- Automation & ControlView all 2,895 products
- Automation Accessories
- AC Drives
- PLC & I/O Cards
- Industrial Communications
- Vibration Monitoring
- Controllers & Data Acquisition
- Isolator & Barrier
- Safety Switches & Guard Locking
- Motor Soft Starters
- CCTV & Security
- Encoders
- Pneumatic Cylinders
- HMI Displays
- Servo Drives
- Industrial Computers
- Machine Vision
- Automation Software
- Robotics
Brand: Hanna Instruments
Hanna HI97725 Chlorine, Cyanuric Acid and pH Photometer
Description
Hanna HI97725 Chlorine, cyanuric acid and pH portable photometer
The Hanna HI97725 is a pocket-sized, single-parameter photometer built for pool, spa and recreational water where chlorine, stabiliser and pH are all controlled. One meter covers free chlorine, total chlorine, cyanuric acid and pH, so the typical swimming-pool test routine needs a single instrument and one cuvette type. Seven Star LLC is an independent supplier in Muscat and is not affiliated with the manufacturer; we quote the HI97725 to industrial and laboratory buyers across Oman and the GCC.
The model number follows Hanna’s HI977xx pattern for dedicated photometers: the last digits identify the chemistry, so HI97725 always means chlorine, cyanuric acid and ph and cannot be reconfigured for another analyte. Reagents carry their own HI93xxx codes, listed below, and must be ordered separately from the meter.
HI97725 technical specifications
| Parameter | Specification |
|---|---|
| Manufacturer / part number | Hanna Instruments HI97725 |
| Parameter | Chlorine, cyanuric acid and pH |
| Range | Chlorine 0.00 to 5.00 mg/L; cyanuric acid 0 to 80 mg/L; pH 6.5 to 8.5 |
| Resolution | 0.01 mg/L Cl2; 1 mg/L CYA; 0.1 pH |
| Accuracy at 25 °C | Cl2 ±0.03 mg/L ±3 %; CYA ±1 mg/L ±15 %; pH ±0.1 |
| Method | DPD (EPA 330.5), turbidimetric (CYA), phenol red (pH) |
| Light source | LED with narrow-band interference filter at 525 nm |
| Filter bandwidth / wavelength accuracy | 8 nm / ±1.0 nm |
| Detector | Silicon photocell |
| Cuvette | Round, 24.6 mm diameter (22 mm inside) |
| Data logging | 50 readings (auto logging) |
| Display | 128 × 64 pixel monochrome LCD with backlight |
| Power | Three 1.5 V AA alkaline batteries; more than 800 measurements without backlight |
| Operating environment | 0 to 50 °C; 0 to 100 % RH |
| Ingress protection | IP67, floating case |
| Dimensions / weight | 142.5 × 102.5 × 50.5 mm / 380 g with batteries |
| Warranty | Two years |
How the HI97725 measurement works
Chlorine uses the DPD pink colour read at 525 nm; cyanuric acid is read by turbidity formed when the reagent reacts with the stabiliser; pH uses phenol red, whose colour shifts with acidity. The meter handles all three with the same LED optics and a method menu.
Inside the case a reference detector tracks the LED output, so battery sag or a warm ambient does not move the blank. A lens gathers the light leaving the cuvette, which is why scratches and small glass imperfections matter less than with a simple colorimeter, and the cap notch locks the cuvette in the same orientation every time. A built-in tutorial mode walks the operator through sampling, reagent quantity and timers on screen, so the instrument can be handed to a technician who is not a chemist.
Reagents, procedure and interferences for HI97725
Select the method from the menu: free chlorine (powder or liquid), total chlorine (powder or liquid), cyanuric acid or pH. Each follows its own fill, zero, reagent and countdown steps; chlorine countdowns are 1 minute (free) and 2 minutes 30 seconds (total).
| Code | Description | Quantity per test |
|---|---|---|
| HI93701-0 / HI93701A-F, B-F | Free chlorine powder or liquid | 1 packet / 3 + 3 drops |
| HI93711-0 / HI93701A-T, B-T, C-T | Total chlorine powder or liquid | 1 packet / 3 + 3 + 1 drops |
| HI93722-0 | Cyanuric acid reagent | 1 packet |
| HI93710-0 | pH reagent | 5 drops |
Reagent sets are supplied as HI93701-01/-03, HI93701-F, HI93701-T, HI93711-01/-03, HI93722-01/-03 and HI93710-01/-03. Follow the zeroing order given in the manual for this method, and read within the time the manual allows after the countdown ends.
