Hanna HI97737 Silver Portable Photometer 0-1.000 mg/L
Description
Hanna HI97737 Silver portable photometer
The Hanna HI97737 is a pocket-sized, single-parameter photometer built for silver in plating baths, rinse waters and silver-based disinfection systems. It reads 0.000 to 1.000 mg/L silver, useful where silver is controlled as a product, a disinfectant or a discharge parameter. Seven Star LLC is an independent supplier in Muscat and is not affiliated with the manufacturer; we quote the Hanna HI97737 to industrial and laboratory buyers across Oman and the GCC.
The model number follows Hanna’s HI977xx pattern for dedicated photometers: the last digits identify the chemistry, so HI97737 always means silver and cannot be reconfigured for another analyte. Reagents carry their own HI93xxx codes, listed below, and must be ordered separately from the meter.
HI97737 technical specifications
| Parameter | Specification |
|---|---|
| Manufacturer / part number | Hanna Instruments HI97737 |
| Parameter | Silver |
| Range | 0.000 to 1.000 mg/L as Ag |
| Resolution | 0.001 mg/L |
| Accuracy at 25 °C | ±0.020 mg/L ±5 % of reading |
| Method | Adaptation of the PAN method |
| Light source | LED with narrow-band interference filter at 575 nm |
| Filter bandwidth / wavelength accuracy | 8 nm / ±1.0 nm |
| Detector | Silicon photocell |
| Cuvette | Round, 24.6 mm diameter (22 mm inside) |
| Data logging | 50 readings (auto logging) |
| Display | 128 × 64 pixel monochrome LCD with backlight |
| Power | Three 1.5 V AA alkaline batteries; more than 800 measurements without backlight |
| Operating environment | 0 to 50 °C; 0 to 100 % RH |
| Ingress protection | IP67, floating case |
| Dimensions / weight | 142.5 × 102.5 × 50.5 mm / 380 g with batteries |
| Warranty | Two years |
How the HI97737 measurement works
Silver forms a coloured complex with PAN indicator in the presence of the buffering and masking reagents; the complex is read at 575 nm against the zeroed sample.
Inside the case a reference detector tracks the LED output, so battery sag or a warm ambient does not move the blank. A lens gathers the light leaving the cuvette, which is why scratches and small glass imperfections matter less than with a simple colorimeter, and the cap notch locks the cuvette in the same orientation every time. A built-in tutorial mode walks the operator through sampling, reagent quantity and timers on screen, so the instrument can be handed to a technician who is not a chemist.
Reagents, procedure and interferences for HI97737
A 25 mL sample is used with four liquid reagents (A 1 mL, B 1 mL, C 2 mL, D 2 mL) plus six drops of HI93703-51 dispersing agent, with several 2 minute countdowns that the meter guides through the screen.
| Code | Description | Quantity per test |
|---|---|---|
| HI93737A-0 | Silver Reagent A | 1 mL |
| HI93737B-0 | Silver Reagent B | 1 mL |
| HI93737C-0 | Silver Reagent C | 2 mL |
| HI93737D-0 | Silver Reagent D | 2 mL |
| HI93703-51 | Dispersing Agent | 6 drops |
Reagent sets are supplied as HI93737-01 (50 tests) and HI93737-03 (150 tests). Follow the zeroing order given in the manual for this method, and read within the time the manual allows after the countdown ends.
Known interferences listed for this method:
- Chloride above 8,000 mg/L and sodium above 5,000 mg/L
- Calcium and magnesium above 1,000 mg/L CaCO3; potassium above 500 mg/L
- Aluminum and zinc above 30 mg/L; chromium(VI) above 40 mg/L; manganese above 25 mg/L
- Cadmium, chromium(III), fluoride and lead above 20 mg/L; copper above 15 mg/L; ferric iron above 10 mg/L
- Cobalt, ferrous iron and nickel above 1.5 mg/L
Where the HI97737 sits among Hanna photometers
The HI97737 is the most reagent-intensive HI977xx model here, with a 25 mL sample and four reagents. Pair it with the HI97731 zinc model if one technician handles both plating and effluent checks.
| Model | Parameter | Range | LED filter |
|---|---|---|---|
| HI97700 | Ammonia low range | 0.00 to 3.00 mg/L as NH3-N | 420 nm |
| HI97715 | Ammonia medium range | 0.00 to 10.00 mg/L as NH3-N | 420 nm |
| HI97733 | Ammonia high range | 0.0 to 100.0 mg/L as NH4+ | 420 nm |
| HI97701 | Free chlorine | 0.00 to 5.00 mg/L as Cl2 | 525 nm |
| HI97762 | Free chlorine ultra low range | 0.000 to 0.500 mg/L as Cl2 | 525 nm |
| HI97734 | Free and total chlorine, high range | 0.00 to 10.00 mg/L as Cl2 | 525 nm |
| HI97725 | Chlorine, cyanuric acid and pH | 0.00 to 5.00 mg/L Cl2; 0 to 80 mg/L CYA; 6.5 to 8.5 pH | 525 nm |
| HI97706 | Phosphorus | 0.0 to 15.0 mg/L as P | 525 nm |
| HI97714 | Cyanide | 0.000 to 0.200 mg/L as CN– | 610 nm |
| HI97718 | Iodine | 0.0 to 12.5 mg/L as I2 | 525 nm |
| HI97731 | Zinc | 0.00 to 3.00 mg/L as Zn | 575 nm |
| HI97737 | Silver | 0.000 to 1.000 mg/L as Ag | 575 nm |
Selection and ordering checklist for HI97737
- Confirm silver is expected between 0.01 and 1.000 mg/L
- Order HI93737-01 (50 tests) or HI93737-03 (150 tests); note the smaller test counts
- Include HI93703-51 dispersing agent in the order if not already stocked
- Review the long interference list for your sample type
- Add HI97737-11 CAL Check cuvettes
Validation, handling and maintenance of HI97737
Use clean 25 mL measuring glassware and keep the reagent order exact. Validate with HI97737-11, and clean cuvettes thoroughly because silver complexes can leave a film. Dispose of reacted samples as metal-bearing waste.
