Hanna HI97725 Chlorine, Cyanuric Acid and pH Photometer

Description

Hanna HI97725 Chlorine, cyanuric acid and pH portable photometer

The Hanna HI97725 is a pocket-sized, single-parameter photometer built for pool, spa and recreational water where chlorine, stabiliser and pH are all controlled. One meter covers free chlorine, total chlorine, cyanuric acid and pH, so the typical swimming-pool test routine needs a single instrument and one cuvette type. Seven Star LLC is an independent supplier in Muscat and is not affiliated with the manufacturer; we quote the HI97725 to industrial and laboratory buyers across Oman and the GCC.

The model number follows Hanna’s HI977xx pattern for dedicated photometers: the last digits identify the chemistry, so HI97725 always means chlorine, cyanuric acid and ph and cannot be reconfigured for another analyte. Reagents carry their own HI93xxx codes, listed below, and must be ordered separately from the meter.

HI97725 technical specifications

Parameter Specification
Manufacturer / part number Hanna Instruments HI97725
Parameter Chlorine, cyanuric acid and pH
Range Chlorine 0.00 to 5.00 mg/L; cyanuric acid 0 to 80 mg/L; pH 6.5 to 8.5
Resolution 0.01 mg/L Cl2; 1 mg/L CYA; 0.1 pH
Accuracy at 25 °C Cl2 ±0.03 mg/L ±3 %; CYA ±1 mg/L ±15 %; pH ±0.1
Method DPD (EPA 330.5), turbidimetric (CYA), phenol red (pH)
Light source LED with narrow-band interference filter at 525 nm
Filter bandwidth / wavelength accuracy 8 nm / ±1.0 nm
Detector Silicon photocell
Cuvette Round, 24.6 mm diameter (22 mm inside)
Data logging 50 readings (auto logging)
Display 128 × 64 pixel monochrome LCD with backlight
Power Three 1.5 V AA alkaline batteries; more than 800 measurements without backlight
Operating environment 0 to 50 °C; 0 to 100 % RH
Ingress protection IP67, floating case
Dimensions / weight 142.5 × 102.5 × 50.5 mm / 380 g with batteries
Warranty Two years

How the HI97725 measurement works

Chlorine uses the DPD pink colour read at 525 nm; cyanuric acid is read by turbidity formed when the reagent reacts with the stabiliser; pH uses phenol red, whose colour shifts with acidity. The meter handles all three with the same LED optics and a method menu.

Inside the case a reference detector tracks the LED output, so battery sag or a warm ambient does not move the blank. A lens gathers the light leaving the cuvette, which is why scratches and small glass imperfections matter less than with a simple colorimeter, and the cap notch locks the cuvette in the same orientation every time. A built-in tutorial mode walks the operator through sampling, reagent quantity and timers on screen, so the instrument can be handed to a technician who is not a chemist.

Reagents, procedure and interferences for HI97725

Select the method from the menu: free chlorine (powder or liquid), total chlorine (powder or liquid), cyanuric acid or pH. Each follows its own fill, zero, reagent and countdown steps; chlorine countdowns are 1 minute (free) and 2 minutes 30 seconds (total).

Code Description Quantity per test
HI93701-0 / HI93701A-F, B-F Free chlorine powder or liquid 1 packet / 3 + 3 drops
HI93711-0 / HI93701A-T, B-T, C-T Total chlorine powder or liquid 1 packet / 3 + 3 + 1 drops
HI93722-0 Cyanuric acid reagent 1 packet
HI93710-0 pH reagent 5 drops

Reagent sets are supplied as HI93701-01/-03, HI93701-F, HI93701-T, HI93711-01/-03, HI93722-01/-03 and HI93710-01/-03. Follow the zeroing order given in the manual for this method, and read within the time the manual allows after the countdown ends.

Known interferences listed for this method:

  • Bromine, iodine, oxidised chromium and manganese, and ozone affect the chlorine methods
  • Hardness above 500 mg/L CaCO3; shake the sample for about 2 minutes after adding powder
  • Alkalinity above 250 mg/L CaCO3 or acidity above 150 mg/L CaCO3 can cause partial or fading colour; neutralise with dilute HCl or NaOH

Where the HI97725 sits among Hanna photometers

Choose the HI97725 when the job is water-chemistry balance for pools and spas. For plain disinfection monitoring the HI97701 (free chlorine) or HI97734 (high range, free and total) are simpler, while the HI97762 is for traces below 0.5 mg/L.