Known interferences listed for this method:
- Bromine, iodine, oxidised chromium and manganese, and ozone affect the chlorine methods
- Hardness above 500 mg/L CaCO3; shake the sample for about 2 minutes after adding powder
- Alkalinity above 250 mg/L CaCO3 or acidity above 150 mg/L CaCO3 can cause partial or fading colour; neutralise with dilute HCl or NaOH
Where the HI97725 sits among Hanna photometers
Choose the HI97725 when the job is water-chemistry balance for pools and spas. For plain disinfection monitoring the HI97701 (free chlorine) or HI97734 (high range, free and total) are simpler, while the HI97762 is for traces below 0.5 mg/L.
| Model | Parameter | Range | LED filter |
|---|---|---|---|
| HI97700 | Ammonia low range | 0.00 to 3.00 mg/L as NH3-N | 420 nm |
| HI97715 | Ammonia medium range | 0.00 to 10.00 mg/L as NH3-N | 420 nm |
| HI97733 | Ammonia high range | 0.0 to 100.0 mg/L as NH4+ | 420 nm |
| HI97701 | Free chlorine | 0.00 to 5.00 mg/L as Cl2 | 525 nm |
| HI97762 | Free chlorine ultra low range | 0.000 to 0.500 mg/L as Cl2 | 525 nm |
| HI97734 | Free and total chlorine, high range | 0.00 to 10.00 mg/L as Cl2 | 525 nm |
| HI97725 | Chlorine, cyanuric acid and pH | 0.00 to 5.00 mg/L Cl2; 0 to 80 mg/L CYA; 6.5 to 8.5 pH | 525 nm |
| HI97706 | Phosphorus | 0.0 to 15.0 mg/L as P | 525 nm |
| HI97714 | Cyanide | 0.000 to 0.200 mg/L as CN– | 610 nm |
| HI97718 | Iodine | 0.0 to 12.5 mg/L as I2 | 525 nm |
| HI97731 | Zinc | 0.00 to 3.00 mg/L as Zn | 575 nm |
| HI97737 | Silver | 0.000 to 1.000 mg/L as Ag | 575 nm |
Selection and ordering checklist for HI97725
- Confirm the four parameters required: free chlorine, total chlorine, cyanuric acid and pH
- Order the reagent sets for each method; none are included
- Check CAL Check kits for chlorine (HI97701-11), cyanuric acid (HI97722-11) and pH (HI97710-11)
- Note that pH is limited to 6.5 to 8.5, so highly alkaline or acidic samples need a pH meter
- Decide whether powder or liquid chlorine reagents suit the site
Validation, handling and maintenance of HI97725
Run validation per parameter with the matching CAL Check kit. Rinse cuvettes between methods because cyanuric acid residue is turbid and phenol red stains; the meter stores GLP data and 50 readings.
Keep cuvettes clean with the cleaning solution and a lint-free cloth, never touch the lower half of the glass, and store the meter in its carrying case. Replace the three AA cells when the battery icon appears; the meter keeps its logged readings and settings. For long storage between field trips remove the batteries and close the reagent bottles tightly.
Typical applications of HI97725 in Oman and the GCC
Resort, hotel and club pools across Oman and the GCC need stabiliser control because strong sun burns off chlorine, and cyanuric acid readings keep dosing efficient. The meter also serves water-park, spa and gymnasium operators and facility-management contractors who service multiple sites.
Heat is a practical factor in the Gulf. Reaction times in the manual are given at 25 °C, so samples that sit in a vehicle at much higher temperature should be allowed to equilibrate, and the meter itself is rated for 0 to 50 °C. The sealed optics and IP67 floating case also cope with dust and spray on plant walkdowns.
