Brand:

Beamex 9010095 PGL Low Pressure Calibration Pump Complete Kit

Description

Beamex 9010095 PGL complete kit with T-hose and case

Beamex 9010095 packages the PGL low pressure calibration pump as a field-ready set. In the hard case you get the pump, a 40 bar (600 psi) pressure T-hose, the user manual, a 4 mm hex wrench, an opening tool for maintenance and the warranty terms. The pump works between -400 and +400 mbar (-160 to +160 iwc) and is adjusted by screws, not by a lever.

This is the order code to choose when a team is buying its first PGL, because the T-hose is part of the deal and the manual states that the published stability figures are valid only with the Beamex 40 bar T-hose.

Kit contents and what they are for

Item Role
PGL pump Coarse adjust handwheel plus fine adjust knob and release valve; G1/8 inch port with Bx G1/8 inch male fitting
40 bar T-hose 1.5 m, listed as code 8009550, joins the reference and the device under test to the pump
Hard case Code 8003311, fitted for the new PGL
User manual Operating steps, specifications, service chapters
Tools 4 mm hex wrench and an opening tool for maintenance
Warranty terms Printed with the delivery

Performance of the 9010095 pump

Parameter Value
Pressure range -400 to 400 mbar (-160 to 160 iwc)
Adjusting sensitivity less than 0.05 mbar (0.02 iwc)
Pressure change after 1 min and readjustment less than 0.3 mbar per minute
Test volume for those figures up to 20 ml
Safety pressure 4 bar (below 60 psi)
Weight and size about 1.8 kg, 283 x 110 x 114 mm
Environment 0 to 50 degrees C operating, below 95 %RH

Temperature effects are minimized by the choice of materials and design, according to the manual, so the pump keeps its pressure while the ambient conditions drift.

First-use checklist for 9010095

  1. Inspect the pump and case for transport damage and contact the purchasing sales office if contents are missing.
  2. Connect the T-hose to the Bx G1/8 inch fitting and run branches to the calibrator pressure module and to the instrument.
  3. Open the release valve and wind the coarse handwheel counterclockwise to the end.
  4. Close the valve with the fine adjust at its mid position and remove the connection offset to read zero.
  5. Raise pressure with the handwheel, finish with the fine adjust and give the system about one minute to settle.
  6. After the last point, open the release valve by turning the fine adjust counterclockwise until the threads show, and turn it one to three more rounds to vent fully.

If the pump does not seem to hold, check that all hose connections are tight, that the fitting types match, and that unused connectors are plugged. A failing coarse adjust usually points to broken piston seals, and difficulty turning the fine adjust to loose connections or wrong fittings, according to the manual troubleshooting tables.

How 9010095 differs from the pump-only 9010090

Both orders deliver the same pump. The 9010095 adds a 40 bar T-hose and a hard case, so it removes the need for separate hose and case orders, and the hose of this kit is the one the specification figures were measured with. The 9010090 is for replacement or already-equipped sites. Neither variant includes a pressure indicator, so a calibrator or another reference module is needed on site.

Compatibility and caution for 9010095

Cases and spare-part sets referring to PGL silver do not fit the improved PGL launched in February 2023, so choose accessories with the new numbers: case 8003311 comes with this kit, and service parts 8003710, 8003715, 8003720, 8003725 and 8003730 are the matching items. Use only the Beamex 40 bar hose family, vent external systems before connecting, and use eye protection and an overpressure relief device where the application could exceed the safety limit.

Reading the 9010095 specification footnotes

Three footnotes in the PGL specification table deserve attention when you plan work with the 9010095. First, the range of -400 to 400 mbar depends on the volume of the test setup; with a small volume of up to 20 ml, ranges of -0.7 to 3.0 bar become possible. Second, the pressure change figure of less than 0.3 mbar per minute is measured into a volume of at most 20 ml after one minute of waiting and readjustment. Third, the stability statement assumes stable ambient conditions. In practice this means the kit performs best on small instruments such as transmitters with short impulse lines, and you should plan extra settling time for long hoses, large vessels or setups standing in sun.

Applications for 9010095 in Gulf industry

Calibration crews from oil terminals, gas processing, power plants and water utilities carry the kit to differential pressure and low range transmitters, furnace draft gauges and HVAC instruments. The hard case protects the pump during transport, and the screw adjustment makes it easy to hold a steady point for a minute while the reading settles. Because the PGL generates both positive and negative pressure from the same body, it also serves vacuum switches and compound gauges that start below atmosphere.

Handling the 9010095 safely

The manual lists warnings that apply to every use of the 9010095: read the user manual before setting up, never connect the pump to an external pressure source, wear eye protection, and consider a relief valve for overpressure protection that is checked from time to time. Vent external systems before connecting, check hoses and connectors for damage and impurities, and depressurize the pump whenever it is left alone. Avoid sudden pressure on low-range gauges with small chambers, and remember that ambient pressure and temperature may restrict the allowed maximum pressure to a lower level than the pump and hose allow.

