Brand:

VEGA VEGAPULS 65 Rod Antenna Radar Level Sensor

Description

Slim rod antenna radar: VEGA VEGAPULS 65

The VEGA VEGAPULS 65 answers a very specific installation problem. Many vessels have only a small opening on top, and a horn antenna simply does not fit. The VEGAPULS 65 uses a slim rod antenna, so it can be lowered into small vessel openings, and it is described by VEGA as a radar sensor for continuous measurement of liquids under simple process conditions. The sensor is intended for the unspectacular but widespread jobs: tanks of water, solvents, lubricants and chemicals where the temperature stays moderate and the pressure low.

The numbers: measuring range up to 35 m (114.8 ft), deviation 8 mm, process pressure -1 to +16 bar, process temperature -40 to +150 °C, and a supply of 9.6 to 36 V DC. The wetted parts are PVDF, 316L or PTFE, sealed with FKM.

VEGAPULS 65 technical specifications

Values come from the 4 to 20 mA/HART two-wire specification sheet.

Parameter Value (datasheet)
Model VEGAPULS 65, radar sensor with slim rod antenna for liquids
Measuring range up to 35 m (114.8 ft)
Deviation 8 mm
Process fitting Thread from G1 1/2; 1 1/2 NPT; flanges from DN 40, 1 1/2 in
Process pressure -1 to +16 bar (-100 to +1600 kPa / -14.5 to +232 psig)
Process temperature -40 to +150 C (-40 to +302 F)
Ambient, storage, transport temperature -40 to +80 C
Operating voltage 9.6 to 36 V DC
SIL qualification Optionally up to SIL2
Wetted materials PVDF, 316L or PTFE
Process seal FKM
Housing Single or double chamber; plastic, stainless steel or aluminium; up to IP 68 (1 bar)
Electronics HART two-wire and four-wire, Profibus PA, Foundation Fieldbus, Modbus, HART with accumulator
Datasheet reference 34508-EN-160119

Why the rod antenna matters

A rod antenna is long and slender, so the sensor can pass through a G1½ or 1½ NPT socket, or a flange from DN 40 (1½ in), without a large horn diameter below the nozzle. Because the antenna is made of PTFE or similar plastic in many versions, the face resists chemical attack and does not collect condensate as easily as a flat plate. The compromise is a looser deviation (8 mm) than the 2 mm of the VEGAPULS 61 and 62, so use the VEGAPULS 65 where an accuracy of several millimetres is acceptable for inventory and control.

Where VEGAPULS 65 is applied in the Gulf region

Typical locations include chemical injection and storage tanks at oil and gas sites, process-water and demineralised-water tanks in power stations, wastewater lift stations, fertilizer and petrochemical solvent tanks, and small IBC or drum stations where the manway is narrow. In manufacturing, the sensor suits washing-liquid tanks, coolant sumps and additive tanks. Because it does not drift with density or conductivity, the same unit can serve alternating products.

VEGAPULS 65 compared with other VEGAPULS models

  • VEGAPULS 61: also plastic, but 3 bar, +80 °C and a 2 mm deviation.
  • VEGAPULS 62: 160 bar, +450 °C, metals and 2 mm deviation, for tough duty.
  • VEGAPULS 64: G¾ thread, 25 bar, -196 to +200 °C, 1 mm deviation, 30 m range.
  • VEGAPULS 66: 160 bar, +400 °C, 8 mm deviation, hygienic and bolted fittings.

This puts the 65 squarely between the plastic-only 61 and the all-rounder 62: more pressure and temperature than the first, far less than the second, but with the slim antenna that neither offers.

Commissioning the VEGAPULS 65

Check that the rod does not touch the nozzle wall or any internal ladder, and keep the antenna end clear of baffles. Terminate the screen at the ground terminal, close the gland, and adjust via PLICSCOM, PACTware with DTM or a HART communicator. Electronics options are the same family as other 6x radars: two-wire and four-wire HART, Profibus PA, Foundation Fieldbus and Modbus, plus a HART variant with an accumulator. With double chamber housings, a radio module or power pack can be added. Optional SIL2 qualification is stated on the data sheet, and ATEX and IEC approvals as well as ship approvals GL, LRS and ABS are listed generally.

Maintenance and quotation notes for VEGAPULS 65

Because the VEGAPULS 65 is non-contact, maintenance is limited to occasional visual checks of the rod for deposits in sticky products. When you enquire, name the process connection, wetted material, housing and electronics so the correct order code can be matched. If the existing instrument is a float, displacer or guided wave device that you want to replace with this rod-antenna radar, tell us the nozzle diameter and height as well, because the slim antenna is chosen precisely for tight openings and the free length below the nozzle matters for a clean echo. Also state whether the vessel is open to atmosphere or vented through a long pipe, since heavy vapour or foam can lower the echo level.

VEGAPULS 65 FAQ summary

  • Why choose the VEGAPULS 65 over a horn radar? Its slim rod antenna fits small vessel openings, which VEGA names as the particular strength of this model.
  • What accuracy does the VEGAPULS 65 give? The data sheet states a deviation of 8 mm, looser than the 2 mm of the VEGAPULS 61 and 62.
  • What are the pressure and temperature limits of the VEGAPULS 65? Process pressure is -1 to +16 bar and process temperature is -40 to +150 C.
  • Which materials are wetted on the VEGAPULS 65? PVDF, 316L or PTFE, with an FKM process seal.
  • Which process fittings fit the VEGAPULS 65? Threads from G1 1/2 and 1 1/2 NPT, and flanges from DN 40 (1 1/2 in).

