the seat from loose parts.

Technical data for the host pump of 8003187

Item Detail
Part number 8003187
Product name PGHH vent valve
Fits Beamex PGHH hydraulic calibration pump
Pump pressure range 0 to 700 bar / 0 to 10,000 psi
Side pressure port G1/4 inch female with pre-installed Bx 1215 male fitting for the 630 bar hose
Top pressure port G1/4 inch B female for an EXT60 to EXT1000 module, closed with a G1/4 inch male plug
Pressure media Low viscosity mineral hydraulic oil (RENOLIN MR 5 VG 22) or distilled water
Reservoir 200 ml / 12.2 cubic inch acrylic, not to be pressurized
Pump weight 1.3 kg / 2.87 lb
Pump size (H x W x D) 265 x 160 x 120 mm
Operating temperature 0 to 50 degrees C
Storage temperature minus 10 to 50 degrees C
Tightening torque 25 Nm for connections, drain screw maximum 15 Nm

Where it matters: field use in Oman and the GCC

A hydraulic pump of this class is used where transmitters and gauges are checked at several hundred bar: wellhead and flowline instrumentation in oil and gas, high-pressure steam and feedwater gauges in power and desalination plants, and hydraulic test circuits in heavy manufacturing. In those jobs the vent valve is operated dozens of times per shift, so it is the first moving part on the pump to show wear.

Keeping a spare assembly in the instrument workshop avoids sending a whole PGHH away for a small fault, which matters when a calibration contract has a fixed turnaround. Shipping glass and liquid-filled pumps to a service centre also carries the rule that the reservoir must be emptied first.

How 8003187 differs from other PGHH service parts

The PGHH family has several separately ordered spare parts, and each addresses a different failure. 8003187 deals only with the venting function. The 8003192 pressure cylinder is a much larger repair for a leaking pump after the cylinder seals have been replaced, the 8003194 reservoir kit replaces the 200 ml acrylic reservoir with two 67 x 3 mm O-rings, and the 8003199 stroke selector kit replaces the prime and high stroke knob and shaft.

If the pump leaks at the handwheel or you cannot fully relieve pressure, 8003187 is the matching part. If pressure falls steadily even though the vent valve is closed, check the seals listed in the manual before ordering anything.

Fitting and commissioning notes

The maker supplies a step-by-step video for this exchange, and the manual states that spare parts are intended for repair by the user or by the maker’s service. Before opening the pump, depressurize it, empty the reservoir and keep it upright. Coat O-rings with a very small amount of silicone grease and tighten connections to 25 Nm.

After fitting 8003187, perform the manual’s leak test. Fill the reservoir, close the vent valve, run the pump to the maximum pressure of the connected module, wait five minutes, return to the set point with the fine adjust and wait two minutes more. At a 700 bar set point the pressure should fall by less than 5 bar in that final period; a larger drop means the pump is still leaking.

Operating cautions that protect the valve

Two rules in the manual are written specifically for this part. First, do not overtighten the vent valve when closing it: the release valve tappet or the top of the pump may be damaged, or the whole valve may break. Second, never unscrew the valve completely, because the tappet inside the pump can fall out and be lost.

For normal pressure release the manual names the fine adjust as the main tool, with the vent valve as an option that needs care to avoid pressure shocks to the measuring system. To dump all pressure, open the vent valve counterclockwise, squeeze the handles fully and only then return the stroke selector to the prime position.

Maintenance and ordering notes

Use only the low viscosity mineral oil or distilled water named by the maker; other fluids such as synthetic hydraulic oil can damage seals and the reservoir. If distilled water is used, empty and dry the pump after each session. Check that the pressure compensation screw is clear, since a blocked screw can break the reservoir.

When you request 8003187, quote the pump serial number and the symptom so the correct kit is confirmed. Related parts that often accompany a vent valve repair are the bonded seal packs and the PGHH full service kit, which should be checked against your pump before ordering.

Documents and quotation

The PGHH user manual is supplied as the datasheet for Beamex 8003187. To receive a quotation, send the part number 8003187, the quantity required and the PGHH serial number in your RFQ to Seven Star LLC. Availability and lead time are confirmed per request.

Other Models in the Pghh Spares Range

Seven Star LLC supplies models from the Pghh Spares range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.