Model Parameter Range LED filter
HI97700 Ammonia low range 0.00 to 3.00 mg/L as NH3-N 420 nm
HI97715 Ammonia medium range 0.00 to 10.00 mg/L as NH3-N 420 nm
HI97733 Ammonia high range 0.0 to 100.0 mg/L as NH4+ 420 nm
HI97701 Free chlorine 0.00 to 5.00 mg/L as Cl2 525 nm
HI97762 Free chlorine ultra low range 0.000 to 0.500 mg/L as Cl2 525 nm
HI97734 Free and total chlorine, high range 0.00 to 10.00 mg/L as Cl2 525 nm
HI97725 Chlorine, cyanuric acid and pH 0.00 to 5.00 mg/L Cl2; 0 to 80 mg/L CYA; 6.5 to 8.5 pH 525 nm
HI97706 Phosphorus 0.0 to 15.0 mg/L as P 525 nm
HI97714 Cyanide 0.000 to 0.200 mg/L as CN– 610 nm
HI97718 Iodine 0.0 to 12.5 mg/L as I2 525 nm
HI97731 Zinc 0.00 to 3.00 mg/L as Zn 575 nm
HI97737 Silver 0.000 to 1.000 mg/L as Ag 575 nm

Selection and ordering checklist for HI97725

  • Confirm the four parameters required: free chlorine, total chlorine, cyanuric acid and pH
  • Order the reagent sets for each method; none are included
  • Check CAL Check kits for chlorine (HI97701-11), cyanuric acid (HI97722-11) and pH (HI97710-11)
  • Note that pH is limited to 6.5 to 8.5, so highly alkaline or acidic samples need a pH meter
  • Decide whether powder or liquid chlorine reagents suit the site

Validation, handling and maintenance of HI97725

Run validation per parameter with the matching CAL Check kit. Rinse cuvettes between methods because cyanuric acid residue is turbid and phenol red stains; the meter stores GLP data and 50 readings.

Keep cuvettes clean with the cleaning solution and a lint-free cloth, never touch the lower half of the glass, and store the meter in its carrying case. Replace the three AA cells when the battery icon appears; the meter keeps its logged readings and settings. For long storage between field trips remove the batteries and close the reagent bottles tightly.

Typical applications of HI97725 in Oman and the GCC

Resort, hotel and club pools across Oman and the GCC need stabiliser control because strong sun burns off chlorine, and cyanuric acid readings keep dosing efficient. The meter also serves water-park, spa and gymnasium operators and facility-management contractors who service multiple sites.

Heat is a practical factor in the Gulf. Reaction times in the manual are given at 25 °C, so samples that sit in a vehicle at much higher temperature should be allowed to equilibrate, and the meter itself is rated for 0 to 50 °C. The sealed optics and IP67 floating case also cope with dust and spray on plant walkdowns.

Information to send with an RFQ for HI97725

  • Quantity of meters and whether reagent sets (the chlorine, HI93722 cyanuric acid and HI93710 pH sets; CAL Check HI97701-11, HI97722-11 and HI97710-11) and CAL Check cuvettes are also required
  • Sample matrix and expected concentration, so we can confirm the range suits the application
  • Delivery location in Oman or the GCC and required lead time
  • Any need for spare cuvettes, carrying case or cleaning solution
  • Project or tender reference, if applicable

Documents and quotation

The instruction manual for the HI97725 is attached on this page and covers specifications, the reagent procedure, CAL Check validation and error messages.

Send the part number HI97725, the quantity and your delivery location and we will reply with availability and a quotation.

Other Models in the Hanna Hi977xx Photometer Range

Seven Star LLC supplies models from the Hanna Hi977xx Photometer range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.

