Brand:

Panametrics oxy.IQ Two-Wire Oxygen Transmitter

Description

Panametrics oxy.IQ overview

The Panametrics oxy.IQ is a compact, two-wire, loop-powered oxygen transmitter with an integrated display and a linearized 4 to 20 mA output. At only 10.4 cm (4.1 in) tall, it measures oxygen in ten ppm ranges and six percentage ranges, all user selectable. The Panametrics oxy.IQ uses an advanced galvanic fuel cell sensor, which the manufacturer describes as stable, accurate and long lived, with no electrolyte to change and no electrodes to clean.

A microprocessor lets the user select the range, trim outputs and carry out calibration from the keypad. The interface includes sensor diagnostics such as a sensor lifetime indication and raw signal, a low reading or sensor failure signal, NAMUR error indication and temperature compensation. With optional Zener barriers or a galvanic isolator, the oxy.IQ can be mounted in a hazardous location, and the keypad remains usable there.

Key features of the oxy.IQ

  • Two-wire, loop-powered 4 to 20 mA transmitter with display and keypad.
  • Ten ppm ranges from 0 to 10 up to 0 to 10,000 ppmv and six percent ranges from 0 to 1 up to 0 to 50 percent.
  • Acid gas compatible sensor versions (OX-2 and OX-4 cells).
  • Intrinsically safe option for Class 1 Division 1 and Division 2 and ATEX/IECEx Zone 0, non-incendive for Division 2.
  • Ex flameproof package option, Ex d IIC T4 IP66.
  • Sensor lifetime indication, sensor failure output and NAMUR error indication.
  • Choice of base adapters including ISO KF/NW40 and ambient air adapter.

oxy.IQ technical specifications

Parameter Value
Sensor technology Galvanic fuel cell
Power 24 to 28 VDC at 50 mA
Accuracy Plus or minus 1 percent of range; plus or minus 2 percent for 0 to 10 ppmv range
Repeatability Plus or minus 1 percent of range
Resolution Plus or minus 0.1 percent of range
Linearity Plus or minus 2 percent of range (OX-1, 2, 3, 5); plus or minus 5 percent (OX-4)
Sensor operating temperature 0 to 45 deg C (32 to 113 deg F)
Sample pressure 10 psig (0.69 barg) maximum operating; 200 psig sample manifold burst
Process connection 1/8 in NPT inlet and outlet
Recommended sample flow 1.0 SCFH (500 cc/min)
Wetted materials 316 stainless steel, Viton O-ring, gold plated contacts and glass; Chemraz option
Dimensions 4.10 x 2.75 x 2.05 in (104.1 x 69.9 x 52.1 mm)
Weight 1.35 lb (612 g)
Pressure effect Plus or minus 0.13 percent of reading per mmHg
IS certification ATEX/IECEx Ex ia IIC Ga T4 Tamb -20 to 60 deg C; US/Canada Class I Div 1 Groups ABCD T4

The ordering code OXY.IQ-BCD-E selects the sensor, package, cable length, Zener barrier or galvanic isolator, base adapter and special options, so state each choice in your RFQ.

Typical applications in Oman and the GCC

Applications named by the manufacturer include glove box purge and leak detection, natural gas, semiconductor wafer machines, coating process machines, membrane air separators, inert welding gases, pure gaseous hydrocarbon streams, monomer processing, and heat treating or bright annealing. For GCC users, the ppm sensors suit nitrogen blanketing checks and hydrogen or hydrocarbon purity monitoring, while the percent sensors suit combustion and inerting checks.

Selection and compatibility notes

Pick the sensor first. OX-1 and OX-5 are standard ppm cells, OX-2 and OX-4 tolerate acid gases, and OX-3 is the standard percent cell. Check the interference table in the datasheet for H2S, SO2, CO2 and HCl levels, because NO2 reads as twice its concentration at ppm levels. Then pick the package: general purpose, intrinsically safe, or one of the aluminium or 316SS flameproof enclosures with 6 mm or 1/4 inch fittings.

