- Automation & ControlView all 2,895 products
- Automation Accessories
- AC Drives
- PLC & I/O Cards
- Industrial Communications
- Vibration Monitoring
- Controllers & Data Acquisition
- Isolator & Barrier
- Safety Switches & Guard Locking
- Motor Soft Starters
- CCTV & Security
- Encoders
- Pneumatic Cylinders
- HMI Displays
- Servo Drives
- Industrial Computers
- Machine Vision
- Automation Software
- Robotics
Brand: Hanna Instruments
Hanna HI97762 Ultra Low Range Free Chlorine Photometer
Description
Hanna HI97762 Free chlorine ultra low range portable photometer
The Hanna HI97762 is a pocket-sized, single-parameter photometer built for very low chlorine residuals where an ordinary 0.01 mg/L resolution is too coarse. It reads 0.000 to 0.500 mg/L with 0.001 mg/L resolution, which is the range that matters downstream of dechlorination and in sensitive process waters. Seven Star LLC is an independent supplier in Muscat and is not affiliated with the manufacturer; we quote the HI97762 to industrial and laboratory buyers across Oman and the GCC.
The model number follows Hanna’s HI977xx pattern for dedicated photometers: the last digits identify the chemistry, so HI97762 always means free chlorine ultra low range and cannot be reconfigured for another analyte. Reagents carry their own HI93xxx codes, listed below, and must be ordered separately from the meter.
HI97762 technical specifications
| Parameter | Specification |
|---|---|
| Manufacturer / part number | Hanna Instruments HI97762 |
| Parameter | Free chlorine ultra low range |
| Range | 0.000 to 0.500 mg/L as Cl2 |
| Resolution | 0.001 mg/L |
| Accuracy at 25 °C | ±0.020 mg/L ±3 % of reading |
| Method | Adaptation of Standard Methods 4500-Cl G, DPD colorimetric |
| Light source | LED with narrow-band interference filter at 525 nm |
| Filter bandwidth / wavelength accuracy | 8 nm / ±1.0 nm |
| Detector | Silicon photocell |
| Cuvette | Round, 24.6 mm diameter (22 mm inside) |
| Data logging | 50 readings (auto logging) |
| Display | 128 × 64 pixel monochrome LCD with backlight |
| Power | Three 1.5 V AA alkaline batteries; more than 800 measurements without backlight |
| Operating environment | 0 to 50 °C; 0 to 100 % RH |
| Ingress protection | IP67, floating case |
| Dimensions / weight | 142.5 × 102.5 × 50.5 mm / 380 g with batteries |
| Warranty | Two years |
How the HI97762 measurement works
Free chlorine and DPD form a pink product measured at 525 nm. The ultra-low-range reagent is formulated for the narrow window, and the meter is zeroed on the sample so that the low signal is read from a clean baseline.
Inside the case a reference detector tracks the LED output, so battery sag or a warm ambient does not move the blank. A lens gathers the light leaving the cuvette, which is why scratches and small glass imperfections matter less than with a simple colorimeter, and the cap notch locks the cuvette in the same orientation every time. A built-in tutorial mode walks the operator through sampling, reagent quantity and timers on screen, so the instrument can be handed to a technician who is not a chemist.
Reagents, procedure and interferences for HI97762
Fill the cuvette with 10 mL of sample, zero the meter, add one HI95762-0 packet, mix, and press Read for a 1 minute countdown before the result appears.
| Code | Description | Quantity per test |
|---|---|---|
| HI95762-0 | Free Chlorine Ultra Low Range Reagent | 1 packet |
Reagent sets are supplied as HI95762-01 (100 tests) and HI95762-03 (300 tests). Follow the zeroing order given in the manual for this method, and read within the time the manual allows after the countdown ends.
Known interferences listed for this method:
- Alkalinity above 1,000 mg/L CaCO3 as bicarbonate (pH below 8.3) or above 25 mg/L as carbonate (pH above 9.0), which may give incomplete or fading colour; neutralise with dilute HCl
- Acidity above 150 mg/L CaCO3; neutralise with dilute NaOH
- Hardness above 500 mg/L CaCO3; shake the sample for about 2 minutes after adding the powder
- Bromine, chlorine dioxide, iodine, oxidised manganese and chromium and ozone give a positive error
Where the HI97762 sits among Hanna photometers
The HI97762 is a verification tool for low residuals. For routine drinking-water control up to 5 mg/L use the HI97701; for high-dose disinfection up to 10 mg/L, with free and total, use the HI97734.
| Model | Parameter | Range | LED filter |
|---|---|---|---|
| HI97700 | Ammonia low range | 0.00 to 3.00 mg/L as NH3-N | 420 nm |
| HI97715 | Ammonia medium range | 0.00 to 10.00 mg/L as NH3-N | 420 nm |
| HI97733 | Ammonia high range | 0.0 to 100.0 mg/L as NH4+ | 420 nm |
| HI97701 | Free chlorine | 0.00 to 5.00 mg/L as Cl2 | 525 nm |
| HI97762 | Free chlorine ultra low range | 0.000 to 0.500 mg/L as Cl2 | 525 nm |
| HI97734 | Free and total chlorine, high range | 0.00 to 10.00 mg/L as Cl2 | 525 nm |
| HI97725 | Chlorine, cyanuric acid and pH | 0.00 to 5.00 mg/L Cl2; 0 to 80 mg/L CYA; 6.5 to 8.5 pH | 525 nm |
| HI97706 | Phosphorus | 0.0 to 15.0 mg/L as P | 525 nm |
| HI97714 | Cyanide | 0.000 to 0.200 mg/L as CN– | 610 nm |
| HI97718 | Iodine | 0.0 to 12.5 mg/L as I2 | 525 nm |
| HI97731 | Zinc | 0.00 to 3.00 mg/L as Zn | 575 nm |
| HI97737 | Silver | 0.000 to 1.000 mg/L as Ag | 575 nm |
Selection and ordering checklist for HI97762
- Confirm the target residual is below 0.500 mg/L
- Order HI95762-01 or HI95762-03 reagent sets separately
- Add CAL Check kit HI97762-11 if low-level validation is part of your QA routine
- Plan scrupulously clean sample bottles and cuvettes, because the range is very small
- Check alkalinity and oxidant interferences
Validation, handling and maintenance of HI97762
At sub-mg/L levels, cuvette cleanliness and chlorine demand of glassware dominate error. Rinse with sample, avoid fingerprints in the light path, and validate periodically with the HI97762-11 CAL Check cuvettes.
