Brand:

JUMO LOGOSCREEN 700 706530 Paperless Recorder

Description

A 144 x 144 mm touchscreen recorder that scales from 0 to 18 measurement inputs

The JUMO LOGOSCREEN 700 (type 706530) is a highly scalable paperless recorder with an icon-based touch interface on a 5.7 inch colour TFT screen. It can be ordered without any measurement input, recording up to 120 process values over Ethernet or Modbus, or equipped with up to 18 universal analog inputs, 3 analog outputs, 18 digital inputs, 24 switchable digital inputs or outputs and 7 relay outputs. Ethernet, two USB ports and an RS232/RS485 interface are standard. Optional extra codes add FDA 21 CFR Part 11 compliant recording with up to 50 users, manipulation detection with a digital certificate, math and logic channels, structured text programming, PROFINET IO device and AMS2750/CQI-9 support. Users can build up to 10 process screens of 100 objects each and record up to 5 batch reports.

Technical specifications

Parameter Specification
Type 706530 paperless recorder, 144 x 144 mm panel mounting (DIN IEC 61554)
Display 5.7 inch (14.5 cm) TFT, 640 x 480 pixel, 65,536 colours; resistive touchscreen
Analog inputs Up to 18 universal inputs: RTD, thermocouple, resistance, potentiometer, 0(4) to 20 mA, voltage; HI inputs to AC 300 V isolation
Slots Three slots for 3 analog + 6 digital + 1 analog output, 3 analog (HI) + 4 digital + 4 I/O, or 6 analog inputs; slot 4 for 12 digital I/O or 6 relays
Analog outputs Up to 3, 0 to 10 V or 0(4) to 20 mA
Digital I/O Up to 18 digital inputs and 24 switchable inputs/outputs, DC 0/24 V; 2 inputs to 12.5 kHz
Relays 1 changeover relay as standard; up to 6 additional changeover relays (slot 4)
Interfaces 1 x Ethernet 10/100, USB host and USB device, RS232/RS485 (Modbus master/slave or barcode scanner); optional PROFINET IO device
External data 120 analog and 120 digital values via Modbus or PROFINET; 64 batch texts and 10 event texts
Recording Sampling 125 ms; 60 analog and 60 digital channels; up to 10 groups; 1 GByte internal memory
Recording time About 42 days at 125 ms, about 8 months at 1 s, about 41 months at 5 s (6 analog channels per group)
Functions 120-channel limit monitoring, 30 counters/integrators, 8 high-speed counters, flow measurement on 8 channels, GPS (NMEA 0183) logging, web server
Optional channels 20 math and 20 logic channels (260); structured text ST code (221)
Supply AC 110 to 240 V +10/-15 % or AC/DC 20 to 30 V, 48 to 63 Hz
Case Zinc steel case, cut-out 138 x 138 mm, panel 2 to 8 mm, depth 120.9 mm, up to 1.75 kg; front IP66, rear IP20
Ambient -20 to +50 °C operation (0 to 40 °C with case 970), storage -20 to +60 °C, up to 85 % humidity
Approvals cULus (E201387); data recording to FDA 21 CFR Part 11 with extra code 888

Order code and extra codes

The code is 706530 / extension – language / options 1 to 4 – supply / extra codes. Extension 0 has no software and 1 includes the setup program with USB cable, PCA3000 and PCC; with extra code 888 it also includes PCS and PCAT. Language 8 is German and English by default. Each of options 1 to 3 is 0 (unused), 1, 2 or 3, and option 4 is 0, 4 (12 digital I/O) or 5 (6 relays). Supply is 23 or 25. Extra codes are 260, 221, 887, 888, 163 (PROFINET), 879 (AMS2750/CQI-9), 949 (TUS/SAT field test device) and 970 (compact carrying case). The order example is 706530/1 8-1 2 3 4-23/260,887,163,970. Later expansion of the options is only possible at the manufacturer.

Recording, security and applications

Measured values are stored in a working memory and copied to flash in 20 kByte blocks, in an encrypted format. A lithium battery buffers memory and clock for more than seven years. Operating modes are standard, time and event operation, in rising priority, each with its own memory cycle and value type (current, average, minimum or maximum). User management supports 5 users, or 50 with extra code 888. Data is read through a USB drive, Ethernet or the PCC software, and PCA3000 is used for evaluation. The built-in web server shows the same visualisation as the screen, and email can be sent through an SMTP server with time sync through SNTP.

Use the model without inputs as a central trend display for a PLC or building system, or fit universal inputs to log furnace, autoclave, cold-chain and water treatment data. Batch recording suits food, pharma and sterilisation processes where documentation is needed. When thermocouples are used for accurate heat-treatment work, the optional TC terminals with integrated cold junction (30053080) improve accuracy, and extra code 879 requires the channels to be specified for the calibration certificate. Related controllers and loggers are in the controllers and data acquisition category.

An enquiry should list the number and type of channels, the digital and relay needs, the supply voltage, the communication interface, and whether FDA 21 CFR Part 11, manipulation detection, math and logic, structured text or PROFINET are required. Mention the evaluation software package so that the right extension code is chosen. The delivery includes the recorder, a quick start guide and four mounting elements, and the terminal cover is available as an accessory.