Documents and quotation

The PGL user manual is the source document for this page. Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer. Send an RFQ listing Beamex 9010095 and the quantity; mention 9010090 if you would like a pump-only alternative quoted together.

Other Models in the Beamex Pgl Range

Seven Star LLC supplies models from the Beamex Pgl range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.

ModelCategory
Beamex 8003715 PGL Coarse Adjust Handwheel Spare PartProcess, Pressure & Temperature CalibratorsView details
Beamex 8003730 PGL Support Leg Kit Spare PartProcess, Pressure & Temperature CalibratorsView details
Beamex 9010095 PGL Low Pressure Calibration Pump Complete KitProcess, Pressure & Temperature CalibratorsYou are viewing this model
Datasheets & Documents
  • PDF Datasheet Beamex 9010095 PGL Low Pressure Calibration Pump Complete Kit Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Beamex is a Finnish maker of calibration equipment and software for process instruments. Four calibration technicians founded the company in 1975 because they could not find the equipment they wanted; their first product was a voltage and current simulator, and the first factory opened in 1977. Beamex is headquartered in Pietarsaari (Jakobstad), Finland, and supplies process industries in more than 100 countries.

Seven Star LLC is an independent supplier listing Beamex documenting calibrators, temperature dry blocks, pressure controllers, calibration pumps and pressure modules for buyers in Oman and the GCC. Equipment is quoted against the model and the pressure or temperature ranges required.

Beamex company profile

Origin Founded in 1975 in Finland by four calibration technicians; first factory 1977
Headquarters Pietarsaari (Jakobstad), Finland; ISO 9001 certified since 1992
Product families MC series multifunction calibrators, temperature dry blocks, automatic and hand pressure generators, external pressure modules, CMX and LOGiCAL calibration software
Typical duty Field and workshop calibration of pressure, temperature and electrical signal instruments, including HART and fieldbus transmitters

Milestones

  • 1975: Four calibration technicians found the company in Finland.
  • 1977: The first factory opens.
  • 1992: ISO 9001 certification.

Beamex calibrators and accessories

Model Type Typical use
MC6 Advanced field calibrator and communicator Pressure, temperature and electrical calibration with a built-in HART, FOUNDATION Fieldbus and PROFIBUS PA communicator, documented results
MC6-Ex Intrinsically safe field calibrator Calibration inside Zone 0 and Zone 1 hazardous areas without a hot work permit
MC6-T Temperature calibrator with integrated dry block Temperature sensors and transmitters, with multifunction electrical and pressure measurement
MC4, MC2 Documenting and hand-held process calibrators Routine loop, transmitter and switch checks in the field
FB, MB Field and metrology temperature dry blocks FB for field calibration; MB for higher-accuracy workshop and laboratory work
ePG, POC8 Electric pressure pump and automatic pressure controller Automated pressure calibration; POC8 for workshop calibration benches
PGC, PGM, PGL, PGPH, PGHH Hand pressure pumps Pressure and vacuum generation in the field, from low pressure (PGL, ±400 mbar) to hydraulic 700 bar (PGHH)
EXT modules External pressure modules Extending MC calibrators to additional ranges, 70 kPa to 100 MPa

Where Beamex equipment is used

Instrument workshops and maintenance teams in oil and gas, refining, petrochemicals, power, water and pharmaceuticals use Beamex calibrators to verify transmitters, switches, gauges and temperature sensors. Documenting calibrators send results to calibration software, which keeps the audit trail required by quality systems and custody transfer metering procedures. Intrinsically safe models are used for in-situ calibration in classified areas.

Ordering Beamex equipment: what we need

State the model and the options you need: internal and external pressure ranges, temperature range, communication protocols (HART, FOUNDATION Fieldbus, PROFIBUS PA), Ex approval, and whether an accredited calibration certificate is required. For spare modules and pumps send the serial number of the calibrator they will be used with, so compatibility can be confirmed.

Facts checked against Beamex company history and product information, October 2026.

FAQ

Questions about the Beamex 9010095 PGL Low Pressure Calibration Pump

Answers from our engineering team on installation, compatibility and integration.

01What is included in Beamex 9010095?
The kit has the PGL pump, a 40 bar pressure T-hose, a hard case, the user manual, a 4 mm hex wrench, an opening tool and warranty terms.
02Why is the T-hose important for 9010095?
The manual states the published specifications are valid only with the Beamex 40 bar pressure T-hose, which is why it is included in the kit.
03What range does the 9010095 pump cover?
It covers -400 to 400 mbar, with extended working range -0.7 to 3.0 bar for small connected volumes of up to 20 ml.
04Is the case compatible with older PGL silver cases?
No. The new PGL launched in February 2023 is not compatible with cases or spare-part sets referring to PGL silver.
05Does the 9010095 include a pressure reference?
No. A calibrator or pressure module must be connected as the reference whenever the pump is used.