ModelCategory
Hanna HI97700 Ammonia Low Range Photometer 0-3 mg/LLiquid Analytical SystemsView details
Hanna HI97715 Ammonia Medium Range Photometer 0-10 mg/LLiquid Analytical SystemsView details
Hanna HI97733 Ammonia High Range Photometer 0-100 mg/LLiquid Analytical SystemsView details
Hanna HI97701 Free Chlorine Portable Photometer 0-5 mg/LLiquid Analytical SystemsView details
Hanna HI97762 Ultra Low Range Free Chlorine PhotometerLiquid Analytical SystemsView details
Hanna HI97734 Free and Total Chlorine HR PhotometerLiquid Analytical SystemsView details
Hanna HI97725 Chlorine, Cyanuric Acid and pH PhotometerLiquid Analytical SystemsYou are viewing this model
Hanna HI97706 Phosphorus Portable Photometer 0-15 mg/LLiquid Analytical SystemsView details
Hanna HI97714 Cyanide Portable Photometer 0-0.200 mg/LLiquid Analytical SystemsView details
Hanna HI97718 Iodine Portable Photometer 0-12.5 mg/LLiquid Analytical SystemsView details
Hanna HI97731 Zinc Portable Photometer 0-3.00 mg/LLiquid Analytical SystemsView details
Hanna HI97737 Silver Portable Photometer 0-1.000 mg/LLiquid Analytical SystemsView details
Datasheets & Documents
  • PDF Manual Hanna HI97725 Instruction Manual Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Hanna Instruments is a maker of analytical instruments for water and liquid testing, founded in 1978 in the Padua area of Italy by Oscar and Anna Nardo. Its early breakthrough was the pHep, an affordable pocket-sized pH tester. The company is now led by the founders' children, Martino and Pamela Nardo, with its headquarters in Woonsocket, Rhode Island, and offers more than 3,000 products.

Seven Star LLC is an independent supplier in Muscat. We list Hanna instruments for water, wastewater and laboratory users in Oman and the GCC.

Hanna Instruments company profile

Founded 1978, near Padua, Italy, by Oscar and Anna Nardo
Headquarters Woonsocket, Rhode Island, USA
Leadership Martino Nardo and Pamela Nardo
Early product The pHep pocket pH tester
Product families pHep and HI98 waterproof testers, edge benchtop meters, HI83 multiparameter photometers, Checker HC handheld colorimeters, HI931 and HI932 titrators, process controllers, reagents and calibration buffers

Hanna product series

Portable meters include the HI98190 pH and ORP meter, the HI98192 EC, TDS and salinity meter, the HI98193 dissolved oxygen and BOD meter, the HI98199 pH, EC and DO meter and the HI9829 multiparameter water quality meter.

The HI98703 is an EPA compliant portable turbidity meter, the HI97711 a free and total chlorine photometer and the HI83414 a benchtop turbidity and chlorine meter.

Waterproof portable meters such as the HI98190 and HI98192 suit field use, with automatic temperature compensation and GLP data for audits. Turbidity measurement with the HI98703 follows EPA method 180.1, and the HI97711 photometer measures free and total chlorine for disinfection control.

Where Hanna instruments are used

Drinking water and desalination plants, wastewater treatment, environmental field surveys, aquaculture, cooling water and quality control labs.

Hanna Instruments products we list

Family Examples
Portable Test & Measurement Hanna HI97711 Free and Total Chlorine Po…, Hanna HI98190 Waterproof Portable pH and…, Hanna HI98192 Waterproof Portable EC, Hanna HI98193 Waterproof Portable Dissol…, Hanna HI98199 Portable pH, Hanna HI9829 Multiparameter Water Qualit…, Hanna HI98703 Portable Turbidity Meter

Ordering Hanna instruments: what we need

Send the model, the parameters and ranges, the probe and cable length and any reagents or calibration solutions required.

Facts checked against Hanna Instruments' own company story and City of Woonsocket business profile, October 2026.

FAQ

Questions about the Hanna HI97725 Chlorine, Cyanuric Acid and pH

Answers from our engineering team on installation, compatibility and integration.

01Which parameters can the HI97725 measure?
Free chlorine, total chlorine (0.00 to 5.00 mg/L), cyanuric acid (0 to 80 mg/L) and pH (6.5 to 8.5), selected from a method menu.
02What methods does it use for each?
DPD adapted from EPA 330.5 for chlorine, a turbidimetric method for cyanuric acid and the phenol red method for pH.
03Are reagents for all four tests included?
No. Reagent sets HI93701, HI93711, HI93722 and HI93710 are separate, in 100 and 300 test sizes depending on the method.
04How accurate is the cyanuric acid reading?
Accuracy is 1 mg/L plus 15 percent of reading at 25 degrees C with 1 mg/L resolution, reflecting the turbidimetric nature of the test.
05Can it replace a pH meter?
It gives pH within 6.5 to 8.5 with 0.1 pH resolution, suitable for pool control, but a probe-based pH meter is better for wider ranges.