Keep cuvettes clean with the cleaning solution and a lint-free cloth, never touch the lower half of the glass, and store the meter in its carrying case. Replace the three AA cells when the battery icon appears; the meter keeps its logged readings and settings. For long storage between field trips remove the batteries and close the reagent bottles tightly.
Typical applications of HI97762 in Oman and the GCC
Operators in Oman use ultra-low-range chlorine checks after dechlorination ahead of reverse-osmosis membranes, in ultrapure and boiler-feed make-up lines, in pharmaceutical and food plants and at hospitals and laboratories that must show chlorine absence. It also suits verifying that discharge water meets local residual limits.
Heat is a practical factor in the Gulf. Reaction times in the manual are given at 25 °C, so samples that sit in a vehicle at much higher temperature should be allowed to equilibrate, and the meter itself is rated for 0 to 50 °C. The sealed optics and IP67 floating case also cope with dust and spray on plant walkdowns.
Information to send with an RFQ for HI97762
- Quantity of meters and whether reagent sets (HI95762-01 or HI95762-03; CAL Check HI97762-11) and CAL Check cuvettes are also required
- Sample matrix and expected concentration, so we can confirm the range suits the application
- Delivery location in Oman or the GCC and required lead time
- Any need for spare cuvettes, carrying case or cleaning solution
- Project or tender reference, if applicable
Documents and quotation
The instruction manual for the HI97762 is attached on this page and covers specifications, the reagent procedure, CAL Check validation and error messages.
Send the part number HI97762, the quantity and your delivery location and we will reply with availability and a quotation.
Other Models in the Hanna Hi977xx Photometer Range
Seven Star LLC supplies models from the Hanna Hi977xx Photometer range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.
Datasheets & Documents
- Manual Hanna HI97762 Instruction Manual Download
Reviews (0)
How would you rate this product? Cancel reply
About brand
Hanna Instruments is a maker of analytical instruments for water and liquid testing, founded in 1978 in the Padua area of Italy by Oscar and Anna Nardo. Its early breakthrough was the pHep, an affordable pocket-sized pH tester. The company is now led by the founders' children, Martino and Pamela Nardo, with its headquarters in Woonsocket, Rhode Island, and offers more than 3,000 products.
Seven Star LLC is an independent supplier in Muscat. We list Hanna instruments for water, wastewater and laboratory users in Oman and the GCC.
Hanna Instruments company profile
| Founded | 1978, near Padua, Italy, by Oscar and Anna Nardo |
|---|---|
| Headquarters | Woonsocket, Rhode Island, USA |
| Leadership | Martino Nardo and Pamela Nardo |
| Early product | The pHep pocket pH tester |
| Product families | pHep and HI98 waterproof testers, edge benchtop meters, HI83 multiparameter photometers, Checker HC handheld colorimeters, HI931 and HI932 titrators, process controllers, reagents and calibration buffers |
Hanna product series
Portable meters include the HI98190 pH and ORP meter, the HI98192 EC, TDS and salinity meter, the HI98193 dissolved oxygen and BOD meter, the HI98199 pH, EC and DO meter and the HI9829 multiparameter water quality meter.
The HI98703 is an EPA compliant portable turbidity meter, the HI97711 a free and total chlorine photometer and the HI83414 a benchtop turbidity and chlorine meter.
Waterproof portable meters such as the HI98190 and HI98192 suit field use, with automatic temperature compensation and GLP data for audits. Turbidity measurement with the HI98703 follows EPA method 180.1, and the HI97711 photometer measures free and total chlorine for disinfection control.
Where Hanna instruments are used
Drinking water and desalination plants, wastewater treatment, environmental field surveys, aquaculture, cooling water and quality control labs.
Hanna Instruments products we list
Ordering Hanna instruments: what we need
Send the model, the parameters and ranges, the probe and cable length and any reagents or calibration solutions required.
Facts checked against Hanna Instruments' own company story and City of Woonsocket business profile, October 2026.
Browse Hanna Instruments products by category: Portable Test & Measurement.
FAQ
Questions about the Hanna HI97762 Ultra Low Range Free Chlorine
Answers from our engineering team on installation, compatibility and integration.
01What makes the HI97762 different from HI97701?
It reads 0.000 to 0.500 mg/L free chlorine with 0.001 mg/L resolution, a far finer window than the 0.00 to 5.00 mg/L range of the HI97701.
02What reagent is required?
One HI95762-0 free chlorine ultra low range packet per test, supplied in sets HI95762-01 for 100 tests and HI95762-03 for 300 tests.
03Which standard method is it based on?
An adaptation of Standard Methods for the Examination of Water and Wastewater, 4500-Cl G, the DPD colorimetric method, read at 525 nm.
04How long is the countdown?
One minute after reagent addition; the operator can start the reading earlier by pressing the Read key a second time.
05Is a validation kit available?
Yes, the HI97762-11 CAL Check cuvette kit lets the user verify the meter and apply a user calibration if the check is out of tolerance.


















Reviews