Documents and quotation

The hosted document is JUMO data sheet 706530 (30 pages, English). Confirm option slots, supply and extra codes before ordering. Seven Star LLC is an independent supplier in Muscat; send the channel list for a quotation.

Datasheets & Documents
  • PDF Datasheet JUMO 706530 data sheet 706530 (EN) Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

JUMO is the German maker of temperature sensors, process controllers, pressure transmitters and liquid analysis instruments. Moritz Kurt Juchheim founded it in Fulda in 1948 as M. K. Juchheim KG, making glass thermometers with six employees. The name JUMO comes from the founder's name. It is still a family company headquartered in Fulda, and the parent company has been called JUMO GmbH & Co. KG since 2003.

Seven Star LLC is an independent supplier in Muscat. We list JUMO RTD and thermocouple probes, thermowells, transmitters, controllers, recorders and water analysis instruments, and supply them to process plants, OEMs, HVAC contractors and water utilities in Oman and the GCC by type number.

JUMO company profile

Established 1948, Fulda, Germany, by Moritz Kurt Juchheim
First products Glass and contact thermometers; platinum glass sensors from 1966
Ownership Family-owned; renamed JUMO GmbH & Co. KG in 2003
Product families RTD and thermocouple probes, thermowells, temperature and pressure transmitters, controllers and thermostats, safety temperature limiters, paperless recorders, liquid analysis and flow and level instruments

Milestones

  • 1948: Moritz Kurt Juchheim founds M. K. Juchheim KG in Fulda, making glass thermometers with six employees.
  • 1966: Platinum glass temperature sensors are added.
  • 2003: The parent company is renamed JUMO GmbH & Co. KG.

JUMO products we list

Family Models Typical use
RTD and thermocouple probes 902020, 902030, 902050, 902120, 902006 2xPt100, PROCESStemp, mineral-insulated RTD, 901020 thermocouple, 901210, 902810 hygienic, 902550 surface, 902520 HVAC, MarineTemp Temperature measurement in pipes, vessels, machines, HVAC and food plants
Thermowells and inserts 909710 thermowells, 909740 measuring inserts Protecting sensors and replacing them without opening the process
Temperature transmitters dTRANS T01, T03, T04, T05, T06 SIL, T07 HART, T09, T1000 IO-Link, DELOS T switch Converting sensor signals to 4-20 mA, HART or IO-Link
Pressure and level dTRANS p20, p30, p31 hygienic, p33 Ex, p02 DELTA, MIDAS S05, MIDAS DP10, DELOS S02, MAERA S25, NESOS R20 Pressure in process, hydraulic and OEM systems; tank and well level
Controllers, limiters and recorders DICON touch, IMAGO 500, dTRON 316, iTRON, cTRON, diraTRON 104, eTRON T, safetyM STB/STW, safetyM TB/TW, LOGOSCREEN 601, mTRON T Furnace, oven and process control, over-temperature protection and recording
Liquid analysis and flow AQUIS touch S, AQUIS touch P, AQUIS 500 pH, AQUIS 500 CR, dTRANS pH 02, pH electrode, digiLine DO, ecoLine NTU, flowTRANS MAG H20, flowTRANS US W01 pH, conductivity, oxygen and turbidity in water treatment; liquid flow

Choosing a temperature sensor starts with the range and accuracy. A Pt100 RTD is accurate and stable from about -200 to 600 °C and is the usual choice in process plants. A thermocouple (type K or J, for example) covers much higher temperatures and responds faster but is less accurate. The wiring also matters: a 3-wire or 4-wire Pt100 connection cancels the resistance of the cable, while a 2-wire connection adds an error that grows with cable length. That is why a head-mounted transmitter is often fitted close to the sensor.

Where JUMO products are used

In Oman and the GCC, JUMO sensors and controllers are used in oil and gas and chemical plants, food and beverage production, plastics and metal heat treatment, HVAC and district cooling, water and wastewater treatment, swimming pools and marine applications.

Ordering JUMO products: what we need

Send the JUMO type number and the full order code from the label (for example 902020/10-...). For probes, add the sensor type (Pt100, Pt1000, type K or J), the number of wires, the insertion length and diameter, and the process connection. For controllers, add the input, outputs and supply voltage.

Facts checked against JUMO's company history, October 2026.

Frequently Asked Questions

How many measuring inputs can the LOGOSCREEN 700 have?
Up to 18 universal analog inputs, or none at all if values come over an interface.
Does it support FDA 21 CFR Part 11?
Yes, extra code 888 provides compliant recording with a digital certificate and up to 50 users.
How long can it record?
About 42 days at 125 ms and about 8 months at 1 s, for a group of six analog channels.
What panel cut-out is needed?
138 x 138 mm for a 144 x 144 mm front, with panel thickness 2 to 8 mm.
Can it join a PROFINET network?
Yes, extra code 163 adds a PROFINET IO device interface.