Brand:

Beamex 8003126 PGC Cylinder Assembly with Selector

Description

What Beamex 8003126 is

Beamex 8003126 is the cylinder assembly of the Beamex PGC calibration pump, supplied together with the preassembled pressure and vacuum selector, the nameplate sticker and the cylinder mounting screws. In the PGC it forms the upper working half of the pump, where air is compressed or evacuated by the piston below. Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer.

The maker explains that a complete replacement of the top part also needs the fine adjust with release valve, 8003122, and a Bx G1/8 male to G1/8 male fitting, 7113270, though existing ones can be reused after replacing the cylinder assembly.

Included items

  • Cylinder with preassembled pressure/vacuum selector.
  • Nameplate sticker.
  • Cylinder mounting screws (the manual lists M3 x 6, four pieces, at this location).

The product image shows a black cylinder with a green band and two side studs, with small screws at the foot.

Data sheet values for the PGC fitted with 8003126

Item Detail
Part number 8003126
Description PGC cylinder assembly including pressure/vacuum selector
Pump range Minus 0.95 to 35 bar / minus 13.7 to 510 psi
Medium Air
Side port G1/8 inch female with Bx G1/8 male fitting for 40 bar hoses
Top port Plugged G3/8 inch B female
Body material Aluminum with brass, ABS and NBR parts
Weight of pump 820 g
Hose working pressure 40 bar / 600 psi, bursting 80 bar
Leak test, pressure 28 bar held 1 minute, fall of no more than 0.09 bar
Leak test, vacuum Minus 0.85 bar held 1 minute, fall of no more than 0.02 bar

Why the selector matters

The PGC switches between pressure and vacuum with a small selector that is moved with a screwdriver. The manual warns repeatedly never to change the selector setting while there is pressure or vacuum in the pump or system, and to make sure it is not left halfway between the two positions: if it is, push it fully into the correct position.

A selector that no longer seats, or a cylinder whose bore is scored, is the usual reason to order 8003126. In both cases the symptom is that repeated pumping produces no pressure change although connections and the release valve are in order.

The selector lets one pump generate either pressure or vacuum from air. If the pump cannot hold pressure or vacuum, the leak test figures above quantify the difference between a healthy and a faulty pump and tell you whether a seal-level or assembly-level repair is justified.

How 8003126 differs from other PGC spare parts

The lower assembly with handles and piston is 8003124, and it changes the pump serial number when fitted. 8003126 instead covers the cylinder side, with the selector. The fine adjust and release valve unit 8003122 is a separate item, and so is the plain fitting 7113270. If handles are broken, you do not need 8003126; if the pump cannot switch to vacuum, you probably do.

Because the cylinder, selector and screws are sold as one item, the repair is predictable: you do not hunt for loose parts. The trade-off is that you replace a more complete block than the problem may strictly need, so confirm the diagnosis with the leak test before ordering.

Fitting guidance

Take the pump out of service and depressurize it. Never connect the pump to an external pressure source. Use the seal location diagram in the manual, which names the lock ring, the PVC 1/8 inch gasket, the brass washer and the cylinder mounting screws, to see what sits around the cylinder. Fit the new assembly with its mounting screws and replace any seal that shows wear.

Do not use PTFE tape on any part, because it can contaminate the pump. Tighten hose fittings by hand and use only the maker’s pressure fittings, since impurities from unsuitable materials may clog the pump.

When any part of the PGC is replaced, check the other maintenance items too: the manual recommends replacing hoses every five years, and an immediate change for hoses that show damage, kinks, cracks or bent connectors. Hoses marked Max. 40 bar are the only ones that should be used.

Checks and operation after replacement

Run the manual’s leak tests for both pressure and vacuum. Then do a normal calibration sequence: check the release valve is open, set the selector to pressure, rotate the fine adjust about halfway counterclockwise, close the release valve clockwise, pump near the next point and trim with the fine adjust. For vacuum, set the selector to its vacuum position before closing the release valve.

After a long period without use the first stroke may feel sluggish; this normally disappears.

Field applications in Oman and the GCC

Technicians use PGC pumps for low-range pressure and vacuum work on transmitters, switches and gauges in oil and gas facilities, water treatment plants, power stations and HVAC contractors. Because the pump uses air and weighs 820 g, it travels easily in a service van or site bag, and the maker lists a hard case for it separately.

Keep a log of the pump serial number, the date of the repair and the leak test readings. Calibration laboratories that work to quality systems can then show that the reference pump was verified after maintenance.

Include the pump serial number in the RFQ so that the cylinder assembly is matched to your PGC.

Documents and quotation

The PGC user manual is the datasheet for Beamex 8003126. To request a quotation, send an RFQ to Seven Star LLC with part number 8003126, the quantity and the pump serial number.