Installation and commissioning

Install the oxy.IQ at the sampling point and vent to atmosphere during operation and calibration. For intrinsically safe installations use a Zener barrier or galvanic isolator, one IS cable and one non-IS cable, and respect the 250 ohm load requirement. Cable part numbers 704-1317 (black jacket) and 704-1318 (blue jacket) are named in the datasheet for the two installation types.

Maintenance and service life

Minimal maintenance is needed because the sensor is self-contained. Recovery from air at low ppm levels takes only a few minutes. Watch the sensor lifetime indication, calibrate at the interval required by your accuracy needs, keep pressure and flow constant during calibration, and avoid exposing the cell to interference gases longer than the intermittent limits in the table.

Related products and ordering notes

The oxy.IQ is also used inside the Panametrics H2pure.IQ hydrogen purity package, and it complements the XMO2 for percent-level duty. For your RFQ for the oxy.IQ, give the sensor type, range, hazardous area package, cable length and base adapter. Seven Star LLC is an independent supplier serving Oman and the GCC and is not affiliated with the manufacturer.

Quick answers about the oxy.IQ

  • How is the oxy.IQ powered? It is a two-wire loop-powered transmitter needing 24 to 28 VDC at 50 mA, with a linearized 4 to 20 mA output.
  • What oxygen ranges can the oxy.IQ measure? Ten ppm ranges from 0 to 10 up to 0 to 10,000 ppmv and six percent ranges from 0 to 1 up to 0 to 50 percent, depending on sensor.
  • Can the oxy.IQ be used in hazardous areas? Yes, with intrinsically safe, non-incendive and flameproof package options, certified to ATEX, IECEx and US and Canada requirements.

Documents and quotation

A manufacturer datasheet for the oxy.IQ is available from Seven Star LLC on request and is attached to this listing where provided. To receive a quotation, send the RFQ with the part number oxy.IQ, the quantity required and your delivery location, together with any hazardous area, range or configuration details from your project specification.

What to Send With Your Panametrics oxy.IQ Enquiry

A fast, accurate quotation for the Panametrics oxy.IQ depends on a few details. Please share the full part number, any suffix codes for options or approvals, the operating conditions at the installation point, and whether certificates of conformity or test reports must ship with the goods. It also helps to know if the unit is for planned maintenance, a shutdown, or an urgent breakdown, because that decides how we source and ship it. With that information we can check compatibility with your existing gas analyzer systems, confirm availability, and reply with a clear delivery estimate for your site.

Other Models in the Panametrics Oxygen Range

Seven Star LLC supplies models from the Panametrics Oxygen range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.

ModelCategory
Panametrics XMO2 Thermoparamagnetic Oxygen TransmitterGas Analyzer SystemsView details
Panametrics oxy.IQ Two-Wire Oxygen TransmitterGas Analyzer SystemsYou are viewing this model
Datasheets & Documents
  • PDF Datasheet Panametrics oxy.IQ Two-Wire Oxygen Transmitter Datasheet Download
Reviews (0)

Reviews

There are no reviews yet.

How would you rate this product?

It takes about 10 seconds — just tap a star. Your email is never published.

About brand

Panametrics is an American maker of ultrasonic flowmeters and moisture, oxygen and gas analysers. It was founded in 1960 and was based in Waltham, Massachusetts, when GE bought it in 2002. It later became a Baker Hughes business, and since 1 January 2026 it has belonged to Crane Company. Its long-running site in Shannon, Ireland, opened in 1973.

Seven Star LLC lists Panametrics flowmeters and analyzers for buyers in Oman: DigitalFlow gas and flare meters, clamp-on and portable liquid meters, and trace moisture analyzers for natural gas and process gases. We are an independent supplier, so each inquiry is checked against the manufacturer's order string before a quotation goes out.