Keep cuvettes clean with the cleaning solution and a lint-free cloth, never touch the lower half of the glass, and store the meter in its carrying case. Replace the three AA cells when the battery icon appears; the meter keeps its logged readings and settings. For long storage between field trips remove the batteries and close the reagent bottles tightly.
Typical applications of HI97737 in Oman and the GCC
In Oman the main uses are plating and jewellery workshops, photographic or electronics waste streams, and silver-ion disinfection in certain water systems. Environmental laboratories also use it for screening industrial discharge.
Heat is a practical factor in the Gulf. Reaction times in the manual are given at 25 °C, so samples that sit in a vehicle at much higher temperature should be allowed to equilibrate, and the meter itself is rated for 0 to 50 °C. The sealed optics and IP67 floating case also cope with dust and spray on plant walkdowns.
Information to send with an RFQ for HI97737
- Quantity of meters and whether reagent sets (HI93737-01 or HI93737-03; CAL Check HI97737-11) and CAL Check cuvettes are also required
- Sample matrix and expected concentration, so we can confirm the range suits the application
- Delivery location in Oman or the GCC and required lead time
- Any need for spare cuvettes, carrying case or cleaning solution
- Project or tender reference, if applicable
Documents and quotation
The instruction manual for the HI97737 is attached on this page and covers specifications, the reagent procedure, CAL Check validation and error messages.
Send the part number HI97737, the quantity and your delivery location and we will reply with availability and a quotation.
Other Models in the Hanna Hi977xx Photometer Range
Seven Star LLC supplies models from the Hanna Hi977xx Photometer range in Oman and the GCC. Compare the related models below and request a quote for the one that matches your application.
Datasheets & Documents
- Manual Hanna HI97737 Instruction Manual Download
Reviews (0)
How would you rate this product? Cancel reply
About brand
Hanna Instruments is a maker of analytical instruments for water and liquid testing, founded in 1978 in the Padua area of Italy by Oscar and Anna Nardo. Its early breakthrough was the pHep, an affordable pocket-sized pH tester. The company is now led by the founders' children, Martino and Pamela Nardo, with its headquarters in Woonsocket, Rhode Island, and offers more than 3,000 products.
Seven Star LLC is an independent supplier in Muscat. We list Hanna instruments for water, wastewater and laboratory users in Oman and the GCC.
Hanna Instruments company profile
| Founded | 1978, near Padua, Italy, by Oscar and Anna Nardo |
|---|---|
| Headquarters | Woonsocket, Rhode Island, USA |
| Leadership | Martino Nardo and Pamela Nardo |
| Early product | The pHep pocket pH tester |
| Product families | pHep and HI98 waterproof testers, edge benchtop meters, HI83 multiparameter photometers, Checker HC handheld colorimeters, HI931 and HI932 titrators, process controllers, reagents and calibration buffers |
Hanna product series
Portable meters include the HI98190 pH and ORP meter, the HI98192 EC, TDS and salinity meter, the HI98193 dissolved oxygen and BOD meter, the HI98199 pH, EC and DO meter and the HI9829 multiparameter water quality meter.
The HI98703 is an EPA compliant portable turbidity meter, the HI97711 a free and total chlorine photometer and the HI83414 a benchtop turbidity and chlorine meter.
Waterproof portable meters such as the HI98190 and HI98192 suit field use, with automatic temperature compensation and GLP data for audits. Turbidity measurement with the HI98703 follows EPA method 180.1, and the HI97711 photometer measures free and total chlorine for disinfection control.
Where Hanna instruments are used
Drinking water and desalination plants, wastewater treatment, environmental field surveys, aquaculture, cooling water and quality control labs.
Hanna Instruments products we list
Ordering Hanna instruments: what we need
Send the model, the parameters and ranges, the probe and cable length and any reagents or calibration solutions required.
Facts checked against Hanna Instruments' own company story and City of Woonsocket business profile, October 2026.
Browse Hanna Instruments products by category: Portable Test & Measurement.
FAQ
Questions about the Hanna HI97737 Silver Portable Photometer 0-1.000
Answers from our engineering team on installation, compatibility and integration.
01What range does the HI97737 cover?
Silver from 0.000 to 1.000 mg/L with 0.001 mg/L resolution and accuracy of 0.020 mg/L plus 5 percent of reading at 25 degrees C.
02Which method does it use?
An adaptation of the PAN method, read at 575 nm with an LED and narrow-band interference filter.
03How many reagents are required?
Four liquid reagents (A to D) plus HI93703-51 dispersing agent per test, supplied as HI93737-01 for 50 tests or HI93737-03 for 150 tests.
04What sample volume is used?
The manual uses a 25 mL sample, larger than the 10 mL used by most of the HI977xx single-parameter photometers.
05Which ions interfere?
Chloride, sodium, calcium, magnesium, potassium and several metals above the thresholds in the manual, including aluminum, zinc, manganese, lead and copper.