ModelCategory
Beamex 8003187 PGHH Vent Valve Spare AssemblyProcess, Pressure & Temperature CalibratorsYou are viewing this model
Beamex 8003192 PGHH Pressure Cylinder Service KitProcess, Pressure & Temperature CalibratorsView details
Beamex 8003194 PGHH Reservoir Replacement KitProcess, Pressure & Temperature CalibratorsView details
Beamex 8003199 PGHH Stroke Selector Service KitProcess, Pressure & Temperature CalibratorsView details
Datasheets & Documents
  • PDF Datasheet Beamex 8003187 PGHH Vent Valve Spare Assembly Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Beamex is a Finnish maker of calibration equipment and software for process instruments. Four calibration technicians founded the company in 1975 because they could not find the equipment they wanted; their first product was a voltage and current simulator, and the first factory opened in 1977. Beamex is headquartered in Pietarsaari (Jakobstad), Finland, and supplies process industries in more than 100 countries.

Seven Star LLC is an independent supplier listing Beamex documenting calibrators, temperature dry blocks, pressure controllers, calibration pumps and pressure modules for buyers in Oman and the GCC. Equipment is quoted against the model and the pressure or temperature ranges required.

Beamex company profile

Origin Founded in 1975 in Finland by four calibration technicians; first factory 1977
Headquarters Pietarsaari (Jakobstad), Finland; ISO 9001 certified since 1992
Product families MC series multifunction calibrators, temperature dry blocks, automatic and hand pressure generators, external pressure modules, CMX and LOGiCAL calibration software
Typical duty Field and workshop calibration of pressure, temperature and electrical signal instruments, including HART and fieldbus transmitters

Milestones

  • 1975: Four calibration technicians found the company in Finland.
  • 1977: The first factory opens.
  • 1992: ISO 9001 certification.

Beamex calibrators and accessories

Model Type Typical use
MC6 Advanced field calibrator and communicator Pressure, temperature and electrical calibration with a built-in HART, FOUNDATION Fieldbus and PROFIBUS PA communicator, documented results
MC6-Ex Intrinsically safe field calibrator Calibration inside Zone 0 and Zone 1 hazardous areas without a hot work permit
MC6-T Temperature calibrator with integrated dry block Temperature sensors and transmitters, with multifunction electrical and pressure measurement
MC4, MC2 Documenting and hand-held process calibrators Routine loop, transmitter and switch checks in the field
FB, MB Field and metrology temperature dry blocks FB for field calibration; MB for higher-accuracy workshop and laboratory work
ePG, POC8 Electric pressure pump and automatic pressure controller Automated pressure calibration; POC8 for workshop calibration benches
PGC, PGM, PGL, PGPH, PGHH Hand pressure pumps Pressure and vacuum generation in the field, from low pressure (PGL, ±400 mbar) to hydraulic 700 bar (PGHH)
EXT modules External pressure modules Extending MC calibrators to additional ranges, 70 kPa to 100 MPa

Where Beamex equipment is used

Instrument workshops and maintenance teams in oil and gas, refining, petrochemicals, power, water and pharmaceuticals use Beamex calibrators to verify transmitters, switches, gauges and temperature sensors. Documenting calibrators send results to calibration software, which keeps the audit trail required by quality systems and custody transfer metering procedures. Intrinsically safe models are used for in-situ calibration in classified areas.

Ordering Beamex equipment: what we need

State the model and the options you need: internal and external pressure ranges, temperature range, communication protocols (HART, FOUNDATION Fieldbus, PROFIBUS PA), Ex approval, and whether an accredited calibration certificate is required. For spare modules and pumps send the serial number of the calibrator they will be used with, so compatibility can be confirmed.

Facts checked against Beamex company history and product information, October 2026.

FAQ

Questions about the Beamex 8003187 PGHH Vent Valve Spare Assembly

Answers from our engineering team on installation, compatibility and integration.

01What does Beamex 8003187 contain?
It contains a preassembled PGHH vent valve mechanism with handwheel, a bonded seal and a 6 mm stainless steel release valve ball.
02Which pump uses the 8003187 vent valve?
It is made for the Beamex PGHH hydraulic calibration pump rated 0 to 700 bar, where the vent valve is item 4 in the manual's overview.
03Can I unscrew the vent valve completely?
No. The manual cautions that the release valve tappet inside the pump may fall out and be lost if the valve is fully unscrewed.
04How do I test the pump after replacing the vent valve?
Run the manual's leak test: at a 700 bar set point the pressure should fall by less than 5 bar in the final two minutes after the five-minute settling period.
05Does 8003187 replace the reservoir or the pressure cylinder?
No. It only replaces the vent valve. The reservoir, pressure cylinder and stroke selector have their own separate kits.