Documents and quotation

The manufacturer VEGA VEGAPULS 65 datasheet is available as a PDF download on this page, so your engineers can check ranges, materials and dimensions against the data sheet before ordering. To request a quotation send the part number VEGAPULS 65 together with the quantity, the required process connection, electronics version and any approval requirement through the RFQ form or by email. Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer; product names and marks belong to their owners.

Other Models in the Vegapuls 6x Range

Seven Star LLC supplies models from the Vegapuls 6x range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.

ModelCategory
VEGA VEGAPULS 62 Radar Level Sensor 160 bar 450 CRadar & Guided Wave Radar LevelView details
VEGA VEGAPULS 61 Radar Level Sensor 35 m PVDFRadar & Guided Wave Radar LevelView details
VEGA VEGAPULS 66 Radar Level Sensor 160 bar 400 CRadar & Guided Wave Radar LevelView details
VEGA VEGAPULS 65 Rod Antenna Radar Level SensorRadar & Guided Wave Radar LevelYou are viewing this model
Datasheets & Documents
  • PDF Datasheet VEGA VEGAPULS 65 Rod Antenna Radar Level Sensor Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

VEGA is a German maker of level and pressure instruments. Bruno Grieshaber founded the company in 1959 in Triberg in the Black Forest, and it has been based in Schiltach since 1975, still family owned as VEGA Grieshaber KG. VEGA is best known for radar level sensors: it introduced the first two-wire radar level sensor in 1997 and the first 80 GHz radar sensor for bulk solids, the VEGAPULS 69, in 2015.

Seven Star LLC is an independent supplier listing VEGA radar and guided wave radar transmitters, point level switches, pressure transmitters and signal conditioning instruments for buyers in Oman and the GCC. Sensors are matched from the product code on the type label.

VEGA at a glance

Origin Founded by Bruno Grieshaber in 1959 in Triberg, Germany
Headquarters Schiltach, Black Forest, Germany (since 1975)
Ownership Family-owned company: VEGA Grieshaber KG
Milestones First two-wire radar level sensor (1997), first 80 GHz radar for bulk solids, VEGAPULS 69 (2015), VEGAPULS 6X single radar platform for liquids and solids
Product families VEGAPULS radar, VEGAFLEX guided wave radar, VEGASWING and VEGAPOINT switches, VEGACAL capacitive probes, VEGABAR and VEGADIF pressure, VEGAMET and VEGADIS signal conditioning

VEGA sensor and instrument series

Series Type Typical use
VEGAPULS 6X, VEGAPULS 69, VEGAPULS 64 Process radar level transmitters (80 GHz) Storage and process tanks, reactors with agitators, silos and bulk solids
VEGAPULS 11, 21, 31, 42 Compact radar sensors Water and wastewater, simple storage tanks; VEGAPULS 42 with IO-Link for machine integration
VEGAPULS C 21, C 11, C 22, C 23 Radar sensors with fixed cable, IP68 Wet wells, pumping stations, open channels and river levels
VEGAFLEX 81, 82, 83, 86 Guided wave radar (TDR) Level and interface in liquids, bulk solids, hygienic and high-temperature or high-pressure service (VEGAFLEX 86)
VEGASWING 61, 63, VEGAPOINT 21 Vibrating and capacitive point level switches Overfill protection, dry-run protection and high or low level alarms
VEGACAL 62 to VEGACAL 67 Capacitive rod and cable probes Point and continuous level in solids and conductive or adhesive liquids
VEGABAR 82, 83, 28, 29, 38 Pressure transmitters, ceramic or metallic cell Process pressure and hydrostatic level; compact 18/19 and 28/29/38 versions for utilities
VEGADIF 85 Differential pressure transmitter DP flow, filter monitoring and level in pressurised vessels
VEGAMET 841, VEGADIS 176 Controller and loop display Level display, pump control and remote indication of 4-20 mA signals

Where VEGA instruments are used

VEGA radar sensors measure level in crude oil and product tanks, separators, chemical reactors, cement and aggregate silos, grain stores and water reservoirs. Compact cable radars are widely used in wastewater pumping stations and flood monitoring. Because 80 GHz radar focuses a narrow beam, it can measure past internal structures in a tank, and it is not affected by the temperature, pressure or vapour above the product, which suits hot Gulf storage tanks.

Ordering VEGA instruments: what we need

The product code on the type label (for example PS64.XXXXXXX) fixes the measuring range, antenna or probe, process connection, electronics and approvals. Send it with the serial number. For a new measurement add the medium and its dielectric constant if known, tank or silo height and internals, process pressure and temperature, process connection, output (4-20 mA HART, PROFIBUS PA, Modbus or IO-Link), and any hazardous-area or overfill protection approval required.

Facts checked against VEGA company and product information, October 2026.

FAQ

Questions about the VEGA VEGAPULS 65 Rod Antenna Radar Level Sensor

Answers from our engineering team on installation, compatibility and integration.

01Why choose the VEGAPULS 65 over a horn radar?
Its slim rod antenna fits small vessel openings, which VEGA names as the particular strength of this model.
02What accuracy does the VEGAPULS 65 give?
The data sheet states a deviation of 8 mm, looser than the 2 mm of the VEGAPULS 61 and 62.
03What are the pressure and temperature limits of the VEGAPULS 65?
Process pressure is -1 to +16 bar and process temperature is -40 to +150 C.
04Which materials are wetted on the VEGAPULS 65?
PVDF, 316L or PTFE, with an FKM process seal.
05Which process fittings fit the VEGAPULS 65?
Threads from G1 1/2 and 1 1/2 NPT, and flanges from DN 40 (1 1/2 in).