Information to send with an RFQ for HI97725
- Quantity of meters and whether reagent sets (the chlorine, HI93722 cyanuric acid and HI93710 pH sets; CAL Check HI97701-11, HI97722-11 and HI97710-11) and CAL Check cuvettes are also required
- Sample matrix and expected concentration, so we can confirm the range suits the application
- Delivery location in Oman or the GCC and required lead time
- Any need for spare cuvettes, carrying case or cleaning solution
- Project or tender reference, if applicable
Documents and quotation
The instruction manual for the HI97725 is attached on this page and covers specifications, the reagent procedure, CAL Check validation and error messages.
Send the part number HI97725, the quantity and your delivery location and we will reply with availability and a quotation.
Other Models in the Hanna Hi977xx Photometer Range
Seven Star LLC supplies models from the Hanna Hi977xx Photometer range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.
Datasheets & Documents
- Manual Hanna HI97725 Instruction Manual Download
Reviews (0)
How would you rate this product? Cancel reply
About brand
Hanna Instruments is a maker of analytical instruments for water and liquid testing, founded in 1978 in the Padua area of Italy by Oscar and Anna Nardo. Its early breakthrough was the pHep, an affordable pocket-sized pH tester. The company is now led by the founders' children, Martino and Pamela Nardo, with its headquarters in Woonsocket, Rhode Island, and offers more than 3,000 products.
Seven Star LLC is an independent supplier in Muscat. We list Hanna instruments for water, wastewater and laboratory users in Oman and the GCC.
Hanna Instruments company profile
| Founded | 1978, near Padua, Italy, by Oscar and Anna Nardo |
|---|---|
| Headquarters | Woonsocket, Rhode Island, USA |
| Leadership | Martino Nardo and Pamela Nardo |
| Early product | The pHep pocket pH tester |
| Product families | pHep and HI98 waterproof testers, edge benchtop meters, HI83 multiparameter photometers, Checker HC handheld colorimeters, HI931 and HI932 titrators, process controllers, reagents and calibration buffers |
Hanna product series
Portable meters include the HI98190 pH and ORP meter, the HI98192 EC, TDS and salinity meter, the HI98193 dissolved oxygen and BOD meter, the HI98199 pH, EC and DO meter and the HI9829 multiparameter water quality meter.
The HI98703 is an EPA compliant portable turbidity meter, the HI97711 a free and total chlorine photometer and the HI83414 a benchtop turbidity and chlorine meter.
Waterproof portable meters such as the HI98190 and HI98192 suit field use, with automatic temperature compensation and GLP data for audits. Turbidity measurement with the HI98703 follows EPA method 180.1, and the HI97711 photometer measures free and total chlorine for disinfection control.
Where Hanna instruments are used
Drinking water and desalination plants, wastewater treatment, environmental field surveys, aquaculture, cooling water and quality control labs.
Hanna Instruments products we list
Ordering Hanna instruments: what we need
Send the model, the parameters and ranges, the probe and cable length and any reagents or calibration solutions required.
Facts checked against Hanna Instruments' own company story and City of Woonsocket business profile, October 2026.
Browse Hanna Instruments products by category: Portable Test & Measurement.
FAQ
Questions about the Hanna HI97725 Chlorine, Cyanuric Acid and pH
Answers from our engineering team on installation, compatibility and integration.
01Which parameters can the HI97725 measure?
Free chlorine, total chlorine (0.00 to 5.00 mg/L), cyanuric acid (0 to 80 mg/L) and pH (6.5 to 8.5), selected from a method menu.
02What methods does it use for each?
DPD adapted from EPA 330.5 for chlorine, a turbidimetric method for cyanuric acid and the phenol red method for pH.
03Are reagents for all four tests included?
No. Reagent sets HI93701, HI93711, HI93722 and HI93710 are separate, in 100 and 300 test sizes depending on the method.
04How accurate is the cyanuric acid reading?
Accuracy is 1 mg/L plus 15 percent of reading at 25 degrees C with 1 mg/L resolution, reflecting the turbidimetric nature of the test.
05Can it replace a pH meter?
It gives pH within 6.5 to 8.5 with 0.1 pH resolution, suitable for pool control, but a probe-based pH meter is better for wider ranges.


















Reviews
There are no reviews yet.