Installation and Commissioning Planning

Planning ahead shortens downtime when a Beamex 8003126 goes into service. Confirm the available mounting space, cable entry or process connection, and the tools and spare seals needed on the day. Keep the existing configuration or calibration data so the new unit can be set up to match, and check that loop drawings and the asset register are updated after the swap. If the installation has to fit a short shutdown window, tell us the date when you request a quotation so that delivery of the Beamex equipment can be planned around it.

Other Models in the Pgc Spares Range

Seven Star LLC supplies models from the Pgc Spares range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.

ModelCategory
Beamex 8003124 PGC Piston and Handle AssemblyProcess, Pressure & Temperature CalibratorsView details
Beamex 8003126 PGC Cylinder Assembly with SelectorProcess, Pressure & Temperature CalibratorsYou are viewing this model
Datasheets & Documents
  • PDF Datasheet Beamex 8003126 PGC Cylinder Assembly with Selector Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Beamex is a Finnish maker of calibration equipment and software for process instruments. Four calibration technicians founded the company in 1975 because they could not find the equipment they wanted; their first product was a voltage and current simulator, and the first factory opened in 1977. Beamex is headquartered in Pietarsaari (Jakobstad), Finland, and supplies process industries in more than 100 countries.

Seven Star LLC is an independent supplier listing Beamex documenting calibrators, temperature dry blocks, pressure controllers, calibration pumps and pressure modules for buyers in Oman and the GCC. Equipment is quoted against the model and the pressure or temperature ranges required.

Beamex company profile

Origin Founded in 1975 in Finland by four calibration technicians; first factory 1977
Headquarters Pietarsaari (Jakobstad), Finland; ISO 9001 certified since 1992
Product families MC series multifunction calibrators, temperature dry blocks, automatic and hand pressure generators, external pressure modules, CMX and LOGiCAL calibration software
Typical duty Field and workshop calibration of pressure, temperature and electrical signal instruments, including HART and fieldbus transmitters

Milestones

  • 1975: Four calibration technicians found the company in Finland.
  • 1977: The first factory opens.
  • 1992: ISO 9001 certification.

Beamex calibrators and accessories

Model Type Typical use
MC6 Advanced field calibrator and communicator Pressure, temperature and electrical calibration with a built-in HART, FOUNDATION Fieldbus and PROFIBUS PA communicator, documented results
MC6-Ex Intrinsically safe field calibrator Calibration inside Zone 0 and Zone 1 hazardous areas without a hot work permit
MC6-T Temperature calibrator with integrated dry block Temperature sensors and transmitters, with multifunction electrical and pressure measurement
MC4, MC2 Documenting and hand-held process calibrators Routine loop, transmitter and switch checks in the field
FB, MB Field and metrology temperature dry blocks FB for field calibration; MB for higher-accuracy workshop and laboratory work
ePG, POC8 Electric pressure pump and automatic pressure controller Automated pressure calibration; POC8 for workshop calibration benches
PGC, PGM, PGL, PGPH, PGHH Hand pressure pumps Pressure and vacuum generation in the field, from low pressure (PGL, ±400 mbar) to hydraulic 700 bar (PGHH)
EXT modules External pressure modules Extending MC calibrators to additional ranges, 70 kPa to 100 MPa

Where Beamex equipment is used

Instrument workshops and maintenance teams in oil and gas, refining, petrochemicals, power, water and pharmaceuticals use Beamex calibrators to verify transmitters, switches, gauges and temperature sensors. Documenting calibrators send results to calibration software, which keeps the audit trail required by quality systems and custody transfer metering procedures. Intrinsically safe models are used for in-situ calibration in classified areas.

Ordering Beamex equipment: what we need

State the model and the options you need: internal and external pressure ranges, temperature range, communication protocols (HART, FOUNDATION Fieldbus, PROFIBUS PA), Ex approval, and whether an accredited calibration certificate is required. For spare modules and pumps send the serial number of the calibrator they will be used with, so compatibility can be confirmed.

Facts checked against Beamex company history and product information, October 2026.

FAQ

Questions about the Beamex 8003126 PGC Cylinder Assembly with Selector

Answers from our engineering team on installation, compatibility and integration.

01What is included with Beamex 8003126?
A cylinder assembly with a preassembled pressure/vacuum selector, a nameplate sticker and the cylinder mounting screws.
02When should the selector be changed?
Never while there is pressure or vacuum in the pump or system, and it must not be left between the pressure and vacuum positions.
03Do I also need the fine adjust and release valve?
Only if you are replacing the entire top part; the maker lists 8003122 and fitting 7113270 for that case, or the existing ones can be reused.
04What leak limits apply after repair?
At 28 bar the fall must not exceed 0.09 bar in one minute, and at minus 0.85 bar vacuum the fall must not exceed 0.02 bar.
05Is 8003126 the lower half of the pump?
No. The lower half with handles and piston is 8003124. This part is the cylinder with the selector.