Panametrics company profile

Founded 1960, Massachusetts, USA
Main sites Billerica, Massachusetts, and Shannon, Ireland (since 1973)
Ownership Crane Company since 1 January 2026; previously GE (from 2002) and Baker Hughes
Key technology Transit-time ultrasonic flow measurement and aluminium oxide moisture sensing
Product families TransPort PT900 portable and DigitalFlow fixed ultrasonic flowmeters for liquid, gas and flare gas, aluminium oxide moisture probes and hygrometers, Aurora laser moisture analysers, XMO2 oxygen and XMTC thermal conductivity transmitters

Milestones

  • 1960: Panametrics is founded.
  • 1973: The Shannon, Ireland, site is established.
  • 2002: GE acquires Panametrics.
  • 2026: Crane Company completes its purchase of Baker Hughes' Precision Sensors & Instrumentation, including Panametrics.

Panametrics product families on this page

The listed items follow the manufacturer's own grouping:

  • Gas and flare flow: DigitalFlow GF868 flare gas meter, GM868 and XGM868i gas flow transmitters
  • Liquid flow: XMT1000 transmitter, AT600 AquaTrans clamp-on meter, TransPort PT900 portable meter
  • Moisture: moisture.IQ multichannel analyzer, HygroPro II aluminum oxide transmitter, Aurora TDLAS analyzer for natural gas

Older units may carry GE Panametrics or GE Sensing on the nameplate. They belong to the same product lines, and the model string is read the same way.

Where Panametrics flowmeters and moisture analyzers are used

Flare meters report flared volumes on production sites, refineries and LNG plants. Clamp-on and portable ultrasonic meters check water, hydrocarbon liquid and utility lines without cutting the pipe. Aluminum oxide and TDLAS moisture analyzers watch dehydration units, gas pipelines, hydrogen and industrial gas streams. The manufacturer also names chemical processing, energy and water and wastewater among its markets, which matches what GCC operators ask us for.

Panametrics products we list

Family Examples
Flowmeters AT600 AquaTrans Clamp-On Ultrasonic Wate…, DigitalFlow GM868 General-Purpose Gas Ul…, GF868 DigitalFlow Flare Gas Ultrasonic F…, PT900 TransPort Portable Ultrasonic Flow…, XGM868i DigitalFlow Gas Ultrasonic Flow…, XMT1000 Liquid Flow Ultrasonic Transmitter
Gas Analyzer Systems Aurora TDLAS Moisture Analyzer for Natur…, HygroPro II Aluminum Oxide Moisture Tran…, moisture.IQ Multichannel Trace Moisture…

Ordering Panametrics equipment: what we need

A Panametrics model is a long configuration string, and the transducers or probes are ordered separately. Please send the full model and serial number of the electronics, the transducer or moisture probe part number, pipe size, schedule and material, fluid, pressure and temperature, flow range, hazardous area approval and output or protocol. For a flare meter add the flare gas composition range and velocity range. For moisture, give the dew point range and whether sample system components are required.

Facts checked against GE's 2002 acquisition release, Crane Company's January 2026 completion release and Irish press, October 2026.

FAQ

Questions about the Panametrics oxy.IQ Two-Wire Oxygen Transmitter

Answers from our engineering team on installation, compatibility and integration.

01How is the oxy.IQ powered?
It is a two-wire loop-powered transmitter needing 24 to 28 VDC at 50 mA, with a linearized 4 to 20 mA output.
02What oxygen ranges can the oxy.IQ measure?
Ten ppm ranges from 0 to 10 up to 0 to 10,000 ppmv and six percent ranges from 0 to 1 up to 0 to 50 percent, depending on sensor.
03Can the oxy.IQ be used in hazardous areas?
Yes, with intrinsically safe, non-incendive and flameproof package options, certified to ATEX, IECEx and US and Canada requirements.
04Does the oxy.IQ need electrolyte changes?
No. The galvanic fuel cell is self-contained, with no electrolyte to change and no electrodes to clean.
05What process connection does the oxy.IQ use?
The process unit has 1/8 in NPT inlet and outlet, and several